Mathematical modelling of scanner-specific bowtie filters for Monte Carlo CT dosimetry

The purpose of bowtie filters in CT scanners is to homogenize the x-ray intensity measured by the detectors in order to improve the image quality and at the same time to reduce the dose to the patient because of the preferential filtering near the periphery of the fan beam. For CT dosimetry, especia...

Full description

Saved in:
Bibliographic Details
Published inPhysics in medicine & biology Vol. 62; no. 3; pp. 781 - 809
Main Authors Kramer, R, Cassola, V F, Andrade, M E A, de Araújo, M W C, Brenner, D J, Khoury, H J
Format Journal Article
LanguageEnglish
Published England IOP Publishing 07.02.2017
Subjects
Online AccessGet full text

Cover

Loading…
Abstract The purpose of bowtie filters in CT scanners is to homogenize the x-ray intensity measured by the detectors in order to improve the image quality and at the same time to reduce the dose to the patient because of the preferential filtering near the periphery of the fan beam. For CT dosimetry, especially for Monte Carlo calculations of organ and tissue absorbed doses to patients, it is important to take the effect of bowtie filters into account. However, material composition and dimensions of these filters are proprietary. Consequently, a method for bowtie filter simulation independent of access to proprietary data and/or to a specific scanner would be of interest to many researchers involved in CT dosimetry. This study presents such a method based on the weighted computer tomography dose index, CTDIw, defined in two cylindrical PMMA phantoms of 16 cm and 32 cm diameter. With an EGSnrc-based Monte Carlo (MC) code, ratios CTDIw/CTDI100,a were calculated for a specific CT scanner using PMMA bowtie filter models based on sigmoid Boltzmann functions combined with a scanner filter factor (SFF) which is modified during calculations until the calculated MC CTDIw/CTDI100,a matches ratios CTDIw/CTDI100,a, determined by measurements or found in publications for that specific scanner. Once the scanner-specific value for an SFF has been found, the bowtie filter algorithm can be used in any MC code to perform CT dosimetry for that specific scanner. The bowtie filter model proposed here was validated for CTDIw/CTDI100,a considering 11 different CT scanners and for CTDI100,c, CTDI100,p and their ratio considering 4 different CT scanners. Additionally, comparisons were made for lateral dose profiles free in air and using computational anthropomorphic phantoms. CTDIw/CTDI100,a determined with this new method agreed on average within 0.89% (max. 3.4%) and 1.64% (max. 4.5%) with corresponding data published by CTDosimetry (www.impactscan.org) for the CTDI HEAD and BODY phantoms, respectively. Comparison with results calculated using proprietary data for the PHILIPS Brilliance 64 scanner showed agreement on average within 2.5% (max. 5.8%) and with data measured for that scanner within 2.1% (max. 3.7%). Ratios of CTDI100,c/CTDI100, p for this study and corresponding data published by CTDosimetry (www.impactscan.org) agree on average within about 11% (max. 28.6%). Lateral dose profiles calculated with the proposed bowtie filter and with proprietary data agreed within 2% (max. 5.9%), and both calculated data agreed within 5.4% (max. 11.2%) with measured results. Application of the proposed bowtie filter and of the exactly modelled filter to human phantom Monte Carlo calculations show agreement on the average within less than 5% (max. 7.9%) for organ and tissue absorbed doses.
