Comparison of dose at an interventional reference point between the displayed estimated value and measured value
Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in re...
Saved in:
Published in | Radiological physics and technology Vol. 4; no. 2; pp. 189 - 193 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
Japan
Springer Japan
01.07.2011
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Abstract | Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in real time. However, no detailed report has evaluated the accuracy of the reference doses displayed on X-ray equipment. Thus, in this study, we compared the displayed reference dose to the actual measured value in many IR X-ray systems. Although the displayed reference doses of many IR X-ray systems agreed with the measured actual values within approximately 15%, the doses of a few IR units were not close. Furthermore, some X-ray units made in Japan displayed reference doses quite different from the actual measured value, probably because the reference point of these units differs from the International Electrotechnical Commission standard. Thus, IR physicians should pay attention to the location of the IRP of the displayed reference dose in Japan. Furthermore, physicians should be aware of the accuracy of the displayed reference dose of the X-ray system that they use for IR. Thus, regular checks of the displayed reference dose of the X-ray system are important. |
---|---|
AbstractList | Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in real time. However, no detailed report has evaluated the accuracy of the reference doses displayed on X-ray equipment. Thus, in this study, we compared the displayed reference dose to the actual measured value in many IR X-ray systems. Although the displayed reference doses of many IR X-ray systems agreed with the measured actual values within approximately 15%, the doses of a few IR units were not close. Furthermore, some X-ray units made in Japan displayed reference doses quite different from the actual measured value, probably because the reference point of these units differs from the International Electrotechnical Commission standard. Thus, IR physicians should pay attention to the location of the IRP of the displayed reference dose in Japan. Furthermore, physicians should be aware of the accuracy of the displayed reference dose of the X-ray system that they use for IR. Thus, regular checks of the displayed reference dose of the X-ray system are important. |
Author | Yanagawa, Isao Chida, Koichi Ebata, Ayako Zuguchi, Masayuki Morishima, Yoshiaki Taura, Masaaki Inaba, Yohei Takeda, Ken |
