Comparison of the quality of the result of digitising panoramic analogue films using transmission and reflection scanner mode with direct digital imaging standard

Introduction: The use of panoramic x-ray radiographs with analogue film negatives is still widely used. Contrast differences in the radiograph are determined by the level of light transmitted from the viewer to the film negative and visual perception of the radiologist, which can cause differences a...

Full description

Saved in:
Bibliographic Details
Published inPadjadjaran Journal of Dentistry (Bandung) Vol. 31; no. 2; pp. 99 - 105
Main Authors Azhari, Azhari, Gianto, Gianto, Suprijanto, Suprijanto
Format Journal Article
LanguageEnglish
Published Universitas Padjadjaran 31.07.2019
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Introduction: The use of panoramic x-ray radiographs with analogue film negatives is still widely used. Contrast differences in the radiograph are determined by the level of light transmitted from the viewer to the film negative and visual perception of the radiologist, which can cause differences and errors of interpretation. This study was aimed to compare the quality of scanners produced by the transmission and reflection modes objectively with direct digital imaging standard. Methods: This study used a descriptive comparative design with purposive sampling method. Radiographic analysis was conducted towards 30 analogue radiographs of Radiology laboratory of Universitas Padjadjaran Dental Hospital. Digitisation was performed using a scanner with setting modes (reflection and transmission, 8-bit intensity resolution and spatial resolution of 800 dpi). Digitising the negative radiographs of analogue films into digital radiographs will get a range of percentage pixels. Histogram of radiographs with a percentage of pixels in the grey level range 0 - 64 exceeds 50%, including "under-exposed", in the grey level range 192 - 255 exceeding 50%, including "over-exposed". The evaluation was carried out to observe the radiograph exposure homogeneity by dividing the radiograph into 4 parts. Each part was calculated as the mean pixel value; then the standard deviation was searched for the four mean values (STDMI - standard deviation mean intensity) determined the homogeneity. Results: Digitising type of radiograph reflection was included under-exposed because the percentage of pixels in the grey level range 0 - 64 was 86.94%. STDMI consecutive reflection = 8.03, transmission = 18.05, direct digital imaging = 20.86. Conclusion: Digitisation with scanner transmission mode is the best result objectively because it approaches the quality value of direct digital imaging standard.Keywords: Objective quality, homogeneity, reflection, transmission
ISSN:1979-0201
2549-6212
DOI:10.24198/pjd.vol31no2.21155