AbstractList The purpose of bowtie filters in CT scanners is to homogenize the x-ray intensity measured by the detectors in order to improve the image quality and at the same time to reduce the dose to the patient because of the preferential filtering near the periphery of the fan beam. For CT dosimetry, especially for Monte Carlo calculations of organ and tissue absorbed doses to patients, it is important to take the effect of bowtie filters into account. However, material composition and dimensions of these filters are proprietary. Consequently, a method for bowtie filter simulation independent of access to proprietary data and/or to a specific scanner would be of interest to many researchers involved in CT dosimetry. This study presents such a method based on the weighted computer tomography dose index, CTDI , defined in two cylindrical PMMA phantoms of 16 cm and 32 cm diameter. With an EGSnrc-based Monte Carlo (MC) code, ratios CTDI /CTDI were calculated for a specific CT scanner using PMMA bowtie filter models based on sigmoid Boltzmann functions combined with a scanner filter factor (SFF) which is modified during calculations until the calculated MC CTDI /CTDI matches ratios CTDI /CTDI , determined by measurements or found in publications for that specific scanner. Once the scanner-specific value for an SFF has been found, the bowtie filter algorithm can be used in any MC code to perform CT dosimetry for that specific scanner. The bowtie filter model proposed here was validated for CTDI /CTDI considering 11 different CT scanners and for CTDI , CTDI and their ratio considering 4 different CT scanners. Additionally, comparisons were made for lateral dose profiles free in air and using computational anthropomorphic phantoms. CTDI /CTDI determined with this new method agreed on average within 0.89% (max. 3.4%) and 1.64% (max. 4.5%) with corresponding data published by CTDosimetry (www.impactscan.org) for the CTDI HEAD and BODY phantoms, respectively. Comparison with results calculated using proprietary data for the PHILIPS Brilliance 64 scanner showed agreement on average within 2.5% (max. 5.8%) and with data measured for that scanner within 2.1% (max. 3.7%). Ratios of CTDI /CTDI , for this study and corresponding data published by CTDosimetry (www.impactscan.org) agree on average within about 11% (max. 28.6%). Lateral dose profiles calculated with the proposed bowtie filter and with proprietary data agreed within 2% (max. 5.9%), and both calculated data agreed within 5.4% (max. 11.2%) with measured results. Application of the proposed bowtie filter and of the exactly modelled filter to human phantom Monte Carlo calculations show agreement on the average within less than 5% (max. 7.9%) for organ and tissue absorbed doses.
The purpose of bowtie filters in CT scanners is to homogenize the x-ray intensity measured by the detectors in order to improve the image quality and at the same time to reduce the dose to the patient because of the preferential filtering near the periphery of the fan beam. For CT dosimetry, especially for Monte Carlo calculations of organ and tissue absorbed doses to patients, it is important to take the effect of bowtie filters into account. However, material composition and dimensions of these filters are proprietary. Consequently, a method for bowtie filter simulation independent of access to proprietary data and/or to a specific scanner would be of interest to many researchers involved in CT dosimetry. This study presents such a method based on the weighted computer tomography dose index, CTDIw, defined in two cylindrical PMMA phantoms of 16 cm and 32 cm diameter. With an EGSnrc-based Monte Carlo (MC) code, ratios CTDIw/CTDI100,a were calculated for a specific CT scanner using PMMA bowtie filter models based on sigmoid Boltzmann functions combined with a scanner filter factor (SFF) which is modified during calculations until the calculated MC CTDIw/CTDI100,a matches ratios CTDIw/CTDI100,a, determined by measurements or found in publications for that specific scanner. Once the scanner-specific value for an SFF has been found, the bowtie filter algorithm can be used in any MC code to perform CT dosimetry for that specific scanner. The bowtie filter model proposed here was validated for CTDIw/CTDI100,a considering 11 different CT scanners and for CTDI100,c, CTDI100,p and their ratio considering 4 different CT scanners. Additionally, comparisons were made for lateral dose profiles free in air and using computational anthropomorphic phantoms. CTDIw/CTDI100,a determined with this new method agreed on average within 0.89% (max. 3.4%) and 1.64% (max. 4.5%) with corresponding data published by CTDosimetry (www.impactscan.org) for the CTDI HEAD and BODY phantoms, respectively. Comparison with results calculated using proprietary data for the PHILIPS Brilliance 64 scanner showed agreement on average within 2.5% (max. 5.8%) and with data measured for that scanner within 2.1% (max. 3.7%). Ratios of CTDI100,c/CTDI100, p for this study and corresponding data published by CTDosimetry (www.impactscan.org) agree on average within about 11% (max. 28.6%). Lateral dose profiles calculated with the proposed bowtie filter and with proprietary data agreed within 2% (max. 5.9%), and both calculated data agreed within 5.4% (max. 11.2%) with measured results. Application of the proposed bowtie filter and of the exactly modelled filter to human phantom Monte Carlo calculations show agreement on the average within less than 5% (max. 7.9%) for organ and tissue absorbed doses.
