Active vibration control method of double crystal monochromator of P2 protein crystallography beam line at SSRF

The angular stability of the double-crystal monochromator (DCM) directly affects the positional stability of the focused spot. This paper describes the active vibration control (AVC) method of the DCM of P2 Experimental Station at SSRF based on the adaptive filtering algorithm and the AVC system imp...

Full description

Saved in:
Bibliographic Details
Published inNuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 1019; p. 165906
Main Authors Bai, Yang, Gong, Xuepeng, Lu, Qipeng, Song, Yuan, Wang, Dazhuang, Peng, Zhongqi, Zhang, Zhen, Mao, Qijun, Ma, Tianyu
Format Journal Article
LanguageEnglish
Published Elsevier B.V 11.12.2021
Subjects
Online AccessGet full text

Cover

Loading…
Abstract The angular stability of the double-crystal monochromator (DCM) directly affects the positional stability of the focused spot. This paper describes the active vibration control (AVC) method of the DCM of P2 Experimental Station at SSRF based on the adaptive filtering algorithm and the AVC system improves the performance of by optimizing the filter weight estimation algorithm. The simulation results show that the amplitude of the crystal in the pitch direction is greatly reduced by the AVC method, the RMS value of the angular displacement in the pitch direction of the crystal is 1.23 nrad, which is 91.93% lower than the passive vibration control (the RMS value of the angular displacement in the pitch direction is 15.25 nrad); the RMS value of the angular displacement in the pitch direction of the optimized AVC system is 0.233 nrad, which is 98.47% lower than that of the passive vibration control method. The AVC method can effectively ensure the angular stability of the DCM in the pitch direction, which is of great significance for the key equipment of the new generation of synchrotron radiation light source to achieve nano-radian stability.
AbstractList The angular stability of the double-crystal monochromator (DCM) directly affects the positional stability of the focused spot. This paper describes the active vibration control (AVC) method of the DCM of P2 Experimental Station at SSRF based on the adaptive filtering algorithm and the AVC system improves the performance of by optimizing the filter weight estimation algorithm. The simulation results show that the amplitude of the crystal in the pitch direction is greatly reduced by the AVC method, the RMS value of the angular displacement in the pitch direction of the crystal is 1.23 nrad, which is 91.93% lower than the passive vibration control (the RMS value of the angular displacement in the pitch direction is 15.25 nrad); the RMS value of the angular displacement in the pitch direction of the optimized AVC system is 0.233 nrad, which is 98.47% lower than that of the passive vibration control method. The AVC method can effectively ensure the angular stability of the DCM in the pitch direction, which is of great significance for the key equipment of the new generation of synchrotron radiation light source to achieve nano-radian stability.
ArticleNumber 165906
Author Wang, Dazhuang
Song, Yuan
Gong, Xuepeng
Lu, Qipeng
Zhang, Zhen
Mao, Qijun
Bai, Yang
Peng, Zhongqi
Ma, Tianyu
Author_xml – sequence: 1
  givenname: Yang
  surname: Bai
  fullname: Bai, Yang
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 2
  givenname: Xuepeng
  orcidid: 0000-0002-9761-605X
  surname: Gong
  fullname: Gong, Xuepeng
  email: gongxuepeng120@foxmail.com
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 3
  givenname: Qipeng
  surname: Lu
  fullname: Lu, Qipeng
  email: luqipeng@126.com
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 4
  givenname: Yuan
  surname: Song
  fullname: Song, Yuan
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 5
  givenname: Dazhuang
  surname: Wang
  fullname: Wang, Dazhuang
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 6
  givenname: Zhongqi
  surname: Peng
  fullname: Peng, Zhongqi
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 7
  givenname: Zhen
  surname: Zhang
  fullname: Zhang, Zhen
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 8
  givenname: Qijun
  surname: Mao
  fullname: Mao, Qijun
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
– sequence: 9
  givenname: Tianyu
  surname: Ma
  fullname: Ma, Tianyu
  organization: State key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
BookMark eNp9kFFLwzAUhYNMcFP_gE_5A51J1iYN-DKGU2GgOH0OaXrrMtrckcbB_r0t01fvy-VyzrkcvhmZBAxAyB1nc864vN_Pg-_sXDDB51wWmskLMuWlEpkulJyQ6WAqM82YuCKzvt-zYbQqpwSXLvkj0KOvok0eA3UYUsSWdpB2WFNsaI3fVQvUxVOf7CBgQLeL2NmEcdTfBD1ETODDn6fFr2gPuxOtwHa09QGoTXS7fV_fkMvGtj3c_u5r8rl-_Fg9Z5vXp5fVcpM5UfCUAYOF1U7ompcVMCekEq4oXO44CKEWeSMVlHWupWqA8aZq6kblmkuomByuxTUR578uYt9HaMwhDoTiyXBmRmRmb0ZkZkRmzsiG0MM5BEOzo4doeuchOKh9BJdMjf6_-A9P63ko
Cites_doi 10.1109/LSP.2003.821745
10.1107/S0909049507067118
ContentType Journal Article
Copyright 2021 Elsevier B.V.
