A comparative evaluation of the CN‐6000 haemostasis analyser using coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry assays

Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. Aims To...

Full description

Saved in:
Bibliographic Details
Published inInternational journal of laboratory hematology Vol. 42; no. 5; pp. 643 - 649
Main Authors Gardiner, Chris, Lane, Philip, Langley, Katy, Tailor, Hitesh, Machin, Samuel J., Mackie, Ian J.
Format Journal Article
LanguageEnglish
Published England Wiley Subscription Services, Inc 01.10.2020
John Wiley and Sons Inc
Subjects
Online AccessGet full text
ISSN1751-5521
1751-553X
1751-553X
DOI10.1111/ijlh.13271

Cover

Loading…
Abstract Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. Aims To evaluate the performance of the CN‐6000 against a predicate device. Methods The CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). Results Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34). Conclusions The CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.
AbstractList The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.BACKGROUNDThe CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.To evaluate the performance of the CN-6000 against a predicate device.AIMSTo evaluate the performance of the CN-6000 against a predicate device.The CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).METHODSThe CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34).RESULTSAcceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34).The CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.CONCLUSIONSThe CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.
The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. To evaluate the performance of the CN-6000 against a predicate device. The CN-6000 was evaluated against the CS-5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between-day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN-6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN-6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in "vote-outs" due to low light transmission (24 vs 34). The CN-6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.
BackgroundThe CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source.AimsTo evaluate the performance of the CN‐6000 against a predicate device.MethodsThe CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days).ResultsAcceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34).ConclusionsThe CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.
Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry (LTA) capabilities. Transmitted light is monitored at multiple wavelengths (340, 405, 575, 660, 800 nm), from an LED light source. Aims To evaluate the performance of the CN‐6000 against a predicate device. Methods The CN‐6000 was evaluated against the CS‐5100 (Sysmex) for 14 different tests, using 880 samples from normal subjects, anticoagulated patients, critically ill patients, plasmas with high or low fibrinogen content or abnormal levels of interfering substances. Between‐day assay imprecision was assessed using commercial QC materials (n = 10 replicates on each of 5 days). Results Acceptable levels of imprecision were obtained for all assays. Agreement between the two analysers was excellent for all assays. Throughput was 35% higher using the CN‐6000 (337 vs 250 tests per hour for PT, aPTT and fibrinogen). The CN‐6000 also demonstrated improved clot detection in plasmas with high levels of interfering substances as demonstrated by a 29% reduction in “vote‐outs” due to low light transmission (24 vs 34). Conclusions The CN‐6000 demonstrated excellent comparability with the predicate instrument and acceptable levels of imprecision in all assays. Improvements in throughput and clot detection in the presence of interfering substances were also shown.
Author Mackie, Ian J.
Langley, Katy
Gardiner, Chris
Machin, Samuel J.
Tailor, Hitesh
Lane, Philip
AuthorAffiliation 1 Haemostasis Research Unit University College London London UK
2 Haematology Evaluations Unit HSL (Analytics) LLP London UK
AuthorAffiliation_xml – name: 1 Haemostasis Research Unit University College London London UK
– name: 2 Haematology Evaluations Unit HSL (Analytics) LLP London UK
Author_xml – sequence: 1
  givenname: Chris
  orcidid: 0000-0002-2318-0062
  surname: Gardiner
  fullname: Gardiner, Chris
  email: chris.gardiner@uclmail.net
  organization: University College London
– sequence: 2
  givenname: Philip
  surname: Lane
  fullname: Lane, Philip
  organization: University College London
– sequence: 3
  givenname: Katy
  surname: Langley
  fullname: Langley, Katy
  organization: University College London
– sequence: 4
  givenname: Hitesh
  orcidid: 0000-0003-3615-7889
  surname: Tailor
  fullname: Tailor, Hitesh
  organization: HSL (Analytics) LLP
– sequence: 5
  givenname: Samuel J.
  surname: Machin
  fullname: Machin, Samuel J.
