Can a partial volume edge effect reduction algorithm improve the repeatability of subject-specific finite element models of femurs obtained from CT data?

The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as...

Full description

Saved in:
Bibliographic Details
Published inComputer methods in biomechanics and biomedical engineering Vol. 17; no. 3; pp. 204 - 209
Main Authors Peleg, Eran, Herblum, Ryan, Beek, Maarten, Joskowicz, Leo, Liebergall, Meir, Mosheiff, Rami, Whyne, Cari
Format Journal Article
LanguageEnglish
Published England Taylor & Francis 17.02.2014
Subjects
Online AccessGet full text
ISSN1025-5842
1476-8259
1476-8259
DOI10.1080/10255842.2012.673595

Cover

Loading…
Abstract The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as a result of the partial volume edge effects (PVEEs). In this study, a new approach is introduced based on an in-house developed algorithm which adjusts the location of the model's surface nodes to a consistent predefined threshold Hounsfield unit value. Three cadaveric human femora specimens were CT scanned, and surface models were created after a semi-automatic segmentation by three different experienced operators. A FE analysis was conducted for each model, with and without applying the surface-adjustment algorithm (a total of 18 models), implementing identical boundary conditions. Maximum principal strain and stress and spatial coordinates were probed at six equivalent surface nodes from the six generated models for each of the three specimens at locations commonly utilised for experimental strain guage measurement validation. A Wilcoxon signed-ranks test was conducted to determine inter-operator variability and the impact of the PVEE-adjustment algorithm. The average inter-operator difference in stress values was significantly reduced after applying the adjustment algorithm (before: 3.32 ± 4.35 MPa, after: 1.47 ± 1.77 MPa, p = 0.025). Strain values were found to be less sensitive to inter-operative variability (p = 0.286). In summary, the new approach as presented in this study may provide a means to improve the repeatability of subject-specific FE models of bone obtained from CT data.
AbstractList The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as a result of the partial volume edge effects (PVEEs). In this study, a new approach is introduced based on an in-house developed algorithm which adjusts the location of the model's surface nodes to a consistent predefined threshold Hounsfield unit value. Three cadaveric human femora specimens were CT scanned, and surface models were created after a semi-automatic segmentation by three different experienced operators. A FE analysis was conducted for each model, with and without applying the surface-adjustment algorithm (a total of 18 models), implementing identical boundary conditions. Maximum principal strain and stress and spatial coordinates were probed at six equivalent surface nodes from the six generated models for each of the three specimens at locations commonly utilised for experimental strain guage measurement validation. A Wilcoxon signed-ranks test was conducted to determine inter-operator variability and the impact of the PVEE-adjustment algorithm. The average inter-operator difference in stress values was significantly reduced after applying the adjustment algorithm (before: 3.32 A- 4.35 MPa, after: 1.47 A- 1.77 MPa, p = 0.025). Strain values were found to be less sensitive to inter-operative variability (p = 0.286). In summary, the new approach as presented in this study may provide a means to improve the repeatability of subject-specific FE models of bone obtained from CT data.
