Electric Field Dependence on Charge Collection of CdZnTe X-Ray Detectors
In this study, the electric field dependence on the charge collection process of CdZnTe detectors, at different x-ray tube settings, within the x-ray diagnostic energy range, is investigated. In addition, the detector contrast at different applied bias voltages and x-ray tube settings have been expe...
Saved in:
Published in | Journal of X-ray science and technology Vol. 7; no. 2; pp. 198 - 210 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
London, England
SAGE Publications
1997
|
Online Access | Get full text |
Cover
Loading…
Summary: | In this study, the electric field dependence on the charge collection process of CdZnTe detectors, at different x-ray tube settings, within the x-ray diagnostic energy range, is investigated. In addition, the detector contrast at different applied bias voltages and x-ray tube settings have been experimentally determined. The experimental results suggest that an efficient charge collection process is obtained by increasing the applied bias voltage. Once the applied bias voltage is sufficiently high, charge collection becomes complete and the detector operates in the saturation region. This is a prerequisite for high contrast and spatial resolution. As a result, the detector contrast is enhanced significantly. Therefore, CdZnTe detectors appear to be potential candidates for digital radiographic applications. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0895-3996 1095-9114 |
DOI: | 10.3233/XST-1997-7210 |