Copyright_xml – notice: 2021 Elsevier B.V.
DBID AAYXX
CITATION
DOI 10.1016/j.nima.2021.165906
DatabaseName CrossRef
DatabaseTitle CrossRef
DatabaseTitleList
DeliveryMethod fulltext_linktorsrc
Discipline Physics
EISSN 1872-9576
ExternalDocumentID 10_1016_j_nima_2021_165906
S0168900221008718
GroupedDBID --K
--M
-~X
.~1
0R~
123
1B1
1RT
1~.
1~5
4.4
457
4G.
7-5
71M
8P~
9JN
AABNK
AACTN
AAEDT
AAEDW
AAIAV
AAIKJ
AAKOC
AALRI
AAOAW
AAXUO
ABMAC
ABNEU
ABYKQ
ACDAQ
ACFVG
ACGFS
ACNCT
ACRLP
ADBBV
ADEZE
AEBSH
AEKER
AFKWA
AFTJW
AGUBO
AGYEJ
AHHHB
AIEXJ
AIKHN
AITUG
AIVDX
AJOXV
ALMA_UNASSIGNED_HOLDINGS
AMFUW
AMRAJ
AXJTR
BKOJK
BLXMC
CS3
EBS
EFJIC
EFLBG
EO8
EO9
EP2
EP3
F5P
FDB
FIRID
FNPLU
FYGXN
G-Q
GBLVA
HME
IHE
J1W
KOM
LZ4
M41
MO0
N9A
O-L
O9-
OAUVE
OGIMB
OZT
P-8
P-9
PC.
Q38
RNS
ROL
RPZ
SDF
SDG
SDP
SES
SHN
SPC
SPCBC
SPD
SSQ
T5K
TN5
~02
~G-
29N
3O-
53G
5VS
6TJ
8WZ
A6W
AAQFI
AAQXK
AATTM
AAXKI
AAYWO
AAYXX
ABFNM
ABWVN
ABXDB
ACNNM
ACRPL
ADMUD
ADNMO
ADVLN
AEIPS
AFJKZ
AGCQF
AGHFR
AGQPQ
AGRNS
AIIUN
ANKPU
APXCP
ASPBG
AVWKF
AZFZN
BNPGV
CITATION
EJD
FEDTE
FGOYB
G-2
HVGLF
HX~
HZ~
H~9
R2-
RIG
SEW
SSH
SSZ
VOH
WUQ
ID FETCH-LOGICAL-c251t-e0e3a9c29d18be0c2672c55c4c1e22734f67e8d4967fe01fbfdf74916eb06fbf3
IEDL.DBID .~1
ISSN 0168-9002
IngestDate Tue Jul 01 02:30:55 EDT 2025
Fri Feb 23 02:41:24 EST 2024
IsPeerReviewed true
IsScholarly true
Keywords Active vibration control
Double-crystal monochromator
Synchrotron radiation
Adaptive filtering algorithm
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c251t-e0e3a9c29d18be0c2672c55c4c1e22734f67e8d4967fe01fbfdf74916eb06fbf3
ORCID 0000-0002-9761-605X
ParticipantIDs crossref_primary_10_1016_j_nima_2021_165906
elsevier_sciencedirect_doi_10_1016_j_nima_2021_165906
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2021-12-11
PublicationDateYYYYMMDD 2021-12-11
PublicationDate_xml – month: 12
  year: 2021
  text: 2021-12-11
  day: 11
PublicationDecade 2020
PublicationTitle Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment
PublicationYear 2021
Publisher Elsevier B.V
Publisher_xml – name: Elsevier B.V
References Chaoyin (b1) 2013
Y. Dabin, L. Zhang, D. Martin, et al. The concrete slab studies for the ESRF upgrade beamlines, in: The 7th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI) 2012, Shanghai, 2012.
