Evaluation of canine and feline leukocyte differential counts obtained with the scil vCell 5 compared to the Advia 2120 hematology analyzer and a manual method

The vCell 5 (scil Animal Care), a point-of-care hematology analyzer (POCA), was recently introduced to veterinary laboratories. This laser- and impedance-based analyzer is capable of providing a CBC with 5-part WBC differential count (Diff) along with WBC cytograms and flags serving as interpretatio...

Full description

Saved in:
Bibliographic Details
Published inJournal of veterinary diagnostic investigation Vol. 35; no. 6; pp. 679 - 697
Main Authors Zelmer, Kim-Lina Charlotte, Moritz, Andreas, Bauer, Natali
Format Journal Article
LanguageEnglish
Published Los Angeles, CA SAGE Publications 01.11.2023
Subjects
Online AccessGet full text

Cover

Loading…
Abstract The vCell 5 (scil Animal Care), a point-of-care hematology analyzer (POCA), was recently introduced to veterinary laboratories. This laser- and impedance-based analyzer is capable of providing a CBC with 5-part WBC differential count (Diff) along with WBC cytograms and flags serving as interpretation aids for numerical results. We compared the scil POCA-Diff to reference methods (i.e., manual differential count, Advia 2120 hematology analyzer [Siemens]) for canine and feline blood samples and considered WBC cytograms and flags. Total observed error (TEo), calculated from CV and bias%, was compared to total allowable error (TEa). Data were analyzed before and after a review process (exclusion of flagged and samples with invalid cytograms). For both species, correlation was good-to-excellent (rs = 0.81–0.97) between both analyzers for all variables, except for feline monocytes (rs = 0.21–0.63) and canine monocyte% (rs = 0.50). Smallest biases were seen for neutrophils (dog: −5.7 to 0.8%; cat: 1.5–9.4%) with both reference methods. Quality requirements (TEo < TEa) were fulfilled for canine and feline neutrophils (TEo = 5.3–10.6%, TEa = 15%) and eosinophils (TEo = 67.1–83%, TEa = (90)–50%) considering at least one reference method. Our review process led to mildly higher rs-values for most variables. Although not completely satisfactory, the scil POCA provides reliable results in compliance with ASVCP quality goals for canine and feline neutrophils and eosinophils. Analyzer flag and cytogram analysis served as useful tools for QA, indicating the necessity for manual review of blood smears, and contributed to improvement of scil POCA performance.
AbstractList The vCell 5 (scil Animal Care), a point-of-care hematology analyzer (POCA), was recently introduced to veterinary laboratories. This laser- and impedance-based analyzer is capable of providing a CBC with 5-part WBC differential count (Diff) along with WBC cytograms and flags serving as interpretation aids for numerical results. We compared the scil POCA-Diff to reference methods (i.e., manual differential count, Advia 2120 hematology analyzer [Siemens]) for canine and feline blood samples and considered WBC cytograms and flags. Total observed error (TEo), calculated from CV and bias%, was compared to total allowable error (TEa). Data were analyzed before and after a review process (exclusion of flagged and samples with invalid cytograms). For both species, correlation was good-to-excellent (rs = 0.81–0.97) between both analyzers for all variables, except for feline monocytes (rs = 0.21–0.63) and canine monocyte% (rs = 0.50). Smallest biases were seen for neutrophils (dog: −5.7 to 0.8%; cat: 1.5–9.4%) with both reference methods. Quality requirements (TEo < TEa) were fulfilled for canine and feline neutrophils (TEo = 5.3–10.6%, TEa = 15%) and eosinophils (TEo = 67.1–83%, TEa = (90)–50%) considering at least one reference method. Our review process led to mildly higher rs-values for most variables. Although not completely satisfactory, the scil POCA provides reliable results in compliance with ASVCP quality goals for canine and feline neutrophils and eosinophils. Analyzer flag and cytogram analysis served as useful tools for QA, indicating the necessity for manual review of blood smears, and contributed to improvement of scil POCA performance.
