Clinical Outcome of the Metal-on-Metal Hybrid Corin Cormet 2000 Hip Resurfacing System
Abstract This report extends the follow-up for the largest center of the first multicenter US Food and Drug Administration investigational device exemption study on metal-on-metal hip resurfacing arthroplasty up to 11 years. A single surgeon performed 373 hip resurfacing arthroplasties using the hyb...
Saved in:
Published in | The Journal of arthroplasty Vol. 27; no. 4; pp. 533 - 538.e1 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Elsevier Inc
01.04.2012
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Abstract This report extends the follow-up for the largest center of the first multicenter US Food and Drug Administration investigational device exemption study on metal-on-metal hip resurfacing arthroplasty up to 11 years. A single surgeon performed 373 hip resurfacing arthroplasties using the hybrid Corin Cormet 2000 system. The Kaplan-Meier survivorship at 11 years was 93% when revision for any reason was used as an end point and 91% if radiographic failures were included. The clinical results demonstrate an acceptable failure rate with use of this system. Loosening of the cemented femoral components was the most common source of failure and occurred at all follow-up intervals. A learning curve that persisted for at least 200 cases was confirmed. All femoral neck fractures occurred before 6 months postoperatively. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2011.06.019 |