Dose Efficiency of Quarter-Millimeter Photon-Counting Computed Tomography: First-in-Human Results
The aim of this study was to assess the clinical feasibility, image quality, and radiation dose implications of 0.25-mm imaging mode in a cohort of humans, achieved by dividing the photon-counting detector (PCD) size in half compared with standard-resolution photon-counting computed tomography (CT)...
Saved in:
Published in | Investigative radiology Vol. 53; no. 6; p. 365 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.06.2018
|
Online Access | Get more information |
Cover
Loading…
Abstract | The aim of this study was to assess the clinical feasibility, image quality, and radiation dose implications of 0.25-mm imaging mode in a cohort of humans, achieved by dividing the photon-counting detector (PCD) size in half compared with standard-resolution photon-counting computed tomography (CT) (0.5 mm).
In this technical feasibility study, a whole-body prototype PCD-CT scanner was studied in the 0.25 mm detector mode (measured at isocenter). A high-resolution PCD-CT protocol was first tested in phantom and canine studies in terms of image noise and spatial resolution. Then, 8 human subjects (mean age, 58 ± 8 years; 2 men) underwent axial PCD 0.25-mm scans of the brain, the thorax, and at the level of the upper left kidney. Filtered backprojection reconstruction was performed with a sharp kernel (B70) for standard-resolution and high-resolution data at 0.5-mm isotropic image voxel. High-resolution data, in addition, were reconstructed with an ultrasharp kernel (U70) at 0.25-mm isotropic voxels.
Image reconstructions from the PCD 0.25-mm detector system led to an improvement in resolution from 9 to 18 line pairs/cm in a line pair phantom. Modulation transfer function improved from 9.5 to 15.8 line pairs/cm at 10% modulation transfer function. When fully exploiting this improvement, image noise increased by 75% compared with dose-matched 0.5-mm slice PCD standard-resolution acquisition. However, when comparing with standard-resolution data at same in-plane resolution and slice thickness, the PCD 0.25-mm detector mode showed 19% less image noise in phantom, animal, and human scans.
High-resolution photon-counting CT in humans showed improved image quality in terms of spatial resolution and image noise compared with standard-resolution photon-counting. |
---|---|
AbstractList | The aim of this study was to assess the clinical feasibility, image quality, and radiation dose implications of 0.25-mm imaging mode in a cohort of humans, achieved by dividing the photon-counting detector (PCD) size in half compared with standard-resolution photon-counting computed tomography (CT) (0.5 mm).
In this technical feasibility study, a whole-body prototype PCD-CT scanner was studied in the 0.25 mm detector mode (measured at isocenter). A high-resolution PCD-CT protocol was first tested in phantom and canine studies in terms of image noise and spatial resolution. Then, 8 human subjects (mean age, 58 ± 8 years; 2 men) underwent axial PCD 0.25-mm scans of the brain, the thorax, and at the level of the upper left kidney. Filtered backprojection reconstruction was performed with a sharp kernel (B70) for standard-resolution and high-resolution data at 0.5-mm isotropic image voxel. High-resolution data, in addition, were reconstructed with an ultrasharp kernel (U70) at 0.25-mm isotropic voxels.
Image reconstructions from the PCD 0.25-mm detector system led to an improvement in resolution from 9 to 18 line pairs/cm in a line pair phantom. Modulation transfer function improved from 9.5 to 15.8 line pairs/cm at 10% modulation transfer function. When fully exploiting this improvement, image noise increased by 75% compared with dose-matched 0.5-mm slice PCD standard-resolution acquisition. However, when comparing with standard-resolution data at same in-plane resolution and slice thickness, the PCD 0.25-mm detector mode showed 19% less image noise in phantom, animal, and human scans.
