A practical method for assessing quantitative scanner accuracy with long-lived radionuclides: The ARTnet insert
To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms...
Saved in:
Published in | Asia Oceania journal of nuclear medicine & biology Vol. 12; no. 1; pp. 27 - 34 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Iran
Mashhad University of Medical Sciences
2024
|
Subjects | |
Online Access | Get full text |
ISSN | 2322-5718 2322-5726 |
DOI | 10.22038/AOJNMB.2023.71860.1503 |
Cover
Loading…
Abstract | To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available "standardised" test objects.
We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical "lung insert" in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with
F,
Ga and
I PET/CT and
Tc,
I and
Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides.
The scanners were found to produce accurate image reconstructions in the insert with 5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement.
The "ARTnet Insert" is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with
I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents. |
---|---|
AbstractList | To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available "standardised" test objects.ObjectivesTo address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available "standardised" test objects.We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical "lung insert" in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with 18F, 68Ga and 124I PET/CT and 99mTc, 131I and 177Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides.MethodsWe have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical "lung insert" in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with 18F, 68Ga and 124I PET/CT and 99mTc, 131I and 177Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides.The scanners were found to produce accurate image reconstructions in the insert with 5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement.ResultsThe scanners were found to produce accurate image reconstructions in the insert with 5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement.The "ARTnet Insert" is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with 131I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents.ConclusionsThe "ARTnet Insert" is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with 131I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents. Objective(s): To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available “standardised” test objects.Methods: We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical “lung insert” in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with 18F, 68Ga and 124I PET/CT and 99mTc, 131I and 177Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides.Results: The scanners were found to produce accurate image reconstructions in the insert with ±5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement.Conclusions: The “ARTnet Insert” is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with 131I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents. To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available "standardised" test objects. We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical "lung insert" in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with F, Ga and I PET/CT and Tc, I and Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides. The scanners were found to produce accurate image reconstructions in the insert with 5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement. The "ARTnet Insert" is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents. Objective(s):To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available “standardised” test objects.Methods:We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical “lung insert” in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with 18F, 68Ga and 124I PET/CT and 99mTc, 131I and 177Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides. Results:The scanners were found to produce accurate image reconstructions in the insert with 5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement. Conclusions: The “ARTnet Insert” is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with 131I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents. |
Author | Muñoz-Ferrada, Carl Bailey, Dale L Willowson, Kathy P |
AuthorAffiliation | 3 Institute of Medical Physics, University of Sydney, Sydney, Australia 4 Globalsonics Institute for Medical Research, Sydney, Australia 1 Department of Nuclear Medicine, Royal North Shore Hospital, Sydney, Australia 2 Faculty of Medicine & Health, University of Sydney, Sydney, Australia |
AuthorAffiliation_xml | – name: 1 Department of Nuclear Medicine, Royal North Shore Hospital, Sydney, Australia – name: 3 Institute of Medical Physics, University of Sydney, Sydney, Australia – name: 2 Faculty of Medicine & Health, University of Sydney, Sydney, Australia – name: 4 Globalsonics Institute for Medical Research, Sydney, Australia |
Author_xml | – sequence: 1 givenname: Dale L surname: Bailey fullname: Bailey, Dale L organization: Faculty of Medicine & Health, University of Sydney, Sydney, Australia – sequence: 2 givenname: Kathy P surname: Willowson fullname: Willowson, Kathy P organization: Institute of Medical Physics, University of Sydney, Sydney, Australia – sequence: 3 givenname: Carl surname: Muñoz-Ferrada fullname: Muñoz-Ferrada, Carl organization: Globalsonics Institute for Medical Research, Sydney, Australia |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/38164228$$D View this record in MEDLINE/PubMed |
BookMark | eNpdkV1vFCEUhompsR_2LyiJN97MyscwMN6YtfGjptrErNeEgTO7NLOwBaam_76srY01XBzCe85z3sM5RgchBkDoNSULxghX75aX3358_7hghPGFpKojCyoIf4aOGGesEZJ1B493qg7Rac5-IIJI1lNBX6BDrmjXMqaOUFziXTK2eGsmvIWyiQ6PMWGTM9SqsMbXswnFF1P8DeBsTQhQZWvnWnaLf_uywVMM62aqusPJOB_DbCfvIL_Hqw3g5c9VgIJ9yJDKS_R8NFOG04d4gn59_rQ6-9pcXH45P1teNI7JrjQGZG8NdApkR1wLLWlNb_vRdWA5FVb0FGwvJKGOyJ6PZBS8VWxQg5F1UMtP0Pk910VzpXfJb0261dF4_echprU2qU49gWaCtYOilHIwLRAyUN7VCIoKboeRVtaHe9ZuHrbgLISSzPQE-lQJfqPX8UZTIqtFwSvh7QMhxesZctFbny1MkwkQ56xZT_ZH8X2zN_-lXsU5hfpXNatuX8hW7YGv_rX06OXvYvkd-KmqSw |
ContentType | Journal Article |
Copyright | 2024 mums.ac.ir All rights reserved. 2024 mums.ac.ir All rights reserved. This work is published under https://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. 2024 mums.ac.ir All rights reserved |
Copyright_xml | – notice: 2024 mums.ac.ir All rights reserved. – notice: 2024 mums.ac.ir All rights reserved. This work is published under https://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. – notice: 2024 mums.ac.ir All rights reserved |
DBID | NPM 3V. 7X7 7XB 8FE 8FH 8FI 8FJ 8FK ABUWG AFKRA AZQEC BBNVY BENPR BHPHI CCPQU DWQXO FYUFA GHDGH GNUQQ HCIFZ K9. LK8 M0S M7P PHGZM PHGZT PIMPY PKEHL PQEST PQGLB PQQKQ PQUKI PRINS 7X8 5PM DOA |
DOI | 10.22038/AOJNMB.2023.71860.1503 |
