Novel Technology Allowing Cone Beam Computed Tomography in 6 Seconds: A Patient Study of Comparative Image Quality

The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical Systems), a platform with enhanced reconstruction algorithms as well as rapid acquisition times. Image quality was compared with both status quo CBCT...

Full description

Saved in:
Bibliographic Details
Published inPractical radiation oncology Vol. 14; no. 3; p. 277
Main Authors Robar, James L, Cherpak, Amanda, MacDonald, Robert Lee, Yashayaeva, Abigail, McAloney, David, McMaster, Natasha, Zhan, Kenny, Cwajna, Slawa, Patil, Nikhilesh, Dahn, Hannah
Format Journal Article
LanguageEnglish
Published United States 01.05.2024
Subjects
Online AccessGet more information
ISSN1879-8519
DOI10.1016/j.prro.2023.10.014

Cover

Loading…
Abstract The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical Systems), a platform with enhanced reconstruction algorithms as well as rapid acquisition times. Image quality was compared with both status quo CBCT for image guidance, and to fan beam CT (FBCT) acquired on a CT simulator (CTsim). In a clinical study, 30 individuals were recruited for whom either deep inspiration (DIBH) or deep exhalation breath hold (DEBH) was used during imaging and radiation treatment of tumors involving liver, lung, breast, abdomen, chest wall, and pancreatic sites. All subjects were imaged during breath hold with CBCT on a standard image guidance platform (TrueBeam 2.7, Varian Medical Systems) and FBCT CT (CTsim, GE Optima). HyperSight imaging with both breath hold (HSBH) and free breathing (HSFB) was performed in a single session. The 4 image sets thus acquired were registered and compared using metrics quantifying artifact index, image nonuniformity, contrast, contrast-to-noise ratio, and difference of Hounsfield unit (HU) from CTsim. HSBH provided less severe artifacts compared with both HSFB and TrueBeam. The severity of artifacts in HSBH images was similar to that in CTsim images, with statistically similar artifact index values. CTsim provided the best image uniformity; however, HSBH provided improved uniformity compared with both HSFB and TrueBeam. CTsim demonstrated elevated contrast compared with HyperSight imaging, but both HSBH and HSFB imaging showed superior contrast-to-noise ratio characteristics compared with TrueBeam. The median HU difference of HSBH from CTsim was within 1 HU for muscle/fat tissue, 12 HU for bone, and 14 HU for lung. The HyperSight system provides 6-second CBCT acquisition with image artifacts that are significantly reduced compared with TrueBeam and comparable to those in CTsim FBCT imaging. HyperSight breath hold imaging was of higher quality compared with free breathing imaging on the same system. The median HU value in HyperSight breath hold imaging is within 15 HU of that in CTsim imaging for muscle, fat, bone, and lung tissue types, indicating the utility of image data for direct dose calculation in adaptive workflows.
AbstractList The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical Systems), a platform with enhanced reconstruction algorithms as well as rapid acquisition times. Image quality was compared with both status quo CBCT for image guidance, and to fan beam CT (FBCT) acquired on a CT simulator (CTsim). In a clinical study, 30 individuals were recruited for whom either deep inspiration (DIBH) or deep exhalation breath hold (DEBH) was used during imaging and radiation treatment of tumors involving liver, lung, breast, abdomen, chest wall, and pancreatic sites. All subjects were imaged during breath hold with CBCT on a standard image guidance platform (TrueBeam 2.7, Varian Medical Systems) and FBCT CT (CTsim, GE Optima). HyperSight imaging with both breath hold (HSBH) and free breathing (HSFB) was performed in a single session. The 4 image sets thus acquired were registered and compared using metrics quantifying artifact index, image nonuniformity, contrast, contrast-to-noise ratio, and difference of Hounsfield unit (HU) from CTsim. HSBH provided less severe artifacts compared