0Optimized Fluoroscopy Setting and Appropriate Project Position Can Reduce X-ray Radiation Doses Rates during Electrophysiology Procedures
Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology procedures.This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dos...
Saved in:
Published in | 中华医学杂志(英文版) Vol. 128; no. 9; pp. 1151 - 1153 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
2015
|
Subjects | |
Online Access | Get full text |
ISSN | 0366-6999 2542-5641 |
DOI | 10.4103/0366-6999.156079 |
Cover
Abstract | Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology procedures.This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dose rates.Methods:From June 2013 to October 2013,105 consecutive patients who underwent complex ablation were enrolled in the study.After the ablation,the radiation dose rates were measured by two different settings (default setting and optimized setting) with three projection positions (posteroanterior [PA] projection;left anterior oblique [LAO] 30° projection;and LAO 45° projection).The parameter of preset voltage,pulse width,critical voltage,peak voltage,noise reduction,edge enhancement,pulse rate,and dose per frame was modified in the optimized setting.Results:The optimized setting reduced radiation dose rates by 87.5% (1.7 Gy/min vs.13.6 Gy/min,P < 0.001) in PA,87.3% (2.5 Gy/min vs.19.7 Gy/min,P<0.001) in LAO 30°,85.9% (3.1 Gy/min vs.22.1 Gy/min,P < 0.001) in LAO 45°.Increase the angle of projection position will increase the radiation dose rate.Conclusions:We can reduce X-ray radiation dose rates by adjusting the parameter setting of X-ray system.Avoiding oblique projection of large angle is another way to reduce X-ray radiation dose rates. |
---|---|
AbstractList | Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology procedures.This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dose rates.Methods:From June 2013 to October 2013,105 consecutive patients who underwent complex ablation were enrolled in the study.After the ablation,the radiation dose rates were measured by two different settings (default setting and optimized setting) with three projection positions (posteroanterior [PA] projection;left anterior oblique [LAO] 30° projection;and LAO 45° projection).The parameter of preset voltage,pulse width,critical voltage,peak voltage,noise reduction,edge enhancement,pulse rate,and dose per frame was modified in the optimized setting.Results:The optimized setting reduced radiation dose rates by 87.5% (1.7 Gy/min vs.13.6 Gy/min,P < 0.001) in PA,87.3% (2.5 Gy/min vs.19.7 Gy/min,P<0.001) in LAO 30°,85.9% (3.1 Gy/min vs.22.1 Gy/min,P < 0.001) in LAO 45°.Increase the angle of projection position will increase the radiation dose rate.Conclusions:We can reduce X-ray radiation dose rates by adjusting the parameter setting of X-ray system.Avoiding oblique projection of large angle is another way to reduce X-ray radiation dose rates. Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology procedures.This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dose rates.Methods:From June 2013 to October 2013,105 consecutive patients who underwent complex ablation were enrolled in the study.After the ablation,the radiation dose rates were measured by two different settings (default setting and optimized setting) with three projection positions (posteroanterior [PA] projection;left anterior oblique [LAO] 30° projection;and LAO 45° projection).The parameter of preset voltage,pulse width,critical voltage,peak voltage,noise reduction,edge enhancement,pulse rate,and dose per frame was modified in the optimized setting.Results:The optimized setting reduced radiation dose rates by 87.5% (1.7 Gy/min vs.13.6 Gy/min,P < 0.001) in PA,87.3% (2.5 Gy/min vs.19.7 Gy/min,P<0.001) in LAO 30°,85.9% (3.1 Gy/min vs.22.1 Gy/min,P < 0.001) in LAO 45°.Increase the angle of projection position will increase the radiation dose rate.Conclusions:We can reduce X-ray radiation dose rates by adjusting the parameter setting of X-ray system.Avoiding oblique projection of large angle is another way to reduce X-ray radiation dose rates. |
