Inter-laboratory difference among eleven clinical laboratories in the Okayama City area
The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 l...
Saved in:
Published in | Acta medica Okayama Vol. 52; no. 5; pp. 261 - 270 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Japan
01.10.1998
|
Subjects | |
Online Access | Get full text |
ISSN | 0386-300X |
DOI | 10.18926/amo/31324 |
Cover
Abstract | The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 laboratories in the Okayama City area. No differences were found in the assay units of 26 tests surveyed. However, considerable differences were observed in test data, reference interval, and clinical level (CL), though most laboratories pointed out that the test data for the normal subject was within the reference intervals and those for the patient with liver cirrhosis showed abnormalities in tests for liver function. The difference in reference intervals was serious in the tests of direct bilirubin (D-Bil), thymol turbidity test (TTT), alkaline phosphatase (ALP), gamma-glutamyltranspeptidase (GGTP) and choline sterase. Marked differences in CLs were found in the tests of D-Bil, TTT, ALP, GGTP, creatine phosphokinase, amylase, heavy density lipoprotein cholesterol and white blood cell count. However, three hepatologists independently suggested that such inter-laboratory differences would not seriously affect a clinical decision on the disease status of the cirrhotic patient. Most tests that showed a trend error in a recent quality control survey appeared to have the same trend in the present study. These results indicate that inter-laboratory differences occur at various levels and control assessment are helpful in establishing, and therefore reducing, the level of inter-laboratory differences. |
---|---|
AbstractList | The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 laboratories in the Okayama City area. No differences were found in the assay units of 26 tests surveyed. However, considerable differences were observed in test data, reference interval, and clinical level (CL), though most laboratories pointed out that the test data for the normal subject was within the reference intervals and those for the patient with liver cirrhosis showed abnormalities in tests for liver function. The difference in reference intervals was serious in the tests of direct bilirubin (D-Bil), thymol turbidity test (TTT), alkaline phosphatase (ALP), gamma-glutamyltranspeptidase (GGTP) and choline sterase. Marked differences in CLs were found in the tests of D-Bil, TTT, ALP, GGTP, creatine phosphokinase, amylase, heavy density lipoprotein cholesterol and white blood cell count. However, three hepatologists independently suggested that such inter-laboratory differences would not seriously affect a clinical decision on the disease status of the cirrhotic patient. Most tests that showed a trend error in a recent quality control survey appeared to have the same trend in the present study. These results indicate that inter-laboratory differences occur at various levels and control assessment are helpful in establishing, and therefore reducing, the level of inter-laboratory differences.The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 laboratories in the Okayama City area. No differences were found in the assay units of 26 tests surveyed. However, considerable differences were observed in test data, reference interval, and clinical level (CL), though most laboratories pointed out that the test data for the normal subject was within the reference intervals and those for the patient with liver cirrhosis showed abnormalities in tests for liver function. The difference in reference intervals was serious in the tests of direct bilirubin (D-Bil), thymol turbidity test (TTT), alkaline phosphatase (ALP), gamma-glutamyltranspeptidase (GGTP) and choline sterase. Marked differences in CLs were found in the tests of D-Bil, TTT, ALP, GGTP, creatine phosphokinase, amylase, heavy density lipoprotein cholesterol and white blood cell count. However, three hepatologists independently suggested that such inter-laboratory differences would not seriously affect a clinical decision