  organization: University College London
– sequence: 6
  givenname: Ian J.
  orcidid: 0000-0002-7393-2365
  surname: Mackie
  fullname: Mackie, Ian J.
  organization: University College London
BackLink https://www.ncbi.nlm.nih.gov/pubmed/32885901$$D View this record in MEDLINE/PubMed
BookMark eNp9ks2OFCEUhYkZ4_zoxgcwJG6MmR75KYqqjcmko86Yjm40cUcoiqqiQ0ELVJva-Qi-ge_mk0h3z0x0YmQDge8czoV7Co6cdxqApxhd4DxembUdLjAlHD8AJ5gzvGCMfjm6WxN8DE5jXCPEeIHqR-CYkqpiNcIn4OclVH7cyCCT2Wqot9JOeekd9B1Mg4bLD7--_ygRQnCQevQxyWgilE7aOeoAp2hcny1kP9m97hzK0bTezsmoc2jGcXI-O6QpNHl71CkYleUttKYfEkxBujiaGHdXyr4Put9DM5Qxyjk-Bg87aaN-cjOfgc9v33xaXi1WH99dLy9XC1UUFV7UlDPKFZdllWtuq1YipQrCUUcRJZqVitCyUXVFmpY3UlWMY6YL0mmO6w5X9Ay8PvhupmbUrdIuJ7NiE8wowyy8NOLvE2cG0futqEmNGauzwYsbg-C_TjomkatS2lrptJ-iIEWBirIsKpLR5_fQtZ9CftE9VdaM03KX6Nmfie6i3P5dBtABUMHHGHQnlEn7P8gBjRUYiV17iF17iH17ZMnLe5Jb13_C-AB_M1bP_yHF9fvV1UHzGzHa0HU
CitedBy_id crossref_primary_10_1111_ijlh_13740
crossref_primary_10_1097_MBC_0000000000001184
crossref_primary_10_1111_ijlh_13656
crossref_primary_10_1111_ijlh_13468
Cites_doi 10.1515/cclm-2013-1094
10.1097/MD.0000000000007976
10.1532/LH96.05031
10.1055/s-0035-1544196
10.1111/ijlh.13129
10.1016/j.plabm.2016.09.002
10.1111/jth.12642
10.1111/ijlh.12498
ContentType Journal Article
Copyright 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd
2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.
2020. This article is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
2020 The Authors. published by John Wiley & Sons Ltd.
Copyright_xml – notice: 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd
– notice: 2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.
– notice: 2020. This article is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
– notice: 2020 The Authors. published by John Wiley & Sons Ltd.
DBID 24P
AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
7T5
H94
K9.
NAPCQ
7X8
5PM
DOI 10.1111/ijlh.13271
DatabaseName Wiley Online Library Open Access
CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
Immunology Abstracts
AIDS and Cancer Research Abstracts
ProQuest Health & Medical Complete (Alumni)
Nursing & Allied Health Premium
MEDLINE - Academic
PubMed Central (Full Participant titles)
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
AIDS and Cancer Research Abstracts
ProQuest Health & Medical Complete (Alumni)
Nursing & Allied Health Premium
Immunology Abstracts
MEDLINE - Academic
DatabaseTitleList MEDLINE - Academic
MEDLINE