Author Moritz, Andreas
Zelmer, Kim-Lina Charlotte
Bauer, Natali
Author_xml – sequence: 1
  givenname: Kim-Lina Charlotte
  orcidid: 0000-0001-7856-7291
  surname: Zelmer
  fullname: Zelmer, Kim-Lina Charlotte
  email: Kim.L.Zelmer@vetmed.uni-giessen.de
  organization: Small Animal Clinic, Internal Medicine, Justus-Liebig-University, Giessen, Germany
– sequence: 2
  givenname: Andreas
  surname: Moritz
  fullname: Moritz, Andreas
  organization: Department of Veterinary Clinical Sciences, Clinical Pathology and Clinical Pathophysiology, Justus-Liebig-University, Giessen, Germany
– sequence: 3
  givenname: Natali
  surname: Bauer
  fullname: Bauer, Natali
  organization: Department of Veterinary Clinical Sciences, Clinical Pathology and Clinical Pathophysiology, Justus-Liebig-University, Giessen, Germany
BookMark eNplkE1OwzAQhS1UJNrCAdj5Ail2bMfOsqrKj1SJTfeRscdNimOjxikql-GquIUdmsWM9N7MG30zNAkxAEL3lCwolfKBEk4qpmTJKFVS1fUVmtKas4LXrJrkOevF2XCDZsOwJ0SUQtIp-l4ftR916mLA0WGjQxcA62CxA38ePYzv0ZwSYNs5BwcIqdMemziGNOD4lnR2WfzZpRanFvBgOo-PK_Aei-zqP_QhyylexKU9dhqXtCS4hV6n6OPulNO0P33B4RKrca_DmBN6SG20t-jaaT_A3V-fo-3jert6LjavTy-r5aYwrFKpYEIJxyomaC6QUjlROuOkoxXNSHitheGcGCi5tdbxTENJS6iCUudFNkeL37OD3kGzj-Mh_zQ0lDRnus0_uuwHcJdvlg
ContentType Journal Article
Copyright 2023 The Author(s)
Copyright_xml – notice: 2023 The Author(s)
DBID AFRWT
DOI 10.1177/10406387231187899
DatabaseName SAGE Journals Open Access
DatabaseTitleList
Database_xml – sequence: 1
  dbid: AFRWT
  name: SAGE Journals Open Access
  url: http://journals.sagepub.com/
  sourceTypes: Publisher
DeliveryMethod fulltext_linktorsrc
Discipline Veterinary Medicine
EISSN 1943-4936
EndPage 697
ExternalDocumentID 10.1177_10406387231187899
GroupedDBID ---
-MK
-TM
.2E
.2J
.2N
01A
0R~
1~K
2WC
53G
54M
5GY
5RE
5VS
AABMB
AACKU
AACMV
AACTG
AADTT
AADUE
AAEWN
AAGGD
AAGMC
AAHBH
AAJIQ
AAJOX
AAJPV
AAKGS
AAMGE
AANSI
AAPEO
AAQXH
AARDL
AARIX
AATAA
AATBZ
AAUAS
AAYTG
ABAWP
ABCCA
ABDWY
ABEIX
ABFWQ
ABHKI
ABJNI
ABKRH
ABLUO
ABPGX
ABPNF
ABQKF
ABQXT
ABRHV
ABVFX
ABYTW
ACARO
ACDSZ
ACDXX
ACFEJ
ACFMA
ACGBL
ACGFS
ACGZU
ACJTF
ACLFY
ACLHI
ACLZU
ACOFE
ACOXC
ACROE
ACSBE
ACSIQ
ACTQU
ACUAV
ACUIR
ACXKE
ACXMB
ADBBV
ADEIA
ADRRZ
ADTBJ
ADUKL
ADZZY
AECGH
AECVZ
AEDTQ
AEKYL
AENEX
AEPTA
AEQLS
AERKM
AESZF
AEUHG
AEUIJ
AEWDL
AEWHI
AEXNY
AFEET
AFKBI
AFKRG
AFMOU
AFQAA
AFRWT
AFUIA
AGKLV
AGNHF
AGWFA
AHJOV
AIOMO
AJEFB
AJMMQ
AJUZI
AJXAJ
ALJHS
ALMA_UNASSIGNED_HOLDINGS
ALTZF
AMCVQ
ANDLU
AOIJS
ARTOV
AUTPY
AUVAJ
AYAKG
B8M
BAWUL
BBRGL
BDDNI
BKIIM
BKSCU
BPACV
BSEHC
BWJAD
CBRKF
CDWPY
CFDXU
CORYS
CQQTX
DB0
DC-
DC.
DD-
DD0
DE-
DE.