on the disease status of the cirrhotic patient. Most tests that showed a trend error in a recent quality control survey appeared to have the same trend in the present study. These results indicate that inter-laboratory differences occur at various levels and control assessment are helpful in establishing, and therefore reducing, the level of inter-laboratory differences. The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 laboratories in the Okayama City area. No differences were found in the assay units of 26 tests surveyed. However, considerable differences were observed in test data, reference interval, and clinical level (CL), though most laboratories pointed out that the test data for the normal subject was within the reference intervals and those for the patient with liver cirrhosis showed abnormalities in tests for liver function. Most tests that showed a trend error in a recent quality control survey appeared to have the same trend in the present study. These results indicate that inter-laboratory differences occur at various levels and control assessment are helpful in establishing, and therefore reducing, the level of inter-laboratory differences. The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to reduce inter-laboratory differences. Blood and serum samples of one healthy subject and one subject with liver cirrhosis were analyzed by 11 laboratories in the Okayama City area. No differences were found in the assay units of 26 tests surveyed. However, considerable differences were observed in test data, reference interval, and clinical level (CL), though most laboratories pointed out that the test data for the normal subject was within the reference intervals and those for the patient with liver cirrhosis showed abnormalities in tests for liver function. The difference in reference intervals was serious in the tests of direct bilirubin (D-Bil), thymol turbidity test (TTT), alkaline phosphatase (ALP), gamma-glutamyltranspeptidase (GGTP) and choline sterase. Marked differences in CLs were found in the tests of D-Bil, TTT, ALP, GGTP, creatine phosphokinase, amylase, heavy density lipoprotein cholesterol and white blood cell count. However, three hepatologists independently suggested that such inter-laboratory differences would not seriously affect a clinical decision on the disease status of the cirrhotic patient. Most tests that showed a trend error in a recent quality control survey appeared to have the same trend in the present study. These results indicate that inter-laboratory differences occur at various levels and control assessment are helpful in establishing, and therefore reducing, the level of inter-laboratory differences. |
Author | Shinji, T Koide, N Matsuda, N Okada, K Mizushima, J Sunami, H |
Author_xml | – sequence: 1 givenname: N surname: Koide fullname: Koide, N organization: First Department of Internal Medicine, Okayama University Medical School, Japan – sequence: 2 givenname: T surname: Shinji fullname: Shinji, T – sequence: 3 givenname: K surname: Okada fullname: Okada, K – sequence: 4 givenname: J surname: Mizushima fullname: Mizushima, J – sequence: 5 givenname: N surname: Matsuda fullname: Matsuda, N – sequence: 6 givenname: H surname: Sunami fullname: Sunami, H |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/9810436$$D View this record in MEDLINE/PubMed |
BookMark | eNqF0D1PwzAUhWEPRaUtLOxInthCfeOvZEQVH5UqdQHBFt0k1-CSOMVJkfLvqUTFynSWR-9w5mwSukCMXYG4hSxPzRLbbilBpmrCZkJmJpFCvJ2zed_vhEhVbsSUTfMMhJJmxl7XYaCYNFh2EYcujrz2zlGkUBE_psI7p4a-KfCq8cFX2PA_66nnPvDhg_j2E0dska_8MHKMhBfszGHT0-VpF-zl4f559ZRsto_r1d0m2UmbDQkZi-TyHKy2CLpCWysyFRhy2lkQunRlrXNUWBpNTmYSylIRoLZOGJHKBbv57e5j93Wgfiha31fUNBioO_SFFUKDAvUvTHOldJrZfyFYkGDT7AivT_BQtlQX--hbjGNx-lb-AInWexY |
ContentType | Journal Article |
DBID | CGR CUY CVF ECM EIF NPM 7T2 C1K 7TB 8FD FR3 7X8 |
DOI | 10.18926/amo/31324 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed Health and Safety Science Abstracts (Full archive) Environmental Sciences and Pollution Management Mechanical & Transportation Engineering Abstracts Technology Research Database Engineering Research Database MEDLINE - Academic |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) Health & Safety Science Abstracts Environmental Sciences and Pollution Management Technology Research Database Mechanical & Transportation Engineering Abstracts Engineering Research Database MEDLINE - Academic |
