Comparison of dose at an interventional reference point between the displayed estimated value and measured value

Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in re...

Full description

Saved in:
Bibliographic Details
Published inRadiological physics and technology Vol. 4; no. 2; pp. 189 - 193
Main Authors Chida, Koichi, Inaba, Yohei, Morishima, Yoshiaki, Taura, Masaaki, Ebata, Ayako, Yanagawa, Isao, Takeda, Ken, Zuguchi, Masayuki
Format Journal Article
LanguageEnglish
Published Japan Springer Japan 01.07.2011
Subjects
Online AccessGet full text

Cover

Loading…
Abstract Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in real time. However, no detailed report has evaluated the accuracy of the reference doses displayed on X-ray equipment. Thus, in this study, we compared the displayed reference dose to the actual measured value in many IR X-ray systems. Although the displayed reference doses of many IR X-ray systems agreed with the measured actual values within approximately 15%, the doses of a few IR units were not close. Furthermore, some X-ray units made in Japan displayed reference doses quite different from the actual measured value, probably because the reference point of these units differs from the International Electrotechnical Commission standard. Thus, IR physicians should pay attention to the location of the IRP of the displayed reference dose in Japan. Furthermore, physicians should be aware of the accuracy of the displayed reference dose of the X-ray system that they use for IR. Thus, regular checks of the displayed reference dose of the X-ray system are important.
AbstractList Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician). The dose displayed at the IRP (the reference dose) of an X-ray unit has been reported to be helpful for characterizing patient exposure in real time. However, no detailed report has evaluated the accuracy of the reference doses displayed on X-ray equipment. Thus, in this study, we compared the displayed reference dose to the actual measured value in many IR X-ray systems. Although the displayed reference doses of many IR X-ray systems agreed with the measured actual values within approximately 15%, the doses of a few IR units were not close. Furthermore, some X-ray units made in Japan displayed reference doses quite different from the actual measured value, probably because the reference point of these units differs from the International Electrotechnical Commission standard. Thus, IR physicians should pay attention to the location of the IRP of the displayed reference dose in Japan. Furthermore, physicians should be aware of the accuracy of the displayed reference dose of the X-ray system that they use for IR. Thus, regular checks of the displayed reference dose of the X-ray system are important.
Author Yanagawa, Isao
Chida, Koichi
Ebata, Ayako
Zuguchi, Masayuki
Morishima, Yoshiaki
Taura, Masaaki
Inaba, Yohei
Takeda, Ken
Author_xml – sequence: 1
  givenname: Koichi
  surname: Chida
  fullname: Chida, Koichi
  email: chida@mail.tains.tohoku.ac.jp
  organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University
– sequence: 2
  givenname: Yohei
  surname: Inaba
  fullname: Inaba, Yohei
  organization: Department of Radiology, Tohoku University Hospital
– sequence: 3
  givenname: Yoshiaki
  surname: Morishima