High-resolution photon-counting CT in humans showed improved image quality in terms of spatial resolution and image noise compared with standard-resolution photon-counting. |
Author | Bluemke, David A Henning, André Pourmorteza, Amir Symons, Rolf Ulzheimer, Stefan |
Author_xml | – sequence: 1 givenname: Amir surname: Pourmorteza fullname: Pourmorteza, Amir – sequence: 2 givenname: Rolf surname: Symons fullname: Symons, Rolf – sequence: 3 givenname: André surname: Henning fullname: Henning, André – sequence: 4 givenname: Stefan surname: Ulzheimer fullname: Ulzheimer, Stefan – sequence: 5 givenname: David A surname: Bluemke fullname: Bluemke, David A |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/29595753$$D View this record in MEDLINE/PubMed |
BookMark | eNpNj8lOwzAARC0Eogv8AUL-ARcv8RJuKLS0UhFQlXPlxHZrlNhRnBz69y0CJOYyozmM3kzAZYjBAnBH8IzgXD5s1qsZ_q9MsAswJpwJhCnBIzBJ6evcU4nZNRjRnOdccjYG-jkmC-fO-crbUB1hdPBj0F1vO_Tq69o39hzh-yH2MaAiDqH3YQ-L2LRDbw3cxibuO90ejo9w4bvUIx_Qcmh0gBubhrpPN-DK6TrZ21-fgs_FfFss0frtZVU8rVHFuGBIKSKNKy1TlkittTTlN6UQWtHMGlVWJsMiy7UkZ_CSGY2tEkwpzqWjrqRTcP-z2w5lY82u7Xyju-Pu7ys9ATucV9E |
CitedBy_id | crossref_primary_10_1097_RLI_0000000000001013 crossref_primary_10_1016_j_jcmg_2018_09_030 crossref_primary_10_1186_s42492_022_00125_0 crossref_primary_10_1177_15266028241230943 crossref_primary_10_1088_1361_6560_ac71f0 crossref_primary_10_1007_s00330_021_08421_8 crossref_primary_10_1007_s10554_018_1499_6 crossref_primary_10_1088_2057_1976_ab37e9 crossref_primary_10_3348_kjr_2024_0096 crossref_primary_10_6009_jjrt_2021_JSRT_77_4_397 crossref_primary_10_2139_ssrn_4196852 crossref_primary_10_2214_AJR_20_22870 crossref_primary_10_1007_s00330_022_08933_x crossref_primary_10_1007_s00330_022_08642_5 crossref_primary_10_1007_s00330_023_10499_1 crossref_primary_10_1016_j_ejrad_2024_111437 crossref_primary_10_1109_JSEN_2023_3301498 crossref_primary_10_1038_s41523_020_00207_3 crossref_primary_10_1016_j_compmedimag_2025_102515 crossref_primary_10_1007_s00117_021_00812_8 crossref_primary_10_2214_AJR_22_28515 crossref_primary_10_3390_diagnostics14242843 crossref_primary_10_1109_TRPMS_2020_3003251 crossref_primary_10_1016_j_ejmp_2023_102683 crossref_primary_10_1016_j_ejmp_2020_10_030 crossref_primary_10_1097_RLI_0000000000000895 crossref_primary_10_1016_j_ejrad_2022_110645 crossref_primary_10_1016_j_jcct_2018_10_008 crossref_primary_10_1148_radiol_2018181156 crossref_primary_10_1002_mp_14047 crossref_primary_10_1002_mp_16983 crossref_primary_10_1016_j_jcct_2022_02_003 crossref_primary_10_1016_j_ejrad_2023_110829 crossref_primary_10_1148_radiol_2020192791 crossref_primary_10_1016_j_jcct_2023_10_009 crossref_primary_10_1007_s00330_022_08925_x crossref_primary_10_1148_radiol_2020203896 crossref_primary_10_1186_s41747_023_00401_5 crossref_primary_10_3390_jcm10245757 crossref_primary_10_1016_j_zemedi_2022_04_001 crossref_primary_10_1109_TRPMS_2020_3019954 crossref_primary_10_1148_radiol_212579 crossref_primary_10_1007_s00330_024_11244_y crossref_primary_10_1097_RLI_0000000000000601 