Capabilities of multi-pinhole SPECT with two stationary detectors for in vivo rat imaging

We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99m Tc-labelled tracers. Sensitivity, spatial resolution, un...

Full description

Saved in:
Bibliographic Details
Published inScientific reports Vol. 10; no. 1; p. 18616
Main Authors Janssen, Jan P., Hoffmann, Jan V., Kanno, Takayuki, Nose, Naoko, Grunz, Jan-Peter, Onoguchi, Masahisa, Chen, Xinyu, Lapa, Constantin, Buck, Andreas K., Higuchi, Takahiro
Format Journal Article
LanguageEnglish
Published London Nature Publishing Group UK 29.10.2020
Nature Publishing Group
Subjects
Online AccessGet full text

Cover

Loading…
Abstract We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99m Tc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with 99m Tc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [ 99m Tc]Tc-MIBI or ~ 150 MBq [ 99m Tc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = “very poor” to 5 = “very good”. Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.
AbstractList We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99m Tc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with 99m Tc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [ 99m Tc]Tc-MIBI or ~ 150 MBq [ 99m Tc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = “very poor” to 5 = “very good”. Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.
We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99mTc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with 99mTc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [99mTc]Tc-MIBI or ~ 150 MBq [99mTc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = "very poor" to 5 = "very good". Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99mTc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with 99mTc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [99mTc]Tc-MIBI or ~ 150 MBq [99mTc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = "very poor" to 5 = "very good". Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.
We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available 99mTc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with 99mTc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [99mTc]Tc-MIBI or ~ 150 MBq [99mTc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = “very poor” to 5 = “very good”. Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.
We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large stationary detectors for visualization of rat hearts and bones using clinically available Tc-labelled tracers. Sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR) of the small-animal SPECT scanner were investigated in phantom studies using an ultra-high-resolution rat and mouse multi-pinhole collimator (UHR-RM). Point source, hot-rod, and uniform phantoms with Tc-solution were scanned for high-count performance assessment and count levels equal to animal scans, respectively. Reconstruction was performed using the similarity-regulated ordered-subsets expectation maximization (SROSEM) algorithm with Gaussian smoothing. Rats were injected with ~ 100 MBq [ Tc]Tc-MIBI or ~ 150 MBq [ Tc]Tc-HMDP and received multi-frame micro-SPECT imaging after tracer distribution. Animal scans were reconstructed for three different acquisition times and post-processed with different sized Gaussian filters. Following reconstruction, CNR was calculated and image quality evaluated by three independent readers on a five-point scale from 1 = "very poor" to 5 = "very good". Point source sensitivity was 567 cps/MBq and radioactive rods as small as 1.2 mm were resolved with the UHR-RM collimator. Collimator-dependent uniformity was 55.5%. Phantom CNR improved with increasing rod size, filter size and activity concentration. Left ventricle and bone structures were successfully visualized in rat experiments. Image quality was strongly affected by the extent of post-filtering, whereas scan time did not have substantial influence on visual assessment. Good image quality was achieved for resolution range greater than 1.8 mm in bone and 2.8 mm in heart. The recently introduced small animal SPECT system with two stationary detectors and UHR-RM collimator is capable to provide excellent image quality in heart and bone scans in a rat using standardized reconstruction parameters and appropriate post-filtering. However, there are still challenges in achieving maximum system resolution in the sub-millimeter range with in vivo settings under limited injection dose and acquisition time.
ArticleNumber 18616
Author Janssen, Jan P.
Higuchi, Takahiro
Kanno, Takayuki
Hoffmann, Jan V.
Nose, Naoko
Onoguchi, Masahisa
Grunz, Jan-Peter
Lapa, Constantin
Buck, Andreas K.
Chen, Xinyu
Author_xml – sequence: 1
  givenname: Jan P.
  surname: Janssen
  fullname: Janssen, Jan P.
  organization: Department of Nuclear Medicine, University Hospital Würzburg, Comprehensive Heart Failure Centre, University Hospital Würzburg
– sequence: 2
  givenname: Jan V.
  surname: Hoffmann
  fullname: Hoffmann, Jan V.
  organization: Department of Nuclear Medicine, University Hospital Würzburg, Comprehensive Heart Failure Centre, University Hospital Würzburg
– sequence: 3
  givenname: Takayuki
  surname: Kanno
  fullname: Kanno, Takayuki
  organization: Comprehensive Heart Failure Centre, University Hospital Würzburg, Department of Quantum Medical Technology, Graduate School of Medical Sciences, Kanazawa University
– sequence: 4
  givenname: Naoko
  surname: Nose
  fullname: Nose, Naoko
  organization: Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University
– sequence: 5
  givenname: Jan-Peter
  surname: Grunz
  fullname: Grunz, Jan-Peter
  organization: Department of Diagnostic and Interventional Radiology, University Hospital Würzburg
– sequence: 6
  givenname: Masahisa
  surname: Onoguchi
  fullname: Onoguchi, Masahisa
  organization: Department of Quantum Medical Technology, Graduate School of Medical Sciences, Kanazawa University
– sequence: 7
  givenname: Xinyu
  surname: Chen
  fullname: Chen, Xinyu
  organization: Comprehensive Heart Failure Centre, University Hospital Würzburg, Nuclear Medicine, Medical Faculty, University of Augsburg
– sequence: 8
  givenname: Constantin
  surname: Lapa
  fullname: Lapa, Constantin
  organization: Nuclear Medicine, Medical Faculty, University of Augsburg
– sequence: 9
  givenname: Andreas K.