  fullname: Morishima, Yoshiaki
  organization: Department of Radiology, Tohoku Employees’ Pension Welfare Hospital
– sequence: 4
  givenname: Masaaki
  surname: Taura
  fullname: Taura, Masaaki
  organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University
– sequence: 5
  givenname: Ayako
  surname: Ebata
  fullname: Ebata, Ayako
  organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University
– sequence: 6
  givenname: Isao
  surname: Yanagawa
  fullname: Yanagawa, Isao
  organization: Department of Radiology, Tohoku University Hospital
– sequence: 7
  givenname: Ken
  surname: Takeda
  fullname: Takeda, Ken
  organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University
– sequence: 8
  givenname: Masayuki
  surname: Zuguchi
  fullname: Zuguchi, Masayuki
  organization: Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University
BackLink https://www.ncbi.nlm.nih.gov/pubmed/21643656$$D View this record in MEDLINE/PubMed
BookMark eNp9kMtOwzAQRS1URB_wAWyQfyDgsR03WaKKl1SJDawtJ5lAqsSO7KSof4-r0C5ZjGY0M_fq6izJzDqLhNwCuwfG1g8BOOQyYQCxOCTqgiwgU2nChITZeRZiTpYh7BhTwDm_InMOSgqVqgXpN67rjW-Cs9TVtHIBqRmosbSxA_o92qFx1rTUY40ebYm0d_FECxx-EC0dvpFWTehbc8CKYhiazgxx2pt2jFa2oh2aMPrT6ppc1qYNePPXV-Tz-elj85ps31_eNo_bpJRCDIkqC6aKLJdsjSLLylpKpsoM11hAnqISqaoZwxwBCiM5Q5XnheFVmpYgVK7EisDkW3oXQkyvex-j-YMGpo_09ERPR3r6SE8fNXeTph-LDquz4oQrPvDpIcST_UKvd270EU_4x_UXwst9ow
CitedBy_id crossref_primary_10_7793_jcoron_23_16_00022
crossref_primary_10_1007_s12194_022_00650_w
crossref_primary_10_3390_bioengineering9080354
crossref_primary_10_3390_diagnostics11091613
crossref_primary_10_1007_s00455_022_10411_x
crossref_primary_10_3390_s23010512
crossref_primary_10_1007_s12194_018_0484_z
crossref_primary_10_6009_jjrt_2013_JSRT_69_10_1173
crossref_primary_10_6009_jjrt_2016_JSRT_72_8_689
crossref_primary_10_3390_s20092741
crossref_primary_10_1007_s00455_016_9718_6
crossref_primary_10_3390_bioengineering10020259
crossref_primary_10_3390_tomography8010036
crossref_primary_10_6009_jjrt_2022_1309
crossref_primary_10_1620_tjem_2022_J001
crossref_primary_10_1007_s12194_022_00660_8
crossref_primary_10_1111_pace_12790
crossref_primary_10_6009_jjrt_2019_JSRT_75_7_646
crossref_primary_10_1007_s12194_020_00575_2
crossref_primary_10_3390_diagnostics10100753
crossref_primary_10_6009_jjrt_2015_JSRT_71_12_1215
crossref_primary_10_1093_jrr_rrx039
crossref_primary_10_1088_1742_6596_694_1_012049
crossref_primary_10_3390_diagnostics10100846
crossref_primary_10_3390_diagnostics13183003
crossref_primary_10_1088_1361_6560_ab0bd5
crossref_primary_10_1088_1361_6498_ad0a9d
crossref_primary_10_1093_rpd_ncae058
Cites_doi 10.2214/AJR.09.2747
10.1093/rpd/ncq286
10.2214/AJR.08.2115
10.2214/AJR.10.6396
10.2214/AJR.04.1653
10.1269/jrr.09112
10.1007/s12194-008-0044-z
10.1016/S0146-6453(01)00004-5
10.2214/AJR.07.2422
10.2214/AJR.10.4466
10.1016/j.jacc.2004.10.014
10.1056/NEJMicm030503
10.3109/02841850903229141
10.1148/radiol.2542082312
10.2214/ajr.183.4.1831111
10.1093/rpd/ncq287
10.6009/jjrt.KJ00004022974
10.1093/rpd/ncp052
10.1161/CIRCULATIONAHA.108.191650
ContentType Journal Article
Copyright Japanese Society of Radiological Technology and Japan Society of Medical Physics 2011
Copyright_xml – notice: Japanese Society of Radiological Technology and Japan Society of Medical Physics 2011
DBID CGR
CUY
CVF
ECM
EIF
