Add-on pacing/sensing lead for successful management of “Defibrillator Shock Storm”: Repair of a Make and Break Circuit
A case of “ICD storm” precipitated by lead noise generation due to fracture in the sensing/pacing circuit component of the ICD lead was successfully repaired by excluding the damaged sensing/pacing circuit component of the ICD lead and putting an additional parallel ventricular lead connected with t...
Saved in:
Published in | IHJ cardiovascular case reports Vol. 2; no. 1; pp. 61 - 63 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
Elsevier B.V
01.01.2018
Elsevier |
Online Access | Get full text |
Cover
Loading…
Summary: | A case of “ICD storm” precipitated by lead noise generation due to fracture in the sensing/pacing circuit component of the ICD lead was successfully repaired by excluding the damaged sensing/pacing circuit component of the ICD lead and putting an additional parallel ventricular lead connected with the existing generator. |
---|---|