Simultaneous Determination Method of Aniline and Quinoline in Drinking Water by LC/MS/MS
In this study, an analysis was conducted for the simultaneous determination method of aniline and quinoline in drinking water using liquid chromatography/tandem mass spectrometry.The calibration curves of aniline and quinoline, obtained from the results of the study, demonstrated good linearity in t...
Saved in:
Published in | Journal of Environmental Conservation Engineering Vol. 47; no. 5; pp. 278 - 285 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | Japanese English |
Published |
Osaka
Society of Environmental Conservation Engineering
20.07.2018
Japan Science and Technology Agency |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In this study, an analysis was conducted for the simultaneous determination method of aniline and quinoline in drinking water using liquid chromatography/tandem mass spectrometry.The calibration curves of aniline and quinoline, obtained from the results of the study, demonstrated good linearity in the range of 1.0 to 20 μg/L of aniline standard and in the range of 0.01 to 0.020 μg/L of quinolone standard. The repeatability of repeated measurements of aniline standard solution (0.02 μg/L, n=5) was observed to be below 0.82% and the repeatability of repeated measurements of quinoline standard solution (0.02 μg/L, n=5) was observed to be below 2.7%. In addition, the repeatability (4.8-5.6%), intermediate precision (4.8-12%), and accuracy (98.8-107%) values obtained via recovery tests for aniline and quinoline in drinking water (aniline: 1.0 μg/L, and quinoline: 0.01 μg/L, n=5, 5days) satisfied the criteria provided in the guidelines for the validation of testing methods for drinking water notified by the Ministry of Health, Labour and Welfare, Japan. |
---|---|
ISSN: | 0388-9459 1882-8590 |
DOI: | 10.5956/jriet.47.278 |