Author_xml | – sequence: 1 givenname: Koichi surname: Chida fullname: Chida, Koichi email: chida@mail.tains.tohoku.ac.jp organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University – sequence: 2 givenname: Yohei surname: Inaba fullname: Inaba, Yohei organization: Department of Radiology, Tohoku University Hospital – sequence: 3 givenname: Yoshiaki surname: Morishima fullname: Morishima, Yoshiaki organization: Department of Radiology, Tohoku Employees’ Pension Welfare Hospital – sequence: 4 givenname: Masaaki surname: Taura fullname: Taura, Masaaki organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University – sequence: 5 givenname: Ayako surname: Ebata fullname: Ebata, Ayako organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University – sequence: 6 givenname: Isao surname: Yanagawa fullname: Yanagawa, Isao organization: Department of Radiology, Tohoku University Hospital – sequence: 7 givenname: Ken surname: Takeda fullname: Takeda, Ken organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University – sequence: 8 givenname: Masayuki surname: Zuguchi fullname: Zuguchi, Masayuki organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/21643656$$D View this record in MEDLINE/PubMed |
BookMark | eNp9kMtOwzAQRS1URB_wAWyQfyDgsR03WaKKl1SJDawtJ5lAqsSO7KSof4-r0C5ZjGY0M_fq6izJzDqLhNwCuwfG1g8BOOQyYQCxOCTqgiwgU2nChITZeRZiTpYh7BhTwDm_InMOSgqVqgXpN67rjW-Cs9TVtHIBqRmosbSxA_o92qFx1rTUY40ebYm0d_FECxx-EC0dvpFWTehbc8CKYhiazgxx2pt2jFa2oh2aMPrT6ppc1qYNePPXV-Tz-elj85ps31_eNo_bpJRCDIkqC6aKLJdsjSLLylpKpsoM11hAnqISqaoZwxwBCiM5Q5XnheFVmpYgVK7EisDkW3oXQkyvex-j-YMGpo_09ERPR3r6SE8fNXeTph-LDquz4oQrPvDpIcST_UKvd270EU_4x_UXwst9ow |
CitedBy_id | crossref_primary_10_7793_jcoron_23_16_00022 crossref_primary_10_1007_s12194_022_00650_w crossref_primary_10_3390_bioengineering9080354 crossref_primary_10_3390_diagnostics11091613 crossref_primary_10_1007_s00455_022_10411_x crossref_primary_10_3390_s23010512 crossref_primary_10_1007_s12194_018_0484_z crossref_primary_10_6009_jjrt_2013_JSRT_69_10_1173 crossref_primary_10_6009_jjrt_2016_JSRT_72_8_689 crossref_primary_10_3390_s20092741 crossref_primary_10_1007_s00455_016_9718_6 crossref_primary_10_3390_bioengineering10020259 crossref_primary_10_3390_tomography8010036 crossref_primary_10_6009_jjrt_2022_1309 crossref_primary_10_1620_tjem_2022_J001 crossref_primary_10_1007_s12194_022_00660_8 crossref_primary_10_1111_pace_12790 crossref_primary_10_6009_jjrt_2019_JSRT_75_7_646 crossref_primary_10_1007_s12194_020_00575_2 crossref_primary_10_3390_diagnostics10100753 crossref_primary_10_6009_jjrt_2015_JSRT_71_12_1215 crossref_primary_10_1093_jrr_rrx039 crossref_primary_10_1088_1742_6596_694_1_012049 crossref_primary_10_3390_diagnostics10100846 crossref_primary_10_3390_diagnostics13183003 crossref_primary_10_1088_1361_6560_ab0bd5 crossref_primary_10_1088_1361_6498_ad0a9d crossref_primary_10_1093_rpd_ncae058 |
Cites_doi | 10.2214/AJR.09.2747 10.1093/rpd/ncq286 10.2214/AJR.08.2115 10.2214/AJR.10.6396 10.2214/AJR.04.1653 10.1269/jrr.09112 10.1007/s12194-008-0044-z 10.1016/S0146-6453(01)00004-5 10.2214/AJR.07.2422 10.2214/AJR.10.4466 10.1016/j.jacc.2004.10.014 10.1056/NEJMicm030503 10.3109/02841850903229141 10.1148/radiol.2542082312 10.2214/ajr.183.4.1831111 10.1093/rpd/ncq287 10.6009/jjrt.KJ00004022974 10.1093/rpd/ncp052 10.1161/CIRCULATIONAHA.108.191650 |
ContentType | Journal Article |
Copyright | Japanese Society of Radiological Technology and Japan Society of Medical Physics 2011 |
Copyright_xml | – notice: Japanese Society of Radiological Technology and Japan Society of Medical Physics 2011 |