DatabaseName | PubMed ProQuest Central (Corporate) ProQuest Health & Medical Collection ProQuest Central (purchase pre-March 2016) ProQuest SciTech Collection ProQuest Natural Science Collection ProQuest Hospital Collection Hospital Premium Collection (Alumni Edition) ProQuest Central (Alumni) (purchase pre-March 2016) ProQuest Central (Alumni) ProQuest Central UK/Ireland ProQuest Central Essentials Biological Science Database ProQuest Central Natural Science Collection ProQuest One Community College ProQuest Central Health Research Premium Collection Health Research Premium Collection (Alumni) ProQuest Central Student SciTech Premium Collection ProQuest Health & Medical Complete (Alumni) Biological Sciences ProQuest Health & Medical Collection Biological Science Database ProQuest Central Premium ProQuest One Academic (New) Publicly Available Content Database ProQuest One Academic Middle East (New) ProQuest One Academic Eastern Edition (DO NOT USE) ProQuest One Applied & Life Sciences ProQuest One Academic ProQuest One Academic UKI Edition ProQuest Central China MEDLINE - Academic PubMed Central (Full Participant titles) Directory of Open Access Journals (DOAJ) |
DatabaseTitle | PubMed Publicly Available Content Database ProQuest Central Student ProQuest One Academic Middle East (New) ProQuest Central Essentials ProQuest Health & Medical Complete (Alumni) ProQuest Central (Alumni Edition) SciTech Premium Collection ProQuest One Community College ProQuest Natural Science Collection ProQuest Central China ProQuest Central ProQuest One Applied & Life Sciences Health Research Premium Collection Health and Medicine Complete (Alumni Edition) Natural Science Collection ProQuest Central Korea Biological Science Collection ProQuest Central (New) ProQuest Biological Science Collection ProQuest One Academic Eastern Edition ProQuest Hospital Collection Health Research Premium Collection (Alumni) Biological Science Database ProQuest SciTech Collection ProQuest Hospital Collection (Alumni) ProQuest Health & Medical Complete ProQuest One Academic UKI Edition ProQuest One Academic ProQuest One Academic (New) ProQuest Central (Alumni) MEDLINE - Academic |
DatabaseTitleList | MEDLINE - Academic PubMed Publicly Available Content Database |
Database_xml | – sequence: 1 dbid: DOA name: DOAJ Directory of Open Access Journals url: https://www.doaj.org/ sourceTypes: Open Website – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: BENPR name: ProQuest Central url: https://www.proquest.com/central sourceTypes: Aggregation Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
EISSN | 2322-5726 |
EndPage | 34 |
ExternalDocumentID | oai_doaj_org_article_2524b81113ea4e00b1364e0e8153cbf1 PMC10757053 38164228 |
Genre | Journal Article |
GeographicLocations | New Zealand Australia United States--US |
GeographicLocations_xml | – name: United States--US – name: Australia – name: New Zealand |
GroupedDBID | ALMA_UNASSIGNED_HOLDINGS M~E NPM 3V. 5VS 7X7 7XB 8FE 8FH 8FI 8FJ 8FK ABDBF ABUWG ACPRK ACUHS ADBBV AFKRA AHMBA ALIPV AOIJS AZQEC BAWUL BBNVY BCNDV BENPR BHPHI BPHCQ BVXVI CCPQU DIK DWQXO EBD EOJEC ESX FYUFA GNUQQ GROUPED_DOAJ HCIFZ HMCUK HYE K9. KQ8 LK8 M48 M7P OBODZ OK1 PHGZM PHGZT PIMPY PKEHL PQEST PQGLB PQQKQ PQUKI PRINS PROAC RPM UKHRP 7X8 5PM PUEGO |
ID | FETCH-LOGICAL-d276t-ae79cae68e760d4e404a9c9fd6ec315c591ec95701d0793f0f53482b8ba7050c3 |
IEDL.DBID | M48 |
ISSN | 2322-5718 |
IngestDate | Wed Aug 27 01:29:39 EDT 2025 Thu Aug 21 18:36:48 EDT 2025 Fri Jul 11 09:33:04 EDT 2025 Fri Jul 25 10:20:58 EDT 2025 Thu Jan 02 22:30:10 EST 2025 |
IsDoiOpenAccess | true |
IsOpenAccess | true |
IsPeerReviewed | false |
IsScholarly | true |
Issue | 1 |
Keywords | Clinical Trial PET SPECT Site validation |
Language | English |