The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as a result of the partial volume edge effects (PVEEs). In this study, a new approach is introduced based on an in-house developed algorithm which adjusts the location of the model's surface nodes to a consistent predefined threshold Hounsfield unit value. Three cadaveric human femora specimens were CT scanned, and surface models were created after a semi-automatic segmentation by three different experienced operators. A FE analysis was conducted for each model, with and without applying the surface-adjustment algorithm (a total of 18 models), implementing identical boundary conditions. Maximum principal strain and stress and spatial coordinates were probed at six equivalent surface nodes from the six generated models for each of the three specimens at locations commonly utilised for experimental strain guage measurement validation. A Wilcoxon signed-ranks test was conducted to determine inter-operator variability and the impact of the PVEE-adjustment algorithm. The average inter-operator difference in stress values was significantly reduced after applying the adjustment algorithm (before: 3.32 ± 4.35 MPa, after: 1.47 ± 1.77 MPa, p = 0.025). Strain values were found to be less sensitive to inter-operative variability (p = 0.286). In summary, the new approach as presented in this study may provide a means to improve the repeatability of subject-specific FE models of bone obtained from CT data.The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as a result of the partial volume edge effects (PVEEs). In this study, a new approach is introduced based on an in-house developed algorithm which adjusts the location of the model's surface nodes to a consistent predefined threshold Hounsfield unit value. Three cadaveric human femora specimens were CT scanned, and surface models were created after a semi-automatic segmentation by three different experienced operators. A FE analysis was conducted for each model, with and without applying the surface-adjustment algorithm (a total of 18 models), implementing identical boundary conditions. Maximum principal strain and stress and spatial coordinates were probed at six equivalent surface nodes from the six generated models for each of the three specimens at locations commonly utilised for experimental strain guage measurement validation. A Wilcoxon signed-ranks test was conducted to determine inter-operator variability and the impact of the PVEE-adjustment algorithm. The average inter-operator difference in stress values was significantly reduced after applying the adjustment algorithm (before: 3.32 ± 4.35 MPa, after: 1.47 ± 1.77 MPa, p = 0.025). Strain values were found to be less sensitive to inter-operative variability (p = 0.286). In summary, the new approach as presented in this study may provide a means to improve the repeatability of subject-specific FE models of bone obtained from CT data.
The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator generated grids can produce variability in strain and stress readings at a desired location, which are magnified at the surface of the model as a result of the partial volume edge effects (PVEEs). In this study, a new approach is introduced based on an in-house developed algorithm which adjusts the location of the model's surface nodes to a consistent predefined threshold Hounsfield unit value. Three cadaveric human femora specimens were CT scanned, and surface models were created after a semi-automatic segmentation by three different experienced operators. A FE analysis was conducted for each model, with and without applying the surface-adjustment algorithm (a total of 18 models), implementing identical boundary conditions. Maximum principal strain and stress and spatial coordinates were probed at six equivalent surface nodes from the six generated models for each of the three specimens at locations commonly utilised for experimental strain guage measurement validation. A Wilcoxon signed-ranks test was conducted to determine inter-operator variability and the impact of the PVEE-adjustment algorithm. The average inter-operator difference in stress values was significantly reduced after applying the adjustment algorithm (before: 3.32 ± 4.35 MPa, after: 1.47 ± 1.77 MPa, p = 0.025). Strain values were found to be less sensitive to inter-operative variability (p = 0.286). In summary, the new approach as presented in this study may provide a means to improve the repeatability of subject-specific FE models of bone obtained from CT data.
