Effect of misalignment of analytical centers on higher-order aberration
[Purpose] Effect of misalignment of the analytical centers on higher-order aberration was examined. [Cases and Methods] The subjects were divided into a group of phakia; 23 eyes in 23 subjects with no disorder other than ametropia, and an IOL group; 23 eyes in 23 subjects implanted with intraocular...
Saved in:
Published in | JAPANESE ORTHOPTIC JOURNAL Vol. 35; pp. 69 - 76 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
JAPANESE ASSOCIATION OF CERTIFIED ORTHOPTISTS
2006
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Abstract | [Purpose] Effect of misalignment of the analytical centers on higher-order aberration was examined. [Cases and Methods] The subjects were divided into a group of phakia; 23 eyes in 23 subjects with no disorder other than ametropia, and an IOL group; 23 eyes in 23 subjects implanted with intraocular lens were evaluated with using a Hartmann-Shack wavefront sensor (KR-9000PW™; Topcon). Analysis was performed at the center of pupil and the center of corneal reflex in order to obtain the corneal aberration and the total ocular aberration (comalike aberration, spherical-like aberration and total higher-order aberration) at pupil diameter of 4mm. (1) Misalignment of the analytical centers (distance between the center of pupil and the center of corneal reflex) (2) aberration amount in analysis at the center of pupil and the center of corneal reflex (3) change quantity of aberration (absolute value of the difference between the center of pupil and the center of corneal reflex) (4) correlation between the misalignment of the analytical centers and the change quantity of aberration were examined. [Results] (1) Misalignment of the analytical centers did not present significant difference between the two groups. (2) Corneal aberration presented significant difference in both groups for comalike aberration and total higher-order aberration, while total ocular aberration presented significant difference in the IOL group for all aberrations. (3) In change quantity of aberration, corneal aberration did not present significant difference between the two groups, while total ocular aberration were significantly larger in the IOL group than in the group of phakia for comalike aberration and total higher-order aberration. (4) Corneal aberration increased in the change quantity of aberration as the misalignment in the centers of analysis intensified, in both groups. Total ocular aberration did not indicate relation with misalignment of the analytical centers in the group of phakia, while positive correlation with the misalignment of the analytical centers was observed in IOL group. [Conclusion] Misalignment of the analytical centers has greater influence on eyes with IOL than on the phakic eyes. Lack of mydriasis or concentrical mydriasis may alter the center of pupil, thus resulting in misalignment of the analytical centers; the measurements should be taken with the pupil uniformly open. |
---|---|
AbstractList | [Purpose] Effect of misalignment of the analytical centers on higher-order aberration was examined. [Cases and Methods] The subjects were divided into a group of phakia; 23 eyes in 23 subjects with no disorder other than ametropia, and an IOL group; 23 eyes in 23 subjects implanted with intraocular lens were evaluated with using a Hartmann-Shack wavefront sensor (KR-9000PW™; Topcon). Analysis was performed at the center of pupil and the center of corneal reflex in order to obtain the corneal aberration and the total ocular aberration (comalike aberration, spherical-like aberration and total higher-order aberration) at pupil diameter of 4mm. (1) Misalignment of the analytical centers (distance between the center of pupil and the center of corneal reflex) (2) aberration amount in analysis at the center of pupil and the center of corneal reflex (3) change quantity of aberration (absolute value of the difference between the center of pupil and the center of corneal