A comparative evaluation of the CN‐6000 haemostasis analyser using coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry assays
Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. Aims To...
Saved in:
Published in | International journal of laboratory hematology Vol. 42; no. 5; pp. 643 - 649 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
England
Wiley Subscription Services, Inc
01.10.2020
John Wiley and Sons Inc |
Subjects | |
Online Access | Get full text |
ISSN | 1751-5521 1751-553X 1751-553X |
DOI | 10.1111/ijlh.13271 |
Cover
Loading…
Abstract | Background
The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.
Aims
To evaluate the performance of the CN‐6000 against a predicate device.
Methods
The CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).
Results
Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34).
Conclusions
The CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown. |
---|---|
AbstractList | The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.BACKGROUNDThe CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.To evaluate the performance of the CN-6000 against a predicate device.AIMSTo evaluate the performance of the CN-6000 against a predicate device.The CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).METHODSThe CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34).RESULTSAcceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34).The CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.CONCLUSIONSThe CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown. The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. To evaluate the performance of the CN-6000 against a predicate device. The CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34). The CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown. BackgroundThe CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.AimsTo evaluate the performance of the CN‐6000 against a predicate device.MethodsThe CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).ResultsAcceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34).ConclusionsThe CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown. Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. Aims To evaluate the performance of the CN‐6000 against a predicate device. Methods The CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). Results Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34). Conclusions The CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown. |
Author | Mackie, Ian J. Langley, Katy Gardiner, Chris Machin, Samuel J. Tailor, Hitesh Lane, Philip |
AuthorAffiliation | 1 Haemostasis Research Unit University College London London UK 2 Haematology Evaluations Unit HSL (Analytics) LLP London UK |
AuthorAffiliation_xml | – name: 1 Haemostasis Research Unit University College London London UK – name: 2 Haematology Evaluations Unit HSL (Analytics) LLP London UK |