Author Joskowicz, Leo
Mosheiff, Rami
Peleg, Eran
Herblum, Ryan
Liebergall, Meir
Whyne, Cari
Beek, Maarten
Author_xml – sequence: 1
  givenname: Eran
  surname: Peleg
  fullname: Peleg, Eran
  email: eran@hadassah.org.il
  organization: Computer Assisted Surgical Laboratory (CASL), Department of Biomedical Engineering, Hadassah University Medical Center
– sequence: 2
  givenname: Ryan
  surname: Herblum
  fullname: Herblum, Ryan
  organization: Orthopaedic Biomechanical Laboratory, Sunnybrook Science Centre
– sequence: 3
  givenname: Maarten
  surname: Beek
  fullname: Beek, Maarten
  organization: Orthopaedic Biomechanical Laboratory, Sunnybrook Science Centre
– sequence: 4
  givenname: Leo
  surname: Joskowicz
  fullname: Joskowicz, Leo
  organization: School of Engineering and Computer Science, The Hebrew University of Jerusalem
– sequence: 5
  givenname: Meir
  surname: Liebergall
  fullname: Liebergall, Meir
  organization: Department of Orthopaedic Surgery, Hadassah University Hospital
– sequence: 6
  givenname: Rami
  surname: Mosheiff
  fullname: Mosheiff, Rami
  organization: Department of Orthopaedic Surgery, Hadassah University Hospital
– sequence: 7
  givenname: Cari
  surname: Whyne
  fullname: Whyne, Cari
  organization: Orthopaedic Biomechanical Laboratory, Sunnybrook Science Centre
BackLink https://www.ncbi.nlm.nih.gov/pubmed/22452517$$D View this record in MEDLINE/PubMed
BookMark eNqFkc1u1TAQhSNURH_gDRDykk0uthMnDpsKXfEnVWJT1tbEGbeu7DjYTtF9FN4WR7d3w4JuZkaa7xxp5lxWZ3OYsareMrpjVNIPjHIhZMt3nDK-6_pGDOJFdcHavqslF8NZmQtSb8x5dZnSA6VUMtm-qs45bwUXrL-o_uxhJkAWiNmCI4_BrR4JTnelGIM6k4jTqrMNBXN3Idp874n1SwyPSPI9lv2CkGG0zuYDCYakdXwowjotqK2xmhg721z8HHqcM_FhQpc20qBfY5nGDHbGiZgYPNnfkqn4Xb-uXhpwCd889avq55fPt_tv9c2Pr9_3n25q3QxdridDje6AAzasXAfjwJE1pu0ko6OephElN0y3vOuFlKwHMIDSoKB9aS00V9X7o2856deKKStvk0bnYMawJsUEpb3oB9k9j7YD7TlrJC3ouyd0HT1OaonWQzyo0-cL8PEI6BhSimiUthm2P-cI1ilG1RazOsWstpjVMeYibv8Rn_yfkV0fZXY2IXr4HaKbVIaDC9FEmLVNqvmvw19uesA-
CitedBy_id crossref_primary_10_1016_j_media_2018_11_001
crossref_primary_10_1016_j_clinbiomech_2022_105790
crossref_primary_10_1016_j_jbiomech_2015_01_013
crossref_primary_10_1007_s10439_022_02924_1
crossref_primary_10_1080_10255842_2022_2119383
crossref_primary_10_1016_j_medengphy_2019_10_013
crossref_primary_10_1016_j_jmbbm_2016_07_004
crossref_primary_10_1016_j_bone_2020_115348
crossref_primary_10_1080_10255842_2020_1789863
crossref_primary_10_1179_1743132814Y_0000000362
crossref_primary_10_1142_S0219519415300033
crossref_primary_10_1016_j_clinbiomech_2013_12_018
crossref_primary_10_1080_10255842_2016_1181173
Cites_doi 10.1016/0141-5425(93)90066-8
10.1016/j.jbiomech.2007.09.009
10.1016/j.jbiomech.2007.02.010
10.1016/j.jbiomech.2010.06.033
10.1016/j.clinbiomech.2008.11.003
10.1016/j.clinbiomech.2007.08.024
10.1016/j.medengphy.2004.10.001
10.1002/jor.1100120610
10.1016/j.medengphy.2006.10.014
10.1016/j.jbiomech.2008.10.039
10.1016/j.jbiomech.2003.12.030
10.1016/j.jbiomech.2009.10.040
10.1016/S1350-4533(03)00081-X
10.1038/nmat832
10.1016/j.jbiomech.2005.07.018
10.1109/TBME.2006.879473
10.1016/S0021-9290(03)00071-X
10.1016/j.jbiomech.2003.08.008
10.1016/j.medengphy.2007.05.006
ContentType Journal Article
Copyright 2012 Taylor & Francis 2014
Copyright_xml – notice: 2012 Taylor & Francis 2014
DBID AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
7X8
7QO
8FD
FR3
P64
DOI 10.1080/10255842.2012.673595
DatabaseName CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
Biotechnology Research Abstracts
Technology Research Database
Engineering Research Database
