Photometer for Continuous Determination of Uranium in Radioactive Process Streams

A modified single-beam filter photometer for continuous determination of uranium in radioactive process streams is described. Because of a unique standardizing system, it is very stable, having a short-term variation of less than 1% and a calibration stability of 3%. The instrument has a remotely lo...

Full description

Saved in:
Bibliographic Details
Published inAnalytical chemistry (Washington) Vol. 32; no. 2; pp. 268 - 272
Main Authors Scott, F. A, Dierks, R. D
Format Journal Article
LanguageEnglish
Published American Chemical Society 01.02.1960
American Chemical Society (ACS)
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A modified single-beam filter photometer for continuous determination of uranium in radioactive process streams is described. Because of a unique standardizing system, it is very stable, having a short-term variation of less than 1% and a calibration stability of 3%. The instrument has a remotely located sensing unit with the ruggedness and simplicity required for production plant applications. The performance of the instrument for monitoring nonradioactive process streams under plant conditions is given.
Bibliography:istex:53AC1BD8DAB32FF96F5EC1C9DAC2D78ADF6D2D89
ark:/67375/TPS-89PCDB71-F
USDOE
ISSN:0003-2700
1520-6882
DOI:10.1021/ac60158a036