Proton beam dosimetry using a TEPC with a 252Cf neutron calibration
A tissue-equivalent proportional counter (TEPC) can measure the linear energy transfer (LET) of incident radiation and directly calculate the equivalent dose to humans in a complicated radiation field. For radiation monitoring, we developed and characterized a TEPC that can simulate a site diameter...
Saved in:
Published in | Journal of the Korean Physical Society Vol. 67; no. 8; pp. 1459 - 1462 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Seoul
The Korean Physical Society
01.10.2015
한국물리학회 |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A tissue-equivalent proportional counter (TEPC) can measure the linear energy transfer (LET) of incident radiation and directly calculate the equivalent dose to humans in a complicated radiation field. For radiation monitoring, we developed and characterized a TEPC that can simulate a site diameter of 2
µ
m for micro-dosimetry. It was calibrated with a
252
Cf neutron standard source at the Korea Research Institute of Standards and Science. The channel to LET calibration factor of the TEPC is about 0.72 keV/
µ
m-channel. Also, we evaluated the possibility of usage the TEPC as a proton dosimeter. |
---|---|
Bibliography: | G704-000411.2015.67.8.019 |
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.67.1459 |