Evaluating Research Grade Bioimpedance Hardware Using Textile Electrodes for Long-Term Fluid Status Monitoring

Fluid overload is a chronic medical condition that affects over six million Americans with conditions such as congestive heart failure, end-stage renal disease, and lymphedema. Remote management of fluid overload continues to be a leading clinical challenge. Bioimpedance is one technique that can be...

Full description

Saved in:
Bibliographic Details
Published inFrontiers in electronics (Online) Vol. 2
Main Authors Delano, Maggie, Ganapati, Vidya, Kamal, Rezhwan, Le, Bryan, Le, Jackie, Mendoza, Rey
Format Journal Article
LanguageEnglish
Published Frontiers Media S.A 21.01.2022
Subjects
Online AccessGet full text
ISSN2673-5857
2673-5857
DOI10.3389/felec.2021.762442

Cover

Loading…
Abstract Fluid overload is a chronic medical condition that affects over six million Americans with conditions such as congestive heart failure, end-stage renal disease, and lymphedema. Remote management of fluid overload continues to be a leading clinical challenge. Bioimpedance is one technique that can be used to estimate the hydration of tissue and track it over time. However, commercially available bioimpedance measurement systems are bulky, expensive, and rely on Ag/AgCl electrodes that dry out and can irritate the skin. The use of bioimpedance today is therefore limited to clinical and research settings, with measurements performed at daily intervals or over short periods of time rather than continuously and long-term. This paper proposes using wearable calf bioimpedance measurements integrated into a compression sock for long-term fluid overload management. A PCB was developed using standard measurement techniques that measures the calf bioimpedance using a custom analog front-end built around an AD8302 gain-phase detection chip. Data is transmitted wirelessly via Bluetooth Low Energy to an iOS device using a custom iOS app. Bioimpedance data were collected both from the wearable system and a commercial measurement system (ImpediMed SFB7) using RRC networks, Ag/AgCl electrodes, and the textile compression sock. Bioimpedance data collected from the wearable system showed close agreement with data from the SFB7 when using RRC networks and in five healthy human subjects with Ag/AgCl electrodes. However, when using the textile compression sock the wearable system had worse precision than the SFB7 (4% run to run compared to < 1 % run to run) and there were larger differences between the two systems than when using the RRC networks and the Ag/AgCl electrodes. Wearable system precision and agreement with the SFB7 was improved by pressure or light wetting of the current electrodes on the sock. Future research should focus on reliable elimination of low-frequency artifacts in research grade hardware to enable long-term calf bioimpedance measurements for fluid overload management.
AbstractList Fluid overload is a chronic medical condition that affects over six million Americans with conditions such as congestive heart failure, end-stage renal disease, and lymphedema. Remote management of fluid overload continues to be a leading clinical challenge. Bioimpedance is one technique that can be used to estimate the hydration of tissue and track it over time. However, commercially available bioimpedance measurement systems are bulky, expensive, and rely on Ag/AgCl electrodes that dry out and can irritate the skin. The use of bioimpedance today is therefore limited to clinical and research settings, with measurements performed at daily intervals or over short periods of time rather than continuously and long-term. This paper proposes using wearable calf bioimpedance measurements integrated into a compression sock for long-term fluid overload management. A PCB was developed using standard measurement techniques that measures the calf bioimpedance using a custom analog front-end built around an AD8302 gain-phase detection chip. Data is transmitted wirelessly via Bluetooth Low Energy to an iOS device using a custom iOS app. Bioimpedance data were collected both from the wearable system and a commercial measurement system (ImpediMed SFB7) using RRC networks, Ag/AgCl electrodes, and the textile compression sock. Bioimpedance data collected from the wearable system showed close agreement with data from the SFB7 when using RRC networks and in five healthy human subjects with Ag/AgCl electrodes. However, when using the textile compression sock the wearable system had worse precision than the SFB7 (4% run to run compared to <1% run to run) and there were larger differences between the two systems than when using the RRC networks and the Ag/AgCl electrodes. Wearable system precision and agreement with the SFB7 was improved by pressure or light wetting of the current electrodes on the sock. Future research should focus on reliable elimination of low-frequency artifacts in research grade hardware to enable long-term calf bioimpedance measurements for fluid overload management.