Biotechnology and BioEngineering Abstracts
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
Engineering Research Database
Biotechnology Research Abstracts
Technology Research Database
Biotechnology and BioEngineering Abstracts
DatabaseTitleList Engineering Research Database
MEDLINE - Academic
MEDLINE

Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Biology
EISSN 1476-8259
EndPage 209
ExternalDocumentID 22452517
10_1080_10255842_2012_673595
673595
Genre Article
Evaluation Studies
Journal Article
GroupedDBID ---
.7F
.QJ
0BK
0R~
29F
2DF
30N
36B
4.4
53G
5GY
5VS
AAENE
AAJMT
AALDU
AAMIU
AAPUL
AAQRR
ABCCY
ABFIM
ABHAV
ABJNI
ABLIJ
ABPAQ
ABPEM
ABTAI
ABXUL
ABXYU
ACGEJ
ACGFS
ACIWK
ACPRK
ACTIO
ADCVX
ADGTB
ADXPE
AEISY
AENEX
AEOZL
AEPSL
AEYOC
AFKVX
AFRAH
AGDLA
AGMYJ
AHDZW
AIJEM
AJWEG
AKBVH
AKOOK
ALMA_UNASSIGNED_HOLDINGS
ALQZU
AQRUH
AVBZW
AWYRJ
BLEHA
CCCUG
CE4
CS3
DGEBU
DKSSO
DU5
EBS
EJD
EMOBN
E~A
E~B
F5P
GTTXZ
H13
HF~
H~P
IPNFZ
J.P
KYCEM
M4Z
NA5
P2P
RIG
RNANH
ROSJB
RTWRZ
S-T
SNACF
TBQAZ
TDBHL
TEN
TFL
TFT
TFW
TN5
TNC
TTHFI
TUROJ
TWF
UT5
UU3
ZGOLN
~S~
AAGDL
AAHIA
AAYXX
ADMLS
ADYSH
AFRVT
AIYEW
AMPGV
CITATION
1TA
ACTTO
ADUMR
AFBWG
AFION
AGVKY
AGWUF
ALRRR
BWMZZ
CAG
CGR
COF
CUY
CVF
CYRSC
DAOYK
ECM
EIF
LJTGL
NPM
OPCYK
TASJS
7X8
7QO
8FD
FR3
P64
ID FETCH-LOGICAL-c396t-df0fc6a2ae31818ab92e13f46810bcddbe82f1c426758817aafae8fe507e8f4a3
ISSN 1025-5842
1476-8259
IngestDate Thu Jul 10 23:58:36 EDT 2025
Fri Jul 11 11:59:27 EDT 2025
Mon Jul 21 05:13:21 EDT 2025
Tue Jul 01 03:32:02 EDT 2025
Thu Apr 24 23:07:08 EDT 2025
Wed Dec 25 09:04:47 EST 2024
IsPeerReviewed true
IsScholarly true
Issue 3
Language English
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c396t-df0fc6a2ae31818ab92e13f46810bcddbe82f1c426758817aafae8fe507e8f4a3
Notes ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Undefined-1
ObjectType-Feature-3
content type line 23
ObjectType-Feature-1
PMID 22452517
PQID 1490721380
PQPubID 23479
PageCount 6
ParticipantIDs proquest_miscellaneous_1490721380
pubmed_primary_22452517
crossref_citationtrail_10_1080_10255842_2012_673595
crossref_primary_10_1080_10255842_2012_673595
informaworld_taylorfrancis_310_1080_10255842_2012_673595
proquest_miscellaneous_1500757986
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2014-02-17
PublicationDateYYYYMMDD 2014-02-17
PublicationDate_xml – month: 02
  year: 2014
  text: 2014-02-17
  day: 17
PublicationDecade 2010
PublicationPlace England
PublicationPlace_xml – name: England
PublicationTitle Computer methods in biomechanics and biomedical engineering
PublicationTitleAlternate Comput Methods Biomech Biomed Engin
PublicationYear 2014
Publisher Taylor & Francis
Publisher_xml – name: Taylor & Francis
References cit0011
cit0001
cit0012
cit0020
cit0010
cit0008
cit0019
cit0006
cit0017
cit0007
cit0018
cit0004
cit0015
cit0005
cit0016
cit0002
cit0013
cit0003
cit0014
References_xml – ident: cit0007
  doi: 10.1016/0141-5425(93)90066-8
– ident: cit0013
  doi: 10.1016/j.jbiomech.2007.09.009
– ident: cit0014
  doi: 10.1016/j.jbiomech.2007.02.010
– ident: cit0011
  doi: 10.1016/j.jbiomech.2010.06.033
– ident: cit0012
  doi: 10.1016/j.clinbiomech.2008.11.003
– ident: cit0004
  doi: 10.1016/j.clinbiomech.2007.08.024
– ident: cit0001
  doi: 10.1016/j.medengphy.2004.10.001
– ident: cit0008
  doi: 10.1002/jor.1100120610
– ident: cit0017
  doi: 10.1016/j.medengphy.2006.10.014
– ident: cit0018
  doi: 10.1016/j.jbiomech.2008.10.039
– ident: cit0019
  doi: 10.1016/j.jbiomech.2003.12.030
– ident: cit0002
  doi: 10.1016/j.jbiomech.2009.10.040
– ident: cit0006
  doi: 10.1016/S1350-4533(03)00081-X
– ident: cit0010
  doi: 10.1038/nmat832
– ident: cit0015
  doi: 10.1016/j.jbiomech.2005.07.018
– ident: cit0016
  doi: 10.1109/TBME.2006.879473
– ident: cit0020
  doi: 10.1016/S0021-9290(03)00071-X
– ident: cit0003
  doi: 10.1016/j.jbiomech.2003.08.008
– ident: cit0005
  doi: 10.1016/j.medengphy.2007.05.006
SSID ssj0008184
Score 2.0619884
Snippet The reliability of patient-specific finite element (FE) modelling is dependent on the ability to provide repeatable analyses. Differences of inter-operator...