DBID | CGR CUY CVF ECM EIF NPM AAYXX CITATION |
DOI | 10.1007/s12194-011-0121-6 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed CrossRef |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) CrossRef |
DatabaseTitleList | MEDLINE |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 2 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
EISSN | 1865-0341 |
EndPage | 193 |
ExternalDocumentID | 10_1007_s12194_011_0121_6 21643656 |
Genre | Journal Article Comparative Study |
GeographicLocations | Japan |
GeographicLocations_xml | – name: Japan |
GroupedDBID | --- -5E -5G -BR -EM -Y2 -~C .86 06C 06D 0R~ 0VY 123 1N0 203 29P 29~ 2JN 2JY 2KG 2VQ 2~H 30V 4.4 406 408 409 40D 5VS 67Z 6NX 8TC 96X AAAVM AABHQ AAFGU AAHNG AAIAL AAJKR AANXM AANZL AAPBV AARHV AARTL AATNV AATVU AAUYE AAWCG AAYFA AAYIU AAYQN AAYTO ABBXA ABDZT ABECU ABFGW ABFTV ABHQN ABJNI ABJOX ABKAS ABKCH ABMNI ABMQK ABPLI ABQBU ABSXP ABTEG ABTKH ABTMW ABULA ABWNU ABXPI ACBMV ACBRV ACBYP ACGFS ACHSB ACIGE ACIPQ ACKNC ACMDZ ACMLO ACOKC ACOMO ACTTH ACVWB ACWMK ADHHG ADHIR ADINQ ADKNI ADKPE ADMDM ADOXG ADRFC ADTPH ADURQ ADYFF ADZKW AEBTG AEFTE AEGAL AEGNC AEJHL AEJRE AEOHA AEPYU AESKC AESTI AETLH AEVLU AEVTX AEXYK AFLOW AFNRJ AFQWF AFWTZ AFZKB AGAYW AGDGC AGGBP AGJBK AGMZJ AGQMX AGWZB AGYKE AHAVH AHBYD AHSBF AHYZX AIAKS AIIXL AILAN AIMYW AITGF AJBLW AJDOV AJRNO AJZVZ AKMHD AKQUC ALFXC ALMA_UNASSIGNED_HOLDINGS ALWAN AMKLP AMXSW AMYLF AMYQR ANMIH AOCGG AUKKA AXYYD B-. BA0 BDATZ BGNMA CAG COF CS3 CSCUP DDRTE DNIVK DPUIP EBLON EBS EIOEI EJD EN4 ESBYG F5P FERAY FFXSO FIGPU FINBP FNLPD FRRFC FSGXE FWDCC G-Y G-Z GGCAI GGRSB GJIRD GNWQR GQ6 GQ7 GQ8 GXS HF~ HG5 HG6 HLICF HMJXF HQYDN HRMNR HZ~ I0C IJ- IKXTQ IWAJR IXC IXD IZIGR IZQ I~X J-C J0Z JBSCW JCJTX JZLTJ KOV KPH LLZTM M4Y MA- NPVJJ NQJWS NU0 O9- O93 O9I O9J OAM P9S PT4 QOR QOS R89 RIG RLLFE RNS ROL RPX RSV S16 S1Z S27 S37 S3B SAP SDH SHX SISQX SMD SNE SNPRN SNX SOHCF SOJ SPISZ SRMVM SSLCW SSXJD STPWE SZ9 T13 TSG TSK TT1 U2A U9L UG4 UOJIU UTJUX UZXMN VC2 VFIZW W48 WJK WK8 YLTOR Z45 Z7X Z83 ZMTXR ZOVNA ~A9 AACDK AAJBT AASML AAYZH ABAKF ACAOD ACDTI ACZOJ AEFQL AEMSY AFBBN AGQEE AGRTI AIGIU CGR CUY CVF ECM EIF H13 NPM SJYHP AAYXX CITATION |
ID | FETCH-LOGICAL-c433t-6cb06b89407e388cf4406c8e7eb195e6356f00e9e11ba420e699ba2d55c136963 |
IEDL.DBID | AGYKE |
ISSN | 1865-0333 |
IngestDate | Thu Sep 12 19:32:40 EDT 2024 Tue Oct 15 23:45:47 EDT 2024 Sat Dec 16 12:02:49 EST 2023 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 2 |
Keywords | Radiation injury Radiation dosimetry Exposure dose Radiation safety Interventional radiology Interventional reference point (IRP) |
Language | English |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-c433t-6cb06b89407e388cf4406c8e7eb195e6356f00e9e11ba420e699ba2d55c136963 |
PMID | 21643656 |
PageCount | 5 |
ParticipantIDs | crossref_primary_10_1007_s12194_011_0121_6 pubmed_primary_21643656 springer_journals_10_1007_s12194_011_0121_6 |
PublicationCentury | 2000 |
PublicationDate | 2011-07-01 |
PublicationDateYYYYMMDD | 2011-07-01 |
PublicationDate_xml | – month: 07 year: 2011 text: 2011-07-01 day: 01 |