  surname: Buck
  fullname: Buck, Andreas K.
  organization: Department of Nuclear Medicine, University Hospital Würzburg
– sequence: 10
  givenname: Takahiro
  surname: Higuchi
  fullname: Higuchi, Takahiro
  email: higuchi_t@ukw.de
  organization: Department of Nuclear Medicine, University Hospital Würzburg, Comprehensive Heart Failure Centre, University Hospital Würzburg, Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University
BackLink https://www.ncbi.nlm.nih.gov/pubmed/33122774$$D View this record in MEDLINE/PubMed
BookMark eNp9kU1rXCEUhqUkNGmaP9BFEbrp5jZ-XnVTKEP6AYEGmi66Eke9M4Y7OlXvhPz7Or1JmmYRERR83nPe4_sKHMQUPQBvMPqAEZVnhWGuZIcI6gTvVd-hF-CYIMY7Qgk5eHQ_AqelXKO2OFEMq5fgiFJMiBDsGPxamK1ZhjHU4AtMA9xMYw3dNsR1Gj38cXm-uII3oa5hvUmwVFNDiibfQuertzXlAoeUYYhwF3YJZlNh2JhViKvX4HAwY_Gnd-cJ-Pn5_Grxtbv4_uXb4tNFZ7lktRNEKecs9tzIvT-37P1gLKJsKa1k3DHnLaNOYCIMphxjSQZljMWDRMhxegI-znW303LjnfWxZjPqbW4-8q1OJuj_X2JY61XaacFVz4RoBd7fFcjp9-RL1ZtQrB9HE32aiiaM9wzvd0PfPUGv05RjG69RAnMiuFSNevvY0YOV-19vAJkBm1Mp2Q8PCEZ6n66e09UtXf03XY2aSD4R2TDH0aYK4_NSOktL6xNXPv-z_YzqDx_zuWI
CitedBy_id crossref_primary_10_1016_j_sab_2023_106806
crossref_primary_10_1021_acsptsci_4c00383
crossref_primary_10_1038_s41598_022_21687_2
crossref_primary_10_1155_2022_9810097
Cites_doi 10.2967/jnumed.112.117572
10.2967/jnumed.108.056606
10.1109/42.97591
10.1002/jcb.10447
10.2967/jnumed.119.226027
10.1109/TNS.2004.830975
10.1007/s12149-012-0681-5
10.2310/7290.2014.00053
10.2967/jnumed.114.147140
10.2967/jnumed.110.087114
10.2967/jnumed.108.055442
10.1242/dmm.002733
10.1162/153535003322556877
10.2967/jnumed.113.136663
10.1186/s40658-015-0111-6
10.1088/0031-9155/50/22/R01
10.1007/s00259-010-1683-y
10.1088/0031-9155/61/11/4300
10.1109/TMI.1982.4307558
10.1007/s12149-014-0828-7
10.1007/s00259-006-0248-6
10.2967/jnumed.111.101899
10.1016/0047-0740(84)90004-4
10.33549/physiolres.931327
ContentType Journal Article
Copyright The Author(s) 2020
The Author(s) 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Copyright_xml – notice: The Author(s) 2020
– notice: The Author(s) 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
DBID C6C
AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
3V.
7X7
7XB
88A
88E
88I
8FE
8FH
8FI
8FJ
8FK
ABUWG
AEUYN
AFKRA
AZQEC
BBNVY
BENPR
BHPHI
CCPQU
DWQXO
FYUFA
GHDGH
GNUQQ
HCIFZ
K9.
