Detecting fixation on a target using time-frequency distributions of a retinal birefringence scanning signal
Background The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such de...
Saved in:
Published in | Biomedical engineering online Vol. 12; no. 1; p. 41 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
London
BioMed Central
13.05.2013
BioMed Central Ltd |
Subjects | |
Online Access | Get full text |
ISSN | 1475-925X 1475-925X |
DOI | 10.1186/1475-925X-12-41 |
Cover
Abstract | Background
The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT.
Method
This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably.
Results
In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used.
Conclusion
Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. |
---|---|
AbstractList | Background
The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT.
Method
This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably.
Results
In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used.
Conclusion
Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. Doc number: 41 Abstract Background: The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT. Method: This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably. Results: In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used. Conclusion: Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT.BACKGROUNDThe fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT.This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably.METHODThis paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably.In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used.RESULTSIn all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used.Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path.CONCLUSIONJoint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT. This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably. In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used. Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT. This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably. In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used. Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. Background: The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT. Method: This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably. Results: In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used. Conclusion: Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. Background The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon reflection. Existing devices use this property to obtain information on the orientation of the fovea and the direction of gaze. Such devices employ specific frequency components that appear during moments of fixation on a target. To detect them, previous methods have used solely the power spectrum of the Fast Fourier Transform (FFT), which, unfortunately, is an integral method, and does not give information as to where exactly the events of interest occur. With