Reduced-dose computed tomography to detect dorsal screw protrusion after distal radius volar plating Reduced-Dose CT to Detect Radius Screw Protrusion [version 1; peer review: 1 approved]

Background: Tenosynovitis and tendon rupture caused by screw penetration of the dorsal cortex are common complications after fixed-angle volar plating of a distal radius fracture. Detecting screw prominence with plain radiography is difficult due to the topography of the distal radius dorsal cortex....

Full description

Saved in:
Bibliographic Details
Published inF1000 research Vol. 7
Main Authors Leffers, Kevin J., Kosty, John W., Garcia, Glenn M., Jupiter, Daniel C, Lindsey, Ronald W., Gugala, Zbigniew
Format Journal Article
LanguageEnglish
Published 2018
Subjects
Online AccessGet full text

Cover

Loading…
Abstract Background: Tenosynovitis and tendon rupture caused by screw penetration of the dorsal cortex are common complications after fixed-angle volar plating of a distal radius fracture. Detecting screw prominence with plain radiography is difficult due to the topography of the distal radius dorsal cortex. Computed tomography (CT) offers more detailed imaging of the bone topography, but is associated with radiation exposure. The present cadaveric study compared reduced-dose and standard-dose CT protocols in the detection of dorsal screw protrusion after fixed-angle volar plating of distal radius fracture. If found equivalent, a reduced-dose protocol could decrease the total radiation exposure to patients. Methods: Standard size distal radius volar locking plates were placed using a standard Henry approach in 3 matched pairs of cadaver wrists. A total of 3 distal locking screws were placed at 3 different lengths for a total of 3 rounds of CT scans per wrist pair. Each wrist pair was imaged by CT using standard-dose and reduced-dose protocols. Dorsal screw penetration was measured in each imaging protocol by 3 radiologists at two time periods to calculate inter- and intra-observer variability. Variability was calculated using the concordance correlation coefficient (CCC), intra-class correlation coefficient (ICC), and Pearson correlation coefficient (PCC). Bland-Altman plots were used and assessed 95% limits of agreement. Results: Intra- and inter-observer variabilities, either with the reduced-dose or standard-dose protocol, were >0.85. Pairwise CCC, ICC, and PCC were >0.91. In the comparison of reduced dose versus standard dose between radiologists, correlations were always >0.95. Conclusions: Comparison of a reduced-dose CT protocol and a standard-dose CT protocol for the detection of dorsal penetrating screws after fixed-angle volar plating showed >0.95 correlation in this cadaveric model. A reduced-dose CT protocol is equivalent to a standard dose CT protocol for orthopedic imaging and should reduce radiation exposure.
AbstractList Background: Tenosynovitis and tendon rupture caused by screw penetration of the dorsal cortex are common complications after fixed-angle volar plating of a distal radius fracture. Detecting screw prominence with plain radiography is difficult due to the topography of the distal radius dorsal cortex. Computed tomography (CT) offers more detailed imaging of the bone topography, but is associated with radiation exposure. The present cadaveric study compared reduced-dose and standard-dose CT protocols in the