crossref_primary_10_1097_RLI_0000000000000887 crossref_primary_10_1097_RLI_0000000000000968 crossref_primary_10_1148_radiol_213260 crossref_primary_10_1109_TRPMS_2020_3007380 crossref_primary_10_1148_radiol_2020202767 crossref_primary_10_1007_s00247_022_05404_9 crossref_primary_10_1097_RLI_0000000000000965 crossref_primary_10_1038_s41598_018_35888_1 crossref_primary_10_1007_s11604_023_01485_0 crossref_primary_10_1016_j_ejrad_2023_111209 crossref_primary_10_1117_1_JMI_11_S1_S12803 crossref_primary_10_1055_a_1308_2693 crossref_primary_10_3390_jcm12113798 crossref_primary_10_3390_diagnostics13132201 crossref_primary_10_1007_s00590_021_02883_8 crossref_primary_10_1007_s00261_024_04298_5 crossref_primary_10_1007_s00330_022_09225_0 crossref_primary_10_1016_j_clinimag_2023_110008 crossref_primary_10_1097_RLI_0000000000000816 crossref_primary_10_1007_s00330_021_08152_w crossref_primary_10_1038_s41598_023_35367_2 crossref_primary_10_1097_RLI_0000000000000933 crossref_primary_10_1097_RLI_0000000000000898 crossref_primary_10_1016_j_bone_2024_117027 crossref_primary_10_1007_s00330_023_09790_y crossref_primary_10_1016_j_yacr_2024_05_004 crossref_primary_10_1055_a_1663_2933 crossref_primary_10_1097_RLI_0000000000000616 crossref_primary_10_1097_RLI_0000000000000814 crossref_primary_10_1097_RLI_0000000000000614 crossref_primary_10_1038_s41598_023_38238_y crossref_primary_10_1055_s_0042_1749180 crossref_primary_10_1007_s00330_023_09616_x crossref_primary_10_1097_RCT_0000000000001562 crossref_primary_10_1097_RLI_0000000000000873 crossref_primary_10_1259_bjr_20220544 crossref_primary_10_1038_s41598_019_56427_6 crossref_primary_10_3390_jimaging8040112 crossref_primary_10_1007_s42058_020_00029_z crossref_primary_10_1148_radiol_222827 crossref_primary_10_1097_RLI_0000000000000948 crossref_primary_10_1088_1361_6560_ac155e crossref_primary_10_1097_RLI_0000000000000949 crossref_primary_10_1016_j_acra_2020_07_033 crossref_primary_10_1097_RLI_0000000000000668 crossref_primary_10_1097_RLI_0000000000000789 crossref_primary_10_1097_RLI_0000000000000787 crossref_primary_10_3174_ajnr_A8350 crossref_primary_10_1097_RLI_0000000000000901 |
ContentType | Journal Article |
DBID | NPM |
DOI | 10.1097/RLI.0000000000000463 |
DatabaseName | PubMed |
DatabaseTitle | PubMed |
DatabaseTitleList | PubMed |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database |
DeliveryMethod | no_fulltext_linktorsrc |
Discipline | Medicine |
EISSN | 1536-0210 |
ExternalDocumentID | 29595753 |
Genre | Research Support, Non-U.S. Gov't Journal Article Research Support, N.I.H., Intramural |