PublicationDecade | 2010 |
PublicationPlace | Japan |
PublicationPlace_xml | – name: Japan |
PublicationTitle | Radiological physics and technology |
PublicationTitleAbbrev | Radiol Phys Technol |
PublicationTitleAlternate | Radiol Phys Technol |
PublicationYear | 2011 |
Publisher | Springer Japan |
Publisher_xml | – name: Springer Japan |
References | ChidaKKatoMSaitoHIshibashiTTakahashiSKohzukiMOptimizing patient radiation dose in intervention proceduresActa Radiol20105133391996139810.3109/02841850903229141 ChidaKSaitoHOtaniHKohzukiMTakahashiSYamadaSRelationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional proceduresAm J Roentgenol200618677477810.2214/AJR.04.1653 TsapakiVAhmedNAAlSuwaidiJSBeganovicABeniderABenOmraneLRadiation exposure to patients during interventional procedures in 20 countries: initial IAEA project resultsAm J Roentgenol200919355956910.2214/AJR.08.2115 ChidaKKagayaYSaitoHTakaiYTakahashiSYamadaSTotal entrance skin dose: an effective indicator of the maximum radiation dose to a patient’s skin during percutaneous coronary interventionAm J Roentgenol2007189W224W22710.2214/AJR.07.2422 HirshfeldJWBalterSBrinkerJAACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures: a report of the American College of Cardiology Foundation/American Heart Association American. College of Physicians Task Force on Clinical Competence and TrainingJ Am Coll Cardiol200444225922821558233510.1016/j.jacc.2004.10.014 KatoMChidaKSatoTOosakaHTosaTKadowakiKEvaluating the maximum patient radiation dose in cardiac interventional proceduresRadiat Prot Dosimetry201114369732094758610.1093/rpd/ncq286 GerberTCCarrJJAraiAEDixonRLFerrariVAGomesASIonizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and InterventionCirculation2009119105610651918851210.1161/CIRCULATIONAHA.108.191650 ChidaKOhnoTKakizakiSTakegawaMYuukiHNakadaMRadiation dose to the pediatric cardiac catheterization and intervention patientAm J Roentgenol20101951175117910.2214/AJR.10.4466 KimKPMillerDLMinimising radiation exposure to physicians performing fluoroscopically guided cardiac catheterisation procedures: a reviewRadiat Prot Dosimetry20091332272331932951110.1093/rpd/ncp052 ChidaKInabaYMasuyamaHYanagawaIMoriISaitoHEvaluating the performance of a MOSFET dosimeter at diagnostic X-ray energies for interventional radiologyRadiol Phys Technol2009258612082113010.1007/s12194-008-0044-z InabaYChidaKShirotoriKShimuraHYanagawaIZuguchiMComparison of the radiation dose in a cardiac IVR X-ray systemRadiat Prot Dosimetry201114374802108151710.1093/rpd/ncq287 International Commission on Radiological Protection. ICRP publication 85: avoidance of radiation injuries from medical interventional procedures. Ann ICRP 2001 (30/2:Publication 85). ChidaKMorishimaYKatahiraYChibaHZuguchiMEvaluation of additional lead shielding in protecting the physician from radiation during cardiac interventional proceduresNippon Hoshasen Gijutsu Gakkai Zasshi2005611632163716395238 Japanese Industrial Standards (JIS). Medical electrical equipment, Particular requirements for the safety