Optoelectronic microfluidic device for point-of-care blood plasma viscosity measurement
Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectro...
Saved in:
Published in | Lab on a chip Vol. 24; no. 13; pp. 335 - 3314 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
England
Royal Society of Chemistry
25.06.2024
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Abstract | Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectronic system using a microfluidic platform for the measurement of blood plasma viscosity. Microchannels were defined onto the filter paper using an available and inexpensive wax crayon, without the need for conventional wax printing equipment. The time required for the 5 μL sample to pass a specific distance was measured using two pairs of infrared sensors. Subsequently, this data was sent to the microcontroller, which automatically calculated the viscosity. Throughout the measurements, sample temperature was maintained at a constant 37 °C through an integrated heater with automated control. The microfluidic platform successfully processed real samples, yielding viscosity measurements in under three minutes. Evaluation with fetal bovine serum, spiked with varying protein concentrations in both native and denatured states, demonstrated a precision exceeding 96% compared to conventional Ostwald viscometer readings. For human subjects exhibiting pathologies affecting serum and plasma viscosity compared to physiological norms, strong correlations were observed between resultant values and clinical diagnoses. The proposed device aims to replace expensive and complex optical equipment, offering a safer alternative for measuring plasma viscosity. Unlike similar devices, it eliminates the risk of component deformation due to chemical contact or unsafe irradiation.
Non-contact, highly accurate and fully automated measurement of the blood plasma viscosity using the optical transparency technique at a constant temperature of 37 °C. |
---|---|
AbstractList | Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectronic system using a microfluidic platform for the measurement of blood plasma viscosity. Microchannels were defined onto the filter paper using an available and inexpensive wax crayon, without the need for conventional wax printing equipment. The time required for the 5 μL sample to pass a specific distance was measured using two pairs of infrared sensors. Subsequently, this data was sent to the microcontroller, which automatically calculated the viscosity. Throughout the measurements, sample temperature was maintained at a constant 37 °C through an integrated heater with automated control. The microfluidic platform successfully processed real samples, yielding viscosity measurements in under three minutes. Evaluation with fetal bovine serum, spiked with varying protein concentrations in both native and denatured states, demonstrated a precision exceeding 96% compared to conventional Ostwald viscometer readings. For human subjects exhibiting pathologies affecting serum and plasma viscosity compared to physiological norms, strong correlations were observed between resultant values and clinical diagnoses. The proposed device aims to replace expensive and complex optical equipment, offering a safer alternative for measuring plasma viscosity. Unlike similar devices, it eliminates the risk of component deformation due to chemical contact or unsafe irradiation. Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectronic system using a microfluidic platform for the measurement of blood plasma viscosity. Microchannels were defined onto the filter paper using an available and inexpensive wax crayon, without the need for conventional wax printing equipment. The time required for the 5 μL sample to pass a specific distance was measured using two pairs of infrared sensors. Subsequently, this data was sent to the microcontroller, which automatically calculated the viscosity. Throughout the measurements, sample temperature was maintained at a constant 37 °C through an integrated heater with automated control. The microfluidic platform successfully processed real samples, yielding viscosity measurements in under three minutes. Evaluation with fetal bovine serum, spiked with varying protein concentrations in both native and denatured states, demonstrated a precision exceeding 96% compared to conventional Ostwald viscometer readings. For human subjects exhibiting pathologies affecting serum and plasma viscosity compared to physiological norms, strong correlations were observed between resultant values and clinical diagnoses. The proposed device aims to replace expensive and complex optical equipment, offering a safer alternative for measuring plasma viscosity. Unlike similar devices, it eliminates the risk of component deformation due to chemical contact or unsafe irradiation.Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectronic system using a microfluidic platform for the measurement of blood plasma viscosity. Microchannels were defined onto the filter paper using an available and inexpensive wax crayon, without the need for conventional wax printing equipment. The time required for the 5 μL sample to pass a specific distance was measured using two pairs of infrared sensors. Subsequently, this data was sent to the microcontroller, which automatically calculated the viscosity. Throughout the measurements, sample temperature was maintained at a constant 37 °C through an integrated heater with automated control. The microfluidic platform successfully processed real samples, yielding viscosity measurements in under three minutes. Evaluation with fetal bovine serum, spiked with varying protein concentrations in both native and denatured states, demonstrated a precision exceeding 96% compared to conventional Ostwald viscometer readings. For human subjects exhibiting pathologies affecting serum and plasma viscosity compared to physiological norms, strong correlations were observed between resultant values and clinical diagnoses. The proposed device aims to replace expensive and complex optical equipment, offering a safer alternative for measuring plasma viscosity. Unlike similar devices, it eliminates the risk of component deformation due to chemical contact or unsafe irradiation. Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled with minimal sample requirement, stimulated the advancement of paper-based viscometers. This study presents a precise, non-contact optoelectronic system using a microfluidic platform for the measurement of blood plasma viscosity. Microchannels were defined onto the filter paper using an available and inexpensive wax crayon, without the need for conventional wax printing equipment. The time required for the 5 μL sample to pass a specific distance was measured using two pairs of infrared sensors. Subsequently, this data was sent to the microcontroller, which automatically calculated the viscosity. Throughout the measurements, sample temperature was maintained at a constant 37 °C through an integrated heater with automated control. The microfluidic platform successfully processed real samples, yielding viscosity measurements in under three minutes. Evaluation with fetal bovine serum, spiked with varying protein concentrations in both native and denatured states, demonstrated a precision exceeding 96% compared to conventional Ostwald viscometer readings. For human subjects exhibiting pathologies affecting serum and plasma viscosity compared to physiological norms, strong correlations were observed between resultant values and clinical diagnoses. The proposed device aims to replace expensive and complex optical equipment, offering a safer alternative for measuring plasma viscosity. Unlike similar devices, it eliminates the risk of component deformation due to chemical contact or unsafe irradiation. Non-contact, highly accurate and fully automated measurement of the blood plasma viscosity using the optical transparency technique at a constant temperature of 37 °C. |