reflex) (4) correlation between the misalignment of the analytical centers and the change quantity of aberration were examined. [Results] (1) Misalignment of the analytical centers did not present significant difference between the two groups. (2) Corneal aberration presented significant difference in both groups for comalike aberration and total higher-order aberration, while total ocular aberration presented significant difference in the IOL group for all aberrations. (3) In change quantity of aberration, corneal aberration did not present significant difference between the two groups, while total ocular aberration were significantly larger in the IOL group than in the group of phakia for comalike aberration and total higher-order aberration. (4) Corneal aberration increased in the change quantity of aberration as the misalignment in the centers of analysis intensified, in both groups. Total ocular aberration did not indicate relation with misalignment of the analytical centers in the group of phakia, while positive correlation with the misalignment of the analytical centers was observed in IOL group. [Conclusion] Misalignment of the analytical centers has greater influence on eyes with IOL than on the phakic eyes. Lack of mydriasis or concentrical mydriasis may alter the center of pupil, thus resulting in misalignment of the analytical centers; the measurements should be taken with the pupil uniformly open. |
Author | Shimamura, Emiko Kuraoka, Sachiko Suto, Chikako Sato, Chiaki Takahashi, Midori |
Author_xml | – sequence: 1 fullname: Kuraoka, Sachiko organization: Department of Ophthalmology, Saiseikai Kurihashi Hospital – sequence: 2 fullname: Suto, Chikako organization: Department of Ophthalmology, Saiseikai Kurihashi Hospital – sequence: 3 fullname: Shimamura, Emiko organization: Department of Ophthalmology, Saiseikai Kurihashi Hospital – sequence: 4 fullname: Sato, Chiaki organization: Department of Ophthalmology, Saiseikai Kurihashi Hospital – sequence: 5 fullname: Takahashi, Midori organization: Department of Ophthalmology, Saiseikai Kurihashi Hospital |
BookMark | eNpNkM1uwjAQhK2KSqWUe495gaRrm_jnWCFKkZB6ac_W4tgkKNjIzoW3bygV7R52Nav55jCPZBJicIQ8U6gWTPCXQ0xDG09DZyteV0LfkSlVipea0XpCpsCVLGsq2QOZ53yAcbjSlNMpWa-8d3Yooi-OXca-24ejCz8aA_bnMRL7wo4vl3IRQ9F2-9alMqbGpQJ3LiUcuhieyL3HPrv5752Rr7fV5_K93H6sN8vXbWkZCF1qEKBrVN4LCjXT9Y6pRrmmkZR6zSwsLGtgIZDJRirplB8pYAAeJVoQfEbgmmtTzDk5b06pO2I6Gwrm0oX568Lw2gg9IpsrcsgD7t0NwDR6evcPoFrKCwTXJfTNY1tMxgX-DRRucXE |
ContentType | Journal Article |
Copyright | JAPANESE ASSOCIATION OF CERTIFIED ORTHOPTISTS |
Copyright_xml | – notice: JAPANESE ASSOCIATION OF CERTIFIED ORTHOPTISTS |
DBID | AAYXX CITATION |
DOI | 10.4263/jorthoptic.35.69 |
DatabaseName | CrossRef |
DatabaseTitle | CrossRef |
DatabaseTitleList | |
DeliveryMethod | fulltext_linktorsrc |
EISSN | 1883-9215 |
EndPage | 76 |
ExternalDocumentID | 10_4263_jorthoptic_35_69 article_jorthoptic1977_35_0_35_0_69_article_char_en |
GroupedDBID | ALMA_UNASSIGNED_HOLDINGS JSF KQ8 RJT AAYXX CITATION |
ID | FETCH-LOGICAL-c2069-906095a8ff6105295b28d8edd711f92c04c2d046a27d787e8f0690200fa7ac063 |
ISSN | 0387-5172 |
IngestDate | Fri Aug 23 04:01:59 EDT 2024 Wed Apr 05 15:04:44 EDT 2023 |
IsDoiOpenAccess | true |
IsOpenAccess | true |
IsPeerReviewed | false |
IsScholarly | true |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c2069-906095a8ff6105295b28d8edd711f92c04c2d046a27d787e8f0690200fa7ac063 |
OpenAccessLink | https://www.jstage.jst.go.jp/article/jorthoptic1977/35/0/35_0_69/_article/-char/en |
PageCount | 8 |
ParticipantIDs | crossref_primary_10_4263_jorthoptic_35_69 jstage_primary_article_jorthoptic1977_35_0_35_0_69_article_char_en |
PublicationCentury | 2000 |
PublicationDate | 2006 |
PublicationDateYYYYMMDD | 2006-01-01 |
PublicationDate_xml | – year: 2006 text: 2006 |
PublicationDecade | 2000 |
PublicationTitle | JAPANESE ORTHOPTIC JOURNAL |
PublicationTitleAlternate | Jpn. Orthopt. J. |
PublicationYear | 2006 |
Publisher | JAPANESE ASSOCIATION OF CERTIFIED ORTHOPTISTS |