Author | Bing-Bo Hou Yan Yao Ling-Min Wu Yu Qiao Li-Hui Zheng Li-Gang Ding Gang Chen |
AuthorAffiliation | State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China |
AuthorAffiliation_xml | – name: State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China |
Author_xml | – sequence: 1 fullname: Bing-Bo Hou Yan Yao Ling-Min Wu Yu Qiao Li-Hui Zheng Li-Gang Ding Gang Chen |
BookMark | eNo9kE9PAjEQxRuDiYDePTYevC32z7bbHgmCmphAkIO3TWm7sATatV1ilo_gp7YE42Umk_febyYzAD3nnQXgHqNRjhF9QpTzjEspR5hxVMgr0CcsJxnjOe6B_r98AwYx7hAijBW8D37QvGnrQ32yBs72Rx981L7p4Idt29ptoHIGjpsm-CbUqrVwEfzO6hYufKzb2js4UQ4urTlqCz-zoDq4VCY5z9KzjzamuU3VHMMZN92ncIJtu1j7vd90Z6BO8WDjLbiu1D7au78-BKvZdDV5zd7nL2-T8XummaBZZdcWKUEwWjNjCK0YKQrErSBGFqgQElcqF9VaW0bXOBeMSimUVrmR2GAi6BA8XrDfylXKbcqdPwaXFpanrT7sCMIMSYRoMj5cjHrr3eYrnV-mJxxU6ErO84ILJhD9BS4GdOk |
ContentType | Journal Article |
Copyright | Copyright © Wanfang Data Co. Ltd. All Rights Reserved. |
Copyright_xml | – notice: Copyright © Wanfang Data Co. Ltd. All Rights Reserved. |
DBID | 2RA 92L CQIGP W91 ~WA 2B. 4A8 92I 93N PSX TCJ |
DOI | 10.4103/0366-6999.156079 |
DatabaseName | 维普期刊资源整合服务平台 中文科技期刊数据库-CALIS站点 中文科技期刊数据库-7.0平台 中文科技期刊数据库-医药卫生 中文科技期刊数据库- 镜像站点 Wanfang Data Journals - Hong Kong WANFANG Data Centre Wanfang Data Journals 万方数据期刊 - 香港版 China Online Journals (COJ) China Online Journals (COJ) |
DatabaseTitleList | |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
DocumentTitleAlternate | 0Optimized Fluoroscopy Setting and Appropriate Project Position Can Reduce X-ray Radiation Doses Rates during Electrophysiology Procedures |
EISSN | 2542-5641 |
EndPage | 1153 |
ExternalDocumentID | zhcmj201509003 664768580 |
GroupedDBID | --- -05 -0E -SE -S~ .55 .GJ 29B 2B. 2C~ 2RA 2WC 3V. 40I 53G 5GY 5RE 5VR 5VS 6J9 7X7 88E 8FI 8FJ 92F 92I 92L 92M 93N 93R 9D9 9DE AAHPQ AASCR ABASU ABCQX ABDIG ABUWG ABVCZ ABXLX ACGFO ACGFS ACILI ADGGA ADHPY ADPDF ADRAZ AENEX AFDTB AFKRA AFUIB AHMBA AHVBC AINUH AJIOK AJNWD ALMA_UNASSIGNED_HOLDINGS ALMTX AMJPA AMKUR AMNEI ANFDX AOHHW BENPR BPHCQ BVXVI C1A CAJEE CAJUS CCEZO CCPQU CHBEP CIEJG CQIGP CW9 DIK DIWNM EBS EEVPB EJD F5P FA0 FCALG FRP FYUFA GNXGY GQDEL GROUPED_DOAJ GX1 HLJTE HMCUK HYE IAO IHR IHW IKREB INH INR IPNFZ ITC JUIAU KQ8 L7B M1P M48 OK1 OPUJH OVD OVDNE OVEED OXXIT P2P P6G PIMPY PQQKQ PROAC PSQYO PV9 Q-- Q-4 R-E RIG RLZ RNS RPM RT5 RZL S.. T8U TCJ TEORI TGQ TR2 TSPGW U1F U1G U5E U5O UKHRP W2D W91 WFFXF X7J X7M XSB ZA5 ZGI ZXP ~WA 0R~ 4A8 AAAAV AAIQE ABZZY ACXJB AFBFQ AHQNM AJCLO AJZMW AKCTQ ALIPV ALKUP AOQMC BQLVK H13 PHGZM PHGZT PMFND PSX |
ID | FETCH-LOGICAL-c583-febe0a8210b5dd23f527706e82d9707891fa48fbce53b14853998aca4d91d1283 |
IEDL.DBID | M48 |
ISSN | 0366-6999 |
IngestDate | Thu May 29 03:55:57 EDT 2025 Wed Feb 14 10:30:34 EST 2024 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 9 |
Keywords | Fluoroscopy Radiation Dose Ablation Projection Position |
Language | English |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-c583-febe0a8210b5dd23f527706e82d9707891fa48fbce53b14853998aca4d91d1283 |