AIDS and Cancer Research Abstracts

Database_xml – sequence: 1
  dbid: 24P
  name: Wiley Online Library Open Access
  url: https://authorservices.wiley.com/open-science/open-access/browse-journals.html
  sourceTypes: Publisher
– sequence: 2
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 3
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
DocumentTitleAlternate GARDINER et al
EISSN 1751-553X
EndPage 649
ExternalDocumentID PMC9291559
32885901
10_1111_ijlh_13271
IJLH13271
Genre article
Journal Article
GroupedDBID ---
.3N
.GA
.Y3
05W
0R~
10A
1OC
24P
29J
31~
33P
36B
3SF
4.4
50Y
50Z
51W
51X
52M
52N
52O
52P
52R
52S
52T
52U
52V
52W
52X
53G
5GY
5HH
5LA
5VS
66C
702
7PT
8-0
8-1
8-3
8-4
8-5
8UM
930
A01
A03
AAESR
AAEVG
AAHHS
AAHQN
AAIPD
AAMNL
AANHP
AANLZ
AAONW
AASGY
AAWTL
AAXRX
AAYCA
AAZKR
ABCQN
ABCUV
ABDBF
ABEML
ABJNI
ABPVW
ABQWH
ABXGK
ACAHQ
ACBWZ
ACCFJ
ACCZN
ACGFS
ACGOF
ACMXC
ACPOU
ACPRK
ACRPL
ACSCC
ACUHS
ACXBN
ACXQS
ACYXJ
ADBBV
ADBTR
ADEOM
ADIZJ
ADKYN
ADMGS
ADNMO
ADOZA
ADXAS
ADZMN
ADZOD
AEEZP
AEIGN
AEIMD
AENEX
AEQDE
AEUQT
AEUYR
AFBPY
AFFPM
AFGKR
AFPWT
AFWVQ
AFZJQ
AHBTC
AIACR
AITYG
AIURR
AIWBW
AJBDE
ALAGY
ALMA_UNASSIGNED_HOLDINGS
ALUQN
ALVPJ
AMBMR
AMYDB
ATUGU
AZBYB
AZFZN
AZVAB
BAFTC
BDRZF
BFHJK
BHBCM
BMXJE
BROTX
BRXPI
BY8
C45
CAG
COF
CS3
D-6
D-7
D-E
D-F
DCZOG
DPXWK
DR2
DRFUL
DRMAN
DRSTM
DU5
EAD
EAP
EAS
EBC
EBD
EBS
EBX
EJD
EMB
EMK
EMOBN
EPT
ESX
EX3
F00
F01
F04
F5P
FEDTE
FUBAC
G-S
G.N
GODZA
H.X
HF~
HGLYW
HVGLF
HZI
HZ~
IHE
IX1
J0M
KBYEO
L7B
LATKE
LC2
LC3
LEEKS
LH4
LITHE
LOXES
LP6
LP7
LUTES
LW6
LYRES
MEWTI
MK4
MRFUL
MRMAN
MRSTM
MSFUL
MSMAN
MSSTM
MXFUL
MXMAN
MXSTM
N04
N05
N9A
NF~
O66
O9-
OIG
OVD
P2P
P2W
P2X
P2Z
P4B
P4D
Q.N
Q11
QB0
Q~Q
R.K
ROL
RX1
SUPJJ
SV3
TEORI
TUS
UB1
W8V
W99
WBKPD
WHWMO
WIH
WIJ
WIK
WOHZO
WOW
WQJ
WRC
WVDHM
WXI
WXSBR
XG1
~IA
~WT
AAYXX
AEYWJ
AGHNM
AGQPQ
AGYGG
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
1OB
7T5
AAMMB
AEFGJ
AGXDD
AIDQK
AIDYY
H94
K9.
NAPCQ
7X8
5PM
ID FETCH-LOGICAL-c4481-937537c7a68751d8da0cc4270f3032e56c236bc982bd7bac85715e42fe719f183
IEDL.DBID DR2
ISSN 1751-5521
1751-553X
IngestDate Thu Aug 21 14:11:39 EDT 2025
Tue Aug 05 10:35:47 EDT 2025
Wed Aug 13 06:25:20 EDT 2025
Wed Feb 19 02:29:31 EST 2025
Tue Jul 01 03:54:10 EDT 2025
Thu Apr 24 23:03:28 EDT 2025
Wed Jan 22 16:59:57 EST 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 5
Keywords coagulation
laboratory automation
laboratory practice
Language English
License Attribution
2020 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.