License | 2024 mums.ac.ir All rights reserved. This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-d276t-ae79cae68e760d4e404a9c9fd6ec315c591ec95701d0793f0f53482b8ba7050c3 |
Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
OpenAccessLink | http://journals.scholarsportal.info/openUrl.xqy?doi=10.22038/AOJNMB.2023.71860.1503 |
PMID | 38164228 |
PQID | 2920357483 |
PQPubID | 2045590 |
PageCount | 8 |
ParticipantIDs | doaj_primary_oai_doaj_org_article_2524b81113ea4e00b1364e0e8153cbf1 pubmedcentral_primary_oai_pubmedcentral_nih_gov_10757053 proquest_miscellaneous_2909090831 proquest_journals_2920357483 pubmed_primary_38164228 |
PublicationCentury | 2000 |
PublicationDate | 2024-00-00 20240101 2024-01-01 |
PublicationDateYYYYMMDD | 2024-01-01 |
PublicationDate_xml | – year: 2024 text: 2024-00-00 |
PublicationDecade | 2020 |
PublicationPlace | Iran |
PublicationPlace_xml | – name: Iran – name: Mashhad – name: Mashhad, Iran |
PublicationTitle | Asia Oceania journal of nuclear medicine & biology |
PublicationTitleAlternate | Asia Ocean J Nucl Med Biol |
PublicationYear | 2024 |
Publisher | Mashhad University of Medical Sciences |
Publisher_xml | – name: Mashhad University of Medical Sciences |
SSID | ssib050729151 ssj0000942046 |
Score | 2.2431061 |
Snippet | To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have... Objective(s):To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems... Objective(s): To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems... |
SourceID | doaj pubmedcentral proquest pubmed |
SourceType | Open Website Open Access Repository Aggregation Database Index Database |
StartPage | 27 |
SubjectTerms | Accuracy Body image clinical trial Computed tomography Design Drug dosages Nuclear medicine pet Positron emission tomography Radiation Radioactivity Radioisotopes Scanners Single photon emission computed tomography site validation Spatial discrimination spect Technical Note |
SummonAdditionalLinks | – databaseName: Directory of Open Access Journals (DOAJ) dbid: DOA link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV1Lb9QwELZQD4gLgvJaaCsjcU3rJH6FW0GtKqRyolJvkR8T2GpxSndz4N8zY4dqFyFxQTlEyjiRPTPOzHjsbxh7N4BARYmm8o0KlYwGKrTrvqqlcWjQvfEZxPXys764kp-u1fVWqS_aE1bggQvjThrVSG-pIDo4CUL4utV4B4tTNfghBz5o87aCKdQkRXjY9Zzvuin75xohS6U5ir7wj1w2ezWNaO2JG2_Sd39MpcSPkZQXXKiMVobx_5vv-ecWyi2bdP6EPZ6dSX5aBvGUPYC0zx5ezunyZ2w85fMpKGxVakVzdFK5y5leNFr8x-RSPmaGPz2-RjYnQHIIE772k9MiLV-N6Wu1Qnrkdy6iGKewWkZYv-eoYhz94QQbvkyU1X_Ors7Pvny8qOYKC1VsjN5UDkwXHGgLRosoQQrputANUUNoaxVUV0PolBF1JCC9QQyKwHC89c4gk0P7gu2lMcErxrXUsg5KDZ20EnTXmTDoBh0MIweKCRfsAzGzvy0gGj3BWucHKOx-Fnb_L2Ev2MFvUfTzXFv3VG-rVUbadsHe3pNxllDqwyUYJ2oj6LItfuJlkdx9Tyh1SkBoC2Z3ZLrT1V1KWn7LSNwYOyNzVPv6fwzuDXuECijL-s4B29vcTXCIHs_GH2Xl_gUNSPp1 priority: 102 providerName: Directory of Open Access Journals – databaseName: ProQuest Central dbid: BENPR link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3daxQxEA96BfFF_Pa0SgRft92PfK0vcictpXCniIW-LfmYbU-ObHsfD_73zuzmqici-7Cwkw0hmcxMZia_YexDCzkyStCZK6XPRNCQoV53WSG0RYXutOtBXGdzdXYhzi_lZXK4rVNa5U4m9oI6dJ585MdUVamSWpjq081tRlWjKLqaSmjcZwcogo0csYPpyfzrtx1HScLFLlLc68eQR1fmYqg4R6cwlMxD0leJvZvjyZfz-Wx6RCXFj5DUO16onFYP5_8vG_TvVMo_dNPpY_YoGZV8MnDBE3YP4lP2YJbC5s9YN-HpNhS2GmpGczRWue0jvqi8-O3Wxv66GQo_vsbpjoBk77f4209Ozlq-7OJVtkR64CsbcDm3frkIsP7IkdU42sURNnwRKbr_nF2cnnz_fJalSgtZKLXaZBZ07S0oA1rlQYDIha193QYFviqkl3UBvpY6LwIB6rV5KwkUxxlnNU6yr16wUewivGJcCSUKL2VbCyNA1bX2rSrR0NCipbPhmE1pMpubAUyjIXjr_kO3umrSbmlKWQpnUAxXYAXkuSsqhW8wKJ-9a4sxO9wtRZP23Lr5zSFj9v6OjLuFQiA2QrelNjk9psIuXg4rdzcSCqESINqYmb013RvqPiUurntEbjxD4-TI6vX_x_WGPUTWEoMH55CNNqstvEWbZuPeJcb9BQVV84A priority: 102 providerName: ProQuest |
