Case Report : Isoflurane cracks the polycarbonate connector of extracorporeal circuit

Cardiopulmonary bypass (CPB) is widely used for cardiac surgery by virtue of its proven safety over the course of its use during the past half century. Even though perfusion is safer, incidents still occur. During the repair of a ventricular-septal defect in an 11-month-old infant, we experienced a...

Full description

Saved in:
Bibliographic Details
Published inKorean journal of anesthesiology Vol. 58; no. 3; pp. 304 - 306
Main Authors Hyung Sun Lim, Sung Hun Cho, Deok Kyu Kim, Sang Kyi Lee, Seong Hoon Ko
Format Journal Article
LanguageKorean
Published 대한마취통증의학회(구 대한마취과학회) 30.03.2010
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Cardiopulmonary bypass (CPB) is widely used for cardiac surgery by virtue of its proven safety over the course of its use during the past half century. Even though perfusion is safer, incidents still occur. During the repair of a ventricular-septal defect in an 11-month-old infant, we experienced a critical incident related to the potential hazardous effect of volatile anesthetics on the polycarbonate connector of extra-corporeal circuit. The damage to the polycarbonate connector had occurred after spillage of isoflurane during the filling of the vaporizer, causing it to crack and leak. The incident was managed by replacement of the cracked connector during a temporary circulatory arrest. The patient was hypothermic and the time off bypass was less than 1.5 min. There were no neurologic sequelae, the patient made an uneventful recovery. In conclusion, the spillage of volatile anesthetics can cause cracks in the polycarbonate connector of the extra-corporeal circuit, leading to potentially interruption of CPB. (Korean J Anesthesiol 2010; 58: 304-306)
Bibliography:The Korean Society of Anesthesiologists
ISSN:2005-6419