This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c4481-937537c7a68751d8da0cc4270f3032e56c236bc982bd7bac85715e42fe719f183
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ORCID 0000-0002-2318-0062
0000-0002-7393-2365
0000-0003-3615-7889
OpenAccessLink https://proxy.k.utb.cz/login?url=https://onlinelibrary.wiley.com/doi/abs/10.1111%2Fijlh.13271
PMID 32885901
PQID 2446957368
PQPubID 2045143
PageCount 7
ParticipantIDs pubmedcentral_primary_oai_pubmedcentral_nih_gov_9291559
proquest_miscellaneous_2440466482
proquest_journals_2446957368
pubmed_primary_32885901
crossref_citationtrail_10_1111_ijlh_13271
crossref_primary_10_1111_ijlh_13271
wiley_primary_10_1111_ijlh_13271_IJLH13271
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate October 2020
PublicationDateYYYYMMDD 2020-10-01
PublicationDate_xml – month: 10
  year: 2020
  text: October 2020
PublicationDecade 2020
PublicationPlace England
PublicationPlace_xml – name: England
– name: Chichester
– name: Hoboken
PublicationTitle International journal of laboratory hematology
PublicationTitleAlternate Int J Lab Hematol
PublicationYear 2020
Publisher Wiley Subscription Services, Inc
John Wiley and Sons Inc
Publisher_xml – name: Wiley Subscription Services, Inc
– name: John Wiley and Sons Inc
References 2008
2016; 6
2017; 96
2014; 52
2006; 12
2020; 42
2016; 38
2015; 61
2014; 12
2015; 41
e_1_2_7_6_1
e_1_2_7_11_1
e_1_2_7_5_1
e_1_2_7_10_1
e_1_2_7_3_1
e_1_2_7_9_1
e_1_2_7_8_1
e_1_2_7_7_1
Chen L (e_1_2_7_4_1) 2015; 61
e_1_2_7_2_1
References_xml – volume: 96
  issue: 38
  year: 2017
  article-title: The effects of hemolysis on plasma prothrombin time and activated partial thromboplastin time tests using photo‐optical method
  publication-title: Medicine (Baltimore)
– volume: 52
  start-page: 1193
  issue: 8
  year: 2014
  end-page: 1202
  article-title: Evaluation of the automated coagulation analyzer CS‐5100 and its utility in high throughput laboratories
  publication-title: Clin Chem Lab Med
– volume: 12
  start-page: 1591
  issue: 10
  year: 2014
  article-title: Iatrogenic anemia (can it be prevented?)
  publication-title: J Thromb Haemost
– volume: 38
  start-page: 375
  issue: 4
  year: 2016
  end-page: 388
  article-title: Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA‐Compact‐Max analyser
  publication-title: Int J Lab Hematol
– volume: 42
  start-page: 140
  issue: 2
  year: 2020
  end-page: 144
  article-title: A practical method for reducing the interference due to lipaemia in coagulation tests
  publication-title: Int J Lab Hematol
– volume: 12
  start-page: 32
  issue: 1
  year: 2006
  end-page: 38
  article-title: Adcock DM. Recommendations for evaluation of coagulation analyzers
  publication-title: Lab Hematol
– volume: 41
  start-page: 294
  issue: 3
  year: 2015
  end-page: 299
  article-title: The changing face of hemostasis testing in modern laboratories: consolidation, automation, and beyond
  publication-title: Semin Thromb Hemost
– year: 2008
– volume: 61
  start-page: 653
  issue: 7
  year: 2015
  end-page: 660
  article-title: Performance evaluation of the sysmex CS‐5100 automated coagulation analyzer
  publication-title: Clin Lab
– volume: 6
  start-page: 38
  year: 2016
  end-page: 47
  article-title: The Sysmex CS‐5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities
  publication-title: Pract Lab Med
– ident: e_1_2_7_6_1
  doi: 10.1515/cclm-2013-1094
– ident: e_1_2_7_9_1
  doi: 10.1097/MD.0000000000007976
– volume: 61
  start-page: 653
  issue: 7
  year: 2015
  ident: e_1_2_7_4_1
  article-title: Performance evaluation of the sysmex CS‐5100 automated coagulation analyzer
  publication-title: Clin Lab
– ident: e_1_2_7_8_1
  doi: 10.1532/LH96.05031
– ident: e_1_2_7_7_1
– ident: e_1_2_7_2_1
  doi: 10.1055/s-0035-1544196
– ident: e_1_2_7_11_1
  doi: 10.1111/ijlh.13129
– ident: e_1_2_7_5_1
  doi: 10.1016/j.plabm.2016.09.002
– ident: e_1_2_7_3_1
  doi: 10.1111/jth.12642
– ident: e_1_2_7_10_1
  doi: 10.1111/ijlh.12498
SSID ssj0057409
Score 2.2658095
Snippet Background The CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission...
