Add-on pacing/sensing lead for successful management of “Defibrillator Shock Storm”: Repair of a Make and Break Circuit

A case of “ICD storm” precipitated by lead noise generation due to fracture in the sensing/pacing circuit component of the ICD lead was successfully repaired by excluding the damaged sensing/pacing circuit component of the ICD lead and putting an additional parallel ventricular lead connected with t...

Full description

Saved in:
Bibliographic Details
Published inIHJ cardiovascular case reports Vol. 2; no. 1; pp. 61 - 63
Main Authors Garg, Naveen, Raja, Deep Chandh
Format Journal Article
LanguageEnglish
Published Elsevier B.V 01.01.2018
Elsevier
Online AccessGet full text

Cover

Loading…
More Information
Summary:A case of “ICD storm” precipitated by lead noise generation due to fracture in the sensing/pacing circuit component of the ICD lead was successfully repaired by excluding the damaged sensing/pacing circuit component of the ICD lead and putting an additional parallel ventricular lead connected with the existing generator.