Author_xml | – sequence: 1 givenname: Chris orcidid: 0000-0002-2318-0062 surname: Gardiner fullname: Gardiner, Chris email: chris.gardiner@uclmail.net organization: University College London – sequence: 2 givenname: Philip surname: Lane fullname: Lane, Philip organization: University College London – sequence: 3 givenname: Katy surname: Langley fullname: Langley, Katy organization: University College London – sequence: 4 givenname: Hitesh orcidid: 0000-0003-3615-7889 surname: Tailor fullname: Tailor, Hitesh organization: HSL (Analytics) LLP – sequence: 5 givenname: Samuel J. surname: Machin fullname: Machin, Samuel J. organization: University College London – sequence: 6 givenname: Ian J. orcidid: 0000-0002-7393-2365 surname: Mackie fullname: Mackie, Ian J. organization: University College London |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/32885901$$D View this record in MEDLINE/PubMed |
BookMark | eNp9ks2OFCEUhYkZ4_zoxgcwJG6MmR75KYqqjcmko86Yjm40cUcoiqqiQ0ELVJva-Qi-ge_mk0h3z0x0YmQDge8czoV7Co6cdxqApxhd4DxembUdLjAlHD8AJ5gzvGCMfjm6WxN8DE5jXCPEeIHqR-CYkqpiNcIn4OclVH7cyCCT2Wqot9JOeekd9B1Mg4bLD7--_ygRQnCQevQxyWgilE7aOeoAp2hcny1kP9m97hzK0bTezsmoc2jGcXI-O6QpNHl71CkYleUttKYfEkxBujiaGHdXyr4Put9DM5Qxyjk-Bg87aaN-cjOfgc9v33xaXi1WH99dLy9XC1UUFV7UlDPKFZdllWtuq1YipQrCUUcRJZqVitCyUXVFmpY3UlWMY6YL0mmO6w5X9Ay8PvhupmbUrdIuJ7NiE8wowyy8NOLvE2cG0futqEmNGauzwYsbg-C_TjomkatS2lrptJ-iIEWBirIsKpLR5_fQtZ9CftE9VdaM03KX6Nmfie6i3P5dBtABUMHHGHQnlEn7P8gBjRUYiV17iF17iH17ZMnLe5Jb13_C-AB_M1bP_yHF9fvV1UHzGzHa0HU |
CitedBy_id | crossref_primary_10_1111_ijlh_13740 crossref_primary_10_1097_MBC_0000000000001184 crossref_primary_10_1111_ijlh_13656 crossref_primary_10_1111_ijlh_13468 |
Cites_doi | 10.1515/cclm-2013-1094 10.1097/MD.0000000000007976 10.1532/LH96.05031 10.1055/s-0035-1544196 10.1111/ijlh.13129 10.1016/j.plabm.2016.09.002 10.1111/jth.12642 10.1111/ijlh.12498 |
ContentType | Journal Article |
Copyright | 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd. 2020. This article is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. 2020 The Authors. published by John Wiley & Sons Ltd. |
Copyright_xml | – notice: 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd – notice: 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd. – notice: 2020. This article is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. – notice: 2020 The Authors. published by John Wiley & Sons Ltd. |
DBID | 24P AAYXX CITATION CGR CUY CVF ECM EIF NPM 7T5 H94 K9. NAPCQ 7X8 5PM |
DOI | 10.1111/ijlh.13271 |
DatabaseName | Wiley Online Library Open Access CrossRef Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed Immunology Abstracts AIDS and Cancer Research Abstracts ProQuest Health & Medical Complete (Alumni) Nursing & Allied Health Premium MEDLINE - Academic PubMed Central (Full Participant titles) |
DatabaseTitle | CrossRef MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) AIDS and Cancer Research Abstracts ProQuest Health & Medical Complete (Alumni) Nursing & Allied Health Premium Immunology Abstracts MEDLINE - Academic |
DatabaseTitleList | MEDLINE - Academic MEDLINE AIDS and Cancer Research Abstracts |
Database_xml | – sequence: 1 dbid: 24P name: Wiley Online Library Open Access url: https://authorservices.wiley.com/open-science/open-access/browse-journals.html sourceTypes: Publisher – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
DocumentTitleAlternate | GARDINER et al |
EISSN | 1751-553X |
EndPage | 649 |
ExternalDocumentID | PMC9291559 32885901 10_1111_ijlh_13271 IJLH13271 |
Genre | article Journal Article |