detection of dorsal screw protrusion after fixed-angle volar plating of distal radius fracture. If found equivalent, a reduced-dose protocol could decrease the total radiation exposure to patients. Methods: Standard size distal radius volar locking plates were placed using a standard Henry approach in 3 matched pairs of cadaver wrists. A total of 3 distal locking screws were placed at 3 different lengths for a total of 3 rounds of CT scans per wrist pair. Each wrist pair was imaged by CT using standard-dose and reduced-dose protocols. Dorsal screw penetration was measured in each imaging protocol by 3 radiologists at two time periods to calculate inter- and intra-observer variability. Variability was calculated using the concordance correlation coefficient (CCC), intra-class correlation coefficient (ICC), and Pearson correlation coefficient (PCC). Bland-Altman plots were used and assessed 95% limits of agreement. Results: Intra- and inter-observer variabilities, either with the reduced-dose or standard-dose protocol, were >0.85. Pairwise CCC, ICC, and PCC were >0.91. In the comparison of reduced dose versus standard dose between radiologists, correlations were always >0.95. Conclusions: Comparison of a reduced-dose CT protocol and a standard-dose CT protocol for the detection of dorsal penetrating screws after fixed-angle volar plating showed >0.95 correlation in this cadaveric model. A reduced-dose CT protocol is equivalent to a standard dose CT protocol for orthopedic imaging and should reduce radiation exposure.
Author Lindsey, Ronald W.
Gugala, Zbigniew
Leffers, Kevin J.
Kosty, John W.
Jupiter, Daniel C
Garcia, Glenn M.
Author_xml – sequence: 1
  givenname: Kevin J.
  orcidid: 0000-0002-6204-4522
  surname: Leffers
  fullname: Leffers, Kevin J.
  organization: Department of Orthopaedic Surgery and Rehabilitation, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-0165, USA
– sequence: 2
  givenname: John W.
  surname: Kosty
  fullname: Kosty, John W.
  organization: Department of Orthopaedic Surgery and Rehabilitation, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-0165, USA
– sequence: 3
  givenname: Glenn M.
  surname: Garcia
  fullname: Garcia, Glenn M.
  organization: Department of Radiology, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-0709, USA
– sequence: 4
  givenname: Daniel C
  surname: Jupiter
  fullname: Jupiter, Daniel C
  organization: Department of Preventive Medicine and Community Health, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-1150, USA
– sequence: 5
  givenname: Ronald W.
  orcidid: 0000-0002-3223-0252
  surname: Lindsey
  fullname: Lindsey, Ronald W.
  organization: Department of Orthopaedic Surgery and Rehabilitation, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-0165, USA
– sequence: 6
  givenname: Zbigniew
  orcidid: 0000-0003-2331-7660
  surname: Gugala
  fullname: Gugala, Zbigniew
  email: zgugala@utmb.edu
  organization: Department of Orthopaedic Surgery and Rehabilitation, University of Texas Medical Branch at Galveston, Galveston, TX, 77555-0165, USA
BookMark eNqlkEFOwzAQRS0EEgV6BTQXSLDTNrRl2YK6RKU7hCzLnrRGaWyNnVQ9G5cjSakQa1Yz0vx5X3o37LJyFTJ2L3gqsnw6fSgE55wwoCK9S8WET_JUXLBBxsd5IsY8u2bDED7bEJ_NRnn2OGBfazS1RpMYFxC02_s6ooHo9m5Lyu-O7QoGI-oIxlFQJQRNeABPLlIdrKtAFREJjA2xvZIytg7QuFIR-FJFW23hXLLsShabjrk8Mden-FvPfP1lvjdI_SKewGOLJ2wsHuYgQPm2u0HzcceuClUGHP7MWzZ_ed4sVkmhdF3GY2dDnnVIwWVvSf6xJHtLUoz-9fwNCSh9uA