Author | Bakhtiaridoost, Somayyeh Musuroi, Cristian Florescu, Monica Volmer, Marius |
AuthorAffiliation | Transilvania University of Brasov Department of Fundamental, Prophylactic and Clinical Disciplines Faculty of Medicine Department of Electrical Engineering and Applied Physics |
AuthorAffiliation_xml | – sequence: 0 name: Faculty of Medicine – sequence: 0 name: Department of Electrical Engineering and Applied Physics – sequence: 0 name: Department of Fundamental, Prophylactic and Clinical Disciplines – sequence: 0 name: Transilvania University of Brasov |
Author_xml | – sequence: 1 givenname: Somayyeh surname: Bakhtiaridoost fullname: Bakhtiaridoost, Somayyeh – sequence: 2 givenname: Cristian surname: Musuroi fullname: Musuroi, Cristian – sequence: 3 givenname: Marius surname: Volmer fullname: Volmer, Marius – sequence: 4 givenname: Monica surname: Florescu fullname: Florescu, Monica |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/38869225$$D View this record in MEDLINE/PubMed |
BookMark | eNpt0c9LHDEUB_BQVnTdeum9ZaAXEUbzY5JMjrKttrDgRelxyGReIJKZTJOM4H_f1F0tSE_Jg897PN73FK2mMAFCnwi-JJipq6HxBmNKiPmA1qSRrMakVau3v5In6DSlR4wJb0R7jE5Y2wpFKV-jX3dzDuDB5BgmZ6rRmRisX9xQigGenIHKhljNwU25DrY2OkLV-xCGavY6jbp6csmE5PJzNYJOS4QRpvwRHVntE5wd3g16uPl-v_1R7-5uf26vd7VhkuVaWtCC8dYIanpCqRo4tUJAwwkFKYXgpbZ9WRUbqXTf9tpio9qeK8O1EGyDzvdz5xh-L5ByN5Z1wHs9QVhSx7CQijRU4EK_vqOPYYlT2a4oSRTjmMuivhzU0o8wdHN0o47P3evJCsB7UA6VUgTbGZd1dmHKUTvfEdz9TaX71uy2L6lsS8vFu5bXqf_Fn_c4JvPm_kXM_gCTWJXv |
CitedBy_id | crossref_primary_10_1063_5_0249103 |
Cites_doi | 10.1063/1.1700096 10.1016/j.aca.2018.05.036 10.1051/matecconf/201819902019 10.3390/diagnostics10080509 10.1039/D1LC00202C 10.1007/10_2020_127 10.1063/1.555688 10.3233/CH-2008-1088 10.1371/journal.pone.0277401 10.1016/j.sna.2023.114497 10.1136/jcp.30.4.345 10.1016/S0950-3536(87)80021-5 10.1039/C4LC00716F 10.1039/D3LC00101F 10.1309/fff7u8rrpk26vapy 10.1088/1742-6596/979/1/012083 10.3390/mi9080375 10.1016/j.snb.2016.04.126 10.1007/s10404-016-1800-8 10.1016/j.aca.2021.338303 10.1053/meta.2000.6264 10.1002/ajh.25254 10.1016/B978-0-7020-6696-2.00006-0 10.1088/1361-6501/aace7d 10.1021/acs.analchem.9b00624 10.1063/1.3088050 10.1109/TED.2020.2989727 10.1016/j.tibtech.2014.04.010 10.3109/02656736.2013.775355 10.1021/acs.langmuir.0c03134 10.1109/LSENS.2019.2894623 10.1016/j.sna.2023.114873 |
ContentType | Journal Article |
Copyright | Copyright Royal Society of Chemistry 2024 |
Copyright_xml | – notice: Copyright Royal Society of Chemistry 2024 |
DBID | AAYXX CITATION NPM 7SP 7TB 7U5 8FD FR3 L7M 7X8 |
DOI | 10.1039/d4lc00211c |
DatabaseName | CrossRef PubMed Electronics & Communications Abstracts Mechanical & Transportation Engineering Abstracts Solid State and Superconductivity Abstracts Technology Research Database Engineering Research Database Advanced Technologies Database with Aerospace MEDLINE - Academic |
DatabaseTitle | CrossRef PubMed Solid State and Superconductivity Abstracts Engineering Research Database Technology Research Database Mechanical & Transportation Engineering Abstracts Advanced Technologies Database with Aerospace Electronics & Communications Abstracts MEDLINE - Academic |
DatabaseTitleList | Solid State and Superconductivity Abstracts CrossRef PubMed MEDLINE - Academic |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Engineering Chemistry Biology |