of X-ray equipment for interventional procedures, Z 4751-2-43, 2005. BalterSHopewellJWMillerDLWagnerLKZelefskyMJFluoroscopically guided interventional procedures: a review of radiation effects on patients’ skin and hairRadiology20102543263412009350710.1148/radiol.2542082312 ChidaKSaitoHZuguchiMShirotoriKKumagaiSNakayamaHDoes digital acquisition reduce patients’ skin dose in cardiac interventional procedures? An experimental studyAm J Roentgenol200418311111114 International Electrotechnical Commission (IEC). Medical electrical equipment. Part 2–43. Particular requirements for the safety of X-ray equipment for interventional procedures. IEC 60601-2-43. Geneva, Switzerland: International Electrotechnical Commission; 2000. Chida K, Takahashi T, Ito D, Shimura H, Takeda K, Zuguchi M. Clarifying and visualizing sources of staff-received scattered radiation in interventional procedures. Am J Roentgenol. 2011 (in press). WongLRehmJImages in clinical medicine: radiation injury from a fluoroscopic procedureN Engl J Med2004350e231520142810.1056/NEJMicm030503 ChidaKKatoMKagayaYZuguchiMSaitoHIshibashiTRadiation dose and radiation protection for patients and physicians during interventional procedureJ Radiat Res201051971052033925310.1269/jrr.09112 ChidaKInabaYSaitoHIshibashiTTakahashiSKohzukiMRadiation dose of interventional radiology system using a flat-panel detectorAm J Roentgenol20091931680168510.2214/AJR.09.2747 15385315 - AJR Am J Roentgenol. 2004 Oct;183(4):1111-4 20339253 - J Radiat Res. 2010;51(2):97-105 15201428 - N Engl J Med. 2004 Jun 17;350(25):e23 16395238 - Nihon Hoshasen Gijutsu Gakkai Zasshi. 2005 Dec 20;61(12):1632-7 20966324 - AJR Am J Roentgenol. 2010 Nov;195(5):1175-9 19933664 - AJR Am J Roentgenol. 2009 Dec;193(6):1680-5 21081517 - Radiat Prot Dosimetry. 2011 Jan;143(1):74-80 19188512 - Circulation. 2009 Feb 24;119(7):1056-65 19620457 - AJR Am J Roentgenol. 2009 Aug;193(2):559-69 17885036 - AJR Am J Roentgenol. 2007 Oct;189(4):W224-7 19329511 - Radiat Prot Dosimetry. 2009 Feb;133(4):227-33 20821130 - Radiol Phys Technol. 2009 Jan;2(1):58-61 20947586 - Radiat Prot Dosimetry. 2011 Jan;143(1):69-73 20093507 - Radiology. 2010 Feb;254(2):326-41 16498105 - AJR Am J Roentgenol. 2006 Mar;186(3):774-8 15582335 - J Am Coll Cardiol. 2004 Dec 7;44(11):2259-82 22021539 - AJR Am J Roentgenol. 2011 Nov;197(5):W900-3 19961398 - Acta Radiol. 2010 Feb;51(1):33-9 K Chida (121_CR6) 2010; 51 K Chida (121_CR12) 2009; 2 M Kato (121_CR8) 2011; 143 K Chida (121_CR14) 2009; 193 K Chida (121_CR5) 2010; 51 K Chida (121_CR10) 2006; 186 121_CR18 S Balter (121_CR7) 2010; 254 L Wong (121_CR3) 2004; 350 121_CR1 Y Inaba (121_CR17) 2011; 143 KP Kim (121_CR11) 2009; 133 121_CR21 K Chida (121_CR9) 2004; 183 V Tsapaki (121_CR15) 2009; 193 K Chida (121_CR16) 2010; 195 121_CR20 JW Hirshfeld (121_CR2) 2004; 44 K Chida (121_CR4) 2005; 61 TC Gerber (121_CR13) 2009; 119 K Chida (121_CR19) 2007; 189 |
