Transcatheter occlusion of the patent ductus arteriosus with Cook detachable coils
OBJECTIVE: To report initial experience with a new occlusion device for native and residual patent ductus arteriosus. DESIGN: Descriptive study of consecutive non-randomised patients undergoing a new method of patent ductus arteriosus closure with detachable coils. SETTING: Tertiary centres for paed...
Saved in:
Published in | Heart (British Cardiac Society) Vol. 76; no. 6; pp. 531 - 535 |
---|---|
Main Authors | , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
London
BMJ Publishing Group Ltd and British Cardiovascular Society
01.12.1996
BMJ BMJ Publishing Group LTD |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | OBJECTIVE: To report initial experience with a new occlusion device for native and residual patent ductus arteriosus. DESIGN: Descriptive study of consecutive non-randomised patients undergoing a new method of patent ductus arteriosus closure with detachable coils. SETTING: Tertiary centres for paediatric cardiology. PATIENTS: 71 consecutive patients, aged 1.2-22 years, with a patent ductus arteriosus (PDA) underwent elective transcatheter closure. 45 had native PDAs (group A) with a minimum diameter of 1.0 mm-5.0 mm (median 2.0 mm). A further 26 had undergone one or more previous occlusion attempts (group B). INTERVENTIONS: A total of 133 detachable (Cook) spring coils were successfully implanted in 70 patients. The procedure was performed transvenously in 51 patients, retrograde arterially in 13, and by both routes in a further 6 patients. One 5 mm coil migrated but was successfully retrieved. MAIN OUTCOME MEASURES: In group A colour flow Doppler echocardiography showed that complete occlusion was achieved in 40/45 (89%) at 24 hours, 41/45 (91%) at 1 month, and 44/45 (98%) by 6 months post procedure. Occlusion rates in residual PDAs were 22/25 (88%) occluded at 24 hours, 23/25 (92%) at 1 month, and 24/25 (96%) at 6 months follow up. CONCLUSIONS: Transcatheter occlusion using detachable (Cook) spring coils is a safe and effective alternative to presently available devices. The delivery system allows full retrieval of the coil until a satisfactory position is obtained. |
---|---|
Bibliography: | local:heartjnl;76/6/531 istex:3DB8114E0F005914D6A207C5B039D40A0B0B26D2 ark:/67375/NVC-HVV49FL3-L href:heartjnl-76-531.pdf PMID:9014804 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1355-6037 1468-201X |
DOI: | 10.1136/hrt.76.6.531 |