GroupedDBID | --- .-D .55 .GJ .Z2 0R~ 3O- 4Q1 4Q2 4Q3 53G 5GY 5RE 5VS 8L- AAAAV AAAXR AAGIX AAHPQ AAIQE AAMOA AAMTA AAQKA AARTV AASCR AAXQO AAYEP ABASU ABBUW ABDIG ABJNI ABVCZ ABXVJ ABZAD ACDDN ACEWG ACGFO ACGFS ACIJW ACILI ACLDA ACWDW ACWRI ACXJB ACXNZ ADFPA ADGGA ADHPY ADNKB AE3 AE6 AEETU AENEX AFDTB AFFNX AFUWQ AGINI AHOMT AHQNM AHRYX AHVBC AIJEX AINUH AJCLO AJIOK AJNWD AJNYG AJZMW AKCTQ AKULP ALKUP ALMA_UNASSIGNED_HOLDINGS ALMTX AMJPA AMKUR AMNEI AOHHW AWKKM BOYCO BQLVK BS7 C45 CS3 DIWNM DU5 DUNZO E.X EBS EEVPB EJD ERAAH EX3 F2K F2L F5P FCALG FL- GNXGY GQDEL H0~ HLJTE HZ~ IKREB IKYAY IN~ IPNFZ JF9 JG8 JK3 JK8 K8S KD2 KMI L-C N9A NPM N~M O9- OAG OAH OCUKA ODA OL1 OLG OLV OLW OLZ OPUJH ORVUJ OUVQU OVD OVDNE OVIDH OVLEI OWU OWV OWW OWX OWY OWZ OXXIT P-K P2P R58 RIG RLZ RXW S4R S4S T8P TAF TEORI TSPGW TWZ V2I VVN W3M WH7 WOQ WOW X3V X3W X7M XXN XYM YFH YOC ZFV ZGI ZXP ZZMQN |
ID | FETCH-LOGICAL-c3563-8817dfbe38e17aaa7db295966a824ed8bcd40649a71753b3da0e86388557f2fb2 |
IngestDate | Wed Feb 19 02:35:05 EST 2025 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 6 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c3563-8817dfbe38e17aaa7db295966a824ed8bcd40649a71753b3da0e86388557f2fb2 |
PMID | 29595753 |
ParticipantIDs | pubmed_primary_29595753 |
PublicationCentury | 2000 |
PublicationDate | 2018-06-01 |
PublicationDateYYYYMMDD | 2018-06-01 |
PublicationDate_xml | – month: 06 year: 2018 text: 2018-06-01 day: 01 |
PublicationDecade | 2010 |
PublicationPlace | United States |
PublicationPlace_xml | – name: United States |
PublicationTitle | Investigative radiology |
PublicationTitleAlternate | Invest Radiol |
PublicationYear | 2018 |
SSID | ssj0002703 |
Score | 2.534164 |
Snippet | The aim of this study was to assess the clinical feasibility, image quality, and radiation dose implications of 0.25-mm imaging mode in a cohort of humans,... |
SourceID | pubmed |
SourceType | Index Database |
StartPage | 365 |
Title | Dose Efficiency of Quarter-Millimeter Photon-Counting Computed Tomography: First-in-Human Results |
URI | https://www.ncbi.nlm.nih.gov/pubmed/29595753 |
Volume | 53 |
hasFullText | |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1LSwMxEA5VQbyI77fk4E2ibvaV9SbaoqIiasGbJJsEF7qtSBXs7_AHO9kk3VoV1B6WdsOGbebr7GT6zTcI7SSZSFUYxoSyXJMo5xERsGsmCY1kLnUcBVVq4PIqOW1H5_fxfaPxPsJaeumLvXzwbV3Jf6wK58Cupkr2D5YdTgon4D3YF45gYTj-ysYnhmzerEQgqgpKCPwcSZOYGr-iNFyX3evHnukTfOy7QrhGDoZCUzrBapMXaBUQCJKiS4Z5_ZeOlXnyweuIJser2n3msviUk7-GWy4Nd3dgc7VlMST-3r7BmticTq-j6_xr1S_J8yrtf_Z-sN0ZPKrC9Xa57SvtYOwyFAGrmVR7yntVw3R2_FXndq1GsIPXqA8NbfOIL77dagbfXJxZzUn_iqyDHDH3U1nZm2ZxBsHoL0bHFLf90ASagL2HaaZqMkDu6U7BRfoSzCzd_-52jMC0m2Jss1IFLXdzaNbtNvCRhc48aqjuApq-dHyKRcQNgnCNINzT-CuC8BiCsEcQrhF0iD_jBzv8LKF2q3l3fEpc0w2Sh3ESEsaCVGqhQqaClHOeSmG-TJJwRiMlmcglxIBRxlOj8SpCyQ8UAyfO4jjVVAu6jCa7va5aRZjTgMN0MgQnD5cGGYOpKHxIaQ7XyTW0Yhfn4ckqqzz4ZVv_cWQDzdQQ20RTGn7Kagviwr7Yrgz1AR92Yks |
linkProvider | National Library of Medicine |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Dose+Efficiency+of+Quarter-Millimeter+Photon-Counting+Computed+Tomography%3A+First-in-Human+Results&rft.jtitle=Investigative+radiology&rft.au=Pourmorteza%2C+Amir&rft.au=Symons%2C+Rolf&rft.au=Henning%2C+Andr%C3%A9&rft.au=Ulzheimer%2C+Stefan&rft.date=2018-06-01&rft.eissn=1536-0210&rft.volume=53&rft.issue=6&rft.spage=365&rft_id=info:doi/10.1097%2FRLI.0000000000000463&rft_id=info%3Apmid%2F29595753&rft_id=info%3Apmid%2F29595753&rft.externalDocID=29595753 |