Clinical Implementation of Lung SABR IMPT Using Two PBS Systems With Different Spot Sizes and Delivery Speeds
Stereotactic Ablative Body Therapy (SABR) is an effective treatment for early-stage lung cancer. Two different proton therapy delivery systems (Proteous-plus with universal nozzle called P+ and Proteus-one called P1) were employed to test the spot size difference, plan dosimetric quality, and delive...
Saved in:
Published in | International journal of radiation oncology, biology, physics Vol. 111; no. 3; pp. e549 - e550 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Elsevier Inc
01.11.2021
|
Online Access | Get full text |
Cover
Loading…
Abstract | Stereotactic Ablative Body Therapy (SABR) is an effective treatment for early-stage lung cancer. Two different proton therapy delivery systems (Proteous-plus with universal nozzle called P+ and Proteus-one called P1) were employed to test the spot size difference, plan dosimetric quality, and delivery time needed for practical and effective clinical treatment.
Total of 9 lung patients with 10 tumors were selected for this study. Seven targets were prescribed 60 Gy in 10 fractions; three patients were 48 Gy in 4 fractions. A treatment planning system was used to generate treatment plans for each nozzle type with exact same optimization objectives. P1 plans used 4cm range-shifter placed at 45cm from isocenter to create large in-air beam spots to mimic P+ system. The spot sizes were compared to the P+ in-air spots without range shifter. P+ plans used 7cm range-shifter placed very close to patients, i.e., 4cm away, to minimize spot size degradation. Target coverage was optimized and normalized to prescription dose cover 100-percent of GTV and 90-percent covers PTV. Dose to the OARs were compared using QUANTEC metrics. Each plan was split into three subsets to perform volumetric dose-repainting to reduce interplay effect. The plan beam-on times of both systems were obtained to evaluate treatment delivery efficiency and impact of patient comfort.
The in-air beam spots were comparable between the P+ without and P1 with range shifter. All PTVs were covered by 90-percent prescription dose. OAR doses were comparable and within the QUANTEC limits. The beam-on time of P+ plans are 3 to 6 time of those of P1 plans. When combined with patient setup and image guidance, patient discomfort could lead to higher uncertainty in P+ treatment.
P1 system with 45cm away from isocenter can effectively simulate a large spot PBS system, P+. Lung SABR IMPT plan qualities between the two systems were dosimetrically comparable. Volumetric dose-repainting were applicable to both systems to reduce interplay effect. However, the plan beam-on and delivery time of P1 system were significantly shorter than that of P+. Therefore, potential less patient movement during treatment leads to less dose uncertainties to patients. |
---|---|
AbstractList | Stereotactic Ablative Body Therapy (SABR) is an effective treatment for early-stage lung cancer. Two different proton therapy delivery systems (Proteous-plus with universal nozzle called P+ and Proteus-one called P1) were employed to test the spot size difference, plan dosimetric quality, and delivery time needed for practical and effective clinical treatment.
Total of 9 lung patients with 10 tumors were selected for this study. Seven targets were prescribed 60 Gy in 10 fractions; three patients were 48 Gy in 4 fractions. A treatment planning system was used to generate treatment plans for each nozzle type with exact same optimization objectives. P1 plans used 4cm range-shifter placed at 45cm from isocenter to create large in-air beam spots to mimic P+ system. The spot sizes were compared to the P+ in-air spots without range shifter. P+ plans used 7cm range-shifter placed very close to patients, i.e., 4cm away, to minimize spot size degradation. Target coverage was optimized and normalized to prescription dose cover 100-percent of GTV and 90-percent covers PTV. Dose to the OARs were compared using QUANTEC metrics. Each plan was split into three subsets to perform volumetric dose-repainting to reduce interplay effect. The plan beam-on times of both systems were obtained to evaluate treatment delivery efficiency and impact of patient comfort.
The in-air beam spots were comparable between the P+ without and P1 with range shifter. All PTVs were covered by 90-percent prescription dose. OAR doses were comparable and within the QUANTEC limits. The beam-on time of P+ plans are 3 to 6 time of those of P1 plans. When combined with patient setup and image guidance, patient discomfort could lead to higher uncertainty in P+ treatment.
