Simultaneous Multislice Accelerated TSE for Improved Spatiotemporal Resolution and Diagnostic Accuracy in Magnetic Resonance Neurography: A Feasibility Study
This study aims to evaluate the utility of simultaneous multislice (SMS) acceleration for routine magnetic resonance neurography (MRN) at 3 T. Patients with multiple sclerosis underwent MRN of the sciatic nerve consisting of a standard fat-saturated T2-weighted turbo spin echo (TSE) sequence using i...
Saved in:
Published in | Investigative radiology Vol. 58; no. 5; p. 363 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.05.2023
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | This study aims to evaluate the utility of simultaneous multislice (SMS) acceleration for routine magnetic resonance neurography (MRN) at 3 T.
Patients with multiple sclerosis underwent MRN of the sciatic nerve consisting of a standard fat-saturated T2-weighted turbo spin echo (TSE) sequence using integrated parallel acquisition technique (PAT2) acceleration and 2 T2 TSE sequences using a combination of PAT-SMS acceleration (1) to reduce scan time (PAT2-SMS2; SMS-TSE FAST ) and (2) for time neutral increase of in-plane resolution (PAT1-SMS2; SMS-TSE HR ). Acquisition times were 5:29 minutes for the standard T2 TSE, 3:12 minutes for the SMS-TSE FAST , and 5:24 minutes for the SMS-TSE HR . Six qualitative imaging parameters were analyzed by 2 blinded readers using a 5-point Likert scale and T2 nerve lesions were quantified, respectively. Qualitative and quantitative image parameters were compared, and both interrater and intrarater reproducibility were statistically assessed. In addition, signal-to-noise ratio/contrast-to-noise ratio (CNR) was obtained in healthy controls using the exact same imaging protocol.
A total of 15 patients with MS (mean age ± standard deviation, 38.1 ± 11 years) and 10 healthy controls (mean age, 29.1 ± 7 years) were enrolled in this study. CNR analysis was highly reliable (intraclass correlation coefficient, 0.755-0.948) and revealed a significant CNR decrease for the sciatic nerve for both SMS protocols compared with standard T2 TSE (SMS-TSE FAST /SMS-TSE HR , -39%/-55%; P ≤ 0.01). Intrarater and interrater reliability of qualitative image review was good to excellent (κ: 0.672-0.971/0.617-0.883). Compared with the standard T2 TSE sequence, both SMS methods were shown to be superior in reducing pulsatile flow artifacts ( P < 0.01). Ratings for muscle border sharpness, detailed muscle structures, nerve border sharpness, and nerve fascicular structure did not differ significantly between the standard T2 TSE and the SMS-TSE FAST ( P > 0.05) and were significantly better for the SMS-TSE HR than for standard T2 TSE ( P < 0.001). Muscle signal homogeneity was mildly inferior for both SMS-TSE FAST ( P > 0.05) and SMS-TSE HR ( P < 0.001). A significantly higher number of T2 nerve lesions were detected by SMS-TSE HR ( P ≤ 0.01) compared with the standard T2 TSE and SMS-TSE FAST , whereas no significant difference was observed between the standard T2 TSE and SMS-TSE FAST .
Implementation of SMS offers either to substantially reduce acquisition time by over 40% without significantly impeding image quality compared with the standard T2 TSE or to increase in-plane resolution for a high-resolution approach and improved depiction of T2 nerve lesions while keeping acquisition times constant. This addresses the specific needs of MRN by providing different imaging approaches for 2D clinical MRN. |
---|---|