Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade
Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade Philip Böhme , MD , Michele Floriot , MD , Marie-Aude Sirveaux , MD , Danielle Durain , RN , Olivier Ziegler , MD, PHD , Pierre Drouin , MD and Bruno Guerci , MD, PHD From the Service de Diabétologie, Maladies Métaboli...
Saved in:
Published in | Diabetes care Vol. 26; no. 4; pp. 1170 - 1175 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Alexandria, VA
American Diabetes Association
01.04.2003
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade
Philip Böhme , MD ,
Michele Floriot , MD ,
Marie-Aude Sirveaux , MD ,
Danielle Durain , RN ,
Olivier Ziegler , MD, PHD ,
Pierre Drouin , MD and
Bruno Guerci , MD, PHD
From the Service de Diabétologie, Maladies Métaboliques & Maladies de la Nutrition, Hôpital Jeanne d’Arc, CHU Nancy, France
Abstract
OBJECTIVE —To assess and compare the technical accuracy of portable glucose meters during the last decade.
RESEARCH DESIGN AND METHODS —One-thousand preprandial (pre) and postprandial (post) capillary whole-blood glucose values measured with meters owned mainly
by diabetic patients were compared with a single laboratory method yearly from 1989 to 1999. A total of 21,950 capillary measurements
and their corresponding laboratory reference values were analyzed at our clinic.
RESULTS —The lowest mean absolute difference was found in 1989 (pre: 2 ± 22 mg/dl, post: 9 ± 31 mg/dl) (mean ± SD). The highest mean
absolute difference was observed in 1993 (pre: 31 ± 33 mg/dl) and 1996 (post: 50 ± 35 mg/dl). The highest mean relative deviation
was observed in 1990 (pre: 16.4%) and 1996 (post: 20.6%). The highest percentage of readings that were within a 5% deviation
limit were observed in 1998 (pre: 44.5%) and in 1997 (post: 36.7%). Based on blood glucose levels within ±5 and ±10% of laboratory
values, the technical accuracy of meters was similar for 1989 and 1999 ( P = 0.27 and 0.52, respectively). The percentage of pre values in zone A of Clarke’s error grid analysis was >90% in 1989,
1997, 1998, and 1999.
CONCLUSIONS —The analytical performance of glucose meters decreased between 1990 and 1996 but was restored between 1997 and 1999. Nevertheless,
our data suggest that the technical accuracy of glucose meters has not significantly improved during the last decade. Complementary
studies taking into account the preanalytical improvements of the recent meters, as well as their calibration method, appear
necessary.
EGA, error grid analysis
HB, home blood glucose meter
RM, reference method
Footnotes
Address correspondence and reprint requests to Dr. Philip Böhme, Service de Diabétologie, Maladies Métaboliques & Maladies
de la Nutrition, Hôpital Jeanne d’Arc, Centre Hospitalo-Universitaire de Nancy, BP 303, 54201 Toul cedex, France. E-mail:
p.bohme{at}chu-nancy.fr .
Received for publication 22 August 2002 and accepted in revised form 15 January 2003.