Igarashi, Ikuta, al (b5) 2008; 15
(b6) 2017
Chunhui, Chaozhu, Liguo (b8) 2009
Lopes, Piedade (b7) 2004; 2
Caliari, Bagnato, De Souza (b2) 2016
(b9) 2020
Guangxu (b10) 2015
R. Geraldes, et al. Mechatronics Concepts for the New High-Dynamics DCM for Sirius, in: presented at MEDSI 2016, Barcelona, Spain, Sep. 2016, this conference.
Igarashi (10.1016/j.nima.2021.165906_b5) 2008; 15
Guangxu (10.1016/j.nima.2021.165906_b10) 2015
10.1016/j.nima.2021.165906_b4
Chaoyin (10.1016/j.nima.2021.165906_b1) 2013
Chunhui (10.1016/j.nima.2021.165906_b8) 2009
Caliari (10.1016/j.nima.2021.165906_b2) 2016
(10.1016/j.nima.2021.165906_b6) 2017
Lopes (10.1016/j.nima.2021.165906_b7) 2004; 2
10.1016/j.nima.2021.165906_b3
(10.1016/j.nima.2021.165906_b9) 2020
References_xml – reference: Y. Dabin, L. Zhang, D. Martin, et al. The concrete slab studies for the ESRF upgrade beamlines, in: The 7th International Conference on Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI) 2012, Shanghai, 2012.
– start-page: 4
  year: 2017
  ident: b6
  article-title: Preliminary design report of SSRF project (phase II)
– reference: R. Geraldes, et al. Mechatronics Concepts for the New High-Dynamics DCM for Sirius, in: presented at MEDSI 2016, Barcelona, Spain, Sep. 2016, this conference.
– start-page: 1
  year: 2020
  ident: b9
  article-title: Dcm stability test report of P2 experimental station at SSRF
– start-page: 363
  year: 2013
  end-page: 365
  ident: b1
  article-title: Optics and Engineering of Synchrotron Radiation
– year: 2016
  ident: b2
  article-title: Studies on flow-induced vibrations for the new high-dynamics DCM for sirius
  publication-title: Proceedings of the 9th Edition of the Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation Conference
– year: 2015
  ident: b10
  article-title: Research on the Influence and Evolution of High-Speed Train Wheel Non-Rounding on Vibration and Noise
– volume: 2
  start-page: 148
  year: 2004
  end-page: 151
  ident: b7
  article-title: The behavior of the modified FX-LMS algorithm with secondary path modeling errors
  publication-title: Signal Process. Lett.
– start-page: 159
  year: 2009
  end-page: 162
  ident: b8
  article-title: Adaptive Signal Processing Technology, Vol. 48-51
– volume: 15
  start-page: 292
  year: 2008
  end-page: 296
  ident: b5
  article-title: X-ray beam stabilization at BL-17a, the protein micro-crystallography beamline of the photon factor
  publication-title: Synchrotron Radiat.
– start-page: 1
  year: 2020
  ident: 10.1016/j.nima.2021.165906_b9
– volume: 2
  start-page: 148
  issue: 11
  year: 2004
  ident: 10.1016/j.nima.2021.165906_b7
  article-title: The behavior of the modified FX-LMS algorithm with secondary path modeling errors
  publication-title: Signal Process. Lett.
  doi: 10.1109/LSP.2003.821745
– start-page: 4
  year: 2017
  ident: 10.1016/j.nima.2021.165906_b6
– year: 2015
  ident: 10.1016/j.nima.2021.165906_b10
– start-page: 363
  year: 2013
  ident: 10.1016/j.nima.2021.165906_b1
– ident: 10.1016/j.nima.2021.165906_b4
– year: 2016
  ident: 10.1016/j.nima.2021.165906_b2
  article-title: Studies on flow-induced vibrations for the new high-dynamics DCM for sirius
– start-page: 159
  year: 2009
  ident: 10.1016/j.nima.2021.165906_b8
– ident: 10.1016/j.nima.2021.165906_b3
– volume: 15
  start-page: 292
  year: 2008
  ident: 10.1016/j.nima.2021.165906_b5
  article-title: X-ray beam stabilization at BL-17a, the protein micro-crystallography beamline of the photon factor
  publication-title: Synchrotron Radiat.
  doi: 10.1107/S0909049507067118
SSID ssj0000978
Score 2.3654501
Snippet The angular stability of the double-crystal monochromator (DCM) directly affects the positional stability of the focused spot. This paper describes the active...
