Liquid chromatography–positive ion electrospray mass spectrometry method for the quantification of citalopram in human plasma
A rapid, sensitive and novel narrow-bore liquid chromatography–mass spectrometric method was developed and fully validated for the quantification of citalopram in human plasma. The analyte and internal standard (imipramine) were extracted by liquid–liquid extraction with a mixture of hexane–heptane–...
Saved in:
Published in | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Vol. 810; no. 2; pp. 235 - 244 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English Japanese |
Published |
Elsevier B.V
25.10.2004
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A rapid, sensitive and novel narrow-bore liquid chromatography–mass spectrometric method was developed and fully validated for the quantification of citalopram in human plasma. The analyte and internal standard (imipramine) were extracted by liquid–liquid extraction with a mixture of hexane–heptane–isopropanol (88:10:2, v/v/v). The use of a Hypersil BDS C
8 micro-bore column (250
mm × 2.1
mm i.d.; 3.5
μm particle size), results in substantial reduction in solvent consumption. The mobile phase consisted of 10
mM ammonium formate–formic acid (pH 4.5) and acetonitrile (30:70, v/v), pumped at a flow rate of 0.15
ml
min
−1. The analytes were detected after positive electrospray ionization using the selected ion-monitoring mode of the species at
m/
z 325 for citalopram and
m/
z 281 for imipramine. The method had a chromatographic run time of 10.0
min and a linear calibration curve over the range 0.50–250
ng
ml
−1 (
r
2 > 0.996). The limit of quantitation was 0.50
ng
ml
−1. Accuracy and precision were below the acceptance limits of 15%. |
---|---|