Effect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study

PURPOSE. The purpose of this study was to compare the accuracy of the implant master cast according to the type (pick-up, transfer) and the length (long, short) of the impression copings. MATERIALS AND METHODS. The metal master cast was fabricated with three internal connection type implant analogs...

Full description

Saved in:
Bibliographic Details
Published inThe journal of advanced prosthodontics Vol. 2; no. 4; pp. 128 - 133
Main Authors Jo, Si-Hoon, Kim, Kyoung-Il, Seo, Jae-Min, Song, Kwang-Yeob, Park, Ju-Mi, Ahn, Seung-Geun
Format Journal Article
LanguageKorean
Published 2010
Subjects
Online AccessGet full text

Cover

Loading…
Abstract PURPOSE. The purpose of this study was to compare the accuracy of the implant master cast according to the type (pick-up, transfer) and the length (long, short) of the impression copings. MATERIALS AND METHODS. The metal master cast was fabricated with three internal connection type implant analogs (Osstem GS III analog), embedded parallel and with $10^{\circ}$ of mesial angulation to the center analog. Four types of impression coping were prepared with different combinations of types (transfer, pick-up) and lengths (long, short) of the coping. The impressions were made using vinyl polysiloxane (one step, heavy + light body) with an individual tray, and 10 impressions were made for each group. Eventually, 40 experimental casts were produced. Then, the difference in the distance between the master cast and the experimental cast were measured, and the error rate was determined. The analysis of variance was performed using the SPSS (v 12.0) program (${\alpha}$= .05), and the statistical significance was set at P < .05. RESULTS. The ANOVA showed that the pick-up type impression coping exhibited a significantly lower error rate than the transfer type. However, no significant difference was observed with respect to the length of the impression coping. Additionally, no significant difference was observed between the parallel and mesial angulated groups. CONCLUSION. Within the limitations of this study, the pick-up type impression coping exhibited a more accurate implant master cast than the transfer type in parallel group. The accuracy of the implant master cast did not differ for different lengths of impression coping of at least 11 mm. Additionally, the accuracy of the implant cast was not different for the parallel and $10^{\circ}$ mesial angulated groups.
AbstractList PURPOSE. The purpose of this study was to compare the accuracy of the implant master cast according to the type (pick-up, transfer) and the length (long, short) of the impression copings. MATERIALS AND METHODS. The metal master cast was fabricated with three internal connection type implant analogs (Osstem GS III analog), embedded parallel and with $10^{\circ}$ of mesial angulation to the center analog. Four types of impression coping were prepared with different combinations of types (transfer, pick-up) and lengths (long, short) of the coping. The impressions were made using vinyl polysiloxane (one step, heavy + light body) with an individual tray, and 10 impressions were made for each group. Eventually, 40 experimental casts were produced. Then, the difference in the distance between the master cast and the experimental cast were measured, and the error rate was determined. The analysis of variance was performed using the SPSS (v 12.0) program (${\alpha}$= .05), and the statistical significance was set at P < .05. RESULTS. The ANOVA showed that the pick-up type impression coping exhibited a significantly lower error rate than the transfer type. However, no significant difference was observed with respect to the length of the impression coping. Additionally, no significant difference was observed between the parallel and mesial angulated groups. CONCLUSION. Within the limitations of this study, the pick-up type impression coping exhibited a more accurate implant master cast than the transfer type in parallel group. The accuracy of the implant master cast did not differ for different lengths of impression coping of at least 11 mm. Additionally, the accuracy of the implant cast was not different for the parallel and $10^{\circ}$ mesial angulated groups.