Author Brenner, D J
Andrade, M E A
de Araújo, M W C
Kramer, R
Khoury, H J
Cassola, V F
Author_xml – sequence: 1
  givenname: R
  surname: Kramer
  fullname: Kramer, R
  email: rkramer@uol.com.br
  organization: Federal University of Pernambuco Department of Nuclear Energy, Avenida Prof Luiz Freire, 1000, CEP 50740-540, Recife, Brazil
– sequence: 2
  givenname: V F
  surname: Cassola
  fullname: Cassola, V F
  organization: Federal University of Pernambuco Department of Nuclear Energy, Avenida Prof Luiz Freire, 1000, CEP 50740-540, Recife, Brazil
– sequence: 3
  givenname: M E A
  surname: Andrade
  fullname: Andrade, M E A
  organization: Federal University of Pernambuco Department of Nuclear Energy, Avenida Prof Luiz Freire, 1000, CEP 50740-540, Recife, Brazil
– sequence: 4
  givenname: M W C
  surname: de Araújo
  fullname: de Araújo, M W C
  organization: Federal University of Pernambuco Department of Nuclear Energy, Avenida Prof Luiz Freire, 1000, CEP 50740-540, Recife, Brazil
– sequence: 5
  givenname: D J
  surname: Brenner
  fullname: Brenner, D J
  organization: Columbia University Medical Centre Center for Radiological Research, New York, NY, USA
– sequence: 6
  givenname: H J
  surname: Khoury
  fullname: Khoury, H J
  organization: Federal University of Pernambuco Department of Nuclear Energy, Avenida Prof Luiz Freire, 1000, CEP 50740-540, Recife, Brazil
BackLink https://www.ncbi.nlm.nih.gov/pubmed/28072578$$D View this record in MEDLINE/PubMed
BookMark eNp9kE1LxDAQhoMoun7cPUkOHjxYzTTNxx5l8QsUL-o1pM1EI21Tky7iv7fLqifxNDA878vMs0s2-9gjIYfAzoBpfQ5cQiGFZOfWCl7xDTL7XW2SGWMcijkIsUN2c35jDECX1TbZKTVTpVB6Rp7v7fiKnR1DY1vaRYdtG_oXGj3Nje17TEUesAk-NLSOH2NA6kM7YsrUx0TvYz8iXdjURrp4pC7m0OGYPvfJlrdtxoPvuUeeri4fFzfF3cP17eLirmi4FGMhpQKOnFdQc6WqRk1XlVXNuJSlm1e64pY7FOBUjdPx4L2rJWNOVCgFgznfIyfr3iHF9yXm0XQhN9MPtse4zAa0UEqAnOsJZWu0STHnhN4MKXQ2fRpgZmXTrNSZlTqztjlFjr7bl3WH7jfwo28CTtdAiIN5i8vUT8_-13f8Bz50tZGl4UZpMIPz_AutyoqF
CODEN PHMBA7
CitedBy_id crossref_primary_10_1002_mp_15306
crossref_primary_10_1088_2057_1976_ac5d0c
crossref_primary_10_1371_journal_pone_0273227
crossref_primary_10_6009_jjrt_2021_JSRT_77_1_75
crossref_primary_10_2478_pjmpe_2020_0002
crossref_primary_10_3934_bioeng_2020019
Cites_doi 10.1088/0031-9155/55/15/014
10.1118/1.4908002
10.1088/0031-9155/55/1/010
10.1118/1.3515864
10.1118/1.3117683
10.1056/NEJMra072149
10.2214/ajr.176.2.1760289
10.1088/0031-9155/55/5/012
10.1117/12.535161
10.1088/0031-9155/52/9/017
10.1118/1.4932222
10.1118/1.4921120
10.1118/1.3017470
10.1118/1.3693052
10.1118/1.3551990
10.1118/1.594995
10.1088/0952-4746/34/2/363
10.1118/1.3539575
10.1118/1.597564
10.1118/1.3264616
10.1148/rg.226025128
10.1118/1.2740466
10.1148/rg.281075024
10.1118/1.4901517
10.1118/1.2713240
10.1118/1.4896196
10.1118/1.4932364
ContentType Journal Article
Copyright 2017 Institute of Physics and Engineering in Medicine
Copyright_xml – notice: 2017 Institute of Physics and Engineering in Medicine
DBID CGR
CUY
CVF
ECM
EIF
NPM
AAYXX
CITATION
7X8
DOI 10.1088/1361-6560/aa5343
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
CrossRef
MEDLINE - Academic
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
CrossRef
MEDLINE - Academic
DatabaseTitleList MEDLINE
MEDLINE - Academic

Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Biology
Physics