very young patients who are not cooperative enough, this presents a problem, because central fixation may be present only during very short-lasting episodes, and can easily be missed by the FFT. Method This paper presents a method for detecting short-lasting moments of central fixation in existing devices for retinal birefringence scanning, with the goal of a reliable detection of eye alignment. Signal analysis is based on the Continuous Wavelet Transform (CWT), which reliably localizes such events in the time-frequency plane. Even though the characteristic frequencies are not always strongly expressed due to possible artifacts, simple topological analysis of the time-frequency distribution can detect fixation reliably. Results In all six subjects tested, the CWT allowed precise identification of both frequency components. Moreover, in four of these subjects, episodes of intermittent but definitely present central fixation were detectable, similar to those in Figure 4. A simple FFT is likely to treat them as borderline cases, or entirely miss them, depending on the thresholds used. Conclusion Joint time-frequency analysis is a powerful tool in the detection of eye alignment, even in a noisy environment. The method is applicable to similar situations, where short-lasting diagnostic events need to be detected in time series acquired by means of scanning some substrate along a specific path. Keywords: Retinal birefringence scanning, Time-frequency distribution, Continuous wavelet transform, Amblyopia, Strabismus, Eye alignment |
Audience | Academic |
Author | Gramatikov, Boris |
AuthorAffiliation | 1 Laboratory of Ophthalmic Optics, Krieger Children’s Eye Center, Pediatric Ophthalmology and Adult Strabismus, Wilmer Ophthalmological Institute, Johns Hopkins University School of Medicine, Rm. 233, 600 N. Wolfe St, Baltimore, Maryland, USA |
AuthorAffiliation_xml | – name: 1 Laboratory of Ophthalmic Optics, Krieger Children’s Eye Center, Pediatric Ophthalmology and Adult Strabismus, Wilmer Ophthalmological Institute, Johns Hopkins University School of Medicine, Rm. 233, 600 N. Wolfe St, Baltimore, Maryland, USA |
Author_xml | – sequence: 1 givenname: Boris surname: Gramatikov fullname: Gramatikov, Boris email: bgramat@jhmi.edu organization: Laboratory of Ophthalmic Optics, Krieger Children’s Eye Center, Pediatric Ophthalmology and Adult Strabismus, Wilmer Ophthalmological Institute, Johns Hopkins University School of Medicine |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/23668264$$D View this record in MEDLINE/PubMed |
BookMark | eNqNkktr3DAUhU1JaR7turti6KZZONHTHm0KafoKBAp9QHdCkq9dBVuaSHJJ_n3kTDqdCUkpNtjofudwfXz2ix3nHRTFS4yOMF7Ux5g1vBKE_6wwqRh-UuytT3Y23neL_RgvECII1eJZsUtoXS9IzfaK4T0kMMm6vuzslUrWuzLfqkwq9JDKKc6jZEeougCXEzhzXbY2pmD1NNOx9F3GA2QPNZTaBuhC1mQQymiUc7NBtH2ePi-edmqI8OLueVD8-Pjh--nn6vzLp7PTk_NK15ynqqXQIKyJ1gtOOqwRQ9ARRpkSiCIBgkML2LSaaqJwzbXpeJM_jtEFwRwJelC8XfkuJz1Ca8CloAa5DHZU4Vp6ZeX2xNlfsve_ZY4FU9Fkg3crA239IwbbE-NHOact57QlJpLhbPLmbovgc3AxydFGA8OgHPgpSkzrhglcC_4fKOfZfSFoRl_fQy_8FHK4K4o3NRL4L9WrAaR1nc9rmtlUnnDKas64mHM6eoDKVwujNblpnc3nW4LDLUFmElylXk0xyrNvX7fZV5s_YR3en-5l4HgFmOBjzLVZIxjJud0PJMrvKYxNt53Ne9vhHzq00sXlbTXDRmaPSG4Aa48MPw |