Fluid overload is a chronic medical condition that affects over six million Americans with conditions such as congestive heart failure, end-stage renal disease, and lymphedema. Remote management of fluid overload continues to be a leading clinical challenge. Bioimpedance is one technique that can be used to estimate the hydration of tissue and track it over time. However, commercially available bioimpedance measurement systems are bulky, expensive, and rely on Ag/AgCl electrodes that dry out and can irritate the skin. The use of bioimpedance today is therefore limited to clinical and research settings, with measurements performed at daily intervals or over short periods of time rather than continuously and long-term. This paper proposes using wearable calf bioimpedance measurements integrated into a compression sock for long-term fluid overload management. A PCB was developed using standard measurement techniques that measures the calf bioimpedance using a custom analog front-end built around an AD8302 gain-phase detection chip. Data is transmitted wirelessly via Bluetooth Low Energy to an iOS device using a custom iOS app. Bioimpedance data were collected both from the wearable system and a commercial measurement system (ImpediMed SFB7) using RRC networks, Ag/AgCl electrodes, and the textile compression sock. Bioimpedance data collected from the wearable system showed close agreement with data from the SFB7 when using RRC networks and in five healthy human subjects with Ag/AgCl electrodes. However, when using the textile compression sock the wearable system had worse precision than the SFB7 (4% run to run compared to < 1 % run to run) and there were larger differences between the two systems than when using the RRC networks and the Ag/AgCl electrodes. Wearable system precision and agreement with the SFB7 was improved by pressure or light wetting of the current electrodes on the sock. Future research should focus on reliable elimination of low-frequency artifacts in research grade hardware to enable long-term calf bioimpedance measurements for fluid overload management.
Author Ganapati, Vidya
Kamal, Rezhwan
Le, Jackie
Le, Bryan
Mendoza, Rey
Delano, Maggie
Author_xml – sequence: 1
  givenname: Maggie
  surname: Delano
  fullname: Delano, Maggie
– sequence: 2
  givenname: Vidya
  surname: Ganapati
  fullname: Ganapati, Vidya
– sequence: 3
  givenname: Rezhwan
  surname: Kamal
  fullname: Kamal, Rezhwan
– sequence: 4
  givenname: Bryan
  surname: Le
  fullname: Le, Bryan
– sequence: 5
  givenname: Jackie
  surname: Le
  fullname: Le, Jackie
– sequence: 6
  givenname: Rey
  surname: Mendoza
  fullname: Mendoza, Rey
BookMark eNp9kNtKAzEQhoMoeOoDeJcX2JrNJpvdS5VqhYqg7XWYJpMa2W4kST28vVsrIl54NQfm-we-Y7Lfhx4JOSvZuKqa9txhh2bMGS_HquZC8D1yxGtVFbKRav9Xf0hGKT0zxrhqRNm2R6SfvEK3gez7FX3AhBDNE72JYJFe-uDXL2ihN0inEO0bRKSLtD2d43v2HdLJ8DjHYDFRFyKdhX5VzDGu6XW38ZY-ZsibRO9C73OIA3hKDhx0CUff9YQsrifzq2kxu7-5vbqYFYY3IhdNs-SidbZ1ssGGW5SCG8ucNNZyZMIxB9yAAmmxVk7VBoWStVLSlIwpUZ2Q212uDfCsX6JfQ_zQAbz-WoS40hCzNx3qZQ3Al1a1VoIAPkx1ZTk4NHUppTRDVrnLMjGkFNH95JVMb_3rL_9661_v_A-M-sMYP7jwoc8RfPcP-QkFU4_5
CitedBy_id crossref_primary_10_3390_jcm14061846
crossref_primary_10_1016_j_aeue_2025_155701
crossref_primary_10_1109_TCSI_2023_3331979
Cites_doi 10.1109/TBME.2016.2641958
10.1161/CIR.0000000000000950
10.1088/2057-1976/ab5b02
10.1053/j.jrn.2014.10.014
10.1088/0967-3334/19/2/012