ContentType Journal Article
Copyright Copyright: © 2018 Leffers KJ et al.
Copyright_xml – notice: Copyright: © 2018 Leffers KJ et al.
DBID C-E
CH4
DOI 10.12688/f1000research.15056.1
DatabaseName F1000Research
Faculty of 1000
DatabaseTitleList
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Women's Studies
EISSN 2046-1402
ExternalDocumentID 10_12688_f1000research_15056_1
GroupedDBID 3V.
53G
5VS
7X7
88I
8FE
8FH
8FI
8FJ
AAFWJ
ABUWG
ACGOD
ACPRK
ADBBV
ADRAZ
AFKRA
AFPKN
AHMBA
ALMA_UNASSIGNED_HOLDINGS
AOIJS
AZQEC
BAWUL
BBNVY
BCNDV
BENPR
BHPHI
BPHCQ
BVXVI
C-E
CCPQU
CH4
DIK
DWQXO
FRP
FYUFA
GNUQQ
GROUPED_DOAJ
GX1
HCIFZ
HMCUK
HYE
KQ8
LK8
M2P
M48
M7P
M~E
OK1
PIMPY
PQQKQ
PROAC
RPM
UKHRP
W2D
ID FETCH-faculty1000_research_10_12688_f1000research_15056_13
IEDL.DBID M48
IngestDate Thu Feb 08 05:46:26 EST 2024
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Keywords distal radius fracture
reduced-dose computed tomography
volar plating
radiation exposure
Language English
License http://creativecommons.org/licenses/by/4.0/: This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
LinkModel DirectLink
MergedId FETCHMERGED-faculty1000_research_10_12688_f1000research_15056_13
ORCID 0000-0003-2331-7660
0000-0002-6204-4522
0000-0002-3223-0252
OpenAccessLink http://journals.scholarsportal.info/openUrl.xqy?doi=10.12688/f1000research.15056.1
ParticipantIDs faculty1000_research_10_12688_f1000research_15056_1
PublicationCentury 2000
PublicationDate 2018
PublicationDateYYYYMMDD 2018-01-01
PublicationDate_xml – year: 2018
  text: 2018
PublicationDecade 2010
PublicationTitle F1000 research
PublicationYear 2018
SSID ssj0000993627
Score 4.1275897
Snippet Background: Tenosynovitis and tendon rupture caused by screw penetration of the dorsal cortex are common complications after fixed-angle volar plating of a...
SourceID faculty1000
SourceType Publisher
Title Reduced-dose computed tomography to detect dorsal screw protrusion after distal radius volar plating Reduced-Dose CT to Detect Radius Screw Protrusion [version 1; peer review: 1 approved]
URI http://dx.doi.org/10.12688/f1000research.15056.1
Volume 7
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3bSsNAEF1KBbEPxSteyzwIPqWQi2lTEZHWWoSK1BYKIiHJbrxQk9q01X6bP-fM7hYr6EtfQiDZWTib7M6ZnT3D2HGMi1joepbhxRESlDA2jZCbnmGFgWUHbjVyZVZl-9Zt9Zyb_mk_x-blUjWA2Z_UjupJ9UaD8uf77AJ_-HOpjeAig4spSK3FcZ7LJi3qZWREK5aDbJ3S-bTL_6o8IpyzK_qw8P_NC6wQB6R-MaOnC2tOc50VtbMIl2p0N1hOJJtsta23wzdZUdafPMlAZwNusa8OSbEKbvA0ExCpkg0cxumblqbGW-CCdg6Ap6MMreO8IT6A9BpGEwqdgSwbDpwcywGMAv4yyWBKMMGQEueSJ5h30qBO6l2y2VA2O-r1e2nz7sfmw1TF5cA8g6FA8-rITA1MkKrmU8Eft1mtedWtt4wFRPw5YD7RCMLR_4WjL3H0TXuH5ZM0EbsMPJtXRIxOQeyFTlS1AiSXTiUStiN4EEZij9lLdLC_VKsDtoauT1UFUw5ZHuEQR-hejMOSpOV4ve6bJfntfANiY9c0
link.rule.ids 315,783,787,867,4031,24330,27935,27936,27937
linkProvider Scholars Portal
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Reduced-dose+computed+tomography+to+detect+dorsal+screw+protrusion+after+distal+radius+volar+plating+Reduced-Dose+CT+to+Detect+Radius+Screw+Protrusion+%5Bversion+1%3B+peer+review%3A+1+approved%5D&rft.jtitle=F1000+research&rft.au=Leffers%2C+Kevin+J.&rft.au=Kosty%2C+John+W.&rft.au=Garcia%2C+Glenn+M.&rft.au=Jupiter%2C+Daniel+C&rft.date=2018&rft.eissn=2046-1402&rft.volume=7&rft_id=info:doi/10.12688%2Ff1000research.15056.1&rft.externalDBID=C-E&rft.externalDocID=10_12688_f1000research_15056_1