CitedBy_id | crossref_primary_10_1016_j_optlastec_2023_109388 crossref_primary_10_1016_j_compbiomed_2020_103672 crossref_primary_10_1016_j_medengphy_2015_06_007 crossref_primary_10_1186_1475_925X_13_104 crossref_primary_10_1080_03091902_2017_1281357 crossref_primary_10_1364_OE_22_007972 crossref_primary_10_1186_s12938_016_0128_7 crossref_primary_10_1186_1475_925X_13_52 |
Cites_doi | 10.1364/AO.38.001273 10.1117/1.1805560 10.1364/JOSAA.16.002103 10.1109/TASSP.1987.1165201 10.1364/JOSAA.5.000049 10.1016/S0169-2607(00)00060-2 10.1016/j.visres.2008.04.031 10.1364/AO.42.003047 10.1007/BF02510534 10.1364/AO.31.003730 10.1364/BOE.2.001955 10.1142/S0218001487000205 |
ContentType | Journal Article |
Copyright | Gramatikov; licensee BioMed Central Ltd. 2013 COPYRIGHT 2013 BioMed Central Ltd. 2013 Gramatikov; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Copyright © 2013 Gramatikov; licensee BioMed Central Ltd. 2013 Gramatikov; licensee BioMed Central Ltd. |
Copyright_xml | – notice: Gramatikov; licensee BioMed Central Ltd. 2013 – notice: COPYRIGHT 2013 BioMed Central Ltd. – notice: 2013 Gramatikov; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. – notice: Copyright © 2013 Gramatikov; licensee BioMed Central Ltd. 2013 Gramatikov; licensee BioMed Central Ltd. |
DBID | C6C AAYXX CITATION CGR CUY CVF ECM EIF NPM ISR 3V. 7QO 7X7 7XB 88E 8FD 8FE 8FG 8FH 8FI 8FJ 8FK ABJCF ABUWG AFKRA AZQEC BBNVY BENPR BGLVJ BHPHI CCPQU DWQXO FR3 FYUFA GHDGH GNUQQ HCIFZ K9. L6V LK8 M0S M1P M7P M7S P64 PHGZM PHGZT PIMPY PJZUB PKEHL PPXIY PQEST PQGLB PQQKQ PQUKI PRINS PTHSS 7X8 5PM |
DOI | 10.1186/1475-925X-12-41 |
DatabaseName | Open Access资源_Springer Nature OA Free Journals CrossRef Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed Gale In Context: Science ProQuest Central (Corporate) Biotechnology Research Abstracts Health & Medical Collection ProQuest Central (purchase pre-March 2016) Medical Database (Alumni Edition) Technology Research Database ProQuest SciTech Collection ProQuest Technology Collection ProQuest Natural Science Collection Hospital Premium Collection Hospital Premium Collection (Alumni Edition) ProQuest Central (Alumni) (purchase pre-March 2016) Materials Science & Engineering Collection ProQuest Central ProQuest Central UK/Ireland ProQuest Central Essentials Biological Science Collection ProQuest Central Database Suite (ProQuest) Technology Collection Natural Science Collection ProQuest One ProQuest Central Korea Engineering Research Database Health Research Premium Collection Health Research Premium Collection (Alumni) ProQuest Central Student SciTech Collection (ProQuest) ProQuest Health & Medical Complete (Alumni) ProQuest Engineering Collection ProQuest Biological Science Collection ProQuest Health & Medical Collection Medical Database Biological