P1 system with 45cm away from isocenter can effectively simulate a large spot PBS system, P+. Lung SABR IMPT plan qualities between the two systems were dosimetrically comparable. Volumetric dose-repainting were applicable to both systems to reduce interplay effect. However, the plan beam-on and delivery time of P1 system were significantly shorter than that of P+. Therefore, potential less patient movement during treatment leads to less dose uncertainties to patients. |
Author | Hsi, W.C. Su, Z. Brooks, E. |
Author_xml | – sequence: 1 givenname: Z. surname: Su fullname: Su, Z. organization: Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, AR – sequence: 2 givenname: W.C. surname: Hsi fullname: Hsi, W.C. organization: Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, FL – sequence: 3 givenname: E. surname: Brooks fullname: Brooks, E. organization: Department of Radiation Oncology, University of Florida College of Medicine, Jacksonville, FL |
BookMark | eNqFkF1LwzAUhoMouE1_gpA_0JqkH1mvZB9-DCYOu6F3oU1ONKVtSlIn89fbMe-9OYdzOM_L4Rmj89a2gNANJSElNL2tQlM5W3YhI4yGhIc0ztgZGtEpz4IoSd7P0YhEKQmi4foSjb2vCCGU8niEmkVtWiOLGq-aroYG2r7ojW2x1Xj91X7gfDZ_xavnzRbvvBnm7bfFm3mO84PvofH4zfSfeGm0BjewOO_sUMwPeFy0Ci-hNntwh2EPoPwVutBF7eH6r0_Q7uF-u3gK1i-Pq8VsHUhGMxaotFCSTxVXqdSETssi4SqGUhEea5UxFlMuoUgojeJMal1ynSVMRwmUlEcKoglKTrnSWe8daNE50xTuICgRR2eiEidn4uhMEC6Ozgbu7sTB8NzegBNeGmglKONA9kJZ80_CL4MTefk |
ContentType | Journal Article |
Copyright | 2021 |
Copyright_xml | – notice: 2021 |
DBID | AAYXX CITATION |
DOI | 10.1016/j.ijrobp.2021.07.1492 |
DatabaseName | CrossRef |
DatabaseTitle | CrossRef |
DatabaseTitleList | |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine |
EISSN | 1879-355X |
EndPage | e550 |
ExternalDocumentID | 10_1016_j_ijrobp_2021_07_1492 S0360301621023622 |
GroupedDBID | --- --K .1- .FO 0R~ 1B1 1P~ 1RT 1~5 4.4 457 4G. 53G 5RE 7-5 AAEDT AAEDW AAIAV AALRI AAWTL AAXUO ABJNI ABLJU ABNEU ABOCM ABUDA ACGFS ACIUM ADBBV AENEX AEVXI AFCTW AFRHN AFTJW AGZHU AHHHB AITUG AJUYK ALMA_UNASSIGNED_HOLDINGS ALXNB AMRAJ BELOY DU5 EBS EFJIC F5P FDB GBLVA HED HMO IHE J1W KOM LX3 M41 MO0 O9- OC~ OO- RNS ROL RPZ SDG SEL SES SSZ UV1 XH2 Z5R ZA5 ~S- .55 .GJ 29J 5VS AAQFI AAQQT AAQXK AAYXX ABEFU ADMUD ADPAM ADVLN AFFNX AFJKZ AGRDE AKRWK ASPBG AVWKF AZFZN CITATION EJD FEDTE FGOYB FIRID G-2 HMK HVGLF HX~ HZ~ NQ- R2- RIG SAE SEW UDS X7M XPP ZGI |
ID | FETCH-LOGICAL-c2192-d6adc78d7d6cf018ba57d4ebd074fd922417cea511349cffb7f952f35eb173de3 |
ISSN | 0360-3016 |
IngestDate | Thu Sep 26 19:19:47 EDT 2024 Fri Feb 23 02:40:35 EST 2024 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 3 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c2192-d6adc78d7d6cf018ba57d4ebd074fd922417cea511349cffb7f952f35eb173de3 |