LK8
M0S
M1P
M2P
M7P
PHGZM
PHGZT
PIMPY
PJZUB
PKEHL
PPXIY
PQEST
PQGLB
PQQKQ
PQUKI
PRINS
Q9U
7X8
5PM
DOI 10.1038/s41598-020-75696-0
DatabaseName Springer Nature OA Free Journals
CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
ProQuest Central (Corporate)
Health & Medical Collection
ProQuest Central (purchase pre-March 2016)
Biology Database (Alumni Edition)
Medical Database (Alumni Edition)
Science Database (Alumni Edition)
ProQuest SciTech Collection
ProQuest Natural Science Collection
Hospital Premium Collection
Hospital Premium Collection (Alumni Edition)
ProQuest Central (Alumni) (purchase pre-March 2016)
ProQuest Central (Alumni)
ProQuest One Sustainability
ProQuest Central UK/Ireland
ProQuest Central Essentials
Biological Science Collection
ProQuest Central
Natural Science Collection
ProQuest One Community College
ProQuest Central
Health Research Premium Collection
Health Research Premium Collection (Alumni)
ProQuest Central Student
SciTech Premium Collection
ProQuest Health & Medical Complete (Alumni)
Biological Sciences
ProQuest Health & Medical Collection
Medical Database
Science Database
Biological Science Database
ProQuest Central Premium
ProQuest One Academic (New)
Publicly Available Content Database
ProQuest Health & Medical Research Collection
ProQuest One Academic Middle East (New)
ProQuest One Health & Nursing
ProQuest One Academic Eastern Edition (DO NOT USE)
ProQuest One Applied & Life Sciences
ProQuest One Academic
ProQuest One Academic UKI Edition
ProQuest Central China
ProQuest Central Basic
MEDLINE - Academic
PubMed Central (Full Participant titles)
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
Publicly Available Content Database
ProQuest Central Student
ProQuest One Academic Middle East (New)
ProQuest Central Essentials
ProQuest Health & Medical Complete (Alumni)
ProQuest Central (Alumni Edition)
SciTech Premium Collection
ProQuest One Community College
ProQuest One Health & Nursing
ProQuest Natural Science Collection
ProQuest Central China
ProQuest Biology Journals (Alumni Edition)
ProQuest Central
ProQuest One Applied & Life Sciences
ProQuest One Sustainability
ProQuest Health & Medical Research Collection
Health Research Premium Collection
Health and Medicine Complete (Alumni Edition)
Natural Science Collection
ProQuest Central Korea
Health & Medical Research Collection
Biological Science Collection
ProQuest Central (New)
ProQuest Medical Library (Alumni)
ProQuest Science Journals (Alumni Edition)
ProQuest Biological Science Collection
ProQuest Central Basic
ProQuest Science Journals
ProQuest One Academic Eastern Edition
ProQuest Hospital Collection
Health Research Premium Collection (Alumni)
Biological Science Database
ProQuest SciTech Collection
ProQuest Hospital Collection (Alumni)
ProQuest Health & Medical Complete
ProQuest Medical Library
ProQuest One Academic UKI Edition
ProQuest One Academic
ProQuest One Academic (New)
ProQuest Central (Alumni)
MEDLINE - Academic
DatabaseTitleList
MEDLINE - Academic

Publicly Available Content Database
CrossRef
MEDLINE
Database_xml – sequence: 1
  dbid: C6C
  name: Springer Nature OA Free Journals
  url: http://www.springeropen.com/
  sourceTypes: Publisher
– sequence: 2
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 3
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
– sequence: 4
  dbid: BENPR
  name: ProQuest Central
  url: https://www.proquest.com/central
  sourceTypes: Aggregation Database
DeliveryMethod fulltext_linktorsrc
Discipline Biology
EISSN 2045-2322
ExternalDocumentID PMC7596477
33122774
10_1038_s41598_020_75696_0
Genre Research Support, Non-U.S. Gov't
Journal Article
GrantInformation_xml – fundername: Projekt DEAL
– fundername: ;
GroupedDBID 0R~
3V.
4.4
53G
5VS
7X7
88A
88E
88I
8FE
8FH
8FI
8FJ
AAFWJ
AAJSJ
AAKDD
ABDBF
ABUWG
ACGFS
ACSMW
ACUHS
ADBBV
ADRAZ
AENEX
AEUYN
AFKRA
AJTQC
ALIPV
ALMA_UNASSIGNED_HOLDINGS
AOIJS
AZQEC
BAWUL
BBNVY
BCNDV
BENPR
BHPHI
BPHCQ
BVXVI
C6C
CCPQU
DIK
DWQXO
EBD
EBLON
EBS
ESX
FYUFA
GNUQQ
GROUPED_DOAJ
GX1
HCIFZ
HH5
HMCUK
HYE
KQ8
LK8
M0L
M1P
M2P
M48
M7P
M~E
NAO
OK1
PIMPY
PQQKQ
PROAC
PSQYO
RNT
RNTTT
RPM
SNYQT
UKHRP
AASML
AAYXX
AFPKN
CITATION
PHGZM
PHGZT
CGR
CUY
CVF
ECM
EIF
NPM
7XB
8FK
AARCD
K9.