NPM
AAYXX
CITATION
DOI 10.1007/s12194-011-0121-6
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
CrossRef
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
CrossRef
DatabaseTitleList
MEDLINE
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
EISSN 1865-0341
EndPage 193
ExternalDocumentID 10_1007_s12194_011_0121_6
21643656
Genre Journal Article
Comparative Study
GeographicLocations Japan
GeographicLocations_xml – name: Japan
GroupedDBID ---
-5E
-5G
-BR
-EM
-Y2
-~C
.86
06C
06D
0R~
0VY
123
1N0
203
29P
29~
2JN
2JY
2KG
2VQ
2~H
30V
4.4
406
408
409
40D
5VS
67Z
6NX
8TC
96X
AAAVM
AABHQ
AAFGU
AAHNG
AAIAL
AAJKR
AANXM
AANZL
AAPBV
AARHV
AARTL
AATNV
AATVU
AAUYE
AAWCG
AAYFA
AAYIU
AAYQN
AAYTO
ABBXA
ABDZT
ABECU
ABFGW
ABFTV
ABHQN
ABJNI
ABJOX
ABKAS
ABKCH
ABMNI
ABMQK
ABPLI
ABQBU
ABSXP
ABTEG
ABTKH
ABTMW
ABULA
ABWNU
ABXPI
ACBMV
ACBRV
ACBYP
ACGFS
ACHSB
ACIGE
ACIPQ
ACKNC
ACMDZ
ACMLO
ACOKC
ACOMO
ACTTH
ACVWB
ACWMK
ADHHG
ADHIR
ADINQ
ADKNI
ADKPE
ADMDM
ADOXG
ADRFC
ADTPH
ADURQ
ADYFF
ADZKW
AEBTG
AEFTE
AEGAL
AEGNC
AEJHL
AEJRE
AEOHA
AEPYU
AESKC
AESTI
AETLH
AEVLU
AEVTX
AEXYK
AFLOW
AFNRJ
AFQWF
AFWTZ
AFZKB
AGAYW
AGDGC
AGGBP
AGJBK
AGMZJ
AGQMX
AGWZB
AGYKE
AHAVH
AHBYD
AHSBF
AHYZX
AIAKS
AIIXL
AILAN
AIMYW
AITGF
AJBLW
AJDOV
AJRNO
AJZVZ
AKMHD
AKQUC
ALFXC
ALMA_UNASSIGNED_HOLDINGS
ALWAN
AMKLP
AMXSW
AMYLF
AMYQR
ANMIH
AOCGG
AUKKA
AXYYD
B-.
BA0
BDATZ
BGNMA
CAG
COF
CS3
CSCUP
DDRTE
DNIVK
DPUIP
EBLON
EBS
EIOEI
EJD
EN4
ESBYG
F5P
FERAY
FFXSO
FIGPU
FINBP
FNLPD
FRRFC
FSGXE
FWDCC
G-Y
G-Z
GGCAI
GGRSB
GJIRD
GNWQR
GQ6
GQ7
GQ8
GXS
HF~
HG5
HG6
HLICF
HMJXF
HQYDN
HRMNR
HZ~
I0C
IJ-
IKXTQ
IWAJR
IXC
IXD
IZIGR
IZQ
I~X
J-C
J0Z
JBSCW
JCJTX
JZLTJ
KOV
KPH
LLZTM
M4Y
MA-
NPVJJ
NQJWS
NU0
O9-
O93
O9I
O9J
OAM
P9S
PT4
QOR
QOS
R89
RIG
RLLFE
RNS
ROL
RPX
RSV
S16
S1Z
S27
S37
S3B
SAP
SDH
SHX
SISQX
SMD
SNE
SNPRN
SNX
SOHCF
SOJ
SPISZ
SRMVM
SSLCW
SSXJD
STPWE
SZ9
T13
TSG
TSK
TT1
U2A
U9L
UG4
UOJIU
UTJUX
UZXMN
VC2
VFIZW
W48
WJK
WK8
YLTOR
Z45
Z7X
Z83
ZMTXR
ZOVNA
~A9
AACDK
AAJBT
AASML
AAYZH
ABAKF
ACAOD
ACDTI
ACZOJ
AEFQL
AEMSY
AFBBN
AGQEE
AGRTI
AIGIU
CGR
CUY
CVF
ECM
EIF
H13
NPM
SJYHP
AAYXX
CITATION
ID FETCH-LOGICAL-c433t-6cb06b89407e388cf4406c8e7eb195e6356f00e9e11ba420e699ba2d55c136963
IEDL.DBID AGYKE
ISSN 1865-0333
IngestDate Thu Sep 12 19:32:40 EDT 2024
Tue Oct 15 23:45:47 EDT 2024
Sat Dec 16 12:02:49 EST 2023
IsPeerReviewed true
IsScholarly true
Issue 2
Keywords Radiation injury
Radiation dosimetry
Exposure dose
Radiation safety
Interventional radiology
Interventional reference point (IRP)
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c433t-6cb06b89407e388cf4406c8e7eb195e6356f00e9e11ba420e699ba2d55c136963
PMID 21643656
PageCount 5
ParticipantIDs crossref_primary_10_1007_s12194_011_0121_6
pubmed_primary_21643656
springer_journals_10_1007_s12194_011_0121_6
PublicationCentury 2000
PublicationDate 2011-07-01
PublicationDateYYYYMMDD 2011-07-01
PublicationDate_xml – month: 07
  year: 2011
  text: 2011-07-01
  day: 01
PublicationDecade 2010
PublicationPlace Japan
PublicationPlace_xml – name: Japan
PublicationTitle Radiological physics and technology
PublicationTitleAbbrev Radiol Phys Technol
PublicationTitleAlternate Radiol Phys Technol
PublicationYear 2011
Publisher Springer Japan
Publisher_xml – name: Springer Japan
References ChidaKKatoMSaitoHIshibashiTTakahashiSKohzukiMOptimizing patient radiation dose in intervention proceduresActa Radiol20105133391996139810.3109/02841850903229141
