Mechanical thrombectomy: results of early experience

To evaluate the clinical application of a mechanical thrombectomy device in occluded lower extremity arteries and bypass grafts. A mechanical thrombectomy device was used in five patients with acutely occluded lower extremity bypass grafts (n = 2) or superficial femoral arteries (n = 3). The thrombu...

Full description

Saved in:
Bibliographic Details
Published inRadiology Vol. 189; no. 3; p. 803
Main Authors Coleman, C C, Krenzel, C, Dietz, C A, Nazarian, G K, Amplatz, K
Format Journal Article
LanguageEnglish
Published United States 01.12.1993
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:To evaluate the clinical application of a mechanical thrombectomy device in occluded lower extremity arteries and bypass grafts. A mechanical thrombectomy device was used in five patients with acutely occluded lower extremity bypass grafts (n = 2) or superficial femoral arteries (n = 3). The thrombus was completely removed in three cases (two bypass grafts and one native superficial femoral artery). In two of the patients with occlusions in the native superficial femoral artery, the thrombus was partially removed. In these two patients, the aspirate demonstrated a more fibrous thrombus that was not easily fragmented. In one of these two patients, the coagulum was thought to be embolic from the heart. The thrombectomy device works well in recent thromboses, particularly in polytetrafluoroethylene grafts in which organization does not occur. Its efficacy in chronically occluded vessels is not certain.