DF0
DN0
DO-
DOPDO
DV7
E3Z
EBS
ECGQY
EJD
EYRJQ
F5P
FHBDP
GROUPED_SAGE_PREMIER_JOURNAL_COLLECTION
HYE
HZ~
J8X
JCYGO
K.F
M4V
O9-
OK1
P.9
P.B
P2P
Q1R
Q7L
Q7U
Q83
ROL
RPM
S01
SCNPE
SFC
SHG
SPQ
SPV
TR2
TVL
W8F
WOQ
ZONMY
ZPPRI
ZRKOI
ZSSAH
~KM
ID FETCH-LOGICAL-c368t-3585f36351515e778f52fcf7f16123149a5c440ce24dddf404087d018e2a5f33
IEDL.DBID AFRWT
ISSN 1040-6387
IngestDate Tue Jul 16 20:52:16 EDT 2024
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 6
Keywords total error
laser flow cytometry
5-part differential
agreement
method comparison
point-of-care
Language English
License This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c368t-3585f36351515e778f52fcf7f16123149a5c440ce24dddf404087d018e2a5f33
ORCID 0000-0001-7856-7291
OpenAccessLink https://journals.sagepub.com/doi/full/10.1177/10406387231187899?utm_source=summon&utm_medium=discovery-provider
PageCount 19
ParticipantIDs sage_journals_10_1177_10406387231187899
PublicationCentury 2000
PublicationDate 2023-11-01
PublicationDateYYYYMMDD 2023-11-01
PublicationDate_xml – month: 11
  year: 2023
  text: 2023-11-01
  day: 01
PublicationDecade 2020
PublicationPlace Los Angeles, CA
PublicationPlace_xml – name: Los Angeles, CA
PublicationTitle Journal of veterinary diagnostic investigation
PublicationYear 2023
Publisher SAGE Publications
Publisher_xml – name: SAGE Publications
References Lee, Kang 2016; 32
Cook 2017; 46
Giori 2011; 40
Riondato 2017; 115
Allison, Meinkoth 2007; 37
Arnold 2019; 48
Bauer 2011; 23
Comar 2017; 39
Rümke 1985; 11
Gilor, Gilor 2011; 26
Irvine 2019; 48
Kratz 2019; 41
Park 2013; 35
Harvey 2017; 19
Tvedten, Lilliehöök 2011; 40
Flatland 2018; 47
Bauer 2012; 24
Ceelie 2007; 60
Bourgès-Abella 2014; 43
Rishniw, Pion 2016; 45
Lilliehöök, Tvedten 2011; 40
Stirn 2014; 10
Cook 2016; 45
Freeman, Flatland 2017; 46
Zelmer 2022; 34
Goldmann 2014; 23
Lilliehöök, Tvedten 2009; 38
Papasouliotis 2006; 35
Tvedten 2017; 31
Lantis 2003; 119
Harris 2005; 11
Grebert 2021; 50
Merino 2018; 40
Goldmann 2013; 22
Whipple 2020; 49
Kritsepi-Konstantinou, Oikonomidis 2016; 5
Welles 2015; 18
Nabity 2018; 47
Goddard 2008; 22
Gelain 2010; 39
Briggs 2008; 30
Kratz 2005; 124
Briggs 2011; 136
Jensen, Kjelgaard-Hansen 2006; 35
Welles 2009; 38
2020
Becker 2008; 37
Kottner 2011; 48
2018
Flatland 2014; 43
Kjelgaard-Hansen, Jensen 2006; 35
Papasouliotis 1999; 28
Briggs 2009; 31
Burton 2014; 28
References_xml – volume: 39
  start-page: 306
  year: 2017
  end-page: 317
  article-title: Evaluation of criteria of manual blood smear review following automated complete blood counts in a large university hospital
  publication-title: Rev Bras Hematol Hemoter
  contributor:
    fullname: Comar
– volume: 136
  start-page: 309
  year: 2011
  end-page: 316
  article-title: Improved flagging rates on the Sysmex XE-5000 compared with the XE-2100 reduce the number of manual film reviews and increase laboratory productivity
  publication-title: Am J Clin Pathol
  contributor:
    fullname: Briggs
– volume: 45
  start-page: 466
  year: 2016
  end-page: 476
  article-title: Quality requirements for veterinary hematology analyzers in small animals—a survey about veterinary experts’ requirements and objective evaluation of analyzer performance based on a meta-analysis of method validation studies: bench top hematology analyzer
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Cook
– volume: 5
  start-page: 62
  year: 2016
  end-page: 68
  article-title: The interpretation of leukogram in dog and cat
  publication-title: Hellenic J Comp Anim Med
  contributor:
    fullname: Oikonomidis
– volume: 48
  start-page: 661
  year: 2011
  end-page: 671
  article-title: Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed
  publication-title: Int J Nurs Stud
  contributor:
    fullname: Kottner
– volume: 48
  start-page: 406
  year: 2019
  end-page: 412
  article-title: Grey eosinophils in a Miniature Schnauzer with a poorly differentiated mast cell tumor
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Irvine
– volume: 50
  start-page: 184
  year: 2021
  end-page: 197
  article-title: Validation of the Sysmex XN-V hematology analyzer for canine specimens
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Grebert
– volume: 35
  start-page: 276
  year: 2006
  end-page: 286
  article-title: Method comparison in the clinical laboratory
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Kjelgaard-Hansen
– volume: 22
  start-page: 309
  year: 2008
  end-page: 316