The CN-6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry (LTA)...
BackgroundThe CN‐6000 (Sysmex Corp.) is a new haemostasis analyser with blood coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry...
SourceID pubmedcentral
proquest
pubmed
crossref
wiley
SourceType Open Access Repository
Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 643
SubjectTerms Blood Coagulation
Blood Coagulation Tests - instrumentation
Blood Coagulation Tests - methods
Blood Coagulation Tests - standards
coagulation
Fibrinogen
Humans
laboratory automation
laboratory practice
Original
ORIGINAL ARTICLES
Reproducibility of Results
Sensitivity and Specificity
Title A comparative evaluation of the CN‐6000 haemostasis analyser using coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry assays
URI https://onlinelibrary.wiley.com/doi/abs/10.1111%2Fijlh.13271
https://www.ncbi.nlm.nih.gov/pubmed/32885901
https://www.proquest.com/docview/2446957368
https://www.proquest.com/docview/2440466482
https://pubmed.ncbi.nlm.nih.gov/PMC9291559
Volume 42
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3NbtQwELZKD4gL_z-BUhnBBdSsEieOHYlLVaiWilYIUakXFDm2sxvYTVCze1hOPAJvwLvxJMw4P9ulCAluUTKJs96ZzDf2zDeEPJOJZMrI1C9YIfw4To0vjYaYB1ynDJgxxpG4Hp8k49P46IyfbZGXfS1Myw8xLLihZbjvNRq4ypsLRl5-mk1HEEu5AvIwSpA4_9X7gTuKi9jld4B7DH0OTqrjJsU0nvWtm97oEsS8nCl5EcE6F3R4g3zsX77NPPk8Wi7ykf76G6_j__66m-R6h03pfqtMt8iWrW6Tq8fd7vsd8mOf6jVZOF0ThdO6oAAk6cHJz2_fAdsEdKrsvAbg2ZQNVY73xJ5TzLGfwCPUpOsZtkfVvDT1bAUD7tESS1VqeAJ4wRxOz7HZl4bbDZ3hEgJdoF8FvcQFPqomEyyvQaEVhQhArZq75PTw9YeDsd81ePA1RIWhD9CIR0ILlUDUFBppVKB1zERQgGNllieaRUmuU8lyI3KlJRchtzErrAjTAj5G98h2VVf2AaHY3j2CeFIDgIklSLPIKpYXqTQ2j7j0yPP-j850x36OTThmWR8F4YxnbsY98nSQ_dJyfvxRaqfXl6yz-yYDsJSkXEQJDPhkuAwzg9swqrL10skESOovmUfut-o1DBMxKbEa2CNiQ_EGAWQD37xSlVPHCg44F7eYPfLC6dVf3jx7c_R27I4e_ovwI3KN4WKDy2TcIduL86V9DIhske-SKyx-t-vs7xexGDgt
linkProvider Wiley-Blackwell