Science Database Engineering Database Biotechnology and BioEngineering Abstracts ProQuest Central Premium ProQuest One Academic (New) Publicly Available Content Database ProQuest Health & Medical Research Collection ProQuest One Academic Middle East (New) ProQuest One Health & Nursing ProQuest One Academic Eastern Edition (DO NOT USE) ProQuest One Applied & Life Sciences ProQuest One Academic ProQuest One Academic UKI Edition ProQuest Central China Engineering Collection MEDLINE - Academic PubMed Central (Full Participant titles) |
DatabaseTitle | CrossRef MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) Publicly Available Content Database ProQuest Central Student Technology Collection Technology Research Database ProQuest One Academic Middle East (New) ProQuest Central Essentials ProQuest Health & Medical Complete (Alumni) ProQuest Central (Alumni Edition) SciTech Premium Collection ProQuest One Community College ProQuest One Health & Nursing ProQuest Natural Science Collection ProQuest Central China ProQuest Central ProQuest One Applied & Life Sciences ProQuest Health & Medical Research Collection ProQuest Engineering Collection Health Research Premium Collection Biotechnology Research Abstracts Health and Medicine Complete (Alumni Edition) Natural Science Collection ProQuest Central Korea Health & Medical Research Collection Biological Science Collection ProQuest Central (New) ProQuest Medical Library (Alumni) Engineering Collection Engineering Database ProQuest Biological Science Collection ProQuest One Academic Eastern Edition ProQuest Hospital Collection ProQuest Technology Collection Health Research Premium Collection (Alumni) Biological Science Database ProQuest SciTech Collection ProQuest Hospital Collection (Alumni) Biotechnology and BioEngineering Abstracts ProQuest Health & Medical Complete ProQuest Medical Library ProQuest One Academic UKI Edition Materials Science & Engineering Collection Engineering Research Database ProQuest One Academic ProQuest One Academic (New) ProQuest Central (Alumni) MEDLINE - Academic |
DatabaseTitleList | Publicly Available Content Database MEDLINE - Academic MEDLINE Engineering Research Database |
Database_xml | – sequence: 1 dbid: C6C name: Springer Nature OA Free Journals url: http://www.springeropen.com/ sourceTypes: Publisher – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database – sequence: 4 dbid: 8FG name: ProQuest Technology Collection url: https://search.proquest.com/technologycollection1 sourceTypes: Aggregation Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine Engineering |
EISSN | 1475-925X |
EndPage | 41 |
ExternalDocumentID | PMC3661397 oai_biomedcentral_com_1475_925X_12_41 2980519801 A534654599 23668264 10_1186_1475_925X_12_41 |
Genre | Research Support, Non-U.S. Gov't Journal Article |
GeographicLocations | United States |
GeographicLocations_xml | – name: United States |
