Time calibration for the end cap TOF system of BESⅢtime-of-flight detector, oiYline calibration, time resolution
The time calibration for end cap TOF system of BESIII is studied in this paper. It has achieved about 110 ps time resolution for muons in dimu events. The pulse height correction using electronic scan curve and the predicted time calculated using Kalman filter method are introduced. This paper also...
Saved in:
Published in | Chinese physics C Vol. 35; no. 1; pp. 72 - 78 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
2011
|
Subjects | |
Online Access | Get full text |
ISSN | 1674-1137 0254-3052 |
Cover
Loading…
Summary: | The time calibration for end cap TOF system of BESIII is studied in this paper. It has achieved about 110 ps time resolution for muons in dimu events. The pulse height correction using electronic scan curve and the predicted time calculated using Kalman filter method are introduced. This paper also describes the study of using electrons and muons as calibration samples. |
---|---|
Bibliography: | TP311 TL817.8 time-of-flight detector, oiYline calibration, time resolution 11-5641/O4 |