10.1038/s41598-020-59817-3
10.1097/MAT.0b013e3181d89160
10.1016/j.cpcardiol.2020.100691
10.1007/s12265-010-9229-z
10.1111/j.1751-7133.2011.00220.x
10.3390/s140610895
10.1101/SQB.1940.008.01.013
10.1088/0957-0233/22/11/115801
10.1515/9783110449822-014
10.1007/s11897-009-0039-z
10.1088/0967-3334/27/12/004
10.1016/S0140-6736(11)60101-3
10.1016/j.measurement.2018.03.079
10.1088/0967-3334/29/6/S42
10.1681/ASN.2016121341
10.1088/0967-3334/35/7/1373
10.1088/0967-3334/31/2/009
10.1007/s11517-013-1064-3
10.1088/0967-3334/30/7/004
10.3390/s18061719
10.5617/jeb.298
10.1088/1361-6579/aaf277
10.1007/s00134-006-0237-z10.1007/s00134-006-0237-z
10.1109/JBHI.2016.2628766
10.1088/0967-3334/29/6/S41
10.1016/j.clnu.2004.09.012
10.1109/TBCAS.2020.3021186
10.3390/nu11040809
ContentType Journal Article
DBID AAYXX
CITATION
DOA
DOI 10.3389/felec.2021.762442
DatabaseName CrossRef
DOAJ Directory of Open Access Journals
DatabaseTitle CrossRef
DatabaseTitleList
CrossRef
Database_xml – sequence: 1
  dbid: DOA
  name: DOAJ Directory of Open Access Journals
  url: https://www.doaj.org/
  sourceTypes: Open Website
DeliveryMethod fulltext_linktorsrc
Discipline Engineering
EISSN 2673-5857
ExternalDocumentID oai_doaj_org_article_b6aa2bd79d5a4a26aa63d2afec61555c
10_3389_felec_2021_762442
GroupedDBID 9T4
AAFWJ
AAYXX
AFPKN
ALMA_UNASSIGNED_HOLDINGS
CITATION
GROUPED_DOAJ
M~E
OK1
ID FETCH-LOGICAL-c284t-88b249fd9f58e82de542cd0f5cdd2e04f0fa2ca7a5de67f76ce4756775c100743
IEDL.DBID DOA
ISSN 2673-5857
IngestDate Wed Aug 27 01:30:37 EDT 2025
Tue Jul 01 01:42:40 EDT 2025
Thu Apr 24 23:01:59 EDT 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c284t-88b249fd9f58e82de542cd0f5cdd2e04f0fa2ca7a5de67f76ce4756775c100743
OpenAccessLink https://doaj.org/article/b6aa2bd79d5a4a26aa63d2afec61555c
ParticipantIDs doaj_primary_oai_doaj_org_article_b6aa2bd79d5a4a26aa63d2afec61555c
crossref_primary_10_3389_felec_2021_762442
crossref_citationtrail_10_3389_felec_2021_762442
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2022-01-21
PublicationDateYYYYMMDD 2022-01-21
PublicationDate_xml – month: 01
  year: 2022
  text: 2022-01-21
  day: 21
PublicationDecade 2020
PublicationTitle Frontiers in electronics (Online)
PublicationYear 2022
Publisher Frontiers Media S.A
Publisher_xml – name: Frontiers Media S.A
References Ferreira (B17) 2017; 21
Bonnet (B9) 2016
Schlebusch (B31) 2010
Abraham (B2); 377
Cole (B11) 1940; 8
Zhu (B40); 29
Abraham (B3); 17
Montalibet (B27) 2018
Schwaiger (B32) 2020; 10
Yang (B39) 2006; 27
Delano (B14)
Wang (B38) 2019; 6
Kyle (B22) 2004; 23
Hersek (B20) 2017; 64
Scharfetter (B30) 1998; 19
Sergi (B34) 2004; 34
Bogónez-Franco (B7) 2009; 30
Delano (B15); 39
Adamson (B4) 2009; 6
B12
Posada-Quintero (B29) 2018; 18
Dheman (B16) 2020; 14
Grimnes (B19) 2015
Zhu (B41); 29
Medrano (B25) 20101992; 56
Beckmann (B5) 2009
Beckmann (B6) 2010; 31
Bogónez-Franco (B8) 2016; 138
Lukaski (B24) 2019; 11
Seoane (B33) 2011; 22
Delano (B13) 2020
Khalil (B21) 2014; 14
Mohamadou (B26) 2018; 123
Montgomery (B28) 2013; 51
Buendia (B10) 2014; 35
Gómez-Sánchez (B18) 2012; 3
La Franca (B23) 2021; 46
Virani (B37) 2021; 143
Zoccali (B42) 2017; 28
Troughton (B36) 2011; 4
Abbas (B1) 2015; 25
Shochat (B35) 2006; 32
References_xml – ident: B12
– volume: 64
  start-page: 2353
  year: 2017
  ident: B20
  article-title: Wearable Vector Electrical Bioimpedance System to Assess Knee Joint Health
  publication-title: IEEE Trans. Biomed. Eng.