GroupedDBID | --- 0R~ 23N 2WC 4.4 53G 5GY 5VS 6J9 6PF 7X7 88E 8FE 8FG 8FH 8FI 8FJ AAFWJ AAJSJ AASML AAWTL ABDBF ABJCF ABUWG ACGFO ACGFS ACIHN ACIWK ACPRK ACUHS ADBBV ADMLS ADRAZ ADUKV AEAQA AENEX AFKRA AFPKN AFRAH AHBYD AHMBA AHSBF AHYZX ALMA_UNASSIGNED_HOLDINGS AMKLP AMTXH AOIJS BAPOH BAWUL BBNVY BCNDV BENPR BFQNJ BGLVJ BHPHI BMC BPHCQ BVXVI C6C CCPQU CS3 DIK E3Z EAD EAP EAS EBD EBLON EBS EJD EMB EMK EMOBN ESX F5P FRP FYUFA GROUPED_DOAJ GX1 H13 HCIFZ HMCUK HYE I-F IAO IGS IHR INH INR IPNFZ ISR ITC KQ8 L6V LK8 M1P M48 M7P M7S MK~ ML~ M~E O5R O5S OK1 OVT P2P PGMZT PHGZM PHGZT PIMPY PJZUB PPXIY PQGLB PQQKQ PROAC PSQYO PTHSS PUEGO RBZ RIG RNS ROL RPM RSV SEG SOJ SV3 TR2 TUS UKHRP W2D WOQ WOW XSB 2VQ AAYXX ALIPV C1A CITATION CGR CUY CVF ECM EIF NPM PMFND 3V. 7QO 7XB 8FD 8FK AZQEC DWQXO FR3 GNUQQ K9. P64 PKEHL PQEST PQUKI PRINS 7X8 -A0 ABVAZ ACRMQ ADINQ AFGXO AFNRJ C24 5PM |
ID | FETCH-LOGICAL-b655t-d3e701b2bb852f1b040ef2434a90309e95ede1cdb3b2a165bcf57002438215093 |
IEDL.DBID | M48 |
ISSN | 1475-925X |
IngestDate | Thu Aug 21 18:20:13 EDT 2025 Wed May 22 07:14:22 EDT 2024 Fri Sep 05 09:26:09 EDT 2025 Fri Sep 05 06:01:34 EDT 2025 Fri Jul 25 19:31:59 EDT 2025 Tue Jun 17 22:04:49 EDT 2025 Tue Jun 10 21:00:54 EDT 2025 Fri Jun 27 05:44:26 EDT 2025 Thu Apr 03 07:06:18 EDT 2025 Tue Jul 01 00:34:34 EDT 2025 Thu Apr 24 22:50:48 EDT 2025 Sat Sep 06 07:30:06 EDT 2025 |
IsDoiOpenAccess | true |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 1 |
Keywords | Retinal birefringence scanning Strabismus Time-frequency distribution Eye alignment Continuous wavelet transform Amblyopia |
Language | English |
License | This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-b655t-d3e701b2bb852f1b040ef2434a90309e95ede1cdb3b2a165bcf57002438215093 |
Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 ObjectType-Article-2 ObjectType-Feature-1 |
OpenAccessLink | https://doi.org/10.1186/1475-925X-12-41 |
PMID | 23668264 |
PQID | 1355576091 |
PQPubID | 42562 |
PageCount | 1 |
ParticipantIDs | pubmedcentral_primary_oai_pubmedcentral_nih_gov_3661397 biomedcentral_primary_oai_biomedcentral_com_1475_925X_12_41 proquest_miscellaneous_1367491695 proquest_miscellaneous_1355475893 proquest_journals_1355576091 gale_infotracmisc_A534654599 gale_infotracacademiconefile_A534654599 gale_incontextgauss_ISR_A534654599 pubmed_primary_23668264 crossref_primary_10_1186_1475_925X_12_41 crossref_citationtrail_10_1186_1475_925X_12_41 springer_journals_10_1186_1475_925X_12_41 |
ProviderPackageCode | CITATION AAYXX |
PublicationCentury | 2000 |
PublicationDate | 2013-05-13 |
PublicationDateYYYYMMDD | 2013-05-13 |
PublicationDate_xml | – month: 05 year: 2013 text: 2013-05-13 day: 13 |
PublicationDecade | 2010 |
PublicationPlace | London |
PublicationPlace_xml | – name: London – name: England |
PublicationTitle | Biomedical engineering online |
PublicationTitleAbbrev | BioMed Eng OnLine |
PublicationTitleAlternate | Biomed Eng Online |
PublicationYear | 2013 |
Publisher | BioMed Central BioMed Central Ltd |
Publisher_xml | – name: BioMed Central – name: BioMed Central Ltd |
