Characterization and qualification of a CRNA eye dosimeter

The aim of this work is to study the calibration methodology of a new eye dosimeter, developed by Dosimetry Laboratory of the Nuclear Research Center of Algiers, using TLD-700 thermoluminescent detectors (LiF: Mg, Ti). Qualification and type test of this dosimeter have been performed in terms of ope...

Full description

Saved in:
Bibliographic Details
Published inPerspectives in science Vol. 12; p. 100402
Main Authors Larabi-Harfouche, K., Dari, F., Herrati, A., Medkour-Boushaki, G., Lounis-Mokrani, Z.
Format Journal Article
LanguageEnglish
Published Elsevier GmbH 01.09.2019
Subjects
Online AccessGet full text
ISSN2213-0209
2213-0209
DOI10.1016/j.pisc.2019.100402

Cover

More Information
Summary:The aim of this work is to study the calibration methodology of a new eye dosimeter, developed by Dosimetry Laboratory of the Nuclear Research Center of Algiers, using TLD-700 thermoluminescent detectors (LiF: Mg, Ti). Qualification and type test of this dosimeter have been performed in terms of operational quantity Hp(3) at Secondary Standard Dosimetry Laboratory (SSDL) using the cylindrical phantom according to ISO 12794, ORAMED 2012 and IEC 62387, 2012. Furthermore, two dosimeter holders have been tested for two positions (frontal and lateral) in order to determine the “best dosimeter position” which allows good estimation of dose received by lens.
ISSN:2213-0209
2213-0209
DOI:10.1016/j.pisc.2019.100402