ChidaKSaitoHOtaniHKohzukiMTakahashiSYamadaSRelationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional proceduresAm J Roentgenol200618677477810.2214/AJR.04.1653
TsapakiVAhmedNAAlSuwaidiJSBeganovicABeniderABenOmraneLRadiation exposure to patients during interventional procedures in 20 countries: initial IAEA project resultsAm J Roentgenol200919355956910.2214/AJR.08.2115
ChidaKKagayaYSaitoHTakaiYTakahashiSYamadaSTotal entrance skin dose: an effective indicator of the maximum radiation dose to a patient’s skin during percutaneous coronary interventionAm J Roentgenol2007189W224W22710.2214/AJR.07.2422
HirshfeldJWBalterSBrinkerJAACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures: a report of the American College of Cardiology Foundation/American Heart Association American. College of Physicians Task Force on Clinical Competence and TrainingJ Am Coll Cardiol200444225922821558233510.1016/j.jacc.2004.10.014
KatoMChidaKSatoTOosakaHTosaTKadowakiKEvaluating the maximum patient radiation dose in cardiac interventional proceduresRadiat Prot Dosimetry201114369732094758610.1093/rpd/ncq286
GerberTCCarrJJAraiAEDixonRLFerrariVAGomesASIonizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and InterventionCirculation2009119105610651918851210.1161/CIRCULATIONAHA.108.191650
ChidaKOhnoTKakizakiSTakegawaMYuukiHNakadaMRadiation dose to the pediatric cardiac catheterization and intervention patientAm J Roentgenol20101951175117910.2214/AJR.10.4466
KimKPMillerDLMinimising radiation exposure to physicians performing fluoroscopically guided cardiac catheterisation procedures: a reviewRadiat Prot Dosimetry20091332272331932951110.1093/rpd/ncp052
ChidaKInabaYMasuyamaHYanagawaIMoriISaitoHEvaluating the performance of a MOSFET dosimeter at diagnostic X-ray energies for interventional radiologyRadiol Phys Technol2009258612082113010.1007/s12194-008-0044-z
InabaYChidaKShirotoriKShimuraHYanagawaIZuguchiMComparison of the radiation dose in a cardiac IVR X-ray systemRadiat Prot Dosimetry201114374802108151710.1093/rpd/ncq287
International Commission on Radiological Protection. ICRP publication 85: avoidance of radiation injuries from medical interventional procedures. Ann ICRP 2001 (30/2:Publication 85).
ChidaKMorishimaYKatahiraYChibaHZuguchiMEvaluation of additional lead shielding in protecting the physician from radiation during cardiac interventional proceduresNippon Hoshasen Gijutsu Gakkai Zasshi2005611632163716395238
Japanese Industrial Standards (JIS). Medical electrical equipment, Particular requirements for the safety of X-ray equipment for interventional procedures, Z 4751-2-43, 2005.
BalterSHopewellJWMillerDLWagnerLKZelefskyMJFluoroscopically guided interventional procedures: a review of radiation effects on patients’ skin and hairRadiology20102543263412009350710.1148/radiol.2542082312
ChidaKSaitoHZuguchiMShirotoriKKumagaiSNakayamaHDoes digital acquisition reduce patients’ skin dose in cardiac interventional procedures? An experimental studyAm J Roentgenol200418311111114
International Electrotechnical Commission (IEC). Medical electrical equipment. Part 2–43. Particular requirements for the safety of X-ray equipment for interventional procedures. IEC 60601-2-43. Geneva, Switzerland: International Electrotechnical Commission; 2000.
