Comparison of 131 I activity measurements at the NCBJ RC POLATOM and the ENEA-INMRI linked to the BIPM SIR system
A bilateral comparison between ENEA-INMRI (Italy) and NCBJ RC POLATOM (Poland) of I-solution activity measurements was organized in the year 2015 and piloted by POLATOM, which provided the sources for the comparison. The I master solution was standardized independently at both institutes by using Li...
Saved in:
Published in | Applied radiation and isotopes Vol. 134; pp. 380 - 384 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
England
01.04.2018
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A bilateral comparison between ENEA-INMRI (Italy) and NCBJ RC POLATOM (Poland) of
I-solution activity measurements was organized in the year 2015 and piloted by POLATOM, which provided the sources for the comparison. The
I master solution was standardized independently at both institutes by using Liquid Scintillation Counting and ionization chamber techniques. The
I master solution was then sent by POLATOM to the BIPM International Reference System (SIR). The comparison was registered as an EURAMET.RI(II)-K2.I-131 key comparison allowing the ENEA-INMRI result to enter in the SIR database. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2017.09.040 |