  article-title: Prognostic usefulness of blood leukocyte changes in canine parvoviral enteritis
  publication-title: J Vet Intern Med
  contributor:
    fullname: Goddard
– volume: 45
  start-page: 604
  year: 2016
  end-page: 614
  article-title: Evaluation of performance of veterinary in-clinic hematology analyzers
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Pion
– volume: 49
  start-page: 240
  year: 2020
  end-page: 248
  article-title: Frequency and classification of errors in laboratory medicine at a veterinary teaching hospital in the United States
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Whipple
– volume: 43
  start-page: 24
  year: 2014
  end-page: 35
  article-title: Changes in hematology measurements in healthy and diseased dog blood stored at room temperature for 24 and 48 hours using the XT-2000iV analyzer
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Bourgès-Abella
– volume: 41
  start-page: 437
  year: 2019
  end-page: 447
  article-title: Digital morphology analyzers in hematology: ICSH review and recommendations
  publication-title: Int J Lab Hematol
  contributor:
    fullname: Kratz
– volume: 115
  start-page: 288
  year: 2017
  end-page: 293
  article-title: Identification of peripheral blood involvement in dogs with large B-cell lymphoma: comparison of different methods
  publication-title: Res Vet Sci
  contributor:
    fullname: Riondato
– year: 2020
  article-title: Icon-5 & Icon-5 OP Hematology Analyzers Case Study Book
  publication-title: Norma Instruments Zrt
– volume: 37
  start-page: 245
  year: 2007
  end-page: 266
  article-title: Hematology without the numbers: in-clinic blood film evaluation
  publication-title: Vet Clin North Am Small Anim Pract
  contributor:
    fullname: Meinkoth
– volume: 40
  start-page: 54
  year: 2018
  end-page: 61
  article-title: Optimizing morphology through blood cell image analysis
  publication-title: Int J Lab Hematol
  contributor:
    fullname: Merino
– volume: 23
  start-page: 283
  year: 2014
  end-page: 296
  article-title: Evaluation of the IDEXX ProCyte Dx analyzer for dogs and cats compared to the Siemens ADVIA 2120 and manual differential
  publication-title: Comp Clin Pathol
  contributor:
    fullname: Goldmann
– volume: 119
  start-page: 656
  year: 2003
  end-page: 662
  article-title: Elimination of instrument-driven reflex manual differential leukocyte counts. Optimization of manual blood smear review criteria in a high-volume automated hematology laboratory
  publication-title: Am J Clin Pathol
  contributor:
    fullname: Lantis
– volume: 10
  start-page: 125
  year: 2014
  article-title: Rate of manual leukocyte differentials in dog, cat and horse blood samples using ADVIA 120 cytograms
  publication-title: BMC Vet Res
  contributor:
    fullname: Stirn
– volume: 26
  start-page: 109
  year: 2011
  end-page: 118
  article-title: Common laboratory artifacts caused by inappropriate sample collection and transport: how to get the most out of a sample
  publication-title: Top Companion Anim Med
  contributor:
    fullname: Gilor
– volume: 11
  start-page: 47
  year: 2005
  end-page: 61
  article-title: The ADVIA 2120 hematology system: flow cytometry-based analysis of blood and body fluids in the routine hematology laboratory
  publication-title: Lab Hematol
  contributor:
    fullname: Harris
– volume: 124
  start-page: 770
  year: 2005
  end-page: 781
  article-title: Performance evaluation of the CellaVision DM96 system: WBC differentials by automated digital image analysis supported by an artificial neural network
  publication-title: Am J Clin Pathol
  contributor:
    fullname: Kratz
– volume: 39
  start-page: 169
  year: 2010
  end-page: 179
  article-title: Identification of neoplastic cells in blood using the Sysmex XT-2000iV: a preliminary step in the diagnosis of canine leukemia
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Gelain
– volume: 35
  start-page: 517
  year: 2013
  end-page: 527
  article-title: Automated digital cell morphology identification system (CellaVision DM96) is very useful for leukocyte differentials in specimens with qualitative or quantitative abnormalities
  publication-title: Int J Lab Hematol
  contributor:
    fullname: Park
– volume: 35
  start-page: 268
  year: 2006
  end-page: 270
  article-title: Is the inherent imprecision of manual leukocyte differential counts acceptable for quantitative purposes?