Chida K, Takahashi T, Ito D, Shimura H, Takeda K, Zuguchi M. Clarifying and visualizing sources of staff-received scattered radiation in interventional procedures. Am J Roentgenol. 2011 (in press).
WongLRehmJImages in clinical medicine: radiation injury from a fluoroscopic procedureN Engl J Med2004350e231520142810.1056/NEJMicm030503
ChidaKKatoMKagayaYZuguchiMSaitoHIshibashiTRadiation dose and radiation protection for patients and physicians during interventional procedureJ Radiat Res201051971052033925310.1269/jrr.09112
ChidaKInabaYSaitoHIshibashiTTakahashiSKohzukiMRadiation dose of interventional radiology system using a flat-panel detectorAm J Roentgenol20091931680168510.2214/AJR.09.2747
15385315 - AJR Am J Roentgenol. 2004 Oct;183(4):1111-4
20339253 - J Radiat Res. 2010;51(2):97-105
15201428 - N Engl J Med. 2004 Jun 17;350(25):e23
16395238 - Nihon Hoshasen Gijutsu Gakkai Zasshi. 2005 Dec 20;61(12):1632-7
20966324 - AJR Am J Roentgenol. 2010 Nov;195(5):1175-9
19933664 - AJR Am J Roentgenol. 2009 Dec;193(6):1680-5
21081517 - Radiat Prot Dosimetry. 2011 Jan;143(1):74-80
19188512 - Circulation. 2009 Feb 24;119(7):1056-65
19620457 - AJR Am J Roentgenol. 2009 Aug;193(2):559-69
17885036 - AJR Am J Roentgenol. 2007 Oct;189(4):W224-7
19329511 - Radiat Prot Dosimetry. 2009 Feb;133(4):227-33
20821130 - Radiol Phys Technol. 2009 Jan;2(1):58-61
20947586 - Radiat Prot Dosimetry. 2011 Jan;143(1):69-73
20093507 - Radiology. 2010 Feb;254(2):326-41
16498105 - AJR Am J Roentgenol. 2006 Mar;186(3):774-8
15582335 - J Am Coll Cardiol. 2004 Dec 7;44(11):2259-82
22021539 - AJR Am J Roentgenol. 2011 Nov;197(5):W900-3
19961398 - Acta Radiol. 2010 Feb;51(1):33-9
K Chida (121_CR6) 2010; 51
K Chida (121_CR12) 2009; 2
M Kato (121_CR8) 2011; 143
K Chida (121_CR14) 2009; 193
K Chida (121_CR5) 2010; 51
K Chida (121_CR10) 2006; 186
121_CR18
S Balter (121_CR7) 2010; 254
L Wong (121_CR3) 2004; 350
121_CR1
Y Inaba (121_CR17) 2011; 143
KP Kim (121_CR11) 2009; 133
121_CR21
K Chida (121_CR9) 2004; 183
V Tsapaki (121_CR15) 2009; 193
K Chida (121_CR16) 2010; 195
121_CR20
JW Hirshfeld (121_CR2) 2004; 44
K Chida (121_CR4) 2005; 61
TC Gerber (121_CR13) 2009; 119
K Chida (121_CR19) 2007; 189
References_xml – volume: 193
  start-page: 1680
  year: 2009
  ident: 121_CR14
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.09.2747
  contributor:
    fullname: K Chida
– volume: 143
  start-page: 69
  year: 2011
  ident: 121_CR8
  publication-title: Radiat Prot Dosimetry
  doi: 10.1093/rpd/ncq286
  contributor:
    fullname: M Kato
– volume: 193
  start-page: 559
  year: 2009
  ident: 121_CR15
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.08.2115
  contributor:
    fullname: V Tsapaki
– ident: 121_CR21
  doi: 10.2214/AJR.10.6396
– ident: 121_CR18
– ident: 121_CR20
– volume: 186
  start-page: 774
  year: 2006
  ident: 121_CR10
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.04.1653
  contributor:
    fullname: K Chida
– volume: 51