GroupedDBID | --- .3N .GA .Y3 05W 0R~ 10A 1OC 24P 29J 31~ 33P 36B 3SF 4.4 50Y 50Z 51W 51X 52M 52N 52O 52P 52R 52S 52T 52U 52V 52W 52X 53G 5GY 5HH 5LA 5VS 66C 702 7PT 8-0 8-1 8-3 8-4 8-5 8UM 930 A01 A03 AAESR AAEVG AAHHS AAHQN AAIPD AAMNL AANHP AANLZ AAONW AASGY AAWTL AAXRX AAYCA AAZKR ABCQN ABCUV ABDBF ABEML ABJNI ABPVW ABQWH ABXGK ACAHQ ACBWZ ACCFJ ACCZN ACGFS ACGOF ACMXC ACPOU ACPRK ACRPL ACSCC ACUHS ACXBN ACXQS ACYXJ ADBBV ADBTR ADEOM ADIZJ ADKYN ADMGS ADNMO ADOZA ADXAS ADZMN ADZOD AEEZP AEIGN AEIMD AENEX AEQDE AEUQT AEUYR AFBPY AFFPM AFGKR AFPWT AFWVQ AFZJQ AHBTC AIACR AITYG AIURR AIWBW AJBDE ALAGY ALMA_UNASSIGNED_HOLDINGS ALUQN ALVPJ AMBMR AMYDB ATUGU AZBYB AZFZN AZVAB BAFTC BDRZF BFHJK BHBCM BMXJE BROTX BRXPI BY8 C45 CAG COF CS3 D-6 D-7 D-E D-F DCZOG DPXWK DR2 DRFUL DRMAN DRSTM DU5 EAD EAP EAS EBC EBD EBS EBX EJD EMB EMK EMOBN EPT ESX EX3 F00 F01 F04 F5P FEDTE FUBAC G-S G.N GODZA H.X HF~ HGLYW HVGLF HZI HZ~ IHE IX1 J0M KBYEO L7B LATKE LC2 LC3 LEEKS LH4 LITHE LOXES LP6 LP7 LUTES LW6 LYRES MEWTI MK4 MRFUL MRMAN MRSTM MSFUL MSMAN MSSTM MXFUL MXMAN MXSTM N04 N05 N9A NF~ O66 O9- OIG OVD P2P P2W P2X P2Z P4B P4D Q.N Q11 QB0 Q~Q R.K ROL RX1 SUPJJ SV3 TEORI TUS UB1 W8V W99 WBKPD WHWMO WIH WIJ WIK WOHZO WOW WQJ WRC WVDHM WXI WXSBR XG1 ~IA ~WT AAYXX AEYWJ AGHNM AGQPQ AGYGG CITATION CGR CUY CVF ECM EIF NPM 1OB 7T5 AAMMB AEFGJ AGXDD AIDQK AIDYY H94 K9. NAPCQ 7X8 5PM |
ID | FETCH-LOGICAL-c4481-937537c7a68751d8da0cc4270f3032e56c236bc982bd7bac85715e42fe719f183 |
IEDL.DBID | DR2 |
ISSN | 1751-5521 1751-553X |
IngestDate | Thu Aug 21 14:11:39 EDT 2025 Tue Aug 05 10:35:47 EDT 2025 Wed Aug 13 06:25:20 EDT 2025 Wed Feb 19 02:29:31 EST 2025 Tue Jul 01 03:54:10 EDT 2025 Thu Apr 24 23:03:28 EDT 2025 Wed Jan 22 16:59:57 EST 2025 |
IsDoiOpenAccess | true |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 5 |
Keywords | coagulation laboratory automation laboratory practice |
Language | English |
License | Attribution 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-c4481-937537c7a68751d8da0cc4270f3032e56c236bc982bd7bac85715e42fe719f183 |
Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
ORCID | 0000-0002-2318-0062 0000-0002-7393-2365 0000-0003-3615-7889 |
OpenAccessLink | https://proxy.k.utb.cz/login?url=https://onlinelibrary.wiley.com/doi/abs/10.1111%2Fijlh.13271 |
PMID | 32885901 |
PQID | 2446957368 |
PQPubID | 2045143 |
PageCount | 7 |
ParticipantIDs | pubmedcentral_primary_oai_pubmedcentral_nih_gov_9291559 proquest_miscellaneous_2440466482 proquest_journals_2446957368 pubmed_primary_32885901 crossref_citationtrail_10_1111_ijlh_13271 crossref_primary_10_1111_ijlh_13271 wiley_primary_10_1111_ijlh_13271_IJLH13271 |
ProviderPackageCode | CITATION AAYXX |
PublicationCentury | 2000 |
PublicationDate | October 2020 |
PublicationDateYYYYMMDD | 2020-10-01 |
PublicationDate_xml | – month: 10 year: 2020 text: October 2020 |
PublicationDecade | 2020 |
PublicationPlace | England |
PublicationPlace_xml | – name: England – name: Chichester – name: Hoboken |
PublicationTitle | International journal of laboratory hematology |
PublicationTitleAlternate | Int J Lab Hematol |
PublicationYear | 2020 |
Publisher | Wiley Subscription Services, Inc John Wiley and Sons Inc |
Publisher_xml | – name: Wiley Subscription Services, Inc – name: John Wiley and Sons Inc |
References | 2008 2016; 6 2017; 96 2014; 52 2006; 12 2020; 42 2016; 38 2015; 61 2014; 12 2015; 41 e_1_2_7_6_1 e_1_2_7_11_1 e_1_2_7_5_1 e_1_2_7_10_1 e_1_2_7_3_1 e_1_2_7_9_1 e_1_2_7_8_1 e_1_2_7_7_1 Chen L (e_1_2_7_4_1) 2015; 61 e_1_2_7_2_1 |
