Characterization, standardization and experiences with KONE ISE for Mg2

The principles of a fully automated potentiometric determination of ionized magnesium in human blood serum are presented. By incorporating a magnesium selective electrode, into a commercially available six channel electrolyte analyzer (Microlyte 6, KONE Instruments, Finland), determination of the bi...

Full description

Saved in:
Bibliographic Details
Published inScandinavian journal of clinical & laboratory investigation. Supplement Vol. 217; p. 37
Main Authors Lewenstam, A, Blomqvist, N, Ost, J
Format Journal Article
LanguageEnglish
Published Norway 1994
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The principles of a fully automated potentiometric determination of ionized magnesium in human blood serum are presented. By incorporating a magnesium selective electrode, into a commercially available six channel electrolyte analyzer (Microlyte 6, KONE Instruments, Finland), determination of the biologically-active, ionized fraction of magnesium has been made possible. In this paper, the analytical performance of the measurement system is presented and determination of ionized magnesium in real patient samples is discussed with special attention to the significance of this new parameter.
ISSN:0085-591X
2166-1030
DOI:10.3109/00365519409095209