Orthogonal image pairs coupled with OSMS for noncoplanar beam angle, intracranial, single-isocenter, SRS treatments with multiple targets on the Varian Edge radiosurgery system

Abstract Purpose To characterize the accuracy of noncoplanar image guided radiation therapy with the Varian Edge radiosurgery system for intracranial stereotactic radiosurgery (SRS) treatments by assessing the accuracy of kV/kV orthogonal pair registration with Optical Surface Monitoring System (OSM...

Full description

Saved in:
Bibliographic Details
Published inAdvances in radiation oncology Vol. 2; no. 3; pp. 494 - 502
Main Authors Oliver, Jasmine A., PhD, Kelly, Patrick, MD, Meeks, Sanford L., PhD, Willoughby, Twyla R., PhD, Shah, Amish P., PhD
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.07.2017
Elsevier
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract Purpose To characterize the accuracy of noncoplanar image guided radiation therapy with the Varian Edge radiosurgery system for intracranial stereotactic radiosurgery (SRS) treatments by assessing the accuracy of kV/kV orthogonal pair registration with Optical Surface Monitoring System (OSMS) monitoring relative to cone beam computed tomography (CT). Methods and materials A Computerized Imaging Reference System head phantom and Encompass SRS Immobilization System were used to determine collision-free space for orthogonal image pairs (kV/kV) for couch rotations (CRs) of 45°, 30°, 15°, 345°, 330°, and 315°. Couch-induced shifts were measured using kV/kV orthogonal image pairs, OSMS, and cone beam CT. The kV/kV image pairs and OSMS localization accuracy was also assessed with respect to cone beam CT. Results Mean orthogonal image pair differences for 315°, 330°, 345°, 15°, 30°, and 45° CRs were ≤±0.60 mm and ±0.37°. OSMS localization accuracy was ≤±0.25 mm and ±0.20°. Correspondingly, kV/kV localization accuracy was ≤±0.30 mm and ±0.5°. Shift differences for various image pairs at all CRs were ≤±1.10 mm and ±0.7°. Cone beam CT deviation was 0.10 mm and 0.00° without patient motion or CR. Conclusion Based on our study, CR-induced shifts with the Varian Edge radiosurgery system will not produce noticeable dosimetric effects for SRS treatments. Thus, replacing cone beam CT with orthogonal kV/kV pairs coupled with OSMS at the treatment couch angle could reduce the number of cone beam CT scans that are acquired during a standard SRS treatment while providing an accurate and safe treatment with negligible dosimetric effects on the treatment plan.
ISSN:2452-1094
2452-1094
DOI:10.1016/j.adro.2017.04.006