PJZUB
PKEHL
PPXIY
PQEST
PQGLB
PQUKI
PRINS
Q9U
7X8
5PM
ID FETCH-LOGICAL-c584t-7299ddc1e5a85294db6efac034b8c845d4dec43d7127a1351182f9aac1f800d53
IEDL.DBID M48
ISSN 2045-2322
IngestDate Thu Aug 21 18:06:34 EDT 2025
Mon Jul 21 09:41:06 EDT 2025
Wed Aug 13 11:12:20 EDT 2025
Thu Apr 03 07:08:10 EDT 2025
Tue Jul 01 02:46:45 EDT 2025
Thu Apr 24 23:12:00 EDT 2025
Fri Feb 21 02:38:45 EST 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 1
Language English
License Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c584t-7299ddc1e5a85294db6efac034b8c845d4dec43d7127a1351182f9aac1f800d53
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
OpenAccessLink http://journals.scholarsportal.info/openUrl.xqy?doi=10.1038/s41598-020-75696-0
PMID 33122774
PQID 2471527589
PQPubID 2041939
ParticipantIDs pubmedcentral_primary_oai_pubmedcentral_nih_gov_7596477
proquest_miscellaneous_2456416416
proquest_journals_2471527589
pubmed_primary_33122774
crossref_primary_10_1038_s41598_020_75696_0
crossref_citationtrail_10_1038_s41598_020_75696_0
springer_journals_10_1038_s41598_020_75696_0
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2020-10-29
PublicationDateYYYYMMDD 2020-10-29
PublicationDate_xml – month: 10
  year: 2020
  text: 2020-10-29
  day: 29
PublicationDecade 2020
PublicationPlace London
PublicationPlace_xml – name: London
– name: England
PublicationTitle Scientific reports
PublicationTitleAbbrev Sci Rep
PublicationTitleAlternate Sci Rep
PublicationYear 2020
Publisher Nature Publishing Group UK
Nature Publishing Group
Publisher_xml – name: Nature Publishing Group UK
– name: Nature Publishing Group
References Jones (CR17) 1984; 11
Schramm (CR29) 2007; 48
Iannaccone, Jacob (CR16) 2009; 2
Bayne (CR26) 1996; 39
King, Pretorius, Farncombe, Beekman (CR1) 2002; 39
Meikle, Kench, Kassiou, Banati (CR2) 2005; 50
Bevan, Tofe, Benedict, Francis, Barnett (CR18) 1980; 21
Boisson (CR28) 2013; 54
Loening, Gambhir (CR23) 2003; 2
Lukas, Kluge, Beindorff, Brenner (CR30) 2020; 61
Beekman, van der Have (CR3) 2007; 34
Ogawa, Harata, Ichihara, Kubo, Hashimoto (CR22) 1991; 10
Arsos (CR27) 2009; 58
Vaissier, Beekman, Goorden (CR15) 2016; 61
Walker (CR25) 2014; 55
van der Have (CR10) 2009; 50
Ivashchenko, van der Have, Goorden, Ramakers, Beekman (CR6) 2015; 56
Ivashchenko (CR11) 2014; 13
Harteveld (CR13) 2011; 52
Magota (CR4) 2011; 38
CR24
Higaki (CR12) 2013; 27
Mizutani (CR14) 2015; 2
Franc, Acton, Mari, Hasegawa (CR7) 2008; 49
Furenlid (CR8) 2004; 51
Bevan, Tofe, Benedict, Francis, Barnett (CR19) 1980; 21
Vaissier (CR20) 2012; 53
Shepp, Vardi (CR21) 1982; 1
Vastenhouw, Beekman (CR9) 2007; 48
Matsunari (CR5) 2014; 28
SR Meikle (75696_CR2) 2005; 50
PM Iannaccone (75696_CR16) 2009; 2
I Matsunari (75696_CR5) 2014; 28
N Schramm (75696_CR29) 2007; 48
O Ivashchenko (75696_CR6) 2015; 56
JA Bevan (75696_CR18) 1980; 21
LR Furenlid (75696_CR8) 2004; 51
Y Higaki (75696_CR12) 2013; 27
PE Vaissier (75696_CR20) 2012; 53
75696_CR24
M Lukas (75696_CR30) 2020; 61
F Beekman (75696_CR3) 2007; 34
LA Shepp (75696_CR21) 1982; 1
F Boisson (75696_CR28) 2013; 54
JA Bevan (75696_CR19) 1980; 21
AG Jones (75696_CR17) 1984; 11
K Magota (75696_CR4) 2011; 38
AM Loening (75696_CR23) 2003; 2
A Mizutani (75696_CR14) 2015; 2
MD Walker (75696_CR25) 2014; 55
O Ivashchenko (75696_CR11) 2014; 13
PE Vaissier (75696_CR15) 2016; 61
G Arsos (75696_CR27) 2009; 58
K Bayne (75696_CR26) 1996; 39
F van der Have (75696_CR10) 2009; 50
K Ogawa (75696_CR22) 1991; 10
MA King (75696_CR1) 2002; 39
B Vastenhouw (75696_CR9) 2007; 48
AA Harteveld (75696_CR13) 2011; 52
BL Franc (75696_CR7) 2008; 49
References_xml – volume: 54
  start-page: 1833
  year: 2013
  end-page: 1840
  ident: CR28
  article-title: Imaging capabilities of the Inveon SPECT system using single-and multipinhole collimators
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.112.117572
– volume: 50
  start-page: 599
  year: 2009
  end-page: 605
  ident: CR10
  article-title: U-SPECT-II: An ultra-high-resolution device for molecular small-animal imaging
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.108.056606
– volume: 10
  start-page: 408
  year: 1991
  end-page: 412
  ident: CR22
  article-title: A practical method for position-dependent Compton-Scatter correction in single photon emission CT
  publication-title: IEEE Trans. Med. Imaging
  doi: 10.1109/42.97591
– volume: 39
  start-page: 208
  issue: 199
  year: 1996
  end-page: 111
  ident: CR26
  article-title: Revised guide for the care and use of laboratory animals available. American Physiological Society
  publication-title: Physiologist
– volume: 39
  start-page: 221
  year: 2002
  end-page: 230
  ident: CR1
  article-title: Introduction to the physics of molecular imaging with radioactive tracers in small animals
  publication-title: J. Cell Biochem. Suppl.
  doi: 10.1002/jcb.10447
– volume: 58
  start-page: 21
  year: 2009
  end-page: 28
  ident: CR27
  article-title: (99m)Tc-sestamibi uptake in rat skeletal muscle and heart: Physiological determinants and correlations
  publication-title: Physiol. Res.