References_xml | – volume: 193 start-page: 1680 year: 2009 ident: 121_CR14 publication-title: Am J Roentgenol doi: 10.2214/AJR.09.2747 contributor: fullname: K Chida – volume: 143 start-page: 69 year: 2011 ident: 121_CR8 publication-title: Radiat Prot Dosimetry doi: 10.1093/rpd/ncq286 contributor: fullname: M Kato – volume: 193 start-page: 559 year: 2009 ident: 121_CR15 publication-title: Am J Roentgenol doi: 10.2214/AJR.08.2115 contributor: fullname: V Tsapaki – ident: 121_CR21 doi: 10.2214/AJR.10.6396 – ident: 121_CR18 – ident: 121_CR20 – volume: 186 start-page: 774 year: 2006 ident: 121_CR10 publication-title: Am J Roentgenol doi: 10.2214/AJR.04.1653 contributor: fullname: K Chida – volume: 51 start-page: 97 year: 2010 ident: 121_CR6 publication-title: J Radiat Res doi: 10.1269/jrr.09112 contributor: fullname: K Chida – volume: 2 start-page: 58 year: 2009 ident: 121_CR12 publication-title: Radiol Phys Technol doi: 10.1007/s12194-008-0044-z contributor: fullname: K Chida – ident: 121_CR1 doi: 10.1016/S0146-6453(01)00004-5 – volume: 189 start-page: W224 year: 2007 ident: 121_CR19 publication-title: Am J Roentgenol doi: 10.2214/AJR.07.2422 contributor: fullname: K Chida – volume: 195 start-page: 1175 year: 2010 ident: 121_CR16 publication-title: Am J Roentgenol doi: 10.2214/AJR.10.4466 contributor: fullname: K Chida – volume: 44 start-page: 2259 year: 2004 ident: 121_CR2 publication-title: J Am Coll Cardiol doi: 10.1016/j.jacc.2004.10.014 contributor: fullname: JW Hirshfeld – volume: 350 start-page: e23 year: 2004 ident: 121_CR3 publication-title: N Engl J Med doi: 10.1056/NEJMicm030503 contributor: fullname: L Wong – volume: 51 start-page: 33 year: 2010 ident: 121_CR5 publication-title: Acta Radiol doi: 10.3109/02841850903229141 contributor: fullname: K Chida – volume: 254 start-page: 326 year: 2010 ident: 121_CR7 publication-title: Radiology doi: 10.1148/radiol.2542082312 contributor: fullname: S Balter – volume: 183 start-page: 1111 year: 2004 ident: 121_CR9 publication-title: Am J Roentgenol doi: 10.2214/ajr.183.4.1831111 contributor: fullname: K Chida – volume: 143 start-page: 74 year: 2011 ident: 121_CR17 publication-title: Radiat Prot Dosimetry doi: 10.1093/rpd/ncq287 contributor: fullname: Y Inaba – volume: 61 start-page: 1632 year: 2005 ident: 121_CR4 publication-title: Nippon Hoshasen Gijutsu Gakkai Zasshi doi: 10.6009/jjrt.KJ00004022974 contributor: fullname: K Chida – volume: 133 start-page: 227 year: 2009 ident: 121_CR11 publication-title: Radiat Prot Dosimetry doi: 10.1093/rpd/ncp052 contributor: fullname: KP Kim – volume: 119 start-page: 1056 year: 2009 ident: 121_CR13 publication-title: Circulation doi: 10.1161/CIRCULATIONAHA.108.191650 contributor: fullname: TC Gerber |
SSID | ssj0061222 |
Score | 2.0009162 |
Snippet | Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician).... |
SourceID | crossref pubmed springer |
SourceType | Aggregation Database Index Database Publisher |
StartPage | 189 |
SubjectTerms | Humans Imaging Japan Medical and Radiation Physics Medicine Medicine & Public Health Nuclear Medicine Radiation Dosage Radiation Injuries - etiology Radiation Injuries - prevention & control Radiology Radiology, Interventional - instrumentation Radiology, Interventional - methods Radiology, Interventional - standards Radiotherapy Reference Standards Safety Management - methods Safety Management - standards X-Rays |