References | 10.1186/1475-925X-12-41-B7 10.1186/1475-925X-12-41-B10 10.1186/1475-925X-12-41-B2 10.1186/1475-925X-12-41-B1 10.1186/1475-925X-12-41-B4 10.1186/1475-925X-12-41-B16 - 10.1186/1475-925X-12-41-B17 10.1186/1475-925X-12-41-B5 10764935 - Comput Methods Programs Biomed. 2000 Jun;62(2):87-98 12790456 - Appl Opt. 2003 Jun 1;42(16):3047-53 21750772 - Biomed Opt Express. 2011 Jul 1;2(7):1955-68 7666698 - Med Biol Eng Comput. 1995 May;33(3 Spec No):482-7 10474891 - J Opt Soc Am A Opt Image Sci Vis. 1999 Sep;16(9):2103-11 20725346 - Appl Opt. 1992 Jul 1;31(19):3730-5 15568959 - J Biomed Opt. 2004 Nov-Dec;9(6):1363-8 3351653 - J Opt Soc Am A. 1988 Jan;5(1):49-57 18556041 - Vision Res. 2008 Nov;48(26):2578-85 18305742 - Appl Opt. 1999 Mar 1;38(7):1273-9 7733188 - Am J Ophthalmol. 1995 May;119(5):627-36 |
References_xml | – ident: - doi: 10.1364/AO.38.001273 – ident: 10.1186/1475-925X-12-41-B7 doi: 10.1117/1.1805560 – ident: - doi: 10.1364/JOSAA.16.002103 – ident: - doi: 10.1109/TASSP.1987.1165201 – ident: 10.1186/1475-925X-12-41-B1 doi: 10.1364/JOSAA.5.000049 – ident: 10.1186/1475-925X-12-41-B17 doi: 10.1016/S0169-2607(00)00060-2 – ident: 10.1186/1475-925X-12-41-B4 doi: 10.1016/j.visres.2008.04.031 – ident: 10.1186/1475-925X-12-41-B10 doi: 10.1364/AO.42.003047 – ident: 10.1186/1475-925X-12-41-B16 doi: 10.1007/BF02510534 – ident: 10.1186/1475-925X-12-41-B2 doi: 10.1364/AO.31.003730 – ident: 10.1186/1475-925X-12-41-B5 doi: 10.1364/BOE.2.001955 – ident: - doi: 10.1142/S0218001487000205 – reference: 21750772 - Biomed Opt Express. 2011 Jul 1;2(7):1955-68 – reference: 12790456 - Appl Opt. 2003 Jun 1;42(16):3047-53 – reference: 7733188 - Am J Ophthalmol. 1995 May;119(5):627-36 – reference: 10764935 - Comput Methods Programs Biomed. 2000 Jun;62(2):87-98 – reference: 7666698 - Med Biol Eng Comput. 1995 May;33(3 Spec No):482-7 – reference: 18305742 - Appl Opt. 1999 Mar 1;38(7):1273-9 – reference: 10474891 - J Opt Soc Am A Opt Image Sci Vis. 1999 Sep;16(9):2103-11 – reference: 3351653 - J Opt Soc Am A. 1988 Jan;5(1):49-57 – reference: 15568959 - J Biomed Opt. 2004 Nov-Dec;9(6):1363-8 – reference: 18556041 - Vision Res. 2008 Nov;48(26):2578-85 – reference: 20725346 - Appl Opt. 1992 Jul 1;31(19):3730-5 |
SSID | ssj0020069 |
Score | 2.0313547 |
Snippet | Background
The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light... The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light upon... Background The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of light... Doc number: 41 Abstract Background: The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the... Background: The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of... BACKGROUND: The fovea, which is the most sensitive part of the retina, is known to have birefringent properties, i.e. it changes the polarization state of... |
SourceID | pubmedcentral biomedcentral proquest gale pubmed crossref springer |
SourceType | Open Access Repository Aggregation Database Index Database Enrichment Source Publisher |
StartPage | 41 |