OpenAccessLink | http://www.redjournal.org/article/S0360301621023622/pdf |
ParticipantIDs | crossref_primary_10_1016_j_ijrobp_2021_07_1492 elsevier_sciencedirect_doi_10_1016_j_ijrobp_2021_07_1492 |
PublicationCentury | 2000 |
PublicationDate | 2021-11-01 2021-11-00 |
PublicationDateYYYYMMDD | 2021-11-01 |
PublicationDate_xml | – month: 11 year: 2021 text: 2021-11-01 day: 01 |
PublicationDecade | 2020 |
PublicationTitle | International journal of radiation oncology, biology, physics |
PublicationYear | 2021 |
Publisher | Elsevier Inc |
Publisher_xml | – name: Elsevier Inc |
SSID | ssj0001174 |
Score | 2.4142303 |
Snippet | Stereotactic Ablative Body Therapy (SABR) is an effective treatment for early-stage lung cancer. Two different proton therapy delivery systems (Proteous-plus... |
SourceID | crossref elsevier |
SourceType | Aggregation Database Publisher |
StartPage | e549 |
Title | Clinical Implementation of Lung SABR IMPT Using Two PBS Systems With Different Spot Sizes and Delivery Speeds |
URI | https://dx.doi.org/10.1016/j.ijrobp.2021.07.1492 |
Volume | 111 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Li9swEBYhC6WX0ifdvtCht2A3lh-yjklo2XabsmyybG7GetGk3XjJOpTuf-l_7ciSHIeG0u5FBIETRfMx82k88wmht0lUliWhIiCCZUHCpAoYsNiAK5YKmmodK5OHnH7JTi6ST4t00ev96lQtbWseituDfSV3sSrMgV1Nl-x_WLb9UpiAz2BfGMHCMP6TjSe-rbHR-L1ybUQNAfy8NUrbo_H54OP0bD6wlQHzH9XgbDzzMuXgEuqv4PTsFSn1YHZdwbC8VTeuQPm7Kdr4ae6oV7Yf2NPY_TxiR31iY6QO7BrWom2E4bueGJtJaYn8bNu8HQlbdN001QWX4aSdGgO3_2abJsJujoJErlmvTZz55pm92k6InRAChpFTwrb-N6csAAq02HPQzh0vuwf4xt2q1OqdutCtUiti-0dYsBmKVbhcbSpuVEpJZCRb4XRIdnGwrU6cmZWZhZnjMAR4iPBHBPxY3kdHo9Pzy9M21EdO5tv_k12L2LuDP3aY_HQIzfwheuBOInhkYfUI9dT6Mbo3dbUWT9CVRxfeRxeuNDbowgZd2KALN-jCgC4M6MIOXdigC7fowgZduEEXBnRhjy5s0fUUXXx4P5-cBO5yjkBAkCOBzEopaC6pzIQeRjkvUyoTxSVwUi2ZYYZUqBL4fJwwoTWnmqVExymQAxpLFT9D_XW1Vs8RVlQwwiOeq0Qnca5KpWiuhjIjPGYiE8co9LtWXFsNlsIXJ64Ku82F2eZiSAuzzcco93tbOCJpCWIBgPj7oy_u_uhLdH-H_VeoX2-26jUw1pq_cbj5Df5Bl0w |
link.rule.ids | 315,783,787,27938,27939 |
linkProvider | Library Specific Holdings |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Clinical+Implementation+of+Lung+SABR+IMPT+Using+Two+PBS+Systems+With+Different+Spot+Sizes+and+Delivery+Speeds&rft.jtitle=International+journal+of+radiation+oncology%2C+biology%2C+physics&rft.au=Su%2C+Z.&rft.au=Hsi%2C+W.C.&rft.au=Brooks%2C+E.&rft.date=2021-11-01&rft.pub=Elsevier+Inc&rft.issn=0360-3016&rft.eissn=1879-355X&rft.volume=111&rft.issue=3&rft.spage=e549&rft.epage=e550&rft_id=info:doi/10.1016%2Fj.ijrobp.2021.07.1492&rft.externalDocID=S0360301621023622 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0360-3016&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0360-3016&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0360-3016&client=summon |