Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device
The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month. In this study, participants aged between 21 and 65 years were recruited. The clinical in...
Saved in:
Published in | Blood pressure monitoring |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
England
01.08.2021
|
Online Access | Get more information |
Cover
Loading…
Abstract | The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month.
In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject.
Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP.
After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population. |
---|---|
AbstractList | The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month.
In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject.
Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP.
After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population. |
Author | Vybornova, Anna Fallet, Sibylle Wurzner-Ghajarzadeh, Arlène Polychronopoulou, Erietta Sola, Josep Wuerzner, Gregoire |
Author_xml | – sequence: 1 givenname: Anna surname: Vybornova fullname: Vybornova, Anna organization: Aktiia SA, Neuchâtel (www.aktiia.com) Service of Nephrology and Hypertension, Lausanne University Hospital and University of Lausanne, CHUV, Lausanne, Switzerland – sequence: 2 givenname: Erietta surname: Polychronopoulou fullname: Polychronopoulou, Erietta – sequence: 3 givenname: Arlène surname: Wurzner-Ghajarzadeh fullname: Wurzner-Ghajarzadeh, Arlène – sequence: 4 givenname: Sibylle surname: Fallet fullname: Fallet, Sibylle – sequence: 5 givenname: Josep surname: Sola fullname: Sola, Josep – sequence: 6 givenname: Gregoire surname: Wuerzner fullname: Wuerzner, Gregoire |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/33675592$$D View this record in MEDLINE/PubMed |
BookMark | eNpNUO9KwzAcDKK4P_oGIr8X6EzSpln9to2pg8kE9fPIv7po25Qkne5BfF8rKngc3Ic77uBG6LhxjUHoguAJwQW_up8_TPB_sJQcoSHJeJowTtkAjUJ4xZjkU85P0SBNc85YQYfoc145p6H1JoTOGyi9qyHuDLg2WiUqmL1FawXMvVCmMvEaBJCkdk3cwV5UVotoXQMhdvoAXbDNC4gGzEc0jTYaVo-bKcE5Tmg_4aJTrgKhRRt7r3S-L1NdWVb9OLx7GyJos7fKnKGTUlTBnP_qGD3fLJ8Wd8l6c7tazNaJYqQgiZ5qibmiWZlzUjLCmVJU5EYwraXsWXCqSJbhTMtCappzJVIpv0OCZUTTMbr86W07WRu9bb2thT9s__6hX7dDaog |
CitedBy_id | crossref_primary_10_1097_HJH_0000000000003960 crossref_primary_10_1146_annurev_bioeng_110220_014644 crossref_primary_10_1097_HJH_0000000000003224 crossref_primary_10_3390_s25030640 crossref_primary_10_1109_TBME_2023_3275031 crossref_primary_10_1097_HJH_0000000000003483 crossref_primary_10_1038_s41440_023_01258_2 crossref_primary_10_1055_a_1871_0043 crossref_primary_10_17116_profmed202427121156 crossref_primary_10_1097_HJH_0000000000003827 crossref_primary_10_1080_07435800_2024_2341146 crossref_primary_10_3390_mi14061218 crossref_primary_10_3389_fmedt_2024_1464473 crossref_primary_10_1088_1361_6579_ad33a2 crossref_primary_10_1038_s41440_023_01378_9 crossref_primary_10_1097_HJH_0000000000003428 crossref_primary_10_3390_bios13030321 crossref_primary_10_1007_s12553_023_00726_6 crossref_primary_10_1161_HYPERTENSIONAHA_122_20410 crossref_primary_10_1016_j_ahj_2023_07_005 crossref_primary_10_1016_j_mayocpiqo_2024_06_001 crossref_primary_10_3390_medicina59020404 crossref_primary_10_1080_08037051_2021_1956178 crossref_primary_10_2196_44147 crossref_primary_10_3389_fpubh_2025_1510456 crossref_primary_10_1097_CRD_0000000000000875 crossref_primary_10_3389_fmed_2023_1138051 crossref_primary_10_1080_08037051_2023_2274595 crossref_primary_10_1097_HJH_0000000000003274 crossref_primary_10_1161_HYPERTENSIONAHA_121_17747 crossref_primary_10_2196_57158 crossref_primary_10_4070_kcj_2021_0405 crossref_primary_10_1038_s41598_021_99294_w crossref_primary_10_1097_MBP_0000000000000639 crossref_primary_10_3389_fmed_2024_1396962 crossref_primary_10_1093_ajh_hpae154 crossref_primary_10_3390_bioengineering10010027 crossref_primary_10_1093_ajh_hpaf003 crossref_primary_10_1016_S2589_7500_23_00087_0 crossref_primary_10_1038_s41440_023_01370_3 crossref_primary_10_1186_s40885_024_00268_7 crossref_primary_10_1016_j_isci_2023_108295 crossref_primary_10_1038_s41371_022_00675_z crossref_primary_10_1038_s41440_023_01215_z |
