A Study on Static Calibration Method for Pressure Sensor Based on Static Pressure Calibration Device

In this paper, a static pressure calibration device is designed and the method of static calibration with piezoelectric pressure sensors is discussed. According to the theory of PASCAL law, the static calibration device is developed, whose stress ratio is up to 1:25. And this device has been accredi...

Full description

Saved in:
Bibliographic Details
Published inApplied Mechanics and Materials Vol. 148-149; pp. 121 - 124
Main Authors Shang, Fei, Xiong, Xian Feng, Kong, De Ren
Format Journal Article
LanguageEnglish
Published Zurich Trans Tech Publications Ltd 01.12.2011
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In this paper, a static pressure calibration device is designed and the method of static calibration with piezoelectric pressure sensors is discussed. According to the theory of PASCAL law, the static calibration device is developed, whose stress ratio is up to 1:25. And this device has been accredited by the authorities. To ensure the consistency of pressure calibration, the improvements on calibration device’s output are made. As a result, we can calibrate up to four sensors at the same time. By means of the pressure calibration device, the calibration method of pressure sensor, the calibration theory, system components and calibration procedures is given. As the test results shown, the corresponding sensors’ static calibration linearity is 0.05%, and its uncertainty is 0.42%. It’s practicable and reliable to make static calibration to sensors by the pressure calibration device.
Bibliography:Selected, peer reviewed papers from the 2011 International Conference on Mechanical Engineering, Materials and Energy, (ICMEME 2011), October 19-21, 2011, Dalian, China
ISBN:3037853409
9783037853405
ISSN:1660-9336
1662-7482
1662-7482
DOI:10.4028/www.scientific.net/AMM.148-149.121