Determination of meropenem in serum by high-performance liquid chromatography with column switching
A rapid, accurate and sensitive liquid chromatographic assay with on-line solid-phase extraction for determination of meropenem in serum is described. Sample was directly injected onto the extraction column for sample clean-up and extraction. Thereafter, using an on-line column-switching system the...
Saved in:
Published in | Journal of Chromatography A Vol. 812; no. 1; pp. 249 - 253 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Netherlands
Elsevier B.V
03.07.1998
|
Subjects | |
Online Access | Get full text |
ISSN | 0021-9673 |
DOI | 10.1016/S0021-9673(98)00249-0 |
Cover
Summary: | A rapid, accurate and sensitive liquid chromatographic assay with on-line solid-phase extraction for determination of meropenem in serum is described. Sample was directly injected onto the extraction column for sample clean-up and extraction. Thereafter, using an on-line column-switching system the drug was quantitatively transferred and separated on a C
18 analytical column. Ultraviolet absorption at 298 nm was used for detection. The assay was linear from 1 to 100 μg/ml. Recovery was 98.5%. Based on a 20-μl sample volume (serum–water, 1:1, v/v), detection limit was 0.1 μg/ml. An application of the method to study the pharmacokinetics of meropenem is given. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 |
ISSN: | 0021-9673 |
DOI: | 10.1016/S0021-9673(98)00249-0 |