  doi: 10.1109/TBME.2016.2641958
– volume: 143
  start-page: e254
  year: 2021
  ident: B37
  article-title: Heart Disease and Stroke Statistics-2021 Update: A Report from the American Heart Association
  publication-title: Circulation
  doi: 10.1161/CIR.0000000000000950
– volume: 6
  start-page: 015010
  year: 2019
  ident: B38
  article-title: Textile Band Electrodes as an Alternative to Spot Ag/AgCl Electrodes for Calf Bioimpedance Measurements
  publication-title: Biomed. Phys. Eng. Express
  doi: 10.1088/2057-1976/ab5b02
– volume: 25
  start-page: 234
  year: 2015
  ident: B1
  article-title: Bioimpedance Can Solve Problems of Fluid Overload
  publication-title: J. Ren. Nutr.
  doi: 10.1053/j.jrn.2014.10.014
– volume: 19
  start-page: 247
  year: 1998
  ident: B30
  article-title: A Model of Artefacts Produced by Stray Capacitance during Whole Body or Segmental Bioimpedance Spectroscopy
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/19/2/012
– volume: 10
  start-page: 1
  year: 2020
  ident: B32
  article-title: Bioimpedance Spectroscopy for Fluid Status Assessment in Patients with Decompensated Liver Cirrhosis: Implications for Peritoneal Dialysis
  publication-title: Sci. Rep.
  doi: 10.1038/s41598-020-59817-3
– volume: 56
  start-page: 215
  year: 20101992
  ident: B25
  article-title: A Novel Bioimpedance Technique to Monitor Fluid Volume State during Hemodialysis Treatment
  publication-title: ASAIO J. (American Soc. Artif. Intern. Organs
  doi: 10.1097/MAT.0b013e3181d89160
– volume: 46
  start-page: 100691
  year: 2021
  ident: B23
  article-title: Physiopathology and Diagnosis of Congestive Heart Failure: Consolidated Certainties and New Perspectives
  publication-title: Curr. Probl. Cardiol.
  doi: 10.1016/j.cpcardiol.2020.100691
– volume: 4
  start-page: 3
  year: 2011
  ident: B36
  article-title: Direct Left Atrial Pressure Monitoring in Severe Heart Failure: Long-Term Sensor Performance
  publication-title: J. Cardiovasc. Trans. Res.
  doi: 10.1007/s12265-010-9229-z
– volume: 17
  start-page: 51
  ident: B3
  article-title: Intrathoracic Impedance vs Daily Weight Monitoring for Predicting Worsening Heart Failure Events: Results of the Fluid Accumulation Status Trial (FAST)
  publication-title: Congest. Heart Fail. (Greenwich, Conn.)
  doi: 10.1111/j.1751-7133.2011.00220.x
– volume: 14
  start-page: 10895
  year: 2014
  ident: B21
  article-title: The Theory and Fundamentals of Bioimpedance Analysis in Clinical Status Monitoring and Diagnosis of Diseases
  publication-title: Sensors
  doi: 10.3390/s140610895
– volume: 8
  start-page: 110
  year: 1940
  ident: B11
  article-title: Permeability and Impermeability of Cell Membranes for Ions
  publication-title: Cold Spring Harbor Symposia Quantitative Biol.
  doi: 10.1101/SQB.1940.008.01.013
– volume: 22
  start-page: 115801
  year: 2011
  ident: B33
  article-title: Simple Voltage-Controlled Current Source for Wideband Electrical Bioimpedance Spectroscopy: Circuit Dependences and Limitations
  publication-title: Meas. Sci. Technol.
  doi: 10.1088/0957-0233/22/11/115801
– volume: 138
  start-page: 138
  year: 2016
  ident: B8
  article-title: Problems Encountered during Inappropriate Use of Commercial Bioimpedance Devices in Novel Applications
  publication-title: Prog. Rep. Impedance Spectrosc. Measurements, Model. Appl.
  doi: 10.1515/9783110449822-014
– volume: 6
  start-page: 287
  year: 2009
  ident: B4
  article-title: Pathophysiology of the Transition from Chronic Compensated and Acute Decompensated Heart Failure: New Insights from Continuous Monitoring Devices
  publication-title: Curr. Heart Fail. Rep.