Title | Comparison of dose at an interventional reference point between the displayed estimated value and measured value |
URI | https://link.springer.com/article/10.1007/s12194-011-0121-6 https://www.ncbi.nlm.nih.gov/pubmed/21643656 |
Volume | 4 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3JbsIwELVYpKqX7gtdkA89tTJKYsdJjoCgqBWcikRPke04EqINqIRD-_Udk4WuB25RYo0Uz3jmjWdD6MYTysBoQZgtGWGxpYmQXBNQfdTzI9dT3NQOD0d8MGYPE3dSQU55dZHMWkVEcq2oN7VucLZMwoTxfh2b8Cqq53Wn9fb982Ov0L9gstexA9vnLrEopUUs8y8i36xRaYJ-xETXpqa_n5X_LdcdCk2Gyay1SmVLffzu37jFXxygvRx54nYmKoeoopMjtDPMY-vHaNEtRxLieYyj-VJjkWKR4OmXvEigUE4mwYs5fMJ5qhcGKImj6XLxIt51hE33DkDD8GT6iQOpJMKv2YVk_uoEjfu9p-6A5PMYiGKUpoQraXHpB-ADaur7KmaABpSvPdD3gatNp7vYsnSgbVsK5liaB4EUTuS6yjZjA-kpqiXzRJ8j7HleDMIQUaoZEwAbmKCSO1KAP6e4zxvotuBLuMjaboSbBstmD0PYw9DsYQiLzzLOlUsdcAIpINUGuivYEOZnc_k_nYutVl-i3ex-2aTuXqFa-rbS1wBQUtkEiex3OqNmLplNVB077U9kAN0q |
link.rule.ids | 315,783,787,27936,27937,41093,41535,42162,42604,52123,52246 |
linkProvider | Springer Nature |
linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV07T8MwELagSMCCeFOeHphAlpLYsZOxqqgKtJ1aqVtkO45UCZKIhIF_zzmvAoKBLUpON5zju-985-8QuhVSWxgtCXMVIyxxDJGKGwKuj4og9oXm9u7wdMbHC_a09JfNPe6i7XZvS5KVp15fdoPNZTsmbPrruYRvoi1Lr24J8xfeoHW_ELGr0oEbcJ84lNK2lPmbim_BqItAP0qiVaQZ7aO9BiLiQb2mB2jDpIdoe9oUwY9QPuxmB-IswXFWGCxLLFO8-tLACBq6ESI4z-ATbnqyMGA-HK-K_EV-mBhbmg2ArfBkib9BVRrj1_rksHl1jBajh_lwTJrBCUQzSkvCtXK4CkJI1gwNAp0wCNs6MAIcc-gbS0mXOI4JjesqyTzH8DBU0ot9X7t2vh89Qb00S80ZwkKIBFYtptQwJiG-M0kV95SExEvzgPfRXWvBKK_5MaI1E7I1dwTmjqy5IxA-rW3ciXqQrVGAlH103xo9ajZR8bee839J36Cd8Xw6iSaPs-cLtFsfCtt-20vUK9_ezRWgilJdV3_RJ7wNweU |
linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV07T8MwELagSBUL4k15emACWU1ix0nGqlCVRysGKnWL_IpUCZKIhIF_z7l5FBAMbFFyuuFzfN-d73yH0GUglHWjBWGuZIQljiFCckPA9NEg1H6guL07PJny8Yzdz_15Pee0aKrdm5RkdafBdmlKy36uk_7q4htsNFs9YUNhzyV8HW0AE1Fb0zfzBo0pBvZephHckPvEoZQ2ac3fVHwjppaNfqRHl6wz2kZbtbuIB9X67qA1k-6i7qROiO-hfNjOEcRZgnVWGCxKLFK8-FLMCBracSI4z-ATruuzMPh_WC-K_EV8GI1tyw1wYeHJNgEHVanGr9UpYv1qH81Gt8_DMamHKBDFKC0JV9LhMowgcDM0DFXCgMJVaAIw0pFvbHu6xHFMZFxXCuY5hkeRFJ72feXaWX_0AHXSLDVHCAdBkMAKakoNYwK4ngkquScFBGGKh7yHrhoE47zqlRGvuiJbuGOAO7ZwxyB8WGHcinoQuVFwL3vougE9rjdU8bee439JX6Du080ofrybPpygzep82JbenqJO-fZuzsDBKOX58if6BFkYxio |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Comparison+of+dose+at+an+interventional+reference+point+between+the+displayed+estimated+value+and+measured+value&rft.jtitle=Radiological+physics+and+technology&rft.au=Chida%2C+Koichi&rft.au=Inaba%2C+Yohei&rft.au=Morishima%2C+Yoshiaki&rft.au=Taura%2C+Masaaki&rft.date=2011-07-01&rft.issn=1865-0333&rft.eissn=1865-0341&rft.volume=4&rft.issue=2&rft.spage=189&rft.epage=193&rft_id=info:doi/10.1007%2Fs12194-011-0121-6&rft.externalDBID=n%2Fa&rft.externalDocID=10_1007_s12194_011_0121_6 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1865-0333&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1865-0333&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1865-0333&client=summon |