SubjectTerms | Biomaterials Biomedical Engineering and Bioengineering Biomedical Engineering/Biotechnology Biomedical research Biotechnology Birefringence Child Digital signal processors Engineering Fixation, Ocular Fovea Centralis - physiology Frequency distribution Humans Light Male Medical equipment Noise Pediatrics Physiological apparatus Physiological aspects Retina Retina - physiology Scanners Signal processing Time Factors Wavelet Analysis Wavelet transforms |
SummonAdditionalLinks | – databaseName: ProQuest Technology Collection dbid: 8FG link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3db9MwELdgSIg9IBgfC2zIICTgwWx2YicRD9MElIE0HoBJfbNix94mVcm2tBL899wlTqkLrOpTfa7t3Nl3lzv_jpCXxiOuW80R-tKyzCnFDGhqVtVeuVzY_aKv3nD8VR2dZF-mchpeuHUhrXI8E_uDum4tviPf46AYwTYG9XZwccmwahRGV0MJjZvkFgdNg3JeTD4tHS6E4Q1wPrxQezzLJSuFnGI6ApaAj664zyLNtH4-ryio9eTJtQhqr5gm98jdYFHSw0EE7pMbrtkimys4g1vk9nGIoD8gsw8OwwbwM_XnP3u2UPhWdEgJp5gHf0qx4jzzV0Oa9S9aI7puKIzV0dYDOV5-xGENHJm-nxhID-3sUAKJYlpINXtITiYff7w_YqHiAjNKyjmrU5fvcyOMKaTw3MAOd15kaVaVGIpxpXS147Y2qREVV9JYj_j4AqOJYFmW6SOy0bSN2ybUpVYhvD18bFb5woDlIWtvcqWchH9KyLvo6euLAV1DI9513AJioJF3GnmnudAZT8jbkVfaBjBzrKkx071TU6i_O7xedhhH-i_pC2S-RnSMBtNvTqtF1-nP37_pQ5ki_pwsYfqvApFvYWBbhdsMsHgE1IoodyJK2L42bh5lTIfjo9N_hD0hz5fN2BNT4hrXLgYamDjYm9fRqDwDB6CUCXk8iO1y9QJ4A75llpA8EuiIEXFLc37WA5BDT3QcEvJmFP2Vqf_7oT65fplPyR3R1xmRjKc7ZGN-tXC7YO3NzbN-S_8GwZVT-A priority: 102 providerName: ProQuest – databaseName: Open Access资源_Springer Nature OA Free Journals dbid: C6C link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3daxQxEA_SguiD2Pq12koUQX2IbbJJdpc-lWqpQn1QC_cWNtmkFo496d6B_vfOZHPH5ayCxz5dJptsZpLMMDO_IeSVDYjr1nGEvnRMeq2ZhZuatV3QvhLusI7VG84_67ML-WmiJgkkCXNh1v33vNYHXFaKNUJNMIQAE9S3FS919Mrqk5VlhXi7Cbfnhk4buezT7AraPIjXbqLNKMkNV2m8gU7vk3tJdaTHI693yC3f75K7a4CCu-T2eXKVPyDT9x79A_A3DVc_4_pTeFo6xn5TDHi_pFhanoXrMZ76F-0QRjdVwBroLAA5ZjnisBbOxhAnBmJCBzfWOqIY_9FOH5KL0w_fTs5YKq3ArFZqzrrSV4fcCmtrJQK3sJV9ELKUbYM-F98o33nuOlta0XKtrAsIhC_QbQgqZFM-Ilv9rPdPCPWl04hjDz8n21BbUDFUF2yltVfwpoIcZatvfowwGgaBrfMW2GMGeWeQd4YLI3lB3i15ZVxCLcfiGVMTrZda_9nhzarDcqS_kr5E5huEwegxzuayXQyD-fj1izlWJQLNqQam_zoRhRkM7NqUtgAfj8hZGeVeRgn71OXNSxkz6ZwYDAd1Dyw-UNoK8mLVjD0x9q33s8VIAxMHxfJfNLqSoOk3qiCPR7Fdfb0A3oARKQtSZQKdMSJv6a--R6Rx6IkWQkHeLkV_beo3L-rT_6B9Ru6IWF1EMV7uka359cLvg443t8_j_v4NDmlLcg priority: 102 providerName: Springer Nature |
Title | Detecting fixation on a target using time-frequency distributions of a retinal birefringence scanning signal |
URI | https://link.springer.com/article/10.1186/1475-925X-12-41 https://www.ncbi.nlm.nih.gov/pubmed/23668264 https://www.proquest.com/docview/1355576091 https://www.proquest.com/docview/1355475893 https://www.proquest.com/docview/1367491695 http://dx.doi.org/10.1186/1475-925X-12-41 https://pubmed.ncbi.nlm.nih.gov/PMC3661397 |