with both HSFB and TrueBeam. The severity of artifacts in HSBH images was similar to that in CTsim images, with statistically similar artifact index values. CTsim provided the best image uniformity; however, HSBH provided improved uniformity compared with both HSFB and TrueBeam. CTsim demonstrated elevated contrast compared with HyperSight imaging, but both HSBH and HSFB imaging showed superior contrast-to-noise ratio characteristics compared with TrueBeam. The median HU difference of HSBH from CTsim was within 1 HU for muscle/fat tissue, 12 HU for bone, and 14 HU for lung. The HyperSight system provides 6-second CBCT acquisition with image artifacts that are significantly reduced compared with TrueBeam and comparable to those in CTsim FBCT imaging. HyperSight breath hold imaging was of higher quality compared with free breathing imaging on the same system. The median HU value in HyperSight breath hold imaging is within 15 HU of that in CTsim imaging for muscle, fat, bone, and lung tissue types, indicating the utility of image data for direct dose calculation in adaptive workflows.
Author Dahn, Hannah
Patil, Nikhilesh
MacDonald, Robert Lee
Cwajna, Slawa
Robar, James L
McAloney, David
Cherpak, Amanda
McMaster, Natasha
Yashayaeva, Abigail
Zhan, Kenny
Author_xml – sequence: 1
  givenname: James L
  surname: Robar
  fullname: Robar, James L
  email: james.robar@nshealth.ca
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology; Physics and Atmospheric Science, Dalhousie University, Halifax, Canada. Electronic address: james.robar@nshealth.ca
– sequence: 2
  givenname: Amanda
  surname: Cherpak
  fullname: Cherpak, Amanda
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology; Physics and Atmospheric Science, Dalhousie University, Halifax, Canada
– sequence: 3
  givenname: Robert Lee
  surname: MacDonald
  fullname: MacDonald, Robert Lee
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology; Physics and Atmospheric Science, Dalhousie University, Halifax, Canada
– sequence: 4
  givenname: Abigail
  surname: Yashayaeva
  fullname: Yashayaeva, Abigail
  organization: Physics and Atmospheric Science, Dalhousie University, Halifax, Canada
– sequence: 5
  givenname: David
  surname: McAloney
  fullname: McAloney, David
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada
– sequence: 6
  givenname: Natasha
  surname: McMaster
  fullname: McMaster, Natasha
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada
– sequence: 7
  givenname: Kenny
  surname: Zhan
  fullname: Zhan, Kenny
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada
– sequence: 8
  givenname: Slawa
  surname: Cwajna
  fullname: Cwajna, Slawa
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology
– sequence: 9
  givenname: Nikhilesh
  surname: Patil
  fullname: Patil, Nikhilesh
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology
– sequence: 10
  givenname: Hannah
  surname: Dahn
  fullname: Dahn, Hannah
  organization: Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Canada; Departments of Radiation Oncology
BackLink https://www.ncbi.nlm.nih.gov/pubmed/37939844$$D View this record in MEDLINE/PubMed
BookMark eNo1j9tKw0AURQdRtFZ_wAc5P5A4l2Qm41st3qB4ofW5zExO0kiSCdOkkr83eHnamwV7wT4nx61vkZArRmNGmbz5jLsQfMwpFxOIKUuOyIxlSkdZyvQpORNKC50lyYyEF3_AGjbodq2vfTnCoq79V9WWsJykcIemmVrTDT3msPGNL4PpdiNULUhYo_Ntvr-FBbyZvsK2h3U_5CP44mdkwkQPCM-NKRHeB1NX_XhBTgpT7_HyL-fk4-F-s3yKVq-Pz8vFKnIplX3E0RmFRjnUQlqUVCRaOyVtqilX1sjUFVQ5JtKCCkwVJs5xaqy2mVOWF3xOrn-93WAbzLddqBoTxu3_ef4NqllbZQ
CitedBy_id crossref_primary_10_37549_ARO_D_24_00021
crossref_primary_10_1002_mp_17721
crossref_primary_10_1088_1361_6560_ad4a1c
crossref_primary_10_3390_cancers16223770
crossref_primary_10_1002_mp_17549
crossref_primary_10_1016_j_phro_2024_100687
crossref_primary_10_1016_j_adro_2024_101692
crossref_primary_10_1016_j_phro_2024_100640
crossref_primary_10_1016_j_ijrobp_2024_03_039
crossref_primary_10_1002_acm2_70023
crossref_primary_10_1002_acm2_70060
crossref_primary_10_1016_j_neuroimage_2024_120922
crossref_primary_10_1088_2057_1976_ad3bb9
crossref_primary_10_7759_cureus_68386
