Characterization of a scintillating lithium glass ultra-cold neutron detector

A \(^{6}\)Li glass based scintillation detector developed for the TRIUMF neutron electric dipole moment experiment was characterized using the ultra-cold neutron source at the Paul Scherrer Institute (PSI). The data acquisition system for this detector was demonstrated to perform well at rejecting b...

Full description

Saved in:
Bibliographic Details
Published inarXiv.org
Main Authors Jamieson, Blair, Rebenitsch, Lori A, Hansen-Romu, Sean, Lauss, Bernhard, Lindner, Thomas, Martin, Jeffery W, Pierre, Edgard
Format Paper Journal Article
LanguageEnglish
Published Ithaca Cornell University Library, arXiv.org 08.11.2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A \(^{6}\)Li glass based scintillation detector developed for the TRIUMF neutron electric dipole moment experiment was characterized using the ultra-cold neutron source at the Paul Scherrer Institute (PSI). The data acquisition system for this detector was demonstrated to perform well at rejecting backgrounds. An estimate of the absolute efficiency of background rejection of \(99.7\pm0.1\)% is made. For variable ultra-cold neutron rate (varying from \(<\)1 kHz to approx. 100 kHz per channel) and background rate seen at the Paul Scherrer Institute, we estimate that the absolute detector efficiency is \(89.7^{+1.3}_{-1.9}\)%. Finally a comparison with a commercial Cascade detector was performed for a specific setup at the West-2 beamline of the ultra-cold neutron source at PSI.
ISSN:2331-8422
DOI:10.48550/arxiv.1608.02912