Effects on radiation oncology treatments involving various neuromodulation devices

Object: Where no society-based or manufacturer guidance on radiation limits to neuromodulation devices is available, this research provides the groundwork for neurosurgeons and radiation oncologists who rely on the computerized treatment plan clinically for cancer patients. The focus of the article...

Full description

Saved in:
Bibliographic Details
Published inJournal of X-ray science and technology Vol. 19; no. 4; pp. 443 - 456
Main Authors Gossman, Michael S., Paralikar, Kunal J., Hebb, Adam O., Wilkinson, Jeffrey D., Graves-Calhoun, Alison R., Lawson, Raymond C., Lopez, Jeffrey P., Powell, James S.
Format Journal Article
LanguageEnglish
Published London, England SAGE Publications 2011
Subjects
Online AccessGet full text

Cover

Loading…
Abstract Object: Where no society-based or manufacturer guidance on radiation limits to neuromodulation devices is available, this research provides the groundwork for neurosurgeons and radiation oncologists who rely on the computerized treatment plan clinically for cancer patients. The focus of the article is to characterize radiation parameters of attenuation and scatter when an incident therapeutic x-ray beam is directed upon them. At the time of this writing, manufacturers of Neuromodulation products do not recommend direct exposure of the device in the beam nor provide guidance for the maximum dose for these devices. Methods: Ten neuromodulation models were chosen to represent the finite class of devices marketed by Medtronic before 2011. CT simulations permitted computer treatment modeling for dose distribution analysis as used routinely in radiation oncology for patients. Phantom case results were directly compared to actual clinical patient cases. Radiation detection measurements were then correlated to computational results. Where the x-ray beam passes through the device and is attenuated, dose reduction was identified with Varian Eclipse computer modeling for these posterior locations. Results: Although the computer algorithm did not identify physical processes of side-scatter and back-scatter, these phenomena were proven by radiation measurement to occur. In general, the computer results underestimated the level of change seen by measurement. Conclusions: For these implantable neurostimulators, the spread in dose changes were found to be −6.2% to −12.5% by attenuation, +1.7% to +3.8% by side-scatter, and +1.1% to +3.1% by back-scatter at 6 MV. At 18 MV, these findings were observed to be −1.4% to −7.0% by attenuation, +1.8% to 5.7% by side-scatter, and 0.8% to 2.7% by back-scatter. No pattern for the behavior of these phenomena was deduced to be a direct consequence of device size.
AbstractList Object: Where no society-based or manufacturer guidance on radiation limits to neuromodulation devices is available, this research provides the groundwork for neurosurgeons and radiation oncologists who rely on the computerized treatment plan clinically for cancer patients. The focus of the article is to characterize radiation parameters of attenuation and scatter when an incident therapeutic x-ray beam is directed upon them. At the time of this writing, manufacturers of Neuromodulation products do not recommend direct exposure of the device in the beam nor provide guidance for the maximum dose for these devices. Methods: Ten neuromodulation models were chosen to represent the finite class of devices marketed by Medtronic before 2011. CT simulations permitted computer treatment modeling for dose distribution analysis as used routinely in radiation oncology for patients. Phantom case results were directly compared to actual clinical patient cases. Radiation detection measurements were then correlated to computational results. Where the x-ray beam passes through the device and is attenuated, dose reduction was identified with Varian Eclipse computer modeling for these posterior locations. Results: Although the computer algorithm did not identify physical processes of side-scatter and back-scatter, these phenomena were proven by radiation measurement to occur. In general, the computer results underestimated the level of change seen by measurement. Conclusions: For these implantable neurostimulators, the spread in dose changes were found to be −6.2% to −12.5% by attenuation, +1.7% to +3.8% by side-scatter, and +1.1% to +3.1% by back-scatter at 6 MV. At 18 MV, these findings were observed to be −1.4% to −7.0% by attenuation, +1.8% to 5.7% by side-scatter, and 0.8% to 2.7% by back-scatter. No pattern for the behavior of these phenomena was deduced to be a direct consequence of device size.
Author Graves-Calhoun, Alison R.
Lopez, Jeffrey P.
Paralikar, Kunal J.
Hebb, Adam O.
Wilkinson, Jeffrey D.
Powell, James S.
Lawson, Raymond C.
Gossman, Michael S.