Title | A practical method for assessing quantitative scanner accuracy with long-lived radionuclides: The ARTnet insert |
URI | https://www.ncbi.nlm.nih.gov/pubmed/38164228 https://www.proquest.com/docview/2920357483 https://www.proquest.com/docview/2909090831 https://pubmed.ncbi.nlm.nih.gov/PMC10757053 https://doaj.org/article/2524b81113ea4e00b1364e0e8153cbf1 |
Volume | 12 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3da9swEBdbC2MvY9_L1gUN9upMtvVhD8ZIRkopJBtjgbwZST63GUFekxjWl_3tu7PddhmlGGywbCHuTro73el3jL2vQKCglCZyifKRLA1EqNddFEtjUaE741oQ19lcnyzk6VItb07W9QTc3uraUT2pxWY9-n1x-Rkn_CdKY05Emn0Yfz2dzyYjqgQ-wrW23S8hBNBDVE-aRH12Y_MrgsqO-1DYzy61LhGyK0JHjhl20OWB3dV3j_B_m1n6f3blP-rq-DF71NuZfNwJxhN2D8JT9mDWR9KfsXrM-wNS-FVXRpqj_cptGwRGfcYvGhvaE2i4HvItciAANnvf4G-XnPZv-boOZ9Ea20u-sSVyuPHrVQnbjxylj6OpHGDHV4EC_s_Z4nj648tJ1BdfiMrE6F1kweTegs7AaFFKkELa3OdVqcGnsfIqj8Hnyoi4JIy9SlSKcHJc5qxBIvv0BTsIdYBXjGupZeyVqnKZSdB5bnylE7Q9jKzIXRywCRGz-NXhaxSEeN2-qDdnRT-BikQl0mW4MqdgJQjh4lTjEzJcsr2r4gE7umJFcSVFBZXiSpWRWTpg766bcQJRVMQGqBv6RtCVpdjFy45z1yOhqCphpA1YtsfTvaHut4TVeQvSjW41Ekelr-8e1xv2EEVLdps6R-xgt2ngLZo5Ozdk983SDNnhZDr_9n3YbhYMW1Gm-5_pX_bO-zg |
linkProvider | Scholars Portal |
linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9NAEF6VVgIuiDeBAosER7drex82EkIJtEofCQi1Um9mX25TRes2iYX6p_iNzNhOIQhxq3KIlHGsteebndmd2fkIeVt6BkBxKjKJsBF3ykfg100Uc6XBoRtlmiauo7EcHvP9E3GyRn4uz8JgWeVyTmwmaldZ3CPfRlalVCiepR8vLiNkjcLs6pJCo4XFgb_6AUu2-Ye9z6Dfd0myu3P0aRh1rAKRS5RcRNqr3GovM68kc9xzxnVu89JJb9NYWJHH3uZCsdhh87iSlQIbwJjMaMUEsync9xbZgDAjByvaGOyMv35bIlhgH-64y7Odt3V7CeMtwx2u-sATtEVmCTxNtt3_sj8eDbaQwnwLRM1GD9J3NfQB_4p5_y7d_MMX7t4n97oglvZb1D0gaz48JLdHXZr-Ean6tDt9BVe1HNUUgmOqmwwzOEt6WevQHG-DyZbOQb3Bg9jaGv52RXFzmE6rcBpNQe7oTDuAT22nE-fn7ylAm0IcHvyCTgJWEzwmxzeigydkPVTBPyNUcsljK0SZ84x7mefKljKBwEbxEteiPTLAl1lctM07Cmyn3fxQzU6LzjqLRCTcZDDtp15zz5iJUwnfPgN_YE0Z98jmUhVFZ-Pz4jcie-TNtRisE1MuOviqxmsYfrIUbvG01dz1SDBliw3YeiRb0enKUFclYXLWdACHNTu8HJE-__-4XpM7w6PRYXG4Nz54Qe4CzHi7e7RJ1hez2r-EeGphXnUgpuT7TdvNL_MKMGI |
linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9NAEB6VVKq4IN6kFFgkOLpZ2_uwkRBKaKM-SKgQlXoz9npdUkXrNomF-tf4dcz4UQhC3KocImUca-35Zmd2Z3Y-gDeF5QiUXHtZII0ncm099OuZ5wudokPPdFY3cZ1M1cGpODqTZxvwszsLQ2WV3ZxYT9R5aWiPfECsSqHUIgoHRVsWcbI3_nB55RGDFGVaOzqNBiLH9voHLt-W7w_3UNdvg2C8__XjgdcyDHh5oNXKS62OTWpVZLXiubCCizQ2cZEra0JfGhn71sRScz-nRnIFLyQ1g8miLNVcchPife_ApkavGPVgc7Q_PfnSoVlST26_zbldNDV8ARcN2x2tANErNAVnAT5ZNBh-PppORrtEZ76LonrTh6i8aiqBf8W_f5dx_uEXx_fhXhvQsmGDwAewYd1D2Jq0KftHUA5ZexILr2r4qhkGyiyts83oONlVlbr6qBtOvGyJqnYWxcZU-LdrRhvFbF66c2-O8pwt0hyhVJn5LLfLdwxhzjAmd3bFZo4qCx7D6a3o4An0XOnsM2BKKOEbKYtYRMKqONamUAEGOVoUtC7tw4heZnLZNPJIqLV2_UO5OE9aS00CGYgsQhcQ2lRYzjM_VPhtI_QNJiv8Pux0qkhae18mv9HZh9c3YrRUSr-kzpYVXcPpE4V4i6eN5m5GQulbasbWh2hNp2tDXZe42fe6Gziu3_HlyHD7_-N6BVtoL8mnw-nxc7iLKBPNRtIO9FaLyr7A0GqVvWwxzODbbZvNLwXDNI4 |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+practical+method+for+assessing+quantitative+scanner+accuracy+with+long-lived+radionuclides%3A+The+ARTnet+insert&rft.jtitle=Asia+Oceania+journal+of+nuclear+medicine+%26+biology&rft.au=Bailey%2C+Dale+L&rft.au=Willowson%2C+Kathy+P&rft.au=Mu%C3%B1oz-Ferrada%2C+Carl&rft.date=2024&rft.issn=2322-5718&rft.volume=12&rft.issue=1&rft.spage=27&rft_id=info:doi/10.22038%2FAOJNMB.2023.71860.1503&rft_id=info%3Apmid%2F38164228&rft.externalDocID=38164228 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2322-5718&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2322-5718&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2322-5718&client=summon |