Influence of interchannel spacing on the dynamical properties of multichannel quartz crystal microbalance

For a multichannel quartz crystal microbalance (MQCM) device, the key problem in design is to minimize the frequency interference between adjacent channels. In this paper, the effects of interchannel separation on the resonance frequencies and frequency interference of the device are studied. The re...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on ultrasonics, ferroelectrics, and frequency control Vol. 51; no. 2; pp. 249 - 253
Main Authors Shen, Feng, Lu, Pin
Format Journal Article
LanguageEnglish
Published United States IEEE 01.02.2004
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:For a multichannel quartz crystal microbalance (MQCM) device, the key problem in design is to minimize the frequency interference between adjacent channels. In this paper, the effects of interchannel separation on the resonance frequencies and frequency interference of the device are studied. The results show that, for the channels with the same size, the resonant modes will appear alternatively, and the frequency interference keeps oscillating in a very small scope in the cases of large interchannel separations. Based on the studies, the considerations for MQCM design also are discussed.