Status and development of the TOP-IMPLART Project
The TOP-IMPLART project consists of the design and implementation of a linear proton accelerator, its control and monitoring systems for the treatment of superficial and semi-deep tumors. The energy of 150 MeV (corresponding to a penetration in tissue of about 15 cm) is a milestone in design being u...
Saved in:
Main Authors | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
---|---|
Format | Journal Article |
Language | English |
Published |
12.10.2020
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The TOP-IMPLART project consists of the design and implementation of a linear
proton accelerator, its control and monitoring systems for the treatment of
superficial and semi-deep tumors. The energy of 150 MeV (corresponding to a
penetration in tissue of about 15 cm) is a milestone in design being useful for
the proton therapy treatment of almost 50% of tumors based on their position
and depth (including ocular melanoma, head-neck tumors, pediatric tumors, and
more superficial tumors). The capability to vary the intensity on a
pulse-to-pulse basis combined with an electronic feedback system allows to get
the required dose uniformity (2.5%) reducing the number of re-paintings. In
this paper the state of the art and the objectives of the TOP-IMPLART project
are described within the framework of the progress of Protontherapy. |
---|---|