References_xml | – volume: 96 issue: 38 year: 2017 article-title: The effects of hemolysis on plasma prothrombin time and activated partial thromboplastin time tests using photo‐optical method publication-title: Medicine (Baltimore) – volume: 52 start-page: 1193 issue: 8 year: 2014 end-page: 1202 article-title: Evaluation of the automated coagulation analyzer CS‐5100 and its utility in high throughput laboratories publication-title: Clin Chem Lab Med – volume: 12 start-page: 1591 issue: 10 year: 2014 article-title: Iatrogenic anemia (can it be prevented?) publication-title: J Thromb Haemost – volume: 38 start-page: 375 issue: 4 year: 2016 end-page: 388 article-title: Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA‐Compact‐Max analyser publication-title: Int J Lab Hematol – volume: 42 start-page: 140 issue: 2 year: 2020 end-page: 144 article-title: A practical method for reducing the interference due to lipaemia in coagulation tests publication-title: Int J Lab Hematol – volume: 12 start-page: 32 issue: 1 year: 2006 end-page: 38 article-title: Adcock DM. Recommendations for evaluation of coagulation analyzers publication-title: Lab Hematol – volume: 41 start-page: 294 issue: 3 year: 2015 end-page: 299 article-title: The changing face of hemostasis testing in modern laboratories: consolidation, automation, and beyond publication-title: Semin Thromb Hemost – year: 2008 – volume: 61 start-page: 653 issue: 7 year: 2015 end-page: 660 article-title: Performance evaluation of the sysmex CS‐5100 automated coagulation analyzer publication-title: Clin Lab – volume: 6 start-page: 38 year: 2016 end-page: 47 article-title: The Sysmex CS‐5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities publication-title: Pract Lab Med – ident: e_1_2_7_6_1 doi: 10.1515/cclm-2013-1094 – ident: e_1_2_7_9_1 doi: 10.1097/MD.0000000000007976 – volume: 61 start-page: 653 issue: 7 year: 2015 ident: e_1_2_7_4_1 article-title: Performance evaluation of the sysmex CS‐5100 automated coagulation analyzer publication-title: Clin Lab – ident: e_1_2_7_8_1 doi: 10.1532/LH96.05031 – ident: e_1_2_7_7_1 – ident: e_1_2_7_2_1 doi: 10.1055/s-0035-1544196 – ident: e_1_2_7_11_1 doi: 10.1111/ijlh.13129 – ident: e_1_2_7_5_1 doi: 10.1016/j.plabm.2016.09.002 – ident: e_1_2_7_3_1 doi: 10.1111/jth.12642 – ident: e_1_2_7_10_1 doi: 10.1111/ijlh.12498 |
SSID | ssj0057409 |
Score | 2.2658095 |
Snippet | Background
The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission... The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA)... BackgroundThe CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry... |
SourceID | pubmedcentral proquest pubmed crossref wiley |
SourceType | Open Access Repository Aggregation Database Index Database Enrichment Source Publisher |
StartPage | 643 |
SubjectTerms | Blood Coagulation Blood Coagulation Tests - instrumentation Blood Coagulation Tests - methods Blood Coagulation Tests - standards coagulation Fibrinogen Humans laboratory automation laboratory practice Original ORIGINAL ARTICLES Reproducibility of Results Sensitivity and Specificity |
Title | A comparative evaluation of the CN‐6000 haemostasis analyser using coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry assays |