DocumentTitleAlternate Mathematical modelling of scanner-specific bowtie filters for Monte Carlo CT dosimetry
EISSN 1361-6560
EndPage 809
ExternalDocumentID 10_1088_1361_6560_aa5343
28072578
pmbaa5343
Genre Research Support, Non-U.S. Gov't
Journal Article
GrantInformation_xml – fundername: Conselho Nacional de Desenvolvimento Científico e Tecnológico - CNPq
  grantid: 312384/2014-9
GroupedDBID ---
-DZ
-~X
123
1JI
4.4
5B3
5RE
5VS
5ZH
7.M
7.Q
AAGCD
AAJIO
AAJKP
AALHV
AATNI
ABCXL
ABHWH
ABJNI
ABLJU
ABQJV
ABVAM
ACAFW
ACGFS
ACHIP
AEFHF
AENEX
AFYNE
AKPSB
ALMA_UNASSIGNED_HOLDINGS
AOAED
ASPBG
ATQHT
AVWKF
AZFZN
CBCFC
CEBXE
CJUJL
CRLBU
CS3
DU5
EBS
EDWGO
EJD
EMSAF
EPQRW
EQZZN
F5P
HAK
IHE
IJHAN
IOP
IZVLO
KOT
LAP
M45
N5L
N9A
NT-
NT.
P2P
PJBAE
R4D
RIN
RNS
RO9
ROL
RPA
SY9
TN5
UCJ
W28
XPP
CGR
CUY
CVF
ECM
EIF
NPM
AAYXX
CITATION
7X8
ID FETCH-LOGICAL-c365t-66713e3341b3774c772524b03662d94843a3de51d7be0111ffdb600d54e650193
IEDL.DBID IOP
ISSN 0031-9155
IngestDate Thu Jul 25 09:36:56 EDT 2024
Fri Aug 23 00:49:39 EDT 2024
Sat Sep 28 08:48:44 EDT 2024
Wed Aug 21 03:40:41 EDT 2024
Thu Jan 07 13:50:38 EST 2021
IsPeerReviewed true
IsScholarly true
Issue 3
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c365t-66713e3341b3774c772524b03662d94843a3de51d7be0111ffdb600d54e650193
Notes Institute of Physics and Engineering in Medicine
PMB-104459.R2
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
PMID 28072578
PQID 1857751698
PQPubID 23479
PageCount 29
ParticipantIDs pubmed_primary_28072578
crossref_primary_10_1088_1361_6560_aa5343
proquest_miscellaneous_1857751698
iop_journals_10_1088_1361_6560_aa5343
PublicationCentury 2000
PublicationDate 2017-02-07
PublicationDateYYYYMMDD 2017-02-07
PublicationDate_xml – month: 02
  year: 2017
  text: 2017-02-07
  day: 07
PublicationDecade 2010
PublicationPlace England
PublicationPlace_xml – name: England
PublicationTitle Physics in medicine & biology
PublicationTitleAbbrev PMB
PublicationTitleAlternate Phys. Med. Biol
PublicationYear 2017
Publisher IOP Publishing
Publisher_xml – name: IOP Publishing
References 22
23
24
NCRP (30) 2009
26
27
28
29
Lee E (21) 2014; 34
Kramer R (19) 2010; 55
DeMarco J J (10) 2007; 52
Liu H (25) 2010; 55
Cranley K (9) 1997
32
11
33
12
34
35
36
18
ICRP (14) 1996
Atherton J V (1) 1993
IAEA International Atomic Energy Agency (13) 2007
2
3
4
5
6
7
Jucius R A (17) 1977; 127
Jones D G (16) 1991
NIST National Institute of Standards (31) 2015
Cassola V F (8) 2010; 55
ICRP (15) 2007
20
References_xml – volume: 55
  start-page: 4399
  issn: 0031-9155
  year: 2010
  ident: 8
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/55/15/014
  contributor:
    fullname: Cassola V F
– ident: 18
  doi: 10.1118/1.4908002
– year: 2009
  ident: 30
  contributor:
    fullname: NCRP
– volume: 55
  start-page: 163
  issn: 0031-9155
  year: 2010
  ident: 19
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/55/1/010
  contributor:
    fullname: Kramer R
– ident: 22
  doi: 10.1118/1.3515864
– ident: 34
  doi: 10.1118/1.3117683
– ident: 7
  doi: 10.1056/NEJMra072149
– ident: 6
  doi: 10.2214/ajr.176.2.1760289
– volume: 55
  start-page: 1441
  issn: 0031-9155
  year: 2010
  ident: 25
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/55/5/012
  contributor:
    fullname: Liu H
– ident: 33
  doi: 10.1117/12.535161
– volume: 52
  start-page: 2583
  issn: 0031-9155
  year: 2007
  ident: 10
  publication-title: Phys. Med. Biol.