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Jensen
– volume: 40
  start-page: 450
  year: 2011
  end-page: 458
  article-title: Errors in basophil enumeration with 3 veterinary hematology systems and observations on occurrence of basophils in dogs
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Tvedten
– volume: 30
  start-page: 105
  year: 2008
  end-page: 116
  article-title: ICSH Guideline for worldwide point-of-care testing in haematology with special reference to the complete blood count
  publication-title: Int J Lab Hematol
  contributor:
    fullname: Briggs
– volume: 47
  start-page: 9
  year: 2018
  end-page: 21
  article-title: ASVCP guidelines: allowable total error hematology
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Nabity
– volume: 31
  start-page: 48
  year: 2009
  end-page: 60
  article-title: Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system
  publication-title: Int J Lab Hematol
  contributor:
    fullname: Briggs
– volume: 46
  start-page: 5
  year: 2017
  end-page: 8
  article-title: Studies of instrument/method comparison, as part of validation for use in veterinary species
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Flatland
– volume: 38
  start-page: 175
  year: 2009
  end-page: 182
  article-title: Validation of the Sysmex XT-2000iV hematology system for dogs, cats, and horses. II. Differential leukocyte counts
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Tvedten
– volume: 47
  start-page: 527
  year: 2018
  end-page: 538
  article-title: Analytical quality goals—a review
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Flatland
– volume: 48
  start-page: 542
  year: 2019
  end-page: 618
  article-title: ASVCP guidelines: Principles of Quality Assurance and Standards for Veterinary Clinical Pathology (version 3.0), developed by the American Society for Veterinary Clinical Pathology’s (ASVCP) Quality Assurance and Laboratory Standards (QALS) Committee
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Arnold
– volume: 22
  start-page: 855
  year: 2013
  end-page: 868
  article-title: Comparison of the ProCyte Dx analyzer with the ADVIA 2120 and the manual differential for validation of equine and bovine hemograms
  publication-title: Comp Clin Pathol
  contributor:
    fullname: Goldmann
– volume: 18
  start-page: 147
  year: 2015
  end-page: 155
  article-title: Automated in-clinic hematology instruments for small animal practitioners: what is available, what can they really do, and how do I make a choice?