URI | https://onlinelibrary.wiley.com/doi/abs/10.1111%2Fijlh.13271 https://www.ncbi.nlm.nih.gov/pubmed/32885901 https://www.proquest.com/docview/2446957368 https://www.proquest.com/docview/2440466482 https://pubmed.ncbi.nlm.nih.gov/PMC9291559 |
Volume | 42 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3NbtQwELZKD4gL_z-BUhnBBdSsEieOHYlLVaiWilYIUakXFDm2sxvYTVCze1hOPAJvwLvxJMw4P9ulCAluUTKJs96ZzDf2zDeEPJOJZMrI1C9YIfw4To0vjYaYB1ynDJgxxpG4Hp8k49P46IyfbZGXfS1Myw8xLLihZbjvNRq4ypsLRl5-mk1HEEu5AvIwSpA4_9X7gTuKi9jld4B7DH0OTqrjJsU0nvWtm97oEsS8nCl5EcE6F3R4g3zsX77NPPk8Wi7ykf76G6_j__66m-R6h03pfqtMt8iWrW6Tq8fd7vsd8mOf6jVZOF0ThdO6oAAk6cHJz2_fAdsEdKrsvAbg2ZQNVY73xJ5TzLGfwCPUpOsZtkfVvDT1bAUD7tESS1VqeAJ4wRxOz7HZl4bbDZ3hEgJdoF8FvcQFPqomEyyvQaEVhQhArZq75PTw9YeDsd81ePA1RIWhD9CIR0ILlUDUFBppVKB1zERQgGNllieaRUmuU8lyI3KlJRchtzErrAjTAj5G98h2VVf2AaHY3j2CeFIDgIklSLPIKpYXqTQ2j7j0yPP-j850x36OTThmWR8F4YxnbsY98nSQ_dJyfvxRaqfXl6yz-yYDsJSkXEQJDPhkuAwzg9swqrL10skESOovmUfut-o1DBMxKbEa2CNiQ_EGAWQD37xSlVPHCg44F7eYPfLC6dVf3jx7c_R27I4e_ovwI3KN4WKDy2TcIduL86V9DIhske-SKyx-t-vs7xexGDgt |
linkProvider | Wiley-Blackwell |
linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3NbtswDBaKFth2Kbp_b92mobtsqAdbliz5WBQr0i4JemiB3gxZkhMXiT006SG3PcLeYO-2JxmpOE6DDgV2M2zaCkJR_CiRHwn5pFLFtFVZWLJShpxnNlTWQMwDrlNFzFrrSVwHw7R3yc-uxFWbm4O1MEt-iG7DDS3Dr9do4LghfcfKq-vJ-CsEU1hBvsMBmGNGH-Pnq4VYSO4zPMBBxqEAN9Wyk2Iiz_rdTX90D2Tez5W8i2G9EzrZI7steqRHS3U_JVuufkYeDdrz8efk9xE1azpvuqbypk1JAerR4-Gfn78AfUR0rN20AWg4q2ZUe2YSd0MxC34En9CjtqvXIdXTyjaTBQx4SCssJmngC-CnCrg9xXZcBl63dIJBPp2j54OZg1twVI9GWACDQgsKGF0vZi_I5cm3i-Ne2LZgCA3EbXEI4EUk0kidQlwTW2V1ZAxnMirB9TEnUsOStDCZYoWVhTZKyFg4zkon46yE5eIl2a6b2r0mFBuwJxDxGYAYXIE0S5xmRZkpCwG9UAH5vFJEblp-cmyTMclXcQoqLfdKC8hBJ_tjycrxT6n9lT7z1jJnOcCZNBMySWHAj91j-GfwoETXrrn1MhHS7isWkFdL9XfDJEwprNcNiNyYGJ0A8nVvPqmrseftBiSKh8AB-eKn0AO_PD896_f81Zv_Ef5AHvcuBv28fzr8_pY8Ybg14PMO98n2_ObWvQP8NC_eeyv5Cy8SGlo |
linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3NbtNAEF5VRaq48A91KbAILqA6stc_u5a4VIUoLW2EEJV6QdZ6d52YJnbVJIdw4hF4A96NJ2Fm_ZOGIiS4WfbY62xmPN_sznxDyEsRCya1SNyc5dwNw0S7QiuIecB1Co9prS2J68kwHpyGR2fR2QZ509bC1PwQ3YIbWob9XqOBX-j8ipEXXybjHsRSWEB-I4wBSiAk-tiRR0U8tAke4B99NwIv1ZCTYh7P6t51d3QNY15PlbwKYa0P6t8mn9u3r1NPznuLedZTX38jdvzfn3eH3GrAKd2vteku2TDlPbJ10my_3yc_9qlasYXTFVM4rXIKSJIeDH9--w7gxqNjaaYVIM9ZMaPSEp-YS4pJ9iN4hBw1TcP2qJwWuposYcA9WmCtSgVPADeYwekpdvtScLumE1xDoHN0rKCYuMJH5WiE9TUotKQQAsjl7AE57b_7dDBwmw4ProKw0HcBG0UBV1zGEDb5WmjpKRUy7uXgWZmJYsWCOFOJYJnmmVQi4n5kQpYb7ic5fI0eks2yKs02odjfPYCAUgGCCQVIs8BIluWJ0CYLIuGQV-0fnaqG_hy7cEzSNgzCGU_tjDvkRSd7UZN-_FFqt9WXtDH8WQpoKU4iHsQw4PPuMswM7sPI0lQLK-Mhq79gDnlUq1c3TMCEwHJgh_A1xesEkA58_UpZjC0tOABd3GN2yGurV3958_Tw6Hhgj3b-RfgZ2frwtp8eHw7fPyY3GS482KzGXbI5v1yYJ4DO5tlTa4S_AN_GOiQ |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+comparative+evaluation+of+the+CN-6000+haemostasis+analyser+using+coagulation%2C+amidolytic%2C+immuno-turbidometric+and+light+transmission+aggregometry+assays&rft.jtitle=International+journal+of+laboratory+hematology&rft.au=Gardiner%2C+Chris&rft.au=Lane%2C+Philip&rft.au=Langley%2C+Katy&rft.au=Tailor%2C+Hitesh&rft.date=2020-10-01&rft.issn=1751-553X&rft.eissn=1751-553X&rft.volume=42&rft.issue=5&rft.spage=643&rft_id=info:doi/10.1111%2Fijlh.13271&rft.externalDBID=NO_FULL_TEXT |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1751-5521&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1751-5521&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1751-5521&client=summon |