Notes | 11-2154/R Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology procedures.This study aimed to evaluate the value of optimized parameter setting and different projection position to reduce X-ray radiation dose rates.Methods:From June 2013 to October 2013,105 consecutive patients who underwent complex ablation were enrolled in the study.After the ablation,the radiation dose rates were measured by two different settings (default setting and optimized setting) with three projection positions (posteroanterior [PA] projection;left anterior oblique [LAO] 30° projection;and LAO 45° projection).The parameter of preset voltage,pulse width,critical voltage,peak voltage,noise reduction,edge enhancement,pulse rate,and dose per frame was modified in the optimized setting.Results:The optimized setting reduced radiation dose rates by 87.5% (1.7 Gy/min vs.13.6 Gy/min,P < 0.001) in PA,87.3% (2.5 Gy/min vs.19.7 Gy/min,P<0.001) in LAO 30°,85.9% (3.1 Gy/min vs.22.1 Gy/min,P < 0.001) in LAO 45°.Increase the angle of projection position will increase the radiation dose rate.Conclusions:We can reduce X-ray radiation dose rates by adjusting the parameter setting of X-ray system.Avoiding oblique projection of large angle is another way to reduce X-ray radiation dose rates. Ablation; Fluoroscopy; Projection Position; Radiation Dose |
PageCount | 3 |
ParticipantIDs | wanfang_journals_zhcmj201509003 chongqing_primary_664768580 |
PublicationCentury | 2000 |
PublicationDate | 2015 |
PublicationDateYYYYMMDD | 2015-01-01 |
PublicationDate_xml | – year: 2015 text: 2015 |
PublicationDecade | 2010 |
PublicationTitle | 中华医学杂志(英文版) |
PublicationTitleAlternate | Chinese Medical Journal |
PublicationTitle_FL | Chinese Medical Journal |
PublicationYear | 2015 |
Publisher | State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China |
Publisher_xml | – name: State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China |
SSID | ssj0025576 |
Score | 2.0131075 |
Snippet | Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology... Background:Nonfluoroscopic three-dimensional electroanatomical system is widely used nowadays,but X-ray remains indispensable for complex electrophysiology... |
SourceID | wanfang chongqing |
SourceType | Aggregation Database Publisher |
StartPage | 1151 |
SubjectTerms | X射线系统 优化设置 位置 参数设置 宾夕法尼亚州 电生理 程序 辐射剂量率 |
Title | 0Optimized Fluoroscopy Setting and Appropriate Project Position Can Reduce X-ray Radiation Doses Rates during Electrophysiology Procedures |
URI | http://lib.cqvip.com/qk/85656X/201509/664768580.html https://d.wanfangdata.com.cn/periodical/zhcmj201509003 |
Volume | 128 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3dT8IwEG8UE-OL8TMqin3wdbit3deTIQgSE5QgJHsj7dqBBjZgkAh_gn-1126iD8SXJct1e7hee_e76_2K0J1nRoGghBuKKt2gTkRhH6TEkJQLZklBPF09b7-4rT59Dp3wtz26UGC2Fdqp-6T683H1c7Z6gAUP8WuVWia5h03YNVyIdKqqLdgLdtEe-CVXQbE23dQUIHb2isplPjovWm79gyJaGKXJcAbuQ3fzJDFLhn8cT_MIHRYRI67lU3yMdmRygvbbRU38FH2Zr7DsJ-9rKXBzvEwVPWU6XeE3qY80Y5YIXFPM4VOwtYXEnTz3gjvFcS1cZwnuKgZXiUNjzla4q_gKtOgxzWQG7xCP4ryhETfye3N0RkSn5LHuNQCpzM5Qr9no1VtGccOCETk-MWKYQZP5gPq4I4RNYsf2PNOVvi0CxQIUWDGjfswj6RAOuEmx2PosYlQElgDHRs5RKUkTeYGwjBkgE0VmBADL4YEfSF8A8ra5JBxgziUqbzQ6mOZEGgPXpZ7ivwfpbaHjwY8FDNajaPJhq4SMSrZe_ft9GR2okXlq5BqVFvOlvIFgYcErGmTD8ym0KtoevgEx_r5P |
linkProvider | Scholars Portal |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=0Optimized+Fluoroscopy+Setting+and+Appropriate+Project+Position+Can+Reduce+X-ray+Radiation+Doses+Rates+during+Electrophysiology+Procedures&rft.jtitle=%E4%B8%AD%E5%8D%8E%E5%8C%BB%E5%AD%A6%E6%9D%82%E5%BF%97%EF%BC%9A%E8%8B%B1%E6%96%87%E7%89%88&rft.au=Bing-Bo+Hou+Yan+Yao+Ling-Min+Wu+Yu+Qiao+Li-Hui+Zheng+Li-Gang+Ding+Gang+Chen&rft.date=2015&rft.issn=0366-6999&rft.eissn=2542-5641&rft.issue=9&rft.spage=1151&rft.epage=1153&rft_id=info:doi/10.4103%2F0366-6999.156079&rft.externalDocID=664768580 |
thumbnail_s | http://utb.summon.serialssolutions.com/2.0.0/image/custom?url=http%3A%2F%2Fimage.cqvip.com%2Fvip1000%2Fqk%2F85656X%2F85656X.jpg http://utb.summon.serialssolutions.com/2.0.0/image/custom?url=http%3A%2F%2Fwww.wanfangdata.com.cn%2Fimages%2FPeriodicalImages%2Fzhcmj%2Fzhcmj.jpg |