  start-page: 97
  year: 2010
  ident: 121_CR6
  publication-title: J Radiat Res
  doi: 10.1269/jrr.09112
  contributor:
    fullname: K Chida
– volume: 2
  start-page: 58
  year: 2009
  ident: 121_CR12
  publication-title: Radiol Phys Technol
  doi: 10.1007/s12194-008-0044-z
  contributor:
    fullname: K Chida
– ident: 121_CR1
  doi: 10.1016/S0146-6453(01)00004-5
– volume: 189
  start-page: W224
  year: 2007
  ident: 121_CR19
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.07.2422
  contributor:
    fullname: K Chida
– volume: 195
  start-page: 1175
  year: 2010
  ident: 121_CR16
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.10.4466
  contributor:
    fullname: K Chida
– volume: 44
  start-page: 2259
  year: 2004
  ident: 121_CR2
  publication-title: J Am Coll Cardiol
  doi: 10.1016/j.jacc.2004.10.014
  contributor:
    fullname: JW Hirshfeld
– volume: 350
  start-page: e23
  year: 2004
  ident: 121_CR3
  publication-title: N Engl J Med
  doi: 10.1056/NEJMicm030503
  contributor:
    fullname: L Wong
– volume: 51
  start-page: 33
  year: 2010
  ident: 121_CR5
  publication-title: Acta Radiol
  doi: 10.3109/02841850903229141
  contributor:
    fullname: K Chida
– volume: 254
  start-page: 326
  year: 2010
  ident: 121_CR7
  publication-title: Radiology
  doi: 10.1148/radiol.2542082312
  contributor:
    fullname: S Balter
– volume: 183
  start-page: 1111
  year: 2004
  ident: 121_CR9
  publication-title: Am J Roentgenol
  doi: 10.2214/ajr.183.4.1831111
  contributor:
    fullname: K Chida
– volume: 143
  start-page: 74
  year: 2011
  ident: 121_CR17
  publication-title: Radiat Prot Dosimetry
  doi: 10.1093/rpd/ncq287
  contributor:
    fullname: Y Inaba
– volume: 61
  start-page: 1632
  year: 2005
  ident: 121_CR4
  publication-title: Nippon Hoshasen Gijutsu Gakkai Zasshi
  doi: 10.6009/jjrt.KJ00004022974
  contributor:
    fullname: K Chida
– volume: 133
  start-page: 227
  year: 2009
  ident: 121_CR11
  publication-title: Radiat Prot Dosimetry
  doi: 10.1093/rpd/ncp052
  contributor:
    fullname: KP Kim
– volume: 119
  start-page: 1056
  year: 2009
  ident: 121_CR13
  publication-title: Circulation
  doi: 10.1161/CIRCULATIONAHA.108.191650
  contributor:
    fullname: TC Gerber
SSID ssj0061222
Score 2.0009162
Snippet Today, interventional radiology (IR) X-ray units are required for display of doses at an interventional reference point (IRP) for the operator (IR physician)....
SourceID crossref
pubmed
springer
SourceType Aggregation Database
Index Database
Publisher
StartPage 189
SubjectTerms Humans
Imaging
Japan
Medical and Radiation Physics
Medicine
Medicine & Public Health
Nuclear Medicine
Radiation Dosage
Radiation Injuries - etiology
Radiation Injuries - prevention & control
Radiology
Radiology, Interventional - instrumentation
Radiology, Interventional - methods
Radiology, Interventional - standards
Radiotherapy
Reference Standards
Safety Management - methods
Safety Management - standards
