We made an improved system of Centry-2, which enable easily to exchange acetate dialysate for bicarbonate one. The new system is briefly indicated below. Basic dialysate without bicarbonate (Solution A) is diluted by the ordinary system of Centry-2. (Solution A). Solution B (5% HCO-3) is infused int...

Full description

Saved in:
Bibliographic Details
Published inJinko Zoki Vol. 14; no. 1; pp. 104 - 106
Main Authors Ogasawara, M., NAKAMURA, Y., NAOI, T., OHWADA, H., SAKAGAWA, K., TAKAHASHI, M.
Format Journal Article
LanguageJapanese
Published JAPANESE SOCIETY FOR ARTIFICIAL ORGANS 1985
一般社団法人 日本人工臓器学会
Subjects
Online AccessGet full text
ISSN0300-0818
1883-6097
DOI10.11392/jsao1972.14.104

Cover

Loading…
More Information
Summary:We made an improved system of Centry-2, which enable easily to exchange acetate dialysate for bicarbonate one. The new system is briefly indicated below. Basic dialysate without bicarbonate (Solution A) is diluted by the ordinary system of Centry-2. (Solution A). Solution B (5% HCO-3) is infused into Solution A' between the control conductivity cell and the stabilyzer by the additional supply pump. (Solution A'+B). The control converter, which monitors the concentration of both Solution A and Solution A'+B, was newly constructed. Results of running test showed good stability of this system. The nature of Solution A'+B was of pH 7.194±0.009, PCO265.9±1.7mmHg, HCO-3 25.4±0.6mEq/L, Na 139.6±0.7mEq/L and K 2.4±0.0mEq/L. These improvements of Centry-2 are very simple and not so expensive as other bicarbonate infusers. Therefore this system is very useful for the dialysis with bicarbonate.
ISSN:0300-0818
1883-6097
DOI:10.11392/jsao1972.14.104