– volume: 61
  start-page: 152
  year: 2020
  end-page: 161
  ident: CR30
  article-title: Multi-isotope capabilities of a small-animal multi-pinhole SPECT system
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.119.226027
– volume: 51
  start-page: 631
  year: 2004
  end-page: 635
  ident: CR8
  article-title: FastSPECT II: A second-generation high-resolution dynamic SPECT imager
  publication-title: IEEE Trans. Nucl. Sci.
  doi: 10.1109/TNS.2004.830975
– volume: 27
  start-page: 271
  year: 2013
  end-page: 278
  ident: CR12
  article-title: Appropriate collimators in a small animal SPECT scanner with CZT detector
  publication-title: Ann. Nucl. Med.
  doi: 10.1007/s12149-012-0681-5
– volume: 21
  start-page: 967
  year: 1980
  end-page: 970
  ident: CR19
  article-title: Tc-99m HMDP (hydroxymethylene diphosphonate): A radiopharmaceutical for skeletal and acute myocardial infarct imaging. II. Comparison of Tc-99m hydroxymethylene diphosphonate (HMDP) with other technetium-labeled bone-imaging agents in a canine model
  publication-title: J. Nucl. Med.
– volume: 13
  start-page: 10
  year: 2014
  ident: CR11
  article-title: Quarter-millimeter-resolution molecular mouse imaging with U-SPECT(+)
  publication-title: Mol.S Imaging
  doi: 10.2310/7290.2014.00053
– volume: 56
  start-page: 470
  year: 2015
  end-page: 475
  ident: CR6
  article-title: Ultra-high-sensitivity submillimeter mouse SPECT
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.114.147140
– volume: 48
  start-page: 487
  year: 2007
  end-page: 493
  ident: CR9
  article-title: Submillimeter total-body murine imaging with U-SPECT-I
  publication-title: J. Nucl. Med.
– volume: 52
  start-page: 1646
  year: 2011
  end-page: 1653
  ident: CR13
  article-title: Using the NEMA NU 4 PET image quality phantom in multipinhole small-animal SPECT
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.110.087114
– volume: 21
  start-page: 961
  year: 1980
  end-page: 966
  ident: CR18
  article-title: Tc-99m HMDP (hydroxymethylene diphosphonate): A radiopharmaceutical for skeletal and acute myocardial infarct imaging. I. Synthesis and distribution in animals
  publication-title: J. Nucl. Med.
– volume: 49
  start-page: 1651
  year: 2008
  end-page: 1663
  ident: CR7
  article-title: Small-animal SPECT and SPECT/CT: Important tools for preclinical investigation
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.108.055442
– volume: 2
  start-page: 206
  year: 2009
  end-page: 210
  ident: CR16
  article-title: Rats!
  publication-title: Dis. Model Mech.
  doi: 10.1242/dmm.002733
– volume: 2
  start-page: 131
  year: 2003
  end-page: 137
  ident: CR23
  article-title: AMIDE: A free software tool for multimodality medical image analysis
  publication-title: Mol. Imaging
  doi: 10.1162/153535003322556877
– volume: 55
  start-page: 1368
  year: 2014
  end-page: 1374
  ident: CR25
  article-title: Performance assessment of a preclinical PET scanner with pinhole collimation by comparison to a coincidence-based small-animal PET scanner
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.113.136663
– volume: 2
  start-page: 7
  year: 2015
  ident: CR14
  article-title: Impact of injection dose, post-reconstruction filtering, and collimator choice on image quality of myocardial perfusion SPECT using cadmium-zinc telluride detectors in the rat
  publication-title: EJNMMI Phys.
  doi: 10.1186/s40658-015-0111-6
– volume: 50
  start-page: R45
  year: 2005
  end-page: 61
  ident: CR2
  article-title: Small animal SPECT and its place in the matrix of molecular imaging technologies
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/50/22/R01
– volume: 38
  start-page: 742
  year: 2011
  end-page: 752
  ident: CR4
  article-title: Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging
  publication-title: Eur. J. Nucl. Med. Mol. Imaging
  doi: 10.1007/s00259-010-1683-y
– volume: 61
  start-page: 4300
  year: 2016
  end-page: 4315
  ident: CR15
  article-title: Similarity-regulation of OS-EM for accelerated SPECT reconstruction
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/61/11/4300
– volume: 1
  start-page: 113
  year: 1982
  end-page: 122
  ident: CR21
  article-title: Maximum likelihood reconstruction for emission tomography
  publication-title: IEEE Trans Med. Imaging
  doi: 10.1109/TMI.1982.4307558
– volume: 28
  start-page: 484
  year: 2014
  end-page: 497
  ident: CR5
  article-title: Performance evaluation of the eXplore speCZT preclinical imaging system
  publication-title: Ann. Nucl. Med.
  doi: 10.1007/s12149-014-0828-7
– volume: 48
  start-page: 436P
  year: 2007
  ident: CR29
  article-title: Improving resolution, sensitivity and applications for the NanoSPECT/CT: A high-performance SPECT/CT imager for small-animal research
  publication-title: J. Nucl. Med.