crossref_primary_10_1016_j_ejmp_2025_104946
crossref_primary_10_7759_cureus_66943
crossref_primary_10_1002_acm2_14337
crossref_primary_10_1016_j_yao_2025_02_003
crossref_primary_10_1016_j_phro_2024_100611
crossref_primary_10_7759_cureus_75284
crossref_primary_10_1002_acm2_14438
crossref_primary_10_1016_j_adro_2024_101681
crossref_primary_10_1016_j_clon_2024_05_018
crossref_primary_10_1016_j_radonc_2024_110178
crossref_primary_10_1002_acm2_14464
crossref_primary_10_1002_acm2_70038
crossref_primary_10_7759_cureus_66876
crossref_primary_10_1016_j_semradonc_2024_07_005
crossref_primary_10_1088_1361_6560_ad311b
crossref_primary_10_37549_ARO_D_24_00037
crossref_primary_10_7759_cureus_68990
ContentType Journal Article
Copyright Copyright © 2023 The Author(s). Published by Elsevier Inc. All rights reserved.
Copyright_xml – notice: Copyright © 2023 The Author(s). Published by Elsevier Inc. All rights reserved.
DBID CGR
CUY
CVF
ECM
EIF
NPM
DOI 10.1016/j.prro.2023.10.014
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
DatabaseTitleList MEDLINE
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod no_fulltext_linktorsrc
EISSN 1879-8519
ExternalDocumentID 37939844
Genre Journal Article
Comparative Study
GroupedDBID CGR
CUY
CVF
ECM
EIF
NPM
ID FETCH-LOGICAL-c506t-2eca7ea7ce936be603499c76b59027ba65cf07c135f03e57e4cc20ab9b8c7b2f2
IngestDate Thu Jan 02 22:45:08 EST 2025
IsDoiOpenAccess false
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 3
Language English
License Copyright © 2023 The Author(s). Published by Elsevier Inc. All rights reserved.
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c506t-2eca7ea7ce936be603499c76b59027ba65cf07c135f03e57e4cc20ab9b8c7b2f2
OpenAccessLink https://dx.doi.org/10.1016/j.prro.2023.10.014
PMID 37939844
ParticipantIDs pubmed_primary_37939844
PublicationCentury 2000
PublicationDate 2024-05-01
PublicationDateYYYYMMDD 2024-05-01
PublicationDate_xml – month: 05
  year: 2024
  text: 2024-05-01
  day: 01
PublicationDecade 2020
PublicationPlace United States
PublicationPlace_xml – name: United States
PublicationTitle Practical radiation oncology
PublicationTitleAlternate Pract Radiat Oncol
PublicationYear 2024
Score 2.4915156
Snippet The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical...
SourceID pubmed
SourceType Index Database
StartPage 277
SubjectTerms Adult
Algorithms
Breath Holding
Cone-Beam Computed Tomography - methods
Female
Humans
Image Processing, Computer-Assisted - methods
Male
Middle Aged
Neoplasms - diagnostic imaging
Neoplasms - radiotherapy
Title Novel Technology Allowing Cone Beam Computed Tomography in 6 Seconds: A Patient Study of Comparative Image Quality
URI https://www.ncbi.nlm.nih.gov/pubmed/37939844
Volume 14
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1db9MwFLU6kNBeEIjvL_mBtypVFjt2zFuZQAOxaUKbNJ6ma-8aNjVp1ZUh-E38SK7tpE0LSIOXKIqVNMo9ub0-OfeYsZeVBistYia1MpmEnSIDrILWFTAvK6D3L6p8D9TesXx_Up4MBj97qqWvCztyP_7YV_I_UaVjFNfQJfsPkV1elA7QPsWXthRh2l4rxgfTK5z02PHheDKZfktf8ql4fI1QD9tlG4Jgpm7tqQPHoQLTPm3OLlNr-mGyV42ywiTR6LmCv6uDsCe5bax9BU5mRyHK82BxkIrPxq0x9R8pYcyXetzhkmveJbDMIObicQ09YmAfeoR1En4PP-ASf5_g8gt8B7xKhLA9_9ypRFruopArpeAIU76ttMmo6DNrCVn2gCf62TWt-PJb1k8ExAX958xDP2chRkGwly7Tg8GsjjgQlJJMJa8xuuHE3Q1tsS2ak4RFVg_32z6sJBnc_P1tdqs7Z2PWEquXozvsdjvt4OOEobtsgM09No_44Sv88A4_POCHB_zwDj98hR9-3nDFW_y84mPeoodH9PCp5z308Ige3qLnPjt---Zody9r1-DIXJmrRVagA42gHRqhLKpgZ2ScVjbY_mgLqnQ-125HlD4XWGqUzhU5WGMrp23hiwfsRkO3_IhxVMZV3oDSZ14K4YGShzUF-BJQ-FI-Zg_TIzqdJaOV0-7hPfnryFO2vcLVM3bT05uNz6lMXNgXMT6_AN4UbNI
linkProvider National Library of Medicine
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Novel+Technology+Allowing+Cone+Beam+Computed+Tomography+in+6+Seconds%3A+A+Patient+Study+of+Comparative+Image+Quality&rft.jtitle=Practical+radiation+oncology&rft.au=Robar%2C+James+L&rft.au=Cherpak%2C+Amanda&rft.au=MacDonald%2C+Robert+Lee&rft.au=Yashayaeva%2C+Abigail&rft.date=2024-05-01&rft.eissn=1879-8519&rft.volume=14&rft.issue=3&rft.spage=277&rft_id=info:doi/10.1016%2Fj.prro.2023.10.014&rft_id=info%3Apmid%2F37939844&rft_id=info%3Apmid%2F37939844&rft.externalDocID=37939844