Volume | 12 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV3rb9MwELdgkxB8QDBegVEZhAR8yLY4tpMIITTKykDqhAaV-s2KHXtMitKtD2n777lzktKUgqiqSq3P8eN89l3v_DtCXmmHuG5FhNCXJuRWylDDSR3mhZM2YeYg9dkbhifyeMS_jsX4dzqgZgJnG007zCc1mpZ7V5fXH0Dg33uBT-V-xBMRZkyMMcwAL7Fvw7Ek0RIb8qVLAU3nzF81aogbnJ8ND1i7-152jqz1jXvl5FqPqlxzrfoTa3CP3G1UTXpYr4375IatdsidFQDCHXJr2LjWH5Dyk0V_AvxM3fmV5xeFd07rWHGKAfJnFFPRh25ax19f0wJhd5uMWTM6cUCOtyKxWQ17qfMdg2VFZ6bOjUQxXiQvH5LR4OhH_zhsUjGEWgoxD4vYJgeRZlqngrlIg-hbx3jM8wx9NDYTtrCRKXSsWR5JoY1D4HyGbkZQObP4EdmqJpV9QqiNjUTce3gZnrtUg0oiCqcTKa2AJwXkXWf21UUNu6EQCLtbAjKpkHcKeacipngUkL2WV8o0KOeYbKNU3tpJ5Z8V3iwrtC39lfQlMl8hbEaFcTln-WI2U1--n6pDESMwncig-68bIjeBhk3eXHOAwSPSVodyt0MJcm26xe0aU61YqAjUQ7AQQckLyItlMdbEWLnKThY1DXQcFNF_0ciEg2WQiYA8rpftcvQMeANGJw9I0lnQHUZ0S6rznx6ZHGqiRRGQt-3SX-n65kl9-h_jeEZuM5-FRIRRvEu25tOFfQ664Fz3yM1knMBnOvjcI9sfj06-ncK3vuz3_L8rPb8H_ALi_F_L |
linkProvider | Scholars Portal |
linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3db9MwED9NQ-LjAcH4CgwwCAQ8hM1O7KRCCE2MqWXrHmCT-mZixx6TqnQsrWD_FH8jd_nomgJ7W9Wn-lzbufN9xOffAbwwnnDdck7QlzaMnVKhQUsdZrlXLhF2M62qNwz3Vf8w_jySoxX43d6FobTKVidWijqfWHpHvsHRMKJvjObtw8mPkKpG0elqW0KjFotdd_YTQ7by_WAb-ftSiJ1PBx_7YVNVIDRKymmYRy7Z5EYYk0rhuUEpdl7EUZz16LjB9aTLHbe5iYzIuJLGesKAF3Riht4TgS-hyr8S05tx3D_J6DzAI9jfBj6Ip2qDx4kMe0KOKP2BSs53rtSPO5Zw2R4sGMTlZM2lE9vKEO7cgpuNB8u2apG7DSuuWIMbC7iGa3B12JzY34HxtqNjCvyZ-eNflRgw_GasTkFnlHd_xKjCfehP67TuM5YTmm9TiKtkE4_kdNmShjWoon01MZRWVtq65BKjNJRsfBcOL4UX92C1mBTuATAXWUVw-vixceZTg56OzL1JlHIS_ymAd52nr09qNA9N-NrdFhQ7TbzTxDvNhY55AG9bXmnbgKdTDY-xroKoVP3d4fW8QzvSf0mfE_M1oXEUlO5zlM3KUg--ftFbMiK8O9nD6b9qiPwEB7ZZc3sCF08AXh3K9Q4lqgvbbW5lTDfqqtTnmyuAZ_Nm6kkpeIWbzGoanDj6txfRqCTGgKMnA7hfi-189QJ5g7FsHEDSEegOI7otxfH3CvAce1KgEsCbVvQXpv7vh_rw4mU-hWv9g-Ge3hvs7z6C66KqcSJDHq3D6vR05h6jpzk1T6rtzeDbZeuTP4tnj5E |
linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV3db9MwED-hIU3wgGB8BQYYhAQ8ZJsd22nE07RRbcAmBEzqmxU79phUpdPSSvDfc5c4VVMGElWf6nPt5M6-O93d7wBe20C4bhUn6EuXSq91alFTp2UVtM-F2xu13RtOTvXRmfw4UZOYm9P02e59SLKraSCUpnq-e1mF7oiP9C6XuUoLoSaUWEBl6zcl6T2K1eqDpb9FKLwRzeeaSWsV7tOBYlq_nlf003ru5FoAtdVL47twJxqUbL-TgHtww9dbcHsFZnALNk9iAP0-TA89RQ3wZxYufrZcYfgtWZcRzigN_pxRw_k0XHVZ1r9YReC6sS9Ww2YByan2kZa1eGOGdmMoPKxxXQckRlkh5fQBnI0_fD84SmPDhdRqpeZplfl8j1th7UiJwC0ecB-EzGRZUCTGF8pXnrvKZlaUXCvrAsHjCwomomFZZA9ho57V_jEwnzlN6Pb4cbIMI4uGh6qCzbX2Cv8pgfeDt28uO3ANQ3DXwxEUAkO8M8Q7w4WRPIGdnlfGRSxzaqkxNa1PM9J_Tni7nNCv9FfSV8R8Q-AYNWXfnJeLpjHH376afZUR_JwqcPtvIlGY4cKujMUM-PCEpzWg3B5Q4ul1w-Fexky8PRrD0QhEPxBFOoGXy2GaSRlxtZ8tOhrcOJqb_6LRuUT7v1AJPOrEdvn0AnmDrqVMIB8I9IARw5H64keLP44zyW9I4F0v-itbv_6lPvkP2hew-eVwbD4fn356CrdE235EpTzbho351cI_QyNwbp-3R_03hYpWpg |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Detecting+fixation+on+a+target+using+time-frequency+distributions+of+a+retinal+birefringence+scanning+signal&rft.jtitle=Biomedical+engineering+online&rft.au=Gramatikov%2C+Boris&rft.date=2013-05-13&rft.issn=1475-925X&rft.eissn=1475-925X&rft.volume=12&rft.spage=41&rft_id=info:doi/10.1186%2F1475-925X-12-41&rft.externalDBID=NO_FULL_TEXT |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1475-925X&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1475-925X&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1475-925X&client=summon |