– volume: 34
  start-page: 151
  year: 2007
  end-page: 161
  ident: CR3
  article-title: The pinhole: Gateway to ultra-high-resolution three-dimensional radionuclide imaging
  publication-title: Eur. J. Nucl. Med. Mol. Imaging
  doi: 10.1007/s00259-006-0248-6
– volume: 53
  start-page: 1292
  year: 2012
  end-page: 1299
  ident: CR20
  article-title: Fast spiral SPECT with stationary gamma-cameras and focusing pinholes
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.111.101899
– ident: CR24
– volume: 11
  start-page: 225
  year: 1984
  end-page: 234
  ident: CR17
  article-title: Biological studies of a new class of technetium complexes: The hexakis(alkylisonitrile)technetium(I) cations
  publication-title: Int. J. Nucl. Med. Biol.
  doi: 10.1016/0047-0740(84)90004-4
– volume: 61
  start-page: 152
  year: 2020
  ident: 75696_CR30
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.119.226027
– volume: 21
  start-page: 961
  year: 1980
  ident: 75696_CR18
  publication-title: J. Nucl. Med.
– volume: 34
  start-page: 151
  year: 2007
  ident: 75696_CR3
  publication-title: Eur. J. Nucl. Med. Mol. Imaging
  doi: 10.1007/s00259-006-0248-6
– volume: 10
  start-page: 408
  year: 1991
  ident: 75696_CR22
  publication-title: IEEE Trans. Med. Imaging
  doi: 10.1109/42.97591
– volume: 58
  start-page: 21
  year: 2009
  ident: 75696_CR27
  publication-title: Physiol. Res.
  doi: 10.33549/physiolres.931327
– volume: 50
  start-page: R45
  year: 2005
  ident: 75696_CR2
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/50/22/R01
– volume: 2
  start-page: 131
  year: 2003
  ident: 75696_CR23
  publication-title: Mol. Imaging
  doi: 10.1162/153535003322556877
– volume: 39
  start-page: 221
  year: 2002
  ident: 75696_CR1
  publication-title: J. Cell Biochem. Suppl.
  doi: 10.1002/jcb.10447
– volume: 28
  start-page: 484
  year: 2014
  ident: 75696_CR5
  publication-title: Ann. Nucl. Med.
  doi: 10.1007/s12149-014-0828-7
– volume: 50
  start-page: 599
  year: 2009
  ident: 75696_CR10
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.108.056606
– volume: 54
  start-page: 1833
  year: 2013
  ident: 75696_CR28
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.112.117572
– volume: 55
  start-page: 1368
  year: 2014
  ident: 75696_CR25
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.113.136663
– volume: 52
  start-page: 1646
  year: 2011
  ident: 75696_CR13
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.110.087114
– volume: 2
  start-page: 206
  year: 2009
  ident: 75696_CR16
  publication-title: Dis. Model Mech.
  doi: 10.1242/dmm.002733
– volume: 2
  start-page: 7
  year: 2015
  ident: 75696_CR14
  publication-title: EJNMMI Phys.
  doi: 10.1186/s40658-015-0111-6
– volume: 56
  start-page: 470
  year: 2015
  ident: 75696_CR6
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.114.147140
– volume: 21
  start-page: 967
  year: 1980
  ident: 75696_CR19
  publication-title: J. Nucl. Med.
– volume: 48
  start-page: 436P
  year: 2007
  ident: 75696_CR29
  publication-title: J. Nucl. Med.
– volume: 53
  start-page: 1292
  year: 2012
  ident: 75696_CR20
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.111.101899
– volume: 13
  start-page: 10
  year: 2014
  ident: 75696_CR11
  publication-title: Mol.S Imaging
  doi: 10.2310/7290.2014.00053
– volume: 11
  start-page: 225
  year: 1984
  ident: 75696_CR17
  publication-title: Int. J. Nucl. Med. Biol.
  doi: 10.1016/0047-0740(84)90004-4
– volume: 51
  start-page: 631
  year: 2004
  ident: 75696_CR8
  publication-title: IEEE Trans. Nucl. Sci.
  doi: 10.1109/TNS.2004.830975
– volume: 1
  start-page: 113
  year: 1982
  ident: 75696_CR21
  publication-title: IEEE Trans Med. Imaging
  doi: 10.1109/TMI.1982.4307558
– volume: 38
  start-page: 742
  year: 2011
  ident: 75696_CR4
  publication-title: Eur. J. Nucl. Med. Mol. Imaging
  doi: 10.1007/s00259-010-1683-y
– volume: 27
  start-page: 271
  year: 2013
  ident: 75696_CR12
  publication-title: Ann. Nucl. Med.
  doi: 10.1007/s12149-012-0681-5
– volume: 49
  start-page: 1651
  year: 2008
  ident: 75696_CR7
  publication-title: J. Nucl. Med.
  doi: 10.2967/jnumed.108.055442
– ident: 75696_CR24
– volume: 61
  start-page: 4300
  year: 2016
  ident: 75696_CR15
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/61/11/4300
– volume: 39
  start-page: 208
  issue: 199
  year: 1996
  ident: 75696_CR26
  publication-title: Physiologist
– volume: 48
  start-page: 487
  year: 2007
  ident: 75696_CR9
  publication-title: J. Nucl. Med.
SSID ssj0000529419
Score 2.3360796
Snippet We aimed to investigate the image quality of the U-SPECT5/CT E-Class a micro single-photon emission computed tomography (SPECT) system with two large...