EISSN | 1473-0189 |
EndPage | 3314 |
ExternalDocumentID | 38869225 10_1039_D4LC00211C d4lc00211c |
Genre | Journal Article |
GroupedDBID | --- -JG 0-7 0R~ 29L 4.4 5GY 705 70~ 7~J AAEMU AAIWI AAJAE AAMEH AANOJ AAWGC AAXHV AAXPP ABASK ABDVN ABEMK ABJNI ABPDG ABRYZ ABXOH ACGFS ACIWK ACLDK ADMRA ADSRN AEFDR AENEX AENGV AESAV AETIL AFLYV AFOGI AFVBQ AGEGJ AGKEF AGRSR AGSTE AHGCF ALMA_UNASSIGNED_HOLDINGS ANUXI APEMP ASKNT AUDPV BLAPV BSQNT C6K CS3 DU5 EBS ECGLT EE0 EF- EJD F5P GGIMP GNO H13 HZ~ H~N IDZ J3I L-8 M4U N9A O9- R7B RAOCF RCNCU RNS RPMJG RRA RRC RSCEA SKA SLH VH6 AAYXX AFRZK AKMSF CITATION R56 NPM 7SP 7TB 7U5 8FD FR3 L7M 7X8 |
ID | FETCH-LOGICAL-c373t-7fea6358c62cb1229d52f66e4512e7766552ffb2250c79ab8baf0c98b59c5a663 |
ISSN | 1473-0197 1473-0189 |
IngestDate | Fri Jul 11 00:52:14 EDT 2025 Sun Jun 29 16:26:04 EDT 2025 Thu Apr 03 06:59:46 EDT 2025 Thu Apr 24 22:59:32 EDT 2025 Tue Jul 01 01:52:53 EDT 2025 Tue Dec 17 20:58:36 EST 2024 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 13 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c373t-7fea6358c62cb1229d52f66e4512e7766552ffb2250c79ab8baf0c98b59c5a663 |
Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
ORCID | 0000-0002-1160-9650 0000-0002-1467-3113 0000-0003-4878-5815 0000-0003-4764-5263 |
OpenAccessLink | http://pubs.rsc.org/en/content/articlepdf/2024/LC/D4LC00211C |
PMID | 38869225 |
PQID | 3071935057 |
PQPubID | 2047488 |
PageCount | 1 |
ParticipantIDs | pubmed_primary_38869225 rsc_primary_d4lc00211c crossref_citationtrail_10_1039_D4LC00211C crossref_primary_10_1039_D4LC00211C proquest_journals_3071935057 proquest_miscellaneous_3067914260 |
ProviderPackageCode | CITATION AAYXX |
PublicationCentury | 2000 |
PublicationDate | 2024-06-25 |
PublicationDateYYYYMMDD | 2024-06-25 |
PublicationDate_xml | – month: 06 year: 2024 text: 2024-06-25 day: 25 |
PublicationDecade | 2020 |
PublicationPlace | England |
PublicationPlace_xml | – name: England – name: Cambridge |
PublicationTitle | Lab on a chip |
PublicationTitleAlternate | Lab Chip |
PublicationYear | 2024 |
Publisher | Royal Society of Chemistry |
Publisher_xml | – name: Royal Society of Chemistry |
References | Cai (D4LC00211C/cit29/1) 2021; 37 Rosencranz (D4LC00211C/cit9/1) 2006; 125 Suppl Osei-Bimpong (D4LC00211C/cit4/1) 2017 Yunus (D4LC00211C/cit18/1) 2018; 979 Meffan (D4LC00211C/cit27/1) 2023; 359 Kim (D4LC00211C/cit14/1) 2016; 234 Hutchinson (D4LC00211C/cit5/1) 1977; 30 Largman-Chalamish (D4LC00211C/cit6/1) 2022; 17 Doolittle (D4LC00211C/cit8/1) 1951; 22 Somer (D4LC00211C/cit2/1) 1987; 1 Kang (D4LC00211C/cit26/1) 2019; 91 Arshavsky-Graham (D4LC00211C/cit12/1) 2020; 179 Rayaprolu (D4LC00211C/cit22/1) 2018; 1044 Elizalde (D4LC00211C/cit21/1) 2016; 20 Mathew (D4LC00211C/cit1/1) 2023 Inman (D4LC00211C/cit11/1) 2013; 29 Xiang (D4LC00211C/cit15/1) 2024; 365 Jiang (D4LC00211C/cit16/1) 2023; 23 Li (D4LC00211C/cit20/1) 2014; 14 Zaccardi (D4LC00211C/cit30/1) 2018; 199 Abeykoon (D4LC00211C/cit33/1) 2018; 93 Volpatti (D4LC00211C/cit13/1) 2014; 32 Tzeng (D4LC00211C/cit10/1) 2018; 9 Hassan (D4LC00211C/cit28/1) 2020; 10 Otto (D4LC00211C/cit34/1) 2000; 49 Puneeth (D4LC00211C/cit23/1) 2019; 3 Vargaftik (D4LC00211C/cit32/1) 1983; 12 Huber (D4LC00211C/cit31/1) 2009; 38 Puneeth (D4LC00211C/cit24/1) 2021; 1153 Puneeth (D4LC00211C/cit25/1) 2020; 67 Brigden (D4LC00211C/cit7/1) 1999; 60 Shiba (D4LC00211C/cit17/1) 2021; 21 Késmárky (D4LC00211C/cit3/1) 2008 Bamshad (D4LC00211C/cit19/1) 2018; 29 |