  doi: 10.1088/0031-9155/52/9/017
  contributor:
    fullname: DeMarco J J
– volume: 127
  start-page: 286
  year: 1977
  ident: 17
  publication-title: Appl. Opt. Instrum. Eng. Med.
  contributor:
    fullname: Jucius R A
– ident: 24
  doi: 10.1118/1.4932222
– ident: 35
  doi: 10.1118/1.4921120
– year: 2007
  ident: 15
  publication-title: Recommendations of the International Commission on Radiological Protection
  contributor:
    fullname: ICRP
– ident: 26
  doi: 10.1118/1.3017470
– ident: 20
  doi: 10.1118/1.3693052
– ident: 27
  doi: 10.1118/1.3551990
– ident: 32
  doi: 10.1118/1.594995
– year: 1991
  ident: 16
  contributor:
    fullname: Jones D G
– year: 2007
  ident: 13
  publication-title: Dosimetry in Diagnostic Radiology: an International Code of Practice
  contributor:
    fullname: IAEA International Atomic Energy Agency
– year: 1993
  ident: 1
  contributor:
    fullname: Atherton J V
– year: 1996
  ident: 14
  publication-title: Conversion Coefficients for Use in Radiological Protection Against External Radiation
  contributor:
    fullname: ICRP
– volume: 34
  start-page: 363
  issn: 0952-4746
  year: 2014
  ident: 21
  publication-title: J. Radiol. Prot.
  doi: 10.1088/0952-4746/34/2/363
  contributor:
    fullname: Lee E
– ident: 5
  doi: 10.1118/1.3539575
– ident: 12
  doi: 10.1118/1.597564
– ident: 4
  doi: 10.1118/1.3264616
– ident: 29
  doi: 10.1148/rg.226025128
– ident: 11
  doi: 10.1118/1.2740466
– ident: 2
  doi: 10.1148/rg.281075024
– year: 1997
  ident: 9
  contributor:
    fullname: Cranley K
– ident: 28
  doi: 10.1118/1.4901517
– ident: 3
  doi: 10.1118/1.2713240
– ident: 36
  doi: 10.1118/1.4896196
– year: 2015
  ident: 31
  contributor:
    fullname: NIST National Institute of Standards
– ident: 23
  doi: 10.1118/1.4932364
SSID ssj0011824
Score 2.2926886
Snippet The purpose of bowtie filters in CT scanners is to homogenize the x-ray intensity measured by the detectors in order to improve the image quality and at the...