SourceID pubmedcentral
proquest
pubmed
crossref
springer
SourceType Open Access Repository
Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 18616
SubjectTerms 692/308
692/308/2778
692/308/575
692/53
692/53/2421
Animals
Bone and Bones - diagnostic imaging
Computed tomography
Equipment Design - instrumentation
Equipment Design - methods
Heart
Heart Ventricles - diagnostic imaging
Humanities and Social Sciences
Image Processing, Computer-Assisted - instrumentation
Image Processing, Computer-Assisted - methods
Mice
multidisciplinary
Performance assessment
Phantoms, Imaging
Radioactive tracers
Rats
Science
Science (multidisciplinary)
Sensors
Single photon emission computed tomography
Spatial discrimination
Technetium - administration & dosage
Technetium Tc 99m Sestamibi - administration & dosage
Tomography, Emission-Computed, Single-Photon - instrumentation
Tomography, Emission-Computed, Single-Photon - methods
Ventricle
SummonAdditionalLinks – databaseName: Health & Medical Collection
  dbid: 7X7
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3Pa90wDBZrx6CXsd_L1g4XdutCE8eOndMYj5Yy2Bi0hddTcGyHBbbkrUk7-t9PcvJSXssKudkhsSxbnyVZH8BHwzOLQDqJjdYuFrVwcWVEFRc2MbYWSlaW7jt_-56fnIuvS7mcHG79lFa53hPDRu06Sz7yQ467qOSIbovPqz8xsUZRdHWi0NiCx1S6jLRaLdXsY6EolkiL6a5MkunDHu0V3SnDM5OSOeXfbtqjeyDzfq7knYBpsEPHz-DpBCDZl3HGn8Mj376AJyOl5M1LuFig9QsJr3gEZl3NQsZgvGpaIsJlpz-OFmeMnK9s-NuxfgzEm8sb5vwQHPg9QxjLmpZdN9cdQ_1gze_AZPQKzo-PzhYn8USfEFtEFQOR1BbO2dRLo0kQrsp9bWySiUpbLaQTzluROZVyZYioD48adWGMTWtEkU5mr2G77Vr_Fpjntck9HsILq4TwuZYOoQTPrUH4qJMqgnQtxNJOtcWJ4uJXGWLcmS5HwZco-DIIvkwiOJjfWY2VNR7svbuem3JaZX15qxMR7M_NuD4o6GFa311RH5kj6MQngjfjVM6fy7KUc8S_EaiNSZ47UO3tzZa2-RlqcCsZrvBG8GmtDre_9f9RvHt4FO9hh5NqomnkxS5sD5dXfg8xz1B9CIr9D35f_6I
  priority: 102
  providerName: ProQuest
– databaseName: HAS SpringerNature Open Access 2022
  dbid: AAJSJ
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1bS-UwEB5cRdgX8brWGxF802KbJm36eDgockARVNCnkiYpW1jbg6cq_nsn6WU5qysIfcuEppNJ55vMDeBI0kghkA58KYT2WcG0n0uW-6kKpCpYwnNl850vr-KLOza55_cLQPtcGBe070paut90Hx12OkNFY5PB0NhJeGwDZ3_Aki3VjrK9NBpNbibDzYr1XbEw7TJkgkh8MnleC32Alh8jJP9xkzrtc74KKx1sJKN2oWuwYKp1WG4bSb5twMMYdZ4Lc0XDl9QFcXGC_rSsbPtbcnN9Nr4l9sqVNK81mbXud_n0RrRp3LX9jCB4JWVFXsqXmqBUkPLR9S_ahLvzs9vxhd81TfAVYonGtqZNtVah4VJYRug8NoVUQcRyoQTjmmmjWKSTkCbStudDA6NIpVRhgdhR82gLFqu6MttADC1kbND0TlXCmIkF1wggaKwkgkYR5B6EPRMz1VUUt40t_mTOsx2JrGV8hozPHOOzwIPjYc60rafxJfVevzdZd7ZmGUV9yinaOakHh8Mwngrr6pCVqZ8tDY8RauLjwa92K4fXRVFIKaJeD5K5TR4IbMXt-ZGq_O0qbyfcJe56cNKLw99l_f8rdr5Hvgs_qRVVVJA03YPF5unZ7CPyafKDTtTfAVsg_sk
  priority: 102
  providerName: Springer Nature
Title Capabilities of multi-pinhole SPECT with two stationary detectors for in vivo rat imaging
URI https://link.springer.com/article/10.1038/s41598-020-75696-0
https://www.ncbi.nlm.nih.gov/pubmed/33122774
https://www.proquest.com/docview/2471527589
https://www.proquest.com/docview/2456416416
https://pubmed.ncbi.nlm.nih.gov/PMC7596477
Volume 10