References_xml | – issn: 2020 issue: vol. 179 end-page: p 247-265 publication-title: Advances in Biochemical Engineering/Biotechnology doi: Arshavsky-Graham Segal – issn: 2018 issue: vol. 199 publication-title: International Conference on Concrete Repair, Rehabilitation and Retrofitting (ICCRRR 2018) doi: Zaccardi Alderete De Belie – issn: 2023 publication-title: Physiology, Blood Plasma doi: Mathew Sankar Varacallo – issn: 2017 end-page: p 93-111 publication-title: Dacie and Lewis Practical Haematology doi: Osei-Bimpong Burthem – issn: 2008 end-page: p 243-246 publication-title: Clinical Hemorheology and Microcirculation doi: Késmárky Kenyeres Rábai Tóth – volume: 22 start-page: 1031 issue: 8 year: 1951 ident: D4LC00211C/cit8/1 publication-title: J. Appl. Phys. doi: 10.1063/1.1700096 – volume: 1044 start-page: 86 year: 2018 ident: D4LC00211C/cit22/1 publication-title: Anal. Chim. Acta doi: 10.1016/j.aca.2018.05.036 – volume: 199 volume-title: International Conference on Concrete Repair, Rehabilitation and Retrofitting (ICCRRR 2018) year: 2018 ident: D4LC00211C/cit30/1 doi: 10.1051/matecconf/201819902019 – volume: 10 start-page: 509 issue: 8 year: 2020 ident: D4LC00211C/cit28/1 publication-title: Diagnostics doi: 10.3390/diagnostics10080509 – volume: 21 start-page: 2805 issue: 14 year: 2021 ident: D4LC00211C/cit17/1 publication-title: Lab Chip doi: 10.1039/D1LC00202C – volume: 179 start-page: 247 volume-title: Advances in Biochemical Engineering/Biotechnology year: 2020 ident: D4LC00211C/cit12/1 doi: 10.1007/10_2020_127 – volume: 12 start-page: 817 issue: 3 year: 1983 ident: D4LC00211C/cit32/1 publication-title: J. Phys. Chem. Ref. Data doi: 10.1063/1.555688 – start-page: 243 volume-title: Clinical Hemorheology and Microcirculation year: 2008 ident: D4LC00211C/cit3/1 doi: 10.3233/CH-2008-1088 – volume-title: Physiology, Blood Plasma year: 2023 ident: D4LC00211C/cit1/1 – volume: 17 start-page: e0277401 issue: 12 year: 2022 ident: D4LC00211C/cit6/1 publication-title: PLoS One doi: 10.1371/journal.pone.0277401 – volume: 359 start-page: 114497 year: 2023 ident: D4LC00211C/cit27/1 publication-title: Sens. Actuators, A doi: 10.1016/j.sna.2023.114497 – volume: 30 start-page: 345 year: 1977 ident: D4LC00211C/cit5/1 publication-title: J. Clin. Pathol. doi: 10.1136/jcp.30.4.345 – volume: 1 start-page: 695 issue: 3 year: 1987 ident: D4LC00211C/cit2/1 publication-title: Baillieres Clin. Haematol. doi: 10.1016/S0950-3536(87)80021-5 – volume: 14 start-page: 4035 issue: 20 year: 2014 ident: D4LC00211C/cit20/1 publication-title: Lab Chip doi: 10.1039/C4LC00716F – volume: 23 start-page: 2577 issue: 11 year: 2023 ident: D4LC00211C/cit16/1 publication-title: Lab Chip doi: 10.1039/D3LC00101F – volume: 125 Suppl start-page: S78 year: 2006 ident: D4LC00211C/cit9/1 publication-title: Am. J. Clin. Pathol. doi: 10.1309/fff7u8rrpk26vapy – volume: 979 start-page: 012083 year: 2018 ident: D4LC00211C/cit18/1 publication-title: J. Phys.: Conf. Ser. doi: 10.1088/1742-6596/979/1/012083 – volume: 9 start-page: 375 issue: 8 year: 2018 ident: D4LC00211C/cit10/1 publication-title: Micromachines doi: 10.3390/mi9080375 – volume: 60 start-page: 1443 issue: 5 year: 1999 ident: D4LC00211C/cit7/1 publication-title: Am. Fam. Physician – volume: 234 start-page: 239 year: 2016 ident: D4LC00211C/cit14/1 publication-title: Sens. Actuators, B doi: 10.1016/j.snb.2016.04.126 – volume: 20 start-page: 135 issue: 10 year: 2016 ident: D4LC00211C/cit21/1 publication-title: Microfluid. Nanofluid. doi: 10.1007/s10404-016-1800-8 – volume: 1153 start-page: 338303 year: 2021 ident: D4LC00211C/cit24/1 publication-title: Anal. Chim. Acta doi: 10.1016/j.aca.2021.338303 – volume: 49 start-page: 810 issue: 6 year: 2000 ident: D4LC00211C/cit34/1 publication-title: Metabolism doi: 10.1053/meta.2000.6264 – volume: 93 start-page: 1384 issue: 11 year: 2018 ident: D4LC00211C/cit33/1 publication-title: Am. J. Hematol. doi: 10.1002/ajh.25254 – start-page: 93 volume-title: Dacie and Lewis Practical Haematology year: 2017 ident: D4LC00211C/cit4/1 doi: 