  publication-title: Vet Clin North Am Exot Anim Pract
  contributor:
    fullname: Welles
– volume: 23
  start-page: 1168
  year: 2011
  end-page: 1180
  article-title: Evaluation of the automated hematology analyzer Sysmex XT-2000iV™ compared to the ADVIA 2120 for its use in dogs, cats, and horses: part I—precision, linearity, and accuracy of complete blood cell count
  publication-title: J Vet Diagn Invest
  contributor:
    fullname: Bauer
– volume: 37
  start-page: 373
  year: 2008
  end-page: 384
  article-title: Comparative clinical study of canine and feline total blood cell count results with seven in-clinic and two commercial laboratory hematology analyzers
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Becker
– volume: 40
  start-page: 324
  year: 2011
  end-page: 339
  article-title: Canine differential leukocyte counting with the CellaVision DM96Vision, Sysmex XT-2000iV, and Advia 2120 hematology analyzers and a manual method
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Lilliehöök
– volume: 19
  start-page: 529
  year: 2017
  end-page: 540
  article-title: The feline blood film
  publication-title: J Feline Med Surg
  contributor:
    fullname: Harvey
– volume: 34
  start-page: 504
  year: 2022
  end-page: 517
  article-title: Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats
  publication-title: J Vet Diagn Invest
  contributor:
    fullname: Zelmer
– volume: 24
  start-page: 74
  year: 2012
  end-page: 89
  article-title: Evaluation of the automated hematology analyzer Sysmex XT-2000iV™ compared to the ADVIA 2120 for its use in dogs, cats, and horses. Part II: accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging
  publication-title: J Vet Diagn Invest
  contributor:
    fullname: Bauer
– volume: 43
  start-page: 475
  year: 2014
  end-page: 486
  article-title: Differentiating between analytical and diagnostic performance evaluation with a focus on the method comparison study and identification of bias
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Flatland
– volume: 40
  start-page: 475
  year: 2011
  end-page: 483
  article-title: Grey eosinophils in sighthounds: frequency in 3 breeds and comparison of eosinophil counts determined manually and with 2 hematology analyzers
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Giori
– volume: 38
  start-page: 20
  year: 2009
  end-page: 29
  article-title: Canine complete blood counts: a comparison of four in-office instruments with the ADVIA 120 and manual differential counts
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Welles
– volume: 46
  start-page: 248
  year: 2017
  end-page: 261
  article-title: Objective evaluation of analyzer performance based on a retrospective meta-analysis of instrument validation studies: point-of-care hematology analyzers
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Cook
– volume: 32
  start-page: 241
  year: 2016
  end-page: 248
  article-title: Changes of hematological references depends on storage period and temperature conditions in rats and dogs
  publication-title: Lab Anim Res
  contributor:
    fullname: Kang
– volume: 60
  start-page: 72
  year: 2007
  end-page: 79
  article-title: Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster and Cellavision DM96
  publication-title: J Clin Pathol
  contributor:
    fullname: Ceelie
– volume: 28
  start-page: 912
  year: 2014
  end-page: 917
  article-title: Degenerative left shift as a prognostic tool in cats
  publication-title: J Vet Intern Med
  contributor:
    fullname: Burton
– volume: 31
  start-page: 1708
  year: 2017
  end-page: 1716
  article-title: Feline differential leukocyte count with ProCyte Dx: frequency and severity of a neutrophil-lymphocyte error and how to avoid it
  publication-title: J Vet Intern Med
  contributor:
    fullname: Tvedten
– volume: 35
  start-page: 295
  year: 2006
  end-page: 302
  article-title: Comparison of white blood cell differential percentages determined by the in-house LaserCyte hematology analyzer and a manual method
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Papasouliotis
– volume: 28
  start-page: 109
  year: 1999
  end-page: 115
  article-title: Analysis of feline, canine and equine hemograms using the QBC VetAutoread
  publication-title: Vet Clin Pathol
  contributor:
    fullname: Papasouliotis
– year: 2018
  article-title: User manual scil vCell 5
  publication-title: scil animal care company GmbH
– volume: 11
  start-page: 137
  year: 1985
  end-page: 140
  article-title: The imprecision of the ratio of two percentages observed in differential white blood cell counts: a warning
  publication-title: Blood Cells
  contributor:
    fullname: Rümke
SSID ssj0052571
Score 2.4152071
SecondaryResourceType review_article
Snippet The vCell 5 (scil Animal Care), a point-of-care hematology analyzer (POCA), was recently introduced to veterinary laboratories. This laser- and impedance-based...