SourceID crossref
elsevier
SourceType Index Database
Publisher
StartPage 165906
SubjectTerms Active vibration control
Adaptive filtering algorithm
Double-crystal monochromator
Synchrotron radiation
Title Active vibration control method of double crystal monochromator of P2 protein crystallography beam line at SSRF
URI https://dx.doi.org/10.1016/j.nima.2021.165906
Volume 1019
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3PS8MwGA1jIngRf-Kmjhy8SbcmTdL2OIZjKgxxDnYrSZpixbVjdoIX_3bz9QdTEA8e0zRQXpLve6Hve0HoSnDpqkBxx7ID4TANIgDfix3p-YEOuYZ_c6C2mIrJnN0t-KKFRk0tDMgq69hfxfQyWtdPBjWag1WaDmaWrAQh5CDwp7EhFirYmQ-rvP-5lXlAnULl7203tgtqnu5W45WlpfcQJX0ieAi3Hv2WnL4lnPEB2q-ZIh5WH3OIWiY7QrulYlO_HaN8WEYq_A7nXUAX16pzXF0KjfMEx_lGvRqs1x-WA9qOPMv18zpfwkEb-h8oLn0a0qx5p3GwxsrIJQYKimWBZ7PH8Qmaj2-eRhOnvj3B0ZazFI5xjSdDTcOYBMq4mgqfas4108RQMLVJhG-CmIXCT4xLEpXEic8sWzTKFbblnaJ2lmfmDGHO7S4XUrpGxcw19jBOjUdixrTHuJK0g64b2KJVZZIRNeqxlwhAjgDkqAK5g3iDbPRjqiMbxf8Y1_3nuHO0By3QoBBygdrFemMuLZMoVK9cKj20M7y9n0y_ACoCyLc
linkProvider Elsevier
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1dS8MwFA1zIvoifuLmVx58k25pm6Tt4xiOqXOI22BvJUlTrLh2zE3wxd9ubj-YgvjgY5sGyknvvSf05FyErjgTRPqSWYYdcIsqEAF4bmQJ1_NVwBT8mwO1xZD3J_RuyqY11K3OwoCsssz9RU7Ps3V5p12i2Z4nSXtkyIofQA0CfxqTYjfQJjXhC20MWp9rnQccVCgMvk1kE5DzNNcirzTJzYccu2VzFkDbo9-q07eK09tDuyVVxJ3ibfZRTacHaCuXbKq3Q5R18lSF32HDC_DiUnaOi67QOItxlK3kq8Zq8WFIoBnI0kw9L7IZ7LRh_NHBuVFDklbPVBbWWGoxw8BBsVji0eipd4QmvZtxt2-V7RMsZUjL0tJEuyJQThDZvtREOdxzFGOKKls74GoTc0_7EQ24F2tixzKOYo8auqgl4ebKPUb1NEv1CcKMmTDnQhAtI0q02Y072rUjSpVLmRROA11XsIXzwiUjrORjLyGAHALIYQFyA7EK2fDHWocmjf8xr_nPeZdouz9-GISD2-H9KdqBERCk2PYZqi8XK31uaMVSXuSfzRf2DspF
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Active+vibration+control+method+of+double+crystal+monochromator+of+P2+protein+crystallography+beam+line+at+SSRF&rft.jtitle=Nuclear+instruments+%26+methods+in+physics+research.+Section+A%2C+Accelerators%2C+spectrometers%2C+detectors+and+associated+equipment&rft.au=Bai%2C+Yang&rft.au=Gong%2C+Xuepeng&rft.au=Lu%2C+Qipeng&rft.au=Song%2C+Yuan&rft.date=2021-12-11&rft.pub=Elsevier+B.V&rft.issn=0168-9002&rft.eissn=1872-9576&rft.volume=1019&rft_id=info:doi/10.1016%2Fj.nima.2021.165906&rft.externalDocID=S0168900221008718
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0168-9002&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0168-9002&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0168-9002&client=summon