10.1016/B978-0-7020-6696-2.00006-0 – volume: 29 start-page: 095901 issue: 9 year: 2018 ident: D4LC00211C/cit19/1 publication-title: Meas. Sci. Technol. doi: 10.1088/1361-6501/aace7d – volume: 91 start-page: 4868 issue: 7 year: 2019 ident: D4LC00211C/cit26/1 publication-title: Anal. Chem. doi: 10.1021/acs.analchem.9b00624 – volume: 38 start-page: 101 issue: 2 year: 2009 ident: D4LC00211C/cit31/1 publication-title: J. Phys. Chem. Ref. Data doi: 10.1063/1.3088050 – volume: 67 start-page: 2559 issue: 6 year: 2020 ident: D4LC00211C/cit25/1 publication-title: IEEE Trans. Electron Devices doi: 10.1109/TED.2020.2989727 – volume: 32 start-page: 347 issue: 7 year: 2014 ident: D4LC00211C/cit13/1 publication-title: Trends Biotechnol. doi: 10.1016/j.tibtech.2014.04.010 – volume: 29 start-page: 206 issue: 3 year: 2013 ident: D4LC00211C/cit11/1 publication-title: Int. J. Hyperthermia doi: 10.3109/02656736.2013.775355 – volume: 37 start-page: 1623 issue: 5 year: 2021 ident: D4LC00211C/cit29/1 publication-title: Langmuir doi: 10.1021/acs.langmuir.0c03134 – volume: 3 start-page: 1 issue: 3 year: 2019 ident: D4LC00211C/cit23/1 publication-title: IEEE Sens. Lett. doi: 10.1109/LSENS.2019.2894623 – volume: 365 start-page: 114873 year: 2024 ident: D4LC00211C/cit15/1 publication-title: Sens. Actuators, A doi: 10.1016/j.sna.2023.114873 |
SSID | ssj0015468 |
Score | 2.4649432 |
Snippet | Physical properties of blood plasma, such as viscosity, serve as crucial indicators of disease. The inherent capillary effect of paper microchannels, coupled... |
SourceID | proquest pubmed crossref rsc |
SourceType | Aggregation Database Index Database Enrichment Source Publisher |
StartPage | 335 |
SubjectTerms | Automatic control Blood plasma Capillarity Filter paper Infrared detectors Microchannels Microfluidic devices Optical equipment Optoelectronic devices Ostwald ripening Physical properties Plasma Viscometers Viscosity Viscosity measurement Waxes |
Title | Optoelectronic microfluidic device for point-of-care blood plasma viscosity measurement |
URI | https://www.ncbi.nlm.nih.gov/pubmed/38869225 https://www.proquest.com/docview/3071935057 https://www.proquest.com/docview/3067914260 |
Volume | 24 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3Nb9MwFLegE4IdEBQGgYGC4IJQtiZ24uQ4haGBOjjQAbfIdhytoo2rNplU_nqe7XxVGxJwieKPuI3fL8_v2e8DoTcgg4icUulxn8UeiSaxF_u4gCIpZB7gUDLtjXz-OTq7IJ9-hIODduNdUvEj8etGv5L_oSrUAV21l-w_ULYbFCrgHugLV6AwXP-Kxl9WlRrksVlq47piUc9zKORS8wBjRbhS87LyVOEZMy9jqq6zR2-W7N3VfCOUMctY9puFQ4F1yrg-T2Da63vVb3v-vATOsJ7nqvEZ-aqWbLuVlz39YDBlLAXSho20Td9Um7HlHEaoe_-ThQLVX9SmRb8PG-5IBERbTlnv5YaJEqpttKzd7ZEc1tl0QS3ntd7TLcLwgI9iG8OkWZIxto6m19j9BOtoqTlZCC2r-KJf1DpTw77xNtoLQJcIRmjv5HT2cdodNoXEeky2f7uNYouT4_7pXbnlmjICosm6TRljRJPZA3S_0SncEwuQh-iWLMfojs0yuh2ju2mb1G-M9gfxJx-h77sAcocAci2AXACQuwMg1wDItQByOwC5AwA9RhcfTmfpmddk2vAEprjyaCEZSJ6xiALB_SBI8jAookgSEAclpVEUQrngwPsngiaMx5wVE5HEPExEyEBoPUCjUpXyKXIZywnlkaAk4STJCS9AwcY4ljnjhBXCQW_becxEE4ZeZ0NZZMYcAifZezJNzZynDnrd9V3Z4Cs39jpsyZE1H-cmg6ULVBOtfTvoVdcMk63Pw1gpVa37RDTxdYoGBz2xZOx-BsdxlMD7OugA6NpV93h49qeG5-he_1EcolG1ruULEFwr_rIB3m8s8JxU |
linkProvider | Royal Society of Chemistry |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Optoelectronic+microfluidic+device+for+point-of-care+blood+plasma+viscosity+measurement&rft.jtitle=Lab+on+a+chip&rft.au=Bakhtiaridoost%2C+Somayyeh&rft.au=Musuroi%2C+Cristian&rft.au=Volmer%2C+Marius&rft.au=Florescu%2C+Monica&rft.date=2024-06-25&rft.issn=1473-0197&rft.eissn=1473-0189&rft.volume=24&rft.issue=13&rft.spage=335&rft.epage=3314&rft_id=info:doi/10.1039%2Fd4lc00211c&rft.externalDocID=d4lc00211c |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1473-0197&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1473-0197&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1473-0197&client=summon |