Author Ahn, Seung-Geun
Park, Ju-Mi
Seo, Jae-Min
Kim, Kyoung-Il
Song, Kwang-Yeob
Jo, Si-Hoon
Author_xml – sequence: 1
  fullname: Jo, Si-Hoon
– sequence: 2
  fullname: Kim, Kyoung-Il
– sequence: 3
  fullname: Seo, Jae-Min
– sequence: 4
  fullname: Song, Kwang-Yeob
– sequence: 5
  fullname: Park, Ju-Mi
– sequence: 6
  fullname: Ahn, Seung-Geun
BookMark eNqNTLsKwjAUDaJg1f5DFsdC2vSlm0hFdHBxLzFNajAmpUmF_r2pFF2Fy7mH81qAqdKKTYAXIZQEWR7G0y9H6Rz4xogbSjDO0iyKPSALzhm1UHMonk3LnK0VpLoRqoZEVYMqibKO150kdnDd2TuDhNKuJbQfu5_Ub8NsXQUKBV_Cthoa21X9Csw4kYb541-C9aG47o_BQxgrSlUZWZ5250uEQhRinDjc4DzF_-beISxLjA
ContentType Journal Article
DBID JDI
DEWEY 617.69
DatabaseName KoreaScience
DatabaseTitleList
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Dentistry
DocumentTitleAlternate Effect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study
EISSN 2005-7814
EndPage 133
ExternalDocumentID JAKO201013351019386
GroupedDBID .UV
34H
5-W
8JR
8XY
9ZL
AAKDD
ADBBV
ADRAZ
AENEX
ALMA_UNASSIGNED_HOLDINGS
AOIJS
BAWUL
DIK
E3Z
EF.
F5P
GX1
HYE
JDI
KQ8
KVFHK
M48
M~E
O5R
O5S
OK1
RPM
ID FETCH-kisti_ndsl_JAKO2010133510193863
IEDL.DBID M48
ISSN 2005-7806
IngestDate Fri Dec 22 11:59:48 EST 2023
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 4
Keywords Impression coping
Master cast accuracy
Implant angulation
Language Korean
LinkModel DirectLink
MergedId FETCHMERGED-kisti_ndsl_JAKO2010133510193863
Notes KISTI1.1003/JNL.JAKO201013351019386
OpenAccessLink http://journals.scholarsportal.info/openUrl.xqy?issn=2005-7806&volume=2&issue=4&spage=128
ParticipantIDs kisti_ndsl_JAKO201013351019386
PublicationCentury 2000
PublicationDate 2010
PublicationDateYYYYMMDD 2010-01-01
PublicationDate_xml – year: 2010
  text: 2010
PublicationDecade 2010
PublicationTitle The journal of advanced prosthodontics
PublicationTitleAlternate The journal of advanced prosthodontics
PublicationYear 2010
SSID ssib053376724
ssib030186702
ssib042018902
ssj0000447056
Score 3.56214
Snippet PURPOSE. The purpose of this study was to compare the accuracy of the implant master cast according to the type (pick-up, transfer) and the length (long,...
SourceID kisti
SourceType Open Access Repository
StartPage 128
Title Effect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study
URI http://click.ndsl.kr/servlet/LinkingDetailView?cn=JAKO201013351019386&dbt=JAKO&org_code=O481&site_code=SS1481&service_code=01
Volume 2
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwhV1LT8MwDLamHYALggHiOeUAuxV1bfrigiZgmoYKFybtNmVpK1WUVrQdYv8eO80oBwRSL6nyUhLbn2PHBrhEncKXKHYMhOKxwT07MUTMEyOITDvxI2ElghTF8MmdzPh07sw7sHFi1wtY_araUT6pWZldf76vbxWFqDCanm-6A03K1iDVKmpDjsh8iTtbNp39sAX-eKp912sDX3EUg2Rw25QRA3mup99NKJ7OuWeqVLDfYyKaJYiX_hBH4z3Y1TiSjZqN34fOa9GD7Xvy_aH0bT3YCrXV_ACyJkQxKxKWvmnH15xJ9VSKiTyivxmuMKO7y8Y3juGH0JAJKVelkGvdVtVq-6husAlLc_aR1mXBVKjaQ7gaP7zcTQw17UUeVdliOnp8Jns4aqpEnIHtu_YRdPMij4-BWcLhXMplhGyBOzJemoE15MIZxhbSbZCcQP_vvk7_q3AGO405nu40zqFbl6v4AqV8veyrDfsCZEWk0Q
link.rule.ids 230,315,786,790,891,2236,4043
linkProvider Scholars Portal
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Effect+of+impression+coping+and+implant+angulation+on+the+accuracy+of+implant+impressions%3A+an+in+vitro+study&rft.jtitle=The+journal+of+advanced+prosthodontics&rft.au=Jo%2C+Si-Hoon&rft.au=Kim%2C+Kyoung-Il&rft.au=Seo%2C+Jae-Min&rft.au=Song%2C+Kwang-Yeob&rft.date=2010&rft.issn=2005-7806&rft.eissn=2005-7814&rft.volume=2&rft.issue=4&rft.spage=128&rft.epage=133&rft.externalDBID=n%2Fa&rft.externalDocID=JAKO201013351019386
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2005-7806&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2005-7806&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2005-7806&client=summon