SourceID sage
SourceType Publisher
StartPage 679
Title Evaluation of canine and feline leukocyte differential counts obtained with the scil vCell 5 compared to the Advia 2120 hematology analyzer and a manual method
URI https://journals.sagepub.com/doi/full/10.1177/10406387231187899
Volume 35
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1LS8NAEF5qe_EiWhXfzEHwFE2ymyY9SSktRagHqdpb2ewDxDSBNinUP-NfdSYPQfTgKbDZMGH2Md-8Gbvuo3Ijer7GFRDKEVILpx9qjkCOI5bzpVYRJThPH3uTZ_EwD-YtljW5MDUH17cUVoV_VF7WdLrJGn1XOxnxKUjUhghOvChEleG-yJeLytzddNWgEfJPF0tybSsKiNw6TXrbDuv4qPyJNusMxk-vs-byptqgXlXAALUqpFA7Qv8k-iP4q5RH4322VwNJGFQrf8BaJu2y7gtFt5QptjCtveaH7HP0XdIbMgvITBwGmWqwlI1uIDHFe6a2uYGmXwqe-wTKPhJryGIyHxgNZLMFBIyAYjOBzdAkCQTQRLFDnpUvB3rzJgFllQtlQdjSbo_UZLL9MKuSrISlpFKoUPWvPmKz8Wg2nDh1YwZH8V6UOxx1DMsRqhAaMmEY2cC3yobWo2IuqHPJQAnhKuMLrbUVyLgo1K4XGV_ih_yYtdMsNScMYhFz6RtrgkgKrrSMSEOM-7FyY6t0cMpuiMWLZlssvLo8-a_FOPv3zHO2Sz3jq4TCC9bOV4W5RGSRx1f1bvgCyevKsw
link.rule.ids 315,786,790,21994,27884,27955,27956,44978,45366
linkProvider SAGE Publications
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV07T8MwELagHWBBUEC8uQGJKaiJnUfHqmpVoO2AAnSrHD8kREikNq1U_gx_lbOTVEIwsCaxHd3Zvu_su-8Iuemgc8MCT6IGmHAYl8zphJIikKOI5TwuRWQSnMeTYPjMHqb-tIqqNLkwlQQXdyasCv_Ibtab1W3zxJmxsiHiEjcK0VvYJk3m0wBnd7M7eHqN633Y0Hy6JRcBOkjYorrT_LOTH3Fc1rQM9slehQmhWyrxgGyprEVaLyZQxWbLwri6AD8kX_0NOzfkGlAu-Bh4JkGbxHIFqVq-52JdKKhLn-ASTsGWhFhAnpiTACXBHL8CYj9AC5jCqqfSFHyoA9KhyO3Lrly9cUCz0wbL7WqP4HE0nq4_1dwOy-GDG1ZTKEtRH5F40I97Q6eqseAIGkSFQ9Fd0BRRhwE2Kgwj7Xta6FC7hpcF3SfuC8baQnlMSqkZCi4KZduNlMexIT0mjSzP1AmBhCWUe0orP-KMCskj4-wlnUS0Ey2kf0pujYhntYZnbsU0_ksZZ__-8prsDOPxaDa6nzyek11TCr7ME7wgjWK-VJcIGIrkqpoZ305LuBc
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1LS8NAEF60BfEiWhXfzkHwFG2STZMeS22ojxaRqr2VzT5AjI20aaH-Gf-qM-lGED14CiTZJMxkd76ZnfmGsbMmOje84SnUAJcOF4o7zVD5COR8xHKeUDKiAudev9F95DfDYGgDblQLYyU4vaC0KvyiYrGm2f2uzKXdY8QjJ0sbIjZxoxA9hlVW5dQqrcKqrfjheVCuxUT16S75CNBJwhF2X_PPh_zI5SrMS7zJNiwuhNZSkVtsRY9rrPZEySpFxSz07Cb4NvvsfDN0Q2YAZYOnQYwVGCou15Dq2WsmF7mGsv0JTuMUirYQU8gSigZoBRSCBcR_gFYwhXlbpykEUCalQ54VF1tq_iIATU8dCn7XIgyPbxPp4kNPitcKeBPEbArLdtQ7bBB3Bu2uY_ssONJvRLnjo8tgfEQeBG50GEYm8Iw0oXGJmwVdKBFIzutSe1wpZTgKLgpV3Y20J3Cgv8sq42ys9xgkPPGFp40OIsF9qUREDl_STGQ9MVIF--ycRDwqtTxyLdv4L2Uc_PvOU7Z2fxWP7q77t4dsnbrBL0sFj1gln8z0MWKGPDmxP8YXQIG5MA
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Evaluation+of+canine+and+feline+leukocyte+differential+counts+obtained+with+the+scil+vCell+5+compared+to+the+Advia+2120+hematology+analyzer+and+a+manual+method&rft.jtitle=Journal+of+veterinary+diagnostic+investigation&rft.au=Zelmer%2C+Kim-Lina+Charlotte&rft.au=Moritz%2C+Andreas&rft.au=Bauer%2C+Natali&rft.date=2023-11-01&rft.pub=SAGE+Publications&rft.issn=1040-6387&rft.eissn=1943-4936&rft.volume=35&rft.issue=6&rft.spage=679&rft.epage=697&rft_id=info:doi/10.1177%2F10406387231187899&rft.externalDocID=10.1177_10406387231187899
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1040-6387&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1040-6387&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1040-6387&client=summon