DatabaseTitleList | MEDLINE - Academic Technology Research Database Health & Safety Science Abstracts MEDLINE |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 2 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
EndPage | 270 |
ExternalDocumentID | 9810436 |
Genre | Journal Article |
GroupedDBID | --- .55 .GJ 23M 2WC 3O- 53G 5GY 5VS 6J9 7.U ADBBV AENEX ALMA_UNASSIGNED_HOLDINGS BAWUL C1A CGR CUY CVF DIK E3Z EBS ECM EIF EJD F5P GROUPED_DOAJ GX1 IPNFZ KQ8 NPM OK1 P2P P6G RIG RNS SJN TKC TR2 X7M XSB ZGI 7T2 C1K OVT 7TB 8FD FR3 7X8 |
ID | FETCH-LOGICAL-j378t-e67aef991757a15ca7d4e6c16ef5f7105bfbd59a4ab65ef3831bb4e1a57f06023 |
ISSN | 0386-300X |
IngestDate | Fri Jul 11 00:52:10 EDT 2025 Fri Jul 11 08:27:07 EDT 2025 Fri Jul 11 08:35:49 EDT 2025 Thu May 23 23:01:52 EDT 2024 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 5 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-j378t-e67aef991757a15ca7d4e6c16ef5f7105bfbd59a4ab65ef3831bb4e1a57f06023 |
Notes | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Article-1 ObjectType-Feature-2 |
OpenAccessLink | https://ousar.lib.okayama-u.ac.jp/31324 |
PMID | 9810436 |
PQID | 17131728 |
PQPubID | 23462 |
PageCount | 10 |
ParticipantIDs | proquest_miscellaneous_70051414 proquest_miscellaneous_29445287 proquest_miscellaneous_17131728 pubmed_primary_9810436 |
PublicationCentury | 1900 |
PublicationDate | 1998-Oct 19981001 |
PublicationDateYYYYMMDD | 1998-10-01 |
PublicationDate_xml | – month: 10 year: 1998 text: 1998-Oct |
PublicationDecade | 1990 |
PublicationPlace | Japan |
PublicationPlace_xml | – name: Japan |
PublicationTitle | Acta medica Okayama |
PublicationTitleAlternate | Acta Med Okayama |
PublicationYear | 1998 |
SSID | ssj0024960 |
Score | 1.4623779 |
Snippet | The aim of the present study was to find the cause of inter-laboratory differences in laboratory test data and to examine whether control assessment helps to... |
SourceID | proquest pubmed |
SourceType | Aggregation Database Index Database |
StartPage | 261 |
SubjectTerms | Aged Humans Laboratories - standards Liver Cirrhosis - blood Liver Cirrhosis - physiopathology Liver Function Tests Male Middle Aged Quality Control Reference Values Reproducibility of Results |
Title | Inter-laboratory difference among eleven clinical laboratories in the Okayama City area |
URI | https://www.ncbi.nlm.nih.gov/pubmed/9810436 https://www.proquest.com/docview/17131728 https://www.proquest.com/docview/29445287 https://www.proquest.com/docview/70051414 |
Volume | 52 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Lb9QwELagSIgL4lVRnj5wq0I3Xr9yXC1bFdhtL7tib9E4cdQUmkU0e6C_nrGdV4EVj4sVRc4o8mdNPmfmmyHkDSt4AYnBs4k2IuIAOoJipCMrpMlkDoYrp0ZenMqTFf-wFuu2m32jLqnN2-z6t7qS_0EV7yGuTiX7D8h2RvEGXiO-OCLCOP4Vxv53XjQPOLpg-bum3YlTAfguQrMvrkLT4bTVP3ZzS5-I5Wnn2Wf4DpdwOHWEfIIkckhYJ1kNIQAP7cTOR2_KULO3j-acl9VFeSP1Gp_J4cbv1EV5vb06Ly-hD0rlnQyvzV_r1Fba5cyN1kNPKthgx4ihWwwF15svLAutQn5x3jrx_SZxfXB0JSV5_5FqA_OnZ-nxaj5Pl7P18ja5w5QKwfn3H_sKi4nXhndv2BSlddaP0PaRt7z7QOGJxfIBud-cCOgkwPuQ3LLVI3J30eQ8PCaffkaZ9ihTjzINKNMWZTpEmZYVRZRpAx51KFOH8hOyOp4tpydR0w4juhgrXUdWKrAF8nklFMQiA5VzK7NY2kIUSBSFKUwuEuBgpLDFWI9jY7iNQahiJJGb7ZO9alPZp4TKGBiYRCN7ZNzCyBik3Y7pMhBC5uyAvG5XJ0V342JIUNnN9iqNVTx2Lc12z2AJ5wLP4btnKF90P-YHZD8sfPo1VFZJEx27lgjP_mj8ObnX78oXZK_-trUvkR3W5pXfDD8AOKFn9A |
linkProvider | Flying Publisher |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Inter-Laboratory+Difference+among+Eleven+Clinical+Laboratories+in+the+Okayama+City+Area&rft.jtitle=Acta+medica+Okayama&rft.au=Koide%2C+N&rft.au=Shinji%2C+T&rft.au=Okada%2C+K&rft.au=Mizushima%2C+J&rft.date=1998-10-01&rft.issn=0386-300X&rft.volume=52&rft.issue=5&rft.spage=261&rft.epage=270&rft_id=info:doi/10.18926%2Famo%2F31324&rft.externalDBID=NO_FULL_TEXT |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0386-300X&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0386-300X&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0386-300X&client=summon |