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV3da9swED_6wWAvY92nty5osLfNmy1Llv1QRmpSSqClrA1kT0aWZGbo7C5xu-W_70m2M7J2Y2AwWPLX6cT9Trq7H8A7SSOFQDrwZZJon5VM-4VkhZ-qQKqSCV4om-98chofz9h0zudbMNAd9QJc3uvaWT6p2eLy468fq8844Q-6lPHk0xKNkE0UQ0dI8NgG1W7DLlomYRkNTnq439X6pilzXB-2CLuPYIL2eTT3P2bTVt0BoHfjKP_YTHU26ugxPOrBJRl32rAHW6Z-Ag86usnVU_iaoWV0wbDoHpOmJC6a0L-qakuSS87PJtkFsQuzpP3ZkGW3SS8XK6JN6xb3lwQhLqlqclPdNAR1h1TfHcvRM5gdTS6yY7-nVvAVIo7WEtimWqvQcJlYoegiNqVUQcSKRCWMa6aNYpEWIRXSkvihG1KmUqqwRISpefQcduqmNi-BGFrK2KCDnirBmIkTrhFm0FhJhJZJUHgQDkLMVV933NJfXOZu_ztK8k7wOQo-d4LPAw_er--56qpu_LP3_jA2-aBAOUWryyl6Q6kHb9fNOHfshoisTXNt-_AYASkeHrzohnL9uigKKUVs7IHYGOR1B1uXe7Olrr65-tyCu_ReDz4M6vD7s_7-F6_-4zNfw0Nq9RNtJ033YaddXJs3CIraYgTbYi5GsDseT8-neD6cnJ59watZnI3cQsPIzYVbG6MMjA
linkProvider Scholars Portal
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9QwEB5VRQguiDeBAkaCE0RNHDuPA0JoabWlDyGxlZZTcGxHRIJkabat9k_xG5lxHtVS0Vul3Owo8Xjs-cYz4w_gteKRRiAd-CpNjS9KYfxCicLPdKB0KRJZaKp3PjyKp8fi81zON-DPUAtDaZXDnug2atNoOiPf5riLSo7oNvuw-O0TaxRFVwcKjU4t9u3qHF229v3eJ5zfN5zv7swmU79nFfA1GtslcbdmxujQSpVKnglTxLZUOohEkepUSCOM1SIyScgTRfx1iMDLTCkdlgiuDLFE4JZ_Aw1vQM5eMk_GMx2Kmokw62tzgijdbtE-Ug0b-miJjCnfd93-XQK1l3Mz_wnQOru3exfu9ICVfew07B5s2Po-3OwoLFcP4NsEra1LsEWXmzUlcxmK_qKqiXiXff2yM5kxOuxly_OGtV3gX52smLFLFzBoGcJmVtXsrDprGOojq3455qSHcHwtgn0Em3VT2yfALC9VbNHpz3QihI1TaRC68FgrhKtpUHgQDkLMdX-XOVFq_MxdTD1K807wOQo-d4LPAw_eju8sups8ruy9NcxN3q_qNr_QQQ9ejc24HinIomrbnFIfGSPIxceDx91Ujp-LopBzxNseJGuTPHagu77XW-rqh7vzO5GuZNiDd4M6XPzW_0fx9OpRvIRb09nhQX6wd7T_DG5zUlM0yzzbgs3lyal9jnhrWbxwSs7g-3Wvqr_2hDyw
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9QwEB5VW4G4IN6kFDASnCDaxLHzOCAE2121FFYraKVySh3bEZFosjRpq_1r_DrGzqNaKnqrlJsdJZ4Zez57ZvwBvBY0kAikPVfEsXJZzpSbCZa5ifSEzFnEM2nqnb_Ow91D9vmIH23An74WxqRV9muiXahVJc0Z-ZjiKsopottknHdpEYud2Yflb9cwSJlIa0-n0ZrIvl5d4Patfr-3g7p-Q-lsejDZdTuGAVei420Mj2uilPQ1FzGnCVNZqHMhvYBlsYwZV0xpyQIV-TQShssO0XieCCH9HIGWMowRuPxvRmZXNILNT9P54ttwwmNiaMxPukodL4jHNXpLU9GGO7aIhyb7d90bXoG4VzM1_wnXWi84uwd3O_hKPrb2dh82dPkAbrWElquH8GOCvtem2-IGnFQ5sfmK7rIoDQ0v-b6YTg6IOfolzUVF6jYNQJyuiNKNDR_UBEE0KUpyXpxXBK2TFCeWR-kRHN6IaB_DqKxK_RSIprkItc5oIiPGdBhzhUCGhlIgeI29zAG_F2Iqu5vNDcHGr9RG2IM4bQWfouBTK_jUc-Dt8M6yvdfj2t7bvW7Sbo7X6aVFOvBqaMbZaUIuotTVmenDQ4S8-DjwpFXl8Lkg8ClF9O1AtKbkoYO5-Xu9pSx-2hvAI24LiB1415vD5W_9fxRb14_iJdzGGZV-2ZvvP4M71Fgp-miabMOoOT3TzxF8NdmLzsoJHN_0xPoLJF9CSw
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Capabilities+of+multi-pinhole+SPECT+with+two+stationary+detectors+for+in+vivo+rat+imaging&rft.jtitle=Scientific+reports&rft.au=Janssen%2C+Jan+P&rft.au=Hoffmann%2C+Jan+V&rft.au=Kanno%2C+Takayuki&rft.au=Nose%2C+Naoko&rft.date=2020-10-29&rft.issn=2045-2322&rft.eissn=2045-2322&rft.volume=10&rft.issue=1&rft.spage=18616&rft_id=info:doi/10.1038%2Fs41598-020-75696-0&rft.externalDBID=NO_FULL_TEXT
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2045-2322&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2045-2322&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2045-2322&client=summon