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3NbtswDBaKFth2Kbp_b92mobtsqAdbliz5WBQr0i4JemiB3gxZkhMXiT006SG3PcLeYO-2JxmpOE6DDgV2M2zaCkJR_CiRHwn5pFLFtFVZWLJShpxnNlTWQMwDrlNFzFrrSVwHw7R3yc-uxFWbm4O1MEt-iG7DDS3Dr9do4LghfcfKq-vJ-CsEU1hBvsMBmGNGH-Pnq4VYSO4zPMBBxqEAN9Wyk2Iiz_rdTX90D2Tez5W8i2G9EzrZI7steqRHS3U_JVuufkYeDdrz8efk9xE1azpvuqbypk1JAerR4-Gfn78AfUR0rN20AWg4q2ZUe2YSd0MxC34En9CjtqvXIdXTyjaTBQx4SCssJmngC-CnCrg9xXZcBl63dIJBPp2j54OZg1twVI9GWACDQgsKGF0vZi_I5cm3i-Ne2LZgCA3EbXEI4EUk0kidQlwTW2V1ZAxnMirB9TEnUsOStDCZYoWVhTZKyFg4zkon46yE5eIl2a6b2r0mFBuwJxDxGYAYXIE0S5xmRZkpCwG9UAH5vFJEblp-cmyTMclXcQoqLfdKC8hBJ_tjycrxT6n9lT7z1jJnOcCZNBMySWHAj91j-GfwoETXrrn1MhHS7isWkFdL9XfDJEwprNcNiNyYGJ0A8nVvPqmrseftBiSKh8AB-eKn0AO_PD896_f81Zv_Ef5AHvcuBv28fzr8_pY8Ybg14PMO98n2_ObWvQP8NC_eeyv5Cy8SGlo
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3NbtNAEF5VRaq48A91KbAILqA6stc_u5a4VIUoLW2EEJV6QdZ6d52YJnbVJIdw4hF4A96NJ2Fm_ZOGIiS4WfbY62xmPN_sznxDyEsRCya1SNyc5dwNw0S7QiuIecB1Co9prS2J68kwHpyGR2fR2QZ509bC1PwQ3YIbWob9XqOBX-j8ipEXXybjHsRSWEB-I4wBSiAk-tiRR0U8tAke4B99NwIv1ZCTYh7P6t51d3QNY15PlbwKYa0P6t8mn9u3r1NPznuLedZTX38jdvzfn3eH3GrAKd2vteku2TDlPbJ10my_3yc_9qlasYXTFVM4rXIKSJIeDH9--w7gxqNjaaYVIM9ZMaPSEp-YS4pJ9iN4hBw1TcP2qJwWuposYcA9WmCtSgVPADeYwekpdvtScLumE1xDoHN0rKCYuMJH5WiE9TUotKQQAsjl7AE57b_7dDBwmw4ProKw0HcBG0UBV1zGEDb5WmjpKRUy7uXgWZmJYsWCOFOJYJnmmVQi4n5kQpYb7ic5fI0eks2yKs02odjfPYCAUgGCCQVIs8BIluWJ0CYLIuGQV-0fnaqG_hy7cEzSNgzCGU_tjDvkRSd7UZN-_FFqt9WXtDH8WQpoKU4iHsQw4PPuMswM7sPI0lQLK-Mhq79gDnlUq1c3TMCEwHJgh_A1xesEkA58_UpZjC0tOABd3GN2yGurV3958_Tw6Hhgj3b-RfgZ2frwtp8eHw7fPyY3GS482KzGXbI5v1yYJ4DO5tlTa4S_AN_GOiQ
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+comparative+evaluation+of+the+CN-6000+haemostasis+analyser+using+coagulation%2C+amidolytic%2C+immuno-turbidometric+and+light+transmission+aggregometry+assays&rft.jtitle=International+journal+of+laboratory+hematology&rft.au=Gardiner%2C+Chris&rft.au=Lane%2C+Philip&rft.au=Langley%2C+Katy&rft.au=Tailor%2C+Hitesh&rft.date=2020-10-01&rft.issn=1751-553X&rft.eissn=1751-553X&rft.volume=42&rft.issue=5&rft.spage=643&rft_id=info:doi/10.1111%2Fijlh.13271&rft.externalDBID=NO_FULL_TEXT
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1751-5521&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1751-5521&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1751-5521&client=summon