Monte Carlo modeling of Halcyon and Ethos radiotherapy beam using CAD geometry: validation and IAEA-compliant phase space

A Monte Carlo (MC) model of a Halcyon and Ethos (Varian Medical Systems, a Siemens Healthineers Company) radiotherapy beam was validated and field-independent phase space (PHSP) files were recorded above the dual-layer multileaf collimators (MLC). The treatment head geometry was modeled according to...

Full description

Saved in:
Bibliographic Details
Published inPhysics in medicine & biology Vol. 68; no. 4; pp. 44002 - 44013
Main Authors Laakkonen, Linda, Lehtomäki, Jouko, Cahill, Alexander, Constantin, Magdalena, Kulmala, Antti, Harju, Ari
Format Journal Article
LanguageEnglish
Published England IOP Publishing 10.02.2023
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A Monte Carlo (MC) model of a Halcyon and Ethos (Varian Medical Systems, a Siemens Healthineers Company) radiotherapy beam was validated and field-independent phase space (PHSP) files were recorded above the dual-layer multileaf collimators (MLC). The treatment head geometry was modeled according to engineering drawings and the dual-layer MLC was imported from CAD (computer-aided design) files. The information for the incident electron beam was achieved from an iterative electromagnetic solver. The validation of the model was performed by comparing the dose delivered by the square MLC fields as well as complex field measurements. An electron phase space was generated from linac simulations and achieved improved MC results. The output factors for square fields were within 1% and the largest differences of 5% were found in the build-up region of PDDs and the penumbra region of profiles. With the more complicated MLC-shaped field (Fishbone), the largest differences of up to 8% were found in the MLC leaf tip region due to the uncertainty of the MLC positioning and the mechanical leaf gap value. The impact of the collimator rotation on the PHSP solution has been assessed with both small and large fields, confirming negligible effects on in-field and out-of-field dose distributions. A computational model of the Halcyon and Ethos radiotherapy beam with a high accuracy implementation of the MLC was shown to be able to reproduce the radiation beam characteristics with square fields and more complex MLC-shaped fields. The field-independent PHSP files that were produced can be used as an accurate treatment head model above the MLC, and reduce the time to simulate particle transport through treatment head components.
Bibliography:PMB-113991.R2
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0031-9155
1361-6560
DOI:10.1088/1361-6560/acb4d9