X-Rays
Title Comparison of dose at an interventional reference point between the displayed estimated value and measured value
URI https://link.springer.com/article/10.1007/s12194-011-0121-6
https://www.ncbi.nlm.nih.gov/pubmed/21643656
Volume 4
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3JbsIwELVYpKqX7gtdkA89tTJKYsdJjoCgqBWcikRPke04EqINqIRD-_Udk4WuB25RYo0Uz3jmjWdD6MYTysBoQZgtGWGxpYmQXBNQfdTzI9dT3NQOD0d8MGYPE3dSQU55dZHMWkVEcq2oN7VucLZMwoTxfh2b8Cqq53Wn9fb982Ov0L9gstexA9vnLrEopUUs8y8i36xRaYJ-xETXpqa_n5X_LdcdCk2Gyay1SmVLffzu37jFXxygvRx54nYmKoeoopMjtDPMY-vHaNEtRxLieYyj-VJjkWKR4OmXvEigUE4mwYs5fMJ5qhcGKImj6XLxIt51hE33DkDD8GT6iQOpJMKv2YVk_uoEjfu9p-6A5PMYiGKUpoQraXHpB-ADaur7KmaABpSvPdD3gatNp7vYsnSgbVsK5liaB4EUTuS6yjZjA-kpqiXzRJ8j7HleDMIQUaoZEwAbmKCSO1KAP6e4zxvotuBLuMjaboSbBstmD0PYw9DsYQiLzzLOlUsdcAIpINUGuivYEOZnc_k_nYutVl-i3ex-2aTuXqFa-rbS1wBQUtkEiex3OqNmLplNVB077U9kAN0q
link.rule.ids 315,783,787,27936,27937,41093,41535,42162,42604,52123,52246
linkProvider Springer Nature
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV07T8MwELagSMCCeFOeHphAlpLYsZOxqqgKtJ1aqVtkO45UCZKIhIF_zzmvAoKBLUpON5zju-985-8QuhVSWxgtCXMVIyxxDJGKGwKuj4og9oXm9u7wdMbHC_a09JfNPe6i7XZvS5KVp15fdoPNZTsmbPrruYRvoi1Lr24J8xfeoHW_ELGr0oEbcJ84lNK2lPmbim_BqItAP0qiVaQZ7aO9BiLiQb2mB2jDpIdoe9oUwY9QPuxmB-IswXFWGCxLLFO8-tLACBq6ESI4z-ATbnqyMGA-HK-K_EV-mBhbmg2ArfBkib9BVRrj1_rksHl1jBajh_lwTJrBCUQzSkvCtXK4CkJI1gwNAp0wCNs6MAIcc-gbS0mXOI4JjesqyTzH8DBU0ot9X7t2vh89Qb00S80ZwkKIBFYtptQwJiG-M0kV95SExEvzgPfRXWvBKK_5MaI1E7I1dwTmjqy5IxA-rW3ciXqQrVGAlH103xo9ajZR8bee839J36Cd8Xw6iSaPs-cLtFsfCtt-20vUK9_ezRWgilJdV3_RJ7wNweU
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV07T8MwELagSBUL4k15emACWU1ix0nGqlCVRysGKnWL_IpUCZKIhIF_z7l5FBAMbFFyuuFzfN-d73yH0GUglHWjBWGuZIQljiFCckPA9NEg1H6guL07PJny8Yzdz_15Pee0aKrdm5RkdafBdmlKy36uk_7q4htsNFs9YUNhzyV8HW0AE1Fb0zfzBo0pBvZephHckPvEoZQ2ac3fVHwjppaNfqRHl6wz2kZbtbuIB9X67qA1k-6i7qROiO-hfNjOEcRZgnVWGCxKLFK8-FLMCBracSI4z-ATruuzMPh_WC-K_EV8GI1tyw1wYeHJNgEHVanGr9UpYv1qH81Gt8_DMamHKBDFKC0JV9LhMowgcDM0DFXCgMJVaAIw0pFvbHu6xHFMZFxXCuY5hkeRFJ72feXaWX_0AHXSLDVHCAdBkMAKakoNYwK4ngkquScFBGGKh7yHrhoE47zqlRGvuiJbuGOAO7ZwxyB8WGHcinoQuVFwL3vougE9rjdU8bee439JX6Du080ofrybPpygzep82JbenqJO-fZuzsDBKOX58if6BFkYxio
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Comparison+of+dose+at+an+interventional+reference+point+between+the+displayed+estimated+value+and+measured+value&rft.jtitle=Radiological+physics+and+technology&rft.au=Chida%2C+Koichi&rft.au=Inaba%2C+Yohei&rft.au=Morishima%2C+Yoshiaki&rft.au=Taura%2C+Masaaki&rft.date=2011-07-01&rft.issn=1865-0333&rft.eissn=1865-0341&rft.volume=4&rft.issue=2&rft.spage=189&rft.epage=193&rft_id=info:doi/10.1007%2Fs12194-011-0121-6&rft.externalDBID=n%2Fa&rft.externalDocID=10_1007_s12194_011_0121_6
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1865-0333&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1865-0333&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1865-0333&client=summon