Author_xml – sequence: 1
  givenname: Michael S.
  surname: Gossman
  fullname: Gossman, Michael S.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 2
  givenname: Kunal J.
  surname: Paralikar
  fullname: Paralikar, Kunal J.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 3
  givenname: Adam O.
  surname: Hebb
  fullname: Hebb, Adam O.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 4
  givenname: Jeffrey D.
  surname: Wilkinson
  fullname: Wilkinson, Jeffrey D.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 5
  givenname: Alison R.
  surname: Graves-Calhoun
  fullname: Graves-Calhoun, Alison R.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 6
  givenname: Raymond C.
  surname: Lawson
  fullname: Lawson, Raymond C.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 7
  givenname: Jeffrey P.
  surname: Lopez
  fullname: Lopez, Jeffrey P.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
– sequence: 8
  givenname: James S.
  surname: Powell
  fullname: Powell, James S.
  organization: Tri-State Christian Neurosurgical Associates, Ashland, KY, USA
BookMark eNqVjkFrwkAUhB9FobHtH-hp_0D0rauSnIviuXrwtizJS1iJ78FuEui_d4PivZeZYZiBbwEzFiaAb41LszZmdTmd8zVqnaPBHU7yBpnGcpuXWm9mkGGRsinL3TssYrxi2m6LIoPffdNQ1UclrIKrvet9SsKVdNL-qT6Q62_EaeB5lG703KrRBS9DVExDkJvUQ_d41TT6iuInzBvXRfp6-gesDvvzzzGPriV7lSFwqq1GO6HbhG4ndPtCN_9_3AFF-06P
ContentType Journal Article
Copyright IOS Press. All rights reserved
Copyright_xml – notice: IOS Press. All rights reserved
DOI 10.3233/XST-2011-030600306
DatabaseTitleList
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Engineering
Physics
EISSN 1095-9114
EndPage 456
ExternalDocumentID 10.3233_XST-2011-030600306
GroupedDBID ---
--K
0R~
1B1
1RT
1~5
29L
4.4
4G.
53G
5GY
5VS
7-5
AACTN
AAEDT
AAFNC
AALRI
AAQFI
AAQXK
AAXUO
ABDBF
ABEFU
ABJNI
ABMAC
ABUBZ
ACGFS
ACPQW
ADBBV
ADMUD
ADZMO
AENEX
AFFNX
AFRHK
AGIAB
AHHHB
AITUG
AKRWK
ALMA_UNASSIGNED_HOLDINGS
CAG
COF
CS3
DM4
DU5
EAD
EAP
EBD
EBS
EJD
EMK
EMOBN
EPL
EST
ESX
F5P
FDB
FEDTE
FGOYB
FIRID
G-2
HEI
HMK
HMO
HVGLF
HZ~
I-F
IHE
IL9
IOS
LG5
M29
MET
MIO
MV1
NGNOM
O-L
O9-
P2P
R2-
RIG
ROL
RPZ
SAE
SSZ
SV3
TUS
UHS
WUQ
XPP
ID FETCH-sage_journals_10_3233_XST_2011_0306003063
ISSN 0895-3996
IngestDate Sun Sep 29 05:52:22 EDT 2024
IsPeerReviewed true
IsScholarly true
Issue 4
Keywords neuromodulation
Medtronic
neurostimulator
radiation
oncology
Accelerator
Language English
LinkModel OpenURL
MergedId FETCHMERGED-sage_journals_10_3233_XST_2011_0306003063
ParticipantIDs sage_journals_10_3233_XST_2011_030600306
PublicationCentury 2000
PublicationDate 20110000
PublicationDateYYYYMMDD 2011-01-01
PublicationDate_xml – year: 2011
  text: 20110000
PublicationDecade 2010
PublicationPlace London, England
PublicationPlace_xml – name: London, England
PublicationTitle Journal of X-ray science and technology
PublicationYear 2011
Publisher SAGE Publications
Publisher_xml – name: SAGE Publications
SSID ssj0011588
Score 3.7293253
Snippet Object: Where no society-based or manufacturer guidance on radiation limits to neuromodulation devices is available, this research provides the groundwork for...
SourceID sage
SourceType Publisher
StartPage 443
Title Effects on radiation oncology treatments involving various neuromodulation devices
URI https://journals.sagepub.com/doi/full/10.3233/XST-2011-030600306
Volume 19
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3dS8MwEA9zQ9EH0an4TR58EEo326aueRy6ORwqbBP7VtKuexnrYB-C_jv-o14-mtZtwhRGGWkITe_Xy-XudxeErqxBzBiFL83l8X_iRZbJiNM3XRJR1rcdWBMF2-L5tvVKHn3XLxS-cqyl-SysRJ8r80r-I1VoA7nyLNk_SFYPCg3wH-QLV5AwXNeScSMlYyTGhNcYkNZfEsm6SppDzilXoIWE6-Ad9saiKCsvyjEa99XpXTx5iquMX2xV35ywDyNNARKkyyWX_AMst8qdqrj4RreSRai4Q2Uo2dztufA-6pstFRWq99nIeNHNoLCGOjdNZZwZ95W8n8LKPBQyNLbaCelR1wQ4qGLYUgXfiNMjZWqp1tE0h0WSU7hEFnlSazeRRcoXlwXH5m7rpt_tmdIlDPsksVnKFkFNTVTdg-XOG6hk16hLiqhUb3fe2jpYZbnieFM9GZmbxUepLo_ygzEojJjeHtpVEsV1CaV9VIiTMtrJ1aQso60nxbYoo01BD46mB6ijkIbHCdZIwynScIY0rJGGFdLwAtKwQtohqjYbvbuWyZ8zUHCfBrBbTF9MwKcU6Ck5R6iYjJP4GGEaUzf2BqHNbIdQMJV53WMWemDlhzc1yk7Q9bqjnq7f9Qxty5fMf-eoOJvM4wuwFWfhpZLVNw6Laok
link.rule.ids 315,786,790,4043,27954,27955,27956
linkProvider Library Specific Holdings
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Effects+on+radiation+oncology+treatments+involving+various+neuromodulation+devices&rft.jtitle=Journal+of+X-ray+science+and+technology&rft.au=Gossman%2C+Michael+S.&rft.au=Paralikar%2C+Kunal+J.&rft.au=Hebb%2C+Adam+O.&rft.au=Wilkinson%2C+Jeffrey+D.&rft.date=2011&rft.pub=SAGE+Publications&rft.issn=0895-3996&rft.eissn=1095-9114&rft.volume=19&rft.issue=4&rft.spage=443&rft.epage=456&rft_id=info:doi/10.3233%2FXST-2011-030600306&rft.externalDocID=10.3233_XST-2011-030600306
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0895-3996&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0895-3996&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0895-3996&client=summon