SourceID proquest
pubmed
crossref
informaworld
SourceType Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 204
SubjectTerms Algorithms
Femur - diagnostic imaging
Finite Element Analysis
Humans
inter-operative
Models, Biological
partial volume edge effects
patient specific
Reproducibility of Results
strain
stress
Stress, Mechanical
surface adjustment
Tomography, X-Ray Computed
Title Can a partial volume edge effect reduction algorithm improve the repeatability of subject-specific finite element models of femurs obtained from CT data?
URI https://www.tandfonline.com/doi/abs/10.1080/10255842.2012.673595
https://www.ncbi.nlm.nih.gov/pubmed/22452517
https://www.proquest.com/docview/1490721380
https://www.proquest.com/docview/1500757986
Volume 17
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Lj9MwELZKV0hcEG_KS0biFqU0ifM6oaWAqhXi1BUrLpHj2FBt21RpAlr-Cf-HH8ZMJknTZXd5XNLIdtIo88WeGX8zw9iL0M8CWNZSO1VOZgvlaBu0DN_OYqFhOYJVSdYE2Q_B7Fgcnfgng8HPHmupKtOx-n5hXMn_SBXaQK4YJfsPku1uCg1wDvKFI0gYjn8l4yl8nNLaYA-8aZpoLHSQNTQNq8DErFQLfPk5LxbllxXGRRb5V03cQr2BuZhSdddb7dsqRceMjQGYSCKyzAKVUksTy5wK59TsD6NXVQFnKfoWQGut41Smcwspp-fYgm3liKZcdc3ArcP-Meq4zRJNeQBqyOhdjsTdzL3UREMrdnCeASTaOs1nu-bXWp9SGBLyVbvmo3x7mn9bqNpj_l7nfX-HI5AiTeGdY01ztAgDGwzbeG8SD3tg9fozMlU3_m2lIGqlgyZVJDAiz3HHQYhhyv3hIO_NqkaPi1vUfvsgexm6265r7MAFY8UdsoPD2ZtPHzuNAJQi0YZtRpOXF_0pJqVubrOnIe3lz73cCqq1ofktdrMxY_ghYfI2G-j1HXadCpue3WU_AJlc8gaZnN4eR2RyQibvkMk7ZPIGmRyQyfeQyXPDzyOTEzJ5g0xOyMSRhEzeIpMjMvl0zhGZr-6x43dv59OZ3RQBsZUXB6WdmYlRgXQlOuudSKaxqx3PCMyjl6osS3XkGkeBogmWb-SEUhqpI6PBzoEfIb37bLjO1_oh40qnoQhUZnRqhO_5kQL1WWUygHXJVUaNmNe--EQ1GfKxUMsycZpEuq3kEpRcQpIbMbu7akMZYv4wPurLNClrz5yhMjqJd_Wlz1v5J7AK4NaeXOu82oIBH2OiQy-aXDHGR_sgjKNgxB4QeLoHbrH36NKex-zG7mt8woZlUemnoI-X6bMG7r8ArIXirw
linkProvider Library Specific Holdings
