Experimental and theoretical analysis of spurious EPR line features caused by field inhomogeneity
Theory is developed for the detailed lineshapes arising in EPR instruments having inhomogeneous static field. Validity of the theory is established by a comparison with experiment. Results are presented for TE 011 cavities parallel and transverse to the magnetic field and for the TE 102 cavity. The...
Saved in:
Published in | Journal of magnetic resonance (1969) Vol. 37; no. 1; pp. 129 - 139 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
Elsevier Inc
1980
|
Online Access | Get full text |
Cover
Loading…
Summary: | Theory is developed for the detailed lineshapes arising in EPR instruments having inhomogeneous static field. Validity of the theory is established by a comparison with experiment. Results are presented for TE
011 cavities parallel and transverse to the magnetic field and for the TE
102 cavity. The results may be generalized with respect to sample size and amount of magnetic field inhomogeneity. |
---|---|
ISSN: | 0022-2364 1557-8968 |
DOI: | 10.1016/0022-2364(80)90098-0 |