SourceID proquest
crossref
pubmed
iop
SourceType Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 781
SubjectTerms Algorithms
bowtie filter
CT dosimetry
exposure to patients
Filtration - instrumentation
Filtration - methods
Humans
Models, Theoretical
Monte Carlo Method
Phantoms, Imaging
Radiation Dosage
Tomography Scanners, X-Ray Computed - standards
Tomography, X-Ray Computed - methods
Title Mathematical modelling of scanner-specific bowtie filters for Monte Carlo CT dosimetry
URI https://iopscience.iop.org/article/10.1088/1361-6560/aa5343
https://www.ncbi.nlm.nih.gov/pubmed/28072578
https://search.proquest.com/docview/1857751698
Volume 62
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3di9QwEB_uThRf_Di_Vj2JoA8-ZO_aJE2KT7J4nMKqD3dyD0JImhQWvXbZdpHzr3em6S6cqIhvpaRpOvmY33RmfgPwos6FLqtScVWGnEtTRO5DGbmSaHsgoqDsXYq2-FCcnMn35-p8B15vc2Ha5Xj0T_EyEQUnEY4BceYwE0XGiTPm0DklpNiFa6h1j6huw7uPn7YuBATOiYJZZJxI0Ecf5e96uKKTdvG9f4abg9o5vg1fNgNO0SZfp-veT6sfv3A5_ucX3YFbIxxlb1LTu7ATm324ngpUXu7DjfnoesebQ6xo1d2Dz_Mt1Ss-OpTSoZx21tasw4lq4opT_ibFIDHffu8XkdULcsp3DBEymxMfFpu51beWzU5ZaLvFRexXl_fh7Pjt6eyEj-UZeCUK1fOi0PQLFdWgFwgiK8TpKpceVWKRh1IaKZwIUWVB-0gV7es6eIRXQcmIsBCB4wPYa9omPgJmjK8dWaZOo3mWZd6JWpWZjFXptDySE3i1mSC7TCwcdvCeG2NJeJaEZ5PwJvAS5WzHrdj9pd3BlXbLC2-L3AqrTWaXoZ7A880asLjlyI_imtiusUejtCb_opnAw7Q4tqMiciE6BR__4yiewM2cgALFgeunsNev1vEAYU7vnw3L-SdN_-9B
link.rule.ids 315,786,790,27957,27958,38900,53877
linkProvider IOP Publishing
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV3fb9MwED6xISZeBgwGZRsYCR54cEtiO3YeUUe1AR172NDeTJw4UgVLqibVNP763cVppSFASLxFkeM45x_3Xe7uO4DXZSx0mqeKq7SIuTSJ565IPVcSbQ9EFJS9S9EWJ8nRufx4oS76OqddLkw974_-IV4GouAgwj4gzowikUScOGNGWaaEFKN5UW7AXdy5mqyv4y-nazcCgudAwywiTkTovZ_yd73c0ksb-O4_Q85O9UwewLfVoEPEyffhsnXD_OcvfI7_8VUPYbuHpex9aP4I7vhqB-6FQpXXO7A17V3weLOLGc2bx_B1uqZ8xUe7kjqU287qkjU4YZVfcMrjpFgk5uqrduZZOSPnfMMQKbMp8WKxcbb4UbPxGSvqZnbp28X1EziffDgbH_G-TAPPRaJaniSafqWiOnQCwWSOeF3F0qFqTOIilUaKTBReRYV2nirbl2XhEGYVSnqEhwggd2Gzqiv_DJgxrszIQs00mmlR5DJRqjSSPk8zLd_JAbxdTZKdBzYO23nRjbEkQEsCtEGAA3iDsrb9lmz-0u7gVrv5pbNJbIXVJrI4CQN4tVoHFrce-VOyytdL7NEorcnPaAbwNCyQ9aiIZIhOw-f_OIqXsHV6OLGfj08-7cH9mLADhYbrfdhsF0t_gMindS-61X0Dwp30sw
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Mathematical+modelling+of+scanner-specific+bowtie+filters+for+Monte+Carlo+CT+dosimetry&rft.jtitle=Physics+in+medicine+%26+biology&rft.au=Kramer%2C+R&rft.au=Cassola%2C+V+F&rft.au=Andrade%2C+M+E+A&rft.au=de+Ara%C3%BAjo%2C+M+W+C&rft.date=2017-02-07&rft.pub=IOP+Publishing&rft.issn=0031-9155&rft.eissn=1361-6560&rft.volume=62&rft.issue=3&rft.spage=781&rft.epage=809&rft_id=info:doi/10.1088%2F1361-6560%2Faa5343&rft.externalDocID=pmbaa5343
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0031-9155&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0031-9155&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0031-9155&client=summon