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1Lb9NAEF5BEYIL5U3Ka5C4bpT1Y70-oSpqFaDklErcrH3SiNaJEvtQ_gn_lhmvHVGkFglOjuTdyF7P7nw7-803jL0vcifRrRlurHA8s8JzRBk5d2Xm0R2hV9IdQXYuZ6fZp6_5wCbc9rRK2kOHKBTRrdU0uSkYPVDi8IpAWGWURyWSsSwoufQ2u5OXsqAiBulkvluM0R91B8tUtZW6DNlz1_zLFe90Rbv0egTaeaLjfWaGd4gElO_jtjFj--MPecf_esmH7EGPU-EwGtYjdsvXj9ndWLny8gn7OdU1aFiT4WGzuMYBxeYgMkRgQ5qw9NVBn39bbZbN2QUsuwiGBwSdeH-NbiCqhF_CKsC2NRQT4pT7SfwlCEvCw-AjwR26mj1bahn8RbvBX4bCGt4BpcjAdAHEdv3wlJ0eHy2mM95XeeA2LWXDXZgEK3WiKRorlDZl4kUaMhJKM9Y541UShEUkgVsbJQqtg_YqeASyeMl0-ozt1avav2BgvSkyaV3wJmR5miuL-Mg6LXHhSWywI5YOX7eyvQQ6VeI4r0SvlDoMekWDXsVBHzG-67WOEiB_aa9-N5yq6UIvIdZJqdKbu74bjKzCaU5nN7r2q3aLO7SSlOxSNbmhTU4AsCiVHLHn0UJ3D5zQCXsuioN_f7i37N5s8eWkOvk4__yS3cc7GdHXRfGK7TWb1r9GdNaYN938-wWnxC3-
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1Lj9MwELZgEYgL70d5DhJXV3XiOM4JoUK1PFRx2JX2Fvm5VOymVZscln_Cv2UmbqpdpF0kOKVS7Cpxxp7P42--YextWXiFbs1y64Tn0onAEWUU3FcyoDtCr2R6guxc7R_Kz0fF0bksfqJV0h46JqGIfq2myb3ycWDE4RVxsJaURiWysSopt_Q6u6FIO5ySOCbz3VqM7qg_V6airdRlSJ675F8uOKcL0qWXA9DeEc3uMjO8QuKf_Bh3rR27n3-oO_7PO95jd7YoFd4ns7rProXmAbuZ6laePWS_pqYBAysyO2yWVjigyBwkfgisSRGWvjmYk-PletF-P4VFH78IgJAT76_QCSSN8DNYRth0liJCnDI_ib0EcUFoGEKit0NfsWdDLWM47db4y1JQI3igBBmYHgBxXd89YoezjwfTfb6t8cBdXqmW-ziJTpnMUCxWaGOrLIg8SpJJs857G3QWhUMcgRsbLUpjogk6BoSxeJEmf8z2mmUTnjJwwZZSOR-DjbLIC-0QHTlvFC47mYtuxPLh49ZuK4BOdThOarHVSR0GvaZBr9Ogjxjf9VolAZC_tNfn7aZu-8BLTFVS6vzqrm8GG6txktPJjWnCstvg_qwiHbtcT65oUxD8KyutRuxJMtDdA2d0vl6I8tm_P9xrduvbh1n99dP8y3N2G29I4q6L8gXba9ddeInQrLWv-tn3G8jMLKI
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Can+a+partial+volume+edge+effect+reduction+algorithm+improve+the+repeatability+of+subject-specific+finite+element+models+of+femurs+obtained+from+CT+data%3F&rft.jtitle=Computer+methods+in+biomechanics+and+biomedical+engineering&rft.au=Peleg%2C+Eran&rft.au=Herblum%2C+Ryan&rft.au=Beek%2C+Maarten&rft.au=Joskowicz%2C+Leo&rft.date=2014-02-17&rft.eissn=1476-8259&rft.volume=17&rft.issue=3&rft.spage=204&rft_id=info:doi/10.1080%2F10255842.2012.673595&rft_id=info%3Apmid%2F22452517&rft.externalDocID=22452517
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1025-5842&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1025-5842&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1025-5842&client=summon