Publisher_xml | – name: JAPANESE ASSOCIATION OF CERTIFIED ORTHOPTISTS |
References | 12) Stiles WS, Crawford BH: The luminous efficiency of rays entering the pupil at different points. Proc R Soc London Ser B112: 428-450, 1933 5) 酒井幸弘, 宇陀恵子, 山田裕子, 内藤尚久, 小島隆司, 市川一夫, 他:ARK-10000を用いた波面収差解析の評価. 日視会誌31:89-94, 2002 3) 後藤浩也, 前田直之:Hartmann-Shack波面センサー. あたらしい眼科21(3):289-296, 2004 1) 黒田輝仁:Hartmann-Shack波面センサー. 角膜形状解析の基礎と臨床:眼科診療プラクティス89. 453-459, 2002 4) 不二門尚:II波面収差解析 波面光学と幾何光学:角膜トポグラファーと波面センサー. 96-103, メジカルビュー社, 東京, 2002 14) He JC, Marcos S, Webb RH, Burns SA: Measurement of the wave-front aberration of the eye by a fast psychophysical procedure. J Opt Soc Am A15: 2449-2456, 1998 6) 大鹿哲郎, 岡本周子, 園尾純一郎, 平塚健太郎, 川名啓介, 坂田典繁, 他:眼科検査診断法 視覚の質quality of visionを測る. 日眼会誌108:770-808, 2004 11) Uozato H, Guyton DL: Centering corneal surgical procedures. Am J Ophthalmol 103: 264-275, 1987 2) 三橋俊文:波面センサーと収差. 診療に役立つ眼光学. 眼科診療プラクティス71. 111-113, 2001 10) Artal P, Antonio G, Berrio E, Williams D: Compensation of corneal aberration by internal optics in the human eye. Journal of Vision 1: 1-8, 2001 15) 大川賀孝, 不二門尚, 黒田輝仁, 二宮さゆり, 前田直之, 田野保雄, 他:眼内レンズの傾き・偏位の高次収差に及ぼす影響. 日眼会誌106臨増:214, 2002 9) Amano S, Amano Y, Yamagami S, Miyai T, Miyata K, Samejima T et al: Age-related changes in corneal and ocular higher-order wavefront aberration. Am J Ophthalmol. 137 (6): 998-992, 2004 7) Oshika T, Klyce SD, Applegate RA: Changes in corneal wavefront aberrations with aging. Invest Ophtalmol Vis Sci 40: 1351-1355, 1999 8) 伊藤美沙絵, 清水公也, 魚里博, 矢野隆, 関口沙知子, 田村千春, 他:高次波面収差の年齢変化. 日眼会誌108臨増:269, 2004 13) 魚里博:屈折矯正手術におけるセンタリングについて. 視覚の科学15(3):138-148, 1994 |
References_xml | |
SSID | ssj0000389131 ssib053390835 |
Score | 1.6271939 |
Snippet | [Purpose] Effect of misalignment of the analytical centers on higher-order aberration was examined. [Cases and Methods] The subjects were divided into a group... |
SourceID | crossref jstage |
SourceType | Aggregation Database Publisher |
StartPage | 69 |
SubjectTerms | analysis at the center of corneal reflex analysis at the center of pupil Hartmann-Shack wavefront sensor higher-order aberration misalignment of the analytical centers |
Title | Effect of misalignment of analytical centers on higher-order aberration |
URI | https://www.jstage.jst.go.jp/article/jorthoptic1977/35/0/35_0_69/_article/-char/en |
Volume | 35 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
ispartofPNX | JAPANESE ORTHOPTIC JOURNAL, 2006/08/31, Vol.35, pp.69-76 |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Lb9QwELaWwoFLBQJEaUE50AOKsiTOyzlCKVSsVAmplXqLHD9IWG2Cll1V7a9nxs7DXYFEuVgby3aymcnn8TwJecvzPGeVFoGWDA4oEu27rEIXKyFVFktGtfHyPc_OLpOvV-nVbHbtRpdsqrm4_WNcyf9QFfqArhglew_KjotCB_wG-kILFIb2n2jcpx5GE3nzCwTq75NpH3ONWDU1ul9ilC6QuTZOHYFJt-nzSq3XE116AfW8qRFB6qY9PkmP2Sd_sYWbo0-Xv7ixXUve-HDtGIF4t-RWvSzqZtlNhqaNVcRCJ3e662bFV1tT4Mg_XbkT-DgBBNu7CgkHsTBRbxrZWjxzZRGVMSA_tTGbA-TGqYOZtlRLv_vaYjC7uI5Z5RHX0ZTVAZSKeZzOh3luCu2drW10OISjDq5RTiuUcVpmxQPykAJCoS_o4tsIRCACF-GQh85s6caaa6ovDn_Rmrxx0fe7j3VHxHn0A6T8wUPQCC0XT8h-f9rwPljWeUpmqn1Gvli28TrtuWyD1xPbeD3beF3ruWzjTWzznFx-Pr04OQv6ehqBoGFWBEWI2QU50xpkZjTwVpRJpqTMo0gXVISJoDJMMk5zCTiumMY01kBizXMuQJZ9QfbarlUviSd5VGU8VJFmNInhk85okulKYz67Ko3iA_JueAXlT5s2pfwbDQ7IR_uOxpH9B-WMjOCYgqND22TFOAaDEwELXt3jhofk8aRIOyJ7m_VWvQbRclO9MWzwG1mgftk |
link.rule.ids | 315,783,787,4031,27935,27936,27937 |
linkProvider | Colorado Alliance of Research Libraries |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Effect+of+misalignment+of+analytical+centers+on+higher-order+aberration&rft.jtitle=Nihon+Shin%C5%8D+Kunrenshi+Ky%C5%8Dkai+shi&rft.au=Kuraoka%2C+Sachiko&rft.au=Suto%2C+Chikako&rft.au=Shimamura%2C+Emiko&rft.au=Sato%2C+Chiaki&rft.date=2006&rft.issn=0387-5172&rft.eissn=1883-9215&rft.volume=35&rft.spage=69&rft.epage=76&rft_id=info:doi/10.4263%2Fjorthoptic.35.69&rft.externalDBID=n%2Fa&rft.externalDocID=10_4263_jorthoptic_35_69 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0387-5172&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0387-5172&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0387-5172&client=summon |