ContentType | Journal Article |
Copyright | Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved. |
Copyright_xml | – notice: Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved. |
DBID | NPM |
DOI | 10.1097/MBP.0000000000000531 |
DatabaseName | PubMed |
DatabaseTitle | PubMed |
DatabaseTitleList | PubMed |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database |
DeliveryMethod | no_fulltext_linktorsrc |
Discipline | Medicine Anatomy & Physiology |
EISSN | 1473-5725 |
ExternalDocumentID | 33675592 |
Genre | Journal Article |
GroupedDBID | --- .Z2 0R~ 23N 4Q1 4Q2 4Q3 5GY 5VS 6J9 6PF 8L- AAAAV AAHPQ AAIQE AARTV AASCR AAWTL AAYEP ABASU ABBUW ABDIG ABJNI ABOCM ABPXF ABVCZ ABXVJ ABZAD ABZZY ACDDN ACEWG ACGFO ACGFS ACILI ACWDW ACWRI ACXJB ACXNZ ADGGA ADHPY AENEX AFBFQ AFDTB AHQNM AHVBC AINUH AJCLO AJIOK AJNWD AJZMW AKCTQ ALKUP ALMA_UNASSIGNED_HOLDINGS ALMTX AMJPA AMKUR AMNEI AOHHW AOQMC BQLVK C45 CS3 DIWNM DU5 E.X EBS EEVPB EX3 F5P FCALG FL- GNXGY GQDEL H0~ HLJTE HZ~ IKREB IN~ IPNFZ JK3 KD2 L-C NPM O9- OAG OAH OPUJH OVD OVDNE OWW OWY OXXIT P2P RIG RLZ S4R S4S TEORI TSPGW V2I VVN W3M WOQ WOW X3V X3W YFH ZFV |
ID | FETCH-LOGICAL-c5191-d8db07c24f671f5175cc2a6ea5ddbbdbb972c14404db9bd267ca3bbcc2aa541d2 |
IngestDate | Mon Jul 21 06:03:25 EDT 2025 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Language | English |
License | Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved. |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c5191-d8db07c24f671f5175cc2a6ea5ddbbdbb972c14404db9bd267ca3bbcc2aa541d2 |
OpenAccessLink | https://pubmed.ncbi.nlm.nih.gov/PMC8248249 |
PMID | 33675592 |
ParticipantIDs | pubmed_primary_33675592 |
PublicationCentury | 2000 |
PublicationDate | 2021-August-01 |
PublicationDateYYYYMMDD | 2021-08-01 |
PublicationDate_xml | – month: 08 year: 2021 text: 2021-August-01 day: 01 |
PublicationDecade | 2020 |
PublicationPlace | England |
PublicationPlace_xml | – name: England |
PublicationTitle | Blood pressure monitoring |
PublicationTitleAlternate | Blood Press Monit |
PublicationYear | 2021 |
SSID | ssj0016877 |
Score | 2.5203269 |
Snippet | The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP)... |
SourceID | pubmed |
SourceType | Index Database |
Title | Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device |
URI | https://www.ncbi.nlm.nih.gov/pubmed/33675592 |
hasFullText | |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1ba9RAFB62CqUvoq33C-dBfAmjm-tkfduKtQrVQlvoWzkzSejW3WSJCbL7P_wv_jzP3HZrraKGJSyZMAw5H-cy851zGHteDMNKpKg4ppHk5FIPOSKWPCHvVeZVrESss5EPPmb7J8mH0_R0MPh-ibXUd_KlWl6bV_I_UqVnJFedJfsPkl1NSg_oP8mX7iRhuv-VjHc16zwwVFZ9DrBKFWnmdod6_LmbTJDkh4qsS2cTm0NO6-vOA1rIxDZUsjVmg_6LTVgM_MZ48P7oU07B3JBHmsfVNQSaAAucay_VsC8D1VfVVCvLr1pZ6BQsx6VbHRT_vMSZUSGtt5e609eCUKg7s1p6Zb0yE4fNlHRz29TNvOmnTe-0dtl1a0vSt8u6bPm7c7zAdolFaTaJxu3UnP_na8rAnm4ZY-zN0UQupu4zut2OKFxx7chYWQ2diJinwmZL_6L_bV3hg91DW5fSX6m1NJcgMZ8ZTMQxBUyp7cb359ErVbn90AbboPhEN1zVu0Tu9CrLhfBpmiPx6rrlbLFNP8WVgMY4Nse32S0XkcDYwusOG5T1NtsZ19g1swW8AMMRNocv22zzwFExdtg3I1nwkgUNPiDwgQMfWPCBB99rQHDQgzX0wEAPDPQAa_DQgzX0wEMPHPSAoEeTeeiBgR5Y6N1lJ3tvj9_sc9fjgyuKHUJe5IUcChUlVSbCKiVnVqkIsxLTopCSfiMRKU1ASAo5kkWUCYWxlPolTJOwiO6xG3VTlw909YE4EmVMHmwmyS1WssqTiuYoMUlwJIcP2X37nc_mtpDLmZfAo9-OPGZbaxQ-YTcr0hzlU3JDO_nMyPwHYX-LeA |
linkProvider | National Library of Medicine |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Blood+pressure+from+the+optical+Aktiia+Bracelet%3A+a+1-month+validation+study+using+an+extended+ISO81060-2+protocol+adapted+for+a+cuffless+wrist+device&rft.jtitle=Blood+pressure+monitoring&rft.au=Vybornova%2C+Anna&rft.au=Polychronopoulou%2C+Erietta&rft.au=Wurzner-Ghajarzadeh%2C+Arl%C3%A8ne&rft.au=Fallet%2C+Sibylle&rft.date=2021-08-01&rft.eissn=1473-5725&rft_id=info:doi/10.1097%2FMBP.0000000000000531&rft_id=info%3Apmid%2F33675592&rft_id=info%3Apmid%2F33675592&rft.externalDocID=33675592 |