  doi: 10.1007/s11897-009-0039-z
– volume: 27
  start-page: 1293
  year: 2006
  ident: B39
  article-title: Design and Preliminary Evaluation of a Portable Device for the Measurement of Bioimpedance Spectroscopy
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/27/12/004
– volume: 377
  start-page: 658
  ident: B2
  article-title: Wireless Pulmonary Artery Haemodynamic Monitoring in Chronic Heart Failure: A Randomised Controlled Trial
  publication-title: The Lancet
  doi: 10.1016/S0140-6736(11)60101-3
– start-page: 3981
  year: 2020
  ident: B13
  article-title: Band Electrodes Reduce Simulated Calf Bioimpedance Measurement Errors Due to Muscle Anisotropy
– volume: 123
  start-page: 304
  year: 2018
  ident: B26
  article-title: Accuracy Enhancement in Low Frequency Gain and Phase Detector (AD8302) Based Bioimpedance Spectroscopy System
  publication-title: Measurement
  doi: 10.1016/j.measurement.2018.03.079
– volume: 29
  start-page: S503
  ident: B40
  article-title: A Method for the Estimation of Hydration State during Hemodialysis Using a Calf Bioimpedance Technique
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/29/6/S42
– start-page: 5910
  ident: B14
  article-title: Electrode Placement for Calf Bioimpedance Measurements during Hemodialysis
– volume: 28
  start-page: 2491
  year: 2017
  ident: B42
  article-title: Chronic Fluid Overload and Mortality in ESRD
  publication-title: Jasn
  doi: 10.1681/ASN.2016121341
– start-page: 302
  year: 2010
  ident: B31
  article-title: On the Road to a Textile Integrated Bioimpedance Early Warning System for Lung Edema
– volume: 35
  start-page: 1373
  year: 2014
  ident: B10
  article-title: Robustness Study of the Different Immittance Spectra and Frequency Ranges in Bioimpedance Spectroscopy Analysis for Assessment of Total Body Composition
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/35/7/1373
– volume: 31
  start-page: 233
  year: 2010
  ident: B6
  article-title: Characterization of Textile Electrodes and Conductors Using Standardized Measurement Setups
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/31/2/009
– volume: 51
  start-page: 1167
  year: 2013
  ident: B28
  article-title: Monitoring Intracellular, Interstitial, and Intravascular Volume Changes during Fluid Management Procedures
  publication-title: Med. Biol. Eng. Comput.
  doi: 10.1007/s11517-013-1064-3
– start-page: 5196
  year: 2016
  ident: B9
  article-title: Wearable Impedance Monitoring System for Dialysis Patients
– volume: 30
  start-page: 573
  year: 2009
  ident: B7
  article-title: Measurement Errors in Multifrequency Bioelectrical Impedance Analyzers with and without Impedance Electrode Mismatch
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/30/7/004
– volume: 18
  start-page: 1719
  year: 2018
  ident: B29
  article-title: Analysis of Consistency of Transthoracic Bioimpedance Measurements Acquired with Dry Carbon Black PDMS Electrodes, Adhesive Electrodes, and Wet Textile Electrodes
  publication-title: Sensors
  doi: 10.3390/s18061719
– volume: 34
  start-page: 416
  year: 2004
  ident: B34
  article-title: Body Fluid Distribution in Elderly Subjects with Congestive Heart Failure
  publication-title: Ann. Clin. Lab. Sci.
– volume: 3
  start-page: 29
  year: 2012
  ident: B18
  article-title: Description of Corrections on Electrode Polarization Impedance Using Isopotential Interface Factor
  publication-title: J. Electr. Bioimpedance
  doi: 10.5617/jeb.298
– start-page: 5775
  year: 2018
  ident: B27
  article-title: A Practical Method to Reduce Electrode Mismatch Artefacts during 4-electrode BioImpedance Spectroscopy Measurements
– volume: 39
  start-page: 125009
  ident: B15
  article-title: Evaluating Calf Bioimpedance Measurements for Fluid Overload Management in a Controlled Environment
  publication-title: Physiol. Meas.
  doi: 10.1088/1361-6579/aaf277
– volume: 32
  start-page: 1214
  year: 2006
  ident: B35
  article-title: Prediction of Cardiogenic Pulmonary Edema Onset by Monitoring Right Lung Impedance
  publication-title: Intensive Care Med.
  doi: 10.1007/s00134-006-0237-z10.1007/s00134-006-0237-z
– volume: 21
  start-page: 1224
  year: 2017
  ident: B17
  article-title: A Handheld and Textile-Enabled Bioimpedance System for Ubiquitous Body Composition Analysis. An Initial Functional Validation
  publication-title: IEEE J. Biomed. Health Inform.
  doi: 10.1109/JBHI.2016.2628766
– volume: 29
  start-page: S491
  ident: B41
  article-title: Extracellular Fluid Redistribution during Hemodialysis: Bioimpedance Measurement and Model
  publication-title: Physiol. Meas.
  doi: 10.1088/0967-3334/29/6/S41
– volume: 23
  start-page: 1430
  year: 2004
  ident: B22
  article-title: Bioelectrical Impedance Analysis-Part II: Utilization in Clinical Practice
  publication-title: Clin. Nutr.
  doi: 10.1016/j.clnu.2004.09.012
– volume: 14
  start-page: 1122
  year: 2020
  ident: B16
  article-title: Wireless, Artefact Aware Impedance Sensor Node for Continuous Bio-Impedance Monitoring
  publication-title: IEEE Trans. Biomed. Circuits Syst.
  doi: 10.1109/TBCAS.2020.3021186
– start-page: 532
  volume-title: World Congress on Medical Physics and Biomedical Engineering
  year: 2009
  ident: B5
  article-title: Monitoring of Body Fluid in Patients with Chronic Heart Failure Using Bioimpedance - Spectroscopy
– volume: 11
  start-page: 809
  year: 2019
  ident: B24
  article-title: Classification of Hydration in Clinical Conditions: Indirect and Direct Approaches Using Bioimpedance
  publication-title: Nutrients
  doi: 10.3390/nu11040809
– volume-title: Bioimpedance & Bioelectricity Basics
  year: 2015
  ident: B19
SSID ssj0002784199
Score 2.1997948
Snippet Fluid overload is a chronic medical condition that affects over six million Americans with conditions such as congestive heart failure, end-stage renal...
SourceID doaj
crossref
SourceType Open Website
Enrichment Source
Index Database
SubjectTerms bioimpedance spectroscopy
calf bioimpedance
dry electrodes
fluid overload
remote monitoring
textile electrodes
Title Evaluating Research Grade Bioimpedance Hardware Using Textile Electrodes for Long-Term Fluid Status Monitoring
URI https://doaj.org/article/b6aa2bd79d5a4a26aa63d2afec61555c
Volume 2
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV1LS8NAEF6kJz2IT6wv9uBJiE33kU2OVlKLWE8t9BY2-yiVkkofePO3O5OkJV704iWQsEmW2dmZb5jZbwi5U5xrxjkPQsF9IKLYBHGCxlC5JNS587nBs8PDt2gwFi8TOWm0-sKasIoeuBJcJ4-0ZrlViZVaaAZ3EbdMe2cwoyYNWl_weY1g6n2bTkuSKo0JUVjS8dhWBuJB1n2A_S8E--GIGnz9pWPpH5HDGhHSx2omx2TPFSfkoMETeEqKtObkLqZ0WypHn5faOtqbLWaAfC0uHsU0_KdeOloWAtARGF7Y9DStWt1Yt6IAUenropgGI7DItD_fzCxFvLlZ0Wp34w_PyLifjp4GQd0oITDgXdZBHOcQRXmbeBm7mFknBTM29NJYy1wofOg1M1ppaV2kvIqME0pGSknTLTHEOWkVi8JdEAqrxVWsFcfATEcmz7HBJJLAAA4zyrdJuJVaZmoWcWxmMc8gmkBBZ6WgMxR0Vgm6Te53r3xUFBq_De7hUuwGIvt1-QB0Iqt1IvtLJy7_4yNXZJ_hUYewG7DuNWmtlxt3AwBknd-WugbX4Vf6DeRA3Cw
linkProvider Directory of Open Access Journals
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Evaluating+Research+Grade+Bioimpedance+Hardware+Using+Textile+Electrodes+for+Long-Term+Fluid+Status+Monitoring&rft.jtitle=Frontiers+in+electronics+%28Online%29&rft.au=Delano%2C+Maggie&rft.au=Ganapati%2C+Vidya&rft.au=Kamal%2C+Rezhwan&rft.au=Le%2C+Bryan&rft.date=2022-01-21&rft.issn=2673-5857&rft.eissn=2673-5857&rft.volume=2&rft_id=info:doi/10.3389%2Ffelec.2021.762442&rft.externalDBID=n%2Fa&rft.externalDocID=